30 May 2003 Dose optimization tool
Ornit Amir, David Braunstein, Ami Altman
Author Affiliations +
Abstract
A dose optimization tool for CT scanners is presented using patient raw data to calculate noise. The tool uses a single patient image which is modified for various lower doses. Dose optimization is carried out without extra measurements by interactively visualizing the dose-induced changes in this image. This tool can be used either off line, on existing image(s) or, as a pre - requisite for dose optimization for the specific patient, during the patient clinical study. The algorithm of low-dose simulation consists of reconstruction of two images from a single measurement and uses those images to create the various lower dose images. This algorithm enables fast simulation of various low dose (mAs) images on a real patient image.
© (2003) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Ornit Amir, David Braunstein, and Ami Altman "Dose optimization tool", Proc. SPIE 5029, Medical Imaging 2003: Visualization, Image-Guided Procedures, and Display, (30 May 2003); https://doi.org/10.1117/12.482412