This will count as one of your downloads.
You will have access to both the presentation and article (if available).
Preliminary results obtained using a Lissajous trajectory beam-scanning approach coupled with temporal multiplexing by the implementation of an optical delay line demonstrate the ability to acquire 2 distinct focal planes simultaneously at frame rates >450 Hz for full 512 × 512 images. The use of multi-channel data acquisition cards allows for simultaneous multimodal image acquisition with perfect image registry between all imaging modalities. Also discussed here is the implementation of Lissajous trajectory beam-scanning on commercially available microscope hardware.
This will count as one of your downloads.
You will have access to both the presentation and article (if available).
View contact details