Paper
14 April 1989 On-The-Job Calibration Of Tomodensitometric Systems - A Case Study
Sanjib K. Ghosh, Mohamed Bougouss
Author Affiliations +
Proceedings Volume 1030, Biostereometrics '88; (1989) https://doi.org/10.1117/12.950483
Event: Biostereometrics '88: Spatial and Spatiotemporal Analysis of Biological Form and Function, 1988, Basel, Switzerland
Abstract
A new method of calibrating tomodensitometric systems developed at Laval Univer-sity is described. This is based on the basic concept that each image represents parallel projection geometry for the sectional plane it represents. Scale affinity and inherent distortions are mathematically modeled. A series of parallel consecutive images taken of object placed in a specially designed plastic cube gives a cost-effective solution. The obtained results indicate a standard error of around ± 0.2 μm in each of three dimensions.
© (1989) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Sanjib K. Ghosh and Mohamed Bougouss "On-The-Job Calibration Of Tomodensitometric Systems - A Case Study", Proc. SPIE 1030, Biostereometrics '88, (14 April 1989); https://doi.org/10.1117/12.950483
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Data acquisition

Computing systems

Lead

Biostereometrics

Cameras

Imaging systems

Back to Top