27 April 1988 Cryogenic Limb Array Etalon Spectrometer (CLAES)1 S/N001 Cryostat Test Results
W. J. Horsley, L. Naes
Author Affiliations +
Abstract
The S/N001 Cryostat2 tested consists of the originally designed solid hydrogen configuration coupled with an instrument thermal mass simulator (ITMS) for simulating the thermal behavior of the instrument package. The test program was designed to verify the operational and thermal performance of the Cryostat/ITMS system as monitored and serviced by the Data Acquiisition System (DAS) and Ground Support Equipment (GSE). After the DAS and GSE passed functional performance tests, the Cryostat/ITMS was slowly pumped from ambient to 10 torr to minimize any possible damage to the structures and Multilayer Insulation (MLI). The system was then pumped to a hard vacuum and tested for vacuum integrity of all 0-ring seals. During this initial pumpdown phase of the test, it was found that better communication was needed between the various compartments of the vacuum space for efficient evacuation of the cryostat. These improvements have been implemented on this unit as well as on the subsequent Ne/CO2 Cryostat. Liquid Nitrogen flowing through the coolant lines was used to precool the Cryostat to slightly above 77 degress Kelvin, at which point the tank was filled with liquid Helium. The test data includes the rate of temperature decrease of the various parts of the Cryostat/ITMS and the final equilibrium temperatures of the shields, thermal links and tank. These temperatures as well as the boiloff rate of the liquid Helium were compared with the expected values as predicted by the thermal models developed in house. The liquid Helium was allowed to boil off and liquid Neon was introduced into the Cryostat. The Neon was solidified by first circulating liquid Helium through the coolant coils and maintained by pumping on the cooled Neon through the fill line. End of life temperature profiles across the tank were obtained and compared with predicted values. Additional temperature and heat rate values were obtained and compared with predictions. System warmup procedures were verified without problems. The overall test resulted in significant design improvements for the Ne/CO2 Cryostat.
© (1988) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
W. J. Horsley and L. Naes "Cryogenic Limb Array Etalon Spectrometer (CLAES)1 S/N001 Cryostat Test Results", Proc. SPIE 0973, Cryogenic Optical Systems and Instruments III, (27 April 1988); https://doi.org/10.1117/12.948384
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Telescopes

Helium

Liquids

Neon

Sensors

Spectroscopy

Interfaces

RELATED CONTENT

Flight-qualified solid argon cooler for the BBXRT instrument
Proceedings of SPIE (November 01 1990)
Growth of very large sapphire crystals for optical applications
Proceedings of SPIE (September 23 1998)
FIFI LS a far infared 3D spectral imager for...
Proceedings of SPIE (March 03 2003)
Field imaging far-infrared line spectrometer FIFI LS
Proceedings of SPIE (December 06 1999)
ISAAC a 1 to 5 um imager spectrometer...
Proceedings of SPIE (March 21 1997)

Back to Top