28 May 2001 Evaluation of a new method for stenosis quantification from 3D x-ray angiography images
Fabienne Betting, Gilles Moris, Jerome Knoplioch, Yves L. Trousset, Francisco Sureda, Laurent Launay
Author Affiliations +
Abstract
A new method for stenosis quantification from 3D X-ray angiography images has been evaluated on both phantom and clinical data. On phantoms, for the parts larger or equal to 3 mm, the standard deviation of the measurement error has always found to be less or equal to 0.4 mm, and the maximum measurement error less than 0.17 mm. No clear relationship has been observed between the performances of the quantification method and the acquisition FoV. On clinical data, the 3D quantification method proved to be more robust to vessel bifurcations than its 3D equivalent. On a total of 15 clinical cases, the differences between 2D and 3D quantification were always less than 0.7 mm. The conclusion is that stenosis quantification from 3D X-4ay angiography images is an attractive alternative to quantification from 2D X-ray images.
© (2001) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Fabienne Betting, Gilles Moris, Jerome Knoplioch, Yves L. Trousset, Francisco Sureda, and Laurent Launay "Evaluation of a new method for stenosis quantification from 3D x-ray angiography images", Proc. SPIE 4319, Medical Imaging 2001: Visualization, Display, and Image-Guided Procedures, (28 May 2001); https://doi.org/10.1117/12.428054
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
3D metrology

3D image processing

Error analysis

X-rays

X-ray imaging

3D acquisition

Angiography

Back to Top