Paper
12 May 2004 Efficient NUFFT-based direct Fourier algorithm for fan beam CT reconstruction
Author Affiliations +
Abstract
Related to the demand for fast and efficient tomographic reconstruction methods, the interest for Direct Fourier (DF) methods, which have a reduced computational complexity, has been growing. In this paper we present a new NUFFT-based DF reconstruction method which can be directly applied to fan-beam CT data sets avoiding the interpolation in Radon space as well as the interpolation in Fourier space. The performance of the new algorithm, in ideal and noisy conditions, is compared to those of other well known reconstruction methods, revealing an excellent behavior, specially in noisy conditions.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Silvia De Francesco and Augusto Marques Ferreira da Silva "Efficient NUFFT-based direct Fourier algorithm for fan beam CT reconstruction", Proc. SPIE 5370, Medical Imaging 2004: Image Processing, (12 May 2004); https://doi.org/10.1117/12.534189
Lens.org Logo
CITATIONS
Cited by 9 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Reconstruction algorithms

Fluctuations and noise

CT reconstruction

Image filtering

Distance measurement

Feature extraction

Fourier transforms