Abstract
Main goal of this study is to compare the performances of a full line of LCD diagnostic display systems, in terms of white point luminance, accuracy and stability over time, GSDF conformance and luminance uniformity, and to test a web-based service for remote calibration and QA in a large hospital. The display systems under test included 3MP and 5MP grayscale, 2MP and 6MP color and 5MP mammography LCD monitors, all manufactured by BARCO, for a total amount of 119 units. Measured performances were all within the acceptance range proposed by the major international protocols and show a very good stability in time, except for a few cases. The web-based service for remote QA and calibration resulted well suited for the management of a large scale medical facility, where high performance displays are in use and time saving QA programs and a central QA policy are both needed.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
S. Busoni, G. Belli, A. Taddeucci, S. Targetti, I. Menchi, C. Gori, L. Capaccioli, A. Falivene, G. Spanò, A. Guasti, P. Saletti, and F. Rossi "Performance evaluation of a full line of medical diagnostic displays and test of a web-based service for remote calibration and quality assurance", Proc. SPIE 7263, Medical Imaging 2009: Image Perception, Observer Performance, and Technology Assessment, 72630A (12 March 2009); https://doi.org/10.1117/12.812283
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
LCDs

Calibration

Sensors

Diagnostics

Modulation transfer functions

Instrument modeling

Chromium

RELATED CONTENT

Proposal for the development of a repository for in flight...
Proceedings of SPIE (September 18 1997)
Z306 black paint measurements
Proceedings of SPIE (December 01 1990)
Image quality analysis of a color LCD as well as...
Proceedings of SPIE (October 18 2007)

Back to Top