9 March 2018 Scatter correction for multi-slice CT system
Yang Wang, Karl Stierstorfer, Martin Petersilka, Michael Grasruck, Yi Tian
Author Affiliations +
Abstract
Scatter radiation in multi-slice CT system nowadays is playing a more and more important role, as the slice number is getting larger and larger, e.g. from 16 to 64 and even more. Scatter radiation may downgrade image quality e.g. create inhomogeneity and decrease the image contrast. Although in current multi-slice CT, anti-scatter collimation is widely used to reduce the scatter radiation received, as the detector is getting wider, its efficiency is getting weaker. Although beam hardening correction can somehow guarantee image homogeneity, as beam hardening effect and scatter radiation have different physical origin, a delegated scatter correction algorithm is desired. In this paper, we would like to propose a scatter correction algorithm working during data pre-processing before image reconstruction. After we implemented this algorithm in Siemens latest released Somatom Go. CT system, we obtain good image quality especially under certain clinical cases.
© (2018) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Yang Wang, Karl Stierstorfer, Martin Petersilka, Michael Grasruck, and Yi Tian "Scatter correction for multi-slice CT system", Proc. SPIE 10573, Medical Imaging 2018: Physics of Medical Imaging, 105733K (9 March 2018); https://doi.org/10.1117/12.2293117
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Image quality

Monte Carlo methods

Convolution

Data modeling

X-ray computed tomography

Reconstruction algorithms

Scattering compensation

Back to Top