Paper
15 September 2008 Investigation of helix-saddle trajectories for cone-beam CT
Author Affiliations +
Abstract
In this paper, a hybrid helix-saddle trajectory scanning mode is proposed for bolus-chasing CT angiography. By combining the conventional helical trajectory and saddle trajectory appropriately, an optimal curve can be obtained with a capability of localized volumetric imaging at desirable locations. In this context, a condition for the PI-line existence is determined. Then, both filtered-backprojection (FBP), backprojection filtration (BPF) and reduced-scan FBP algorithms are developed. Numerical studies with the 3D Shepp-Logan phantom support the validity and merits of the proposed trajectories and associated algorithms.
© (2008) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Yang Lu, Jun Zhao, Erwei Bai, and Ge Wang "Investigation of helix-saddle trajectories for cone-beam CT", Proc. SPIE 7078, Developments in X-Ray Tomography VI, 70780I (15 September 2008); https://doi.org/10.1117/12.793728
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Computed tomography

Reconstruction algorithms

Algorithm development

Angiography

Detection and tracking algorithms

Biomedical engineering

Back to Top