Paper
21 February 2013 A phantom study to characterize the imaging quality of a phase-contrast tomosynthesis prototype
Author Affiliations +
Proceedings Volume 8582, Biophotonics and Immune Responses VIII; 85820J (2013) https://doi.org/10.1117/12.2008644
Event: SPIE BiOS, 2013, San Francisco, California, United States
Abstract
This research is aimed at studying the advantages of an x-ray phase-contrast tomosynthesis prototype by using phantoms. A prototype system is assembled with a micro-focus x-ray source, a rotating stage and a computed radiography detector mounted on an optical rail. A custom designed bubble wrap phantom is used in experiments. Angular projection images are acquired from -20° to +20° with 2° interval. The in-plane slices are reconstructed. The feature area on the phantom is observed. The prototype system provides an intrinsic way to investigate the potential and imaging quality of a phase-contrast tomosynthesis imaging method. As the result, phase-contrast tomosynthesis imaging method is demonstrated for its advantages in avoiding structure noise and overlapping issues by comparing the results acquired by computed radiography and phase-contrast radiography.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Di Wu, Muhammad U. Ghani, Hui Miao, Yuhua Li, Wei R. Chen, Xizeng Wu, and Hong Liu "A phantom study to characterize the imaging quality of a phase-contrast tomosynthesis prototype", Proc. SPIE 8582, Biophotonics and Immune Responses VIII, 85820J (21 February 2013); https://doi.org/10.1117/12.2008644
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Radiography

Imaging systems

Prototyping

Chromium

X-rays

Sensors

X-ray imaging

Back to Top