Paper
16 September 1985 Multibeam Crt For Direct-View Display Of Photographic Quality Medical Images
Robert G. Quick Jr.
Author Affiliations +
Abstract
A display system has been developed which is capable of full 2048 X 2048 pixel resolution with an eight-bit gray scale. An essential element of this system is a multibeam cathode ray tube (CRT), which enables the flicker-free display of 4,000,000 picture elements without the penalty of unmanageable video and deflection frequencies. Such a tube has been found to deliver 15-20 ft-Lamberts brightness with actual tube resolution somewhat in excess of the requirement above, over a 10 inch square raster. This paper describes the tube, its electron optics, and performance.
© (1985) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Robert G. Quick Jr. "Multibeam Crt For Direct-View Display Of Photographic Quality Medical Images", Proc. SPIE 0536, 3rd Intl Conf on Picture Archiving and Communication Systems, (16 September 1985); https://doi.org/10.1117/12.947329