Paper
15 September 1995 Performance tests of a speckle camera system
Fesseha G. Mariam, Jeff E. Houchard, Mark A. Lundgren, Daniel G. O'Connell, Victor S. Wang, Andrew H. Suzuki
Author Affiliations +
Abstract
A series of tests on a speckle imaging camera system using the Knox-Thompson image reconstruction algorithm was performed. For simulated and real point sources the speckle OTF remains above the noise level for spatial frequencies up to 0.8 of diffraction limit. Laboratory measurements of MTF curves using a resolution chart target were also made with and without simulated atmospheric phase aberrations. Even in the presence of aberrations the KT processed images have consistently higher MTF than those without phase aberrations. We also present a method for subtracting power spectra bias applicable to CCD focal planes with finite electronic noise.
© (1995) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Fesseha G. Mariam, Jeff E. Houchard, Mark A. Lundgren, Daniel G. O'Connell, Victor S. Wang, and Andrew H. Suzuki "Performance tests of a speckle camera system", Proc. SPIE 2540, Current Developments in Optical Design and Engineering V, (15 September 1995); https://doi.org/10.1117/12.219508