16 October 1995 Sensitivity to x-ray polarization of a microgap gas proportional counter
Author Affiliations +
Abstract
We measured the average anisotropy of the primary charge cloud produced by photoelectron when an x-ray beam linearly polarized is absorbed on a Ne-DME gas mixture by using a micro-gap proportional counter. This average anisotropy is not present when an Fe55 unpolarized x-ray source is used. We discuss the results of our measurement in terms of performances of this detector as an x-ray polarimeter.
© (1995) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Paolo Soffitta, Enrico Costa, Ennio Morelli, Ronaldo Bellazzini, Alessandro Brez, R. Raffo, "Sensitivity to x-ray polarization of a microgap gas proportional counter", Proc. SPIE 2517, X-Ray and EUV/FUV Spectroscopy and Polarimetry, (16 October 1995); doi: 10.1117/12.224922; https://doi.org/10.1117/12.224922