Paper
27 June 1997 Spectrometer design for transmission and reflection measurements of full-scale windows
Donald S. Fisher, Darrel Fuller
Author Affiliations +
Abstract
Current spectrometers are unsuitable for measuring transmission of large thick electro-optical windows because the optical path through the instrument is changed by the optic. Customization of a standard FTIR spectrometer capable of making normal and non-normal incidence angle transmission and reflection measurements is presented which addresses this and other problems associated with high precision spectroradiometry. Emphasis is on design features to ensure ordinate accuracy of these measurements. These techniques can replace witness sample testing commonly made in the industry.
© (1997) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Donald S. Fisher and Darrel Fuller "Spectrometer design for transmission and reflection measurements of full-scale windows", Proc. SPIE 3060, Window and Dome Technologies and Materials V, (27 June 1997); https://doi.org/10.1117/12.277068
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Reflectivity

Transmittance

Mirrors

Collimation

Spectroscopy

FT-IR spectroscopy

Back to Top