Paper
28 June 2001 Temporal artifacts in flat dynamic x-ray detectors
Michael Overdick, Torsten Solf, Hans-Aloys Wischmann
Author Affiliations +
Abstract
Flat X-ray detectors based on CsI:Tl scintillators and amorphous silicon photodiodes are known to exhibit temporal artefacts (ghost images) which decay over time. Previously, these temporal artefacts have been attributed mainly to residual signals from the amorphous silicon photodiodes. More detailed experiments presented here show that a second class of effects, the so-called gain effects, also contributes significantly to the observed temporal artefacts. Both the residual signals and the photodiode gain effect have been characterized under various exposure conditions in the study presented here. The results of the experiments quantitatively show the decay of the temporal artefacts. Additionally, the influence of the detector's reset light on both effects in the photodiode has been studied in detail. The data from the measurements is interpreted based on a simple trapping model which suggests a strong link between the photodiode residual signals and the photodiode gain effect. For the residual signal effect a possible correction scheme is described. Furthermore, the relevance of the remaining temporal artefacts for the applications is briefly discussed for both the photodiode residual signals and the photodiode gain effect.
© (2001) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Michael Overdick, Torsten Solf, and Hans-Aloys Wischmann "Temporal artifacts in flat dynamic x-ray detectors", Proc. SPIE 4320, Medical Imaging 2001: Physics of Medical Imaging, (28 June 2001); https://doi.org/10.1117/12.430913
Lens.org Logo
CITATIONS
Cited by 41 scholarly publications and 4 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Photodiodes

Signal detection

X-rays

Amorphous silicon

Scintillators

X-ray detectors

RELATED CONTENT

Latest advancements in state of the art aSi based x...
Proceedings of SPIE (March 09 2018)
Silicon drift chamber with extended energy range
Proceedings of SPIE (December 13 2000)
Solid-State 1024 Pixel Linear X-Ray Detector
Proceedings of SPIE (August 03 1987)
X VIEW a high resolution real time solid state...
Proceedings of SPIE (January 10 2003)

Back to Top