7 March 2022 Effect of air trap in balloon on circumferential laser ablation of urethral stenosis
Author Affiliations +
Abstract
A balloon-integrated diffusing applicator has been developed to treat urethral stenosis. To deliver uniform light, the balloon was inflated by inserting water. However, the air trap in the balloon can cause non-uniform ablation in the tissue due to different thermal boundary conditions between air and water. The current study aims to evaluate optical and thermal effects of the air trap on cylindrical laser treatment of urethral stenosis. Although a siginificant decrease of the air volume in the balloon, at least two deflations were required to entail uniform coagulation in a tubular structure and to avoid the unpredictable optical/thermal effects from the air trap.