7 November 1994 X-ray performance of the engineering prototype Stellar X-Ray Polarimeter
Eric H. Silver, Klaus Ziock, J. Dwyer, Philip E. Kaaret, Robert Novick, Ronald F. Elsner, Martin C. Weisskopf, Enrico Costa, Paolo Soffitta, Giuseppe Manzo, Andrea E. Santangelo, Igor Y. Lapshov, Rashid Sunyaev
Author Affiliations +
Abstract
The performance of the engineering prototype Stellar X-Ray Polarimeter (SXRP) has been evaluated. One hundred percent polarized monochromatic x rays at 2.6 keV and 9.7 keV were used to measure the response of the instrument in the energy bands of the graphite and lithium polarizing elements, respectively. On-line analysis showed that the respective depths of modulation are 96% ad 70% as expected. Irradiating SXRP with broadband unpolarized x rays in the energy band 2 - 17 keV demonstrated that the level of spurious modulation inherent in the instrument is less than 3%. Up-to-date results are presented and compared to the predictions of Monte Carlo simulations.
© (1994) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Eric H. Silver, Klaus Ziock, J. Dwyer, Philip E. Kaaret, Robert Novick, Ronald F. Elsner, Martin C. Weisskopf, Enrico Costa, Paolo Soffitta, Giuseppe Manzo, Andrea E. Santangelo, Igor Y. Lapshov, and Rashid Sunyaev "X-ray performance of the engineering prototype Stellar X-Ray Polarimeter", Proc. SPIE 2283, X-Ray and Ultraviolet Spectroscopy and Polarimetry, (7 November 1994); https://doi.org/10.1117/12.193200
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-rays

Modulation

Monte Carlo methods

Polarizers

Crystals

Lithium

Polarimetry

RELATED CONTENT

Status of the stellar x ray polarimeter for the Spectrum...
Proceedings of SPIE (February 15 1994)
X-Ray Polarimeters
Proceedings of SPIE (October 19 1977)
Calibration of the Stellar X-Ray Polarimeter
Proceedings of SPIE (October 15 1997)
Cylindrical proportional counter for x-ray polarimetry
Proceedings of SPIE (February 01 1991)
Development of bent crystal for imaging polarimetry
Proceedings of SPIE (September 17 2012)

Back to Top