28 October 2020 Identification of aspirated radiolucent foreign bodies in the pediatric airway using digital tomosynthesis: a multireader phantom study
Gentry Russell, Nima Kasraie, Janelle R. Noel-MacDonnell, Amie L. Robinson, Sherwin S. Chan
Author Affiliations +
Abstract

Purpose: Foreign body aspiration is difficult to diagnose because many aspirated foreign bodies are low density or radiolucent. Digital radiographs (DR) are poor at detecting radiolucent foreign bodies. Digital tomosynthesis (DTS) has been shown to be ideally suited for applications where DR is insensitive and the increased dose from computed tomography (CT) is not justified. Our objective was to determine if DTS can be a practical alternative to DR and CT in the diagnosis of foreign body aspiration.

Approach: A phantom approximating the densities of a pediatric chest was constructed. Radiolucent foreign bodies were placed in the airways. Seven pediatric radiologists assessed DTS and DR images with and without simulated breathing motion. Two rounds were performed with fixed exposure techniques and then automatic exposure control techniques. Interobserver agreement was evaluated using Fleiss’ kappa.

Results: DTS and DR images using fixed exposure techniques performed very poorly with accuracies of 42% to 60%. DTS with automatic exposure control techniques increased accuracy to 84% for a stationary phantom, but the accuracy dropped to 70% in a phantom with simulated motion. DTS outperformed DR, with DR accuracies of 60% and 63% for stationary simulations and motion, respectively. Interobserver agreement was poor with Fleiss’ kappa of 0.476.

Conclusion: DTS is superior to DR for radiolucent foreign body detection. However, the overall accuracy and interobserver agreement are likely too low for this modality to be clinically useful.

© 2020 Society of Photo-Optical Instrumentation Engineers (SPIE) 2329-4302/2020/$28.00 © 2020 SPIE
Gentry Russell, Nima Kasraie, Janelle R. Noel-MacDonnell, Amie L. Robinson, and Sherwin S. Chan "Identification of aspirated radiolucent foreign bodies in the pediatric airway using digital tomosynthesis: a multireader phantom study," Journal of Medical Imaging 7(5), 055502 (28 October 2020). https://doi.org/10.1117/1.JMI.7.5.055502
Received: 22 July 2020; Accepted: 14 October 2020; Published: 28 October 2020
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Radiography

Chest

Bronchoscopy

Diagnostics

3D modeling

Medicine

3D printing