Handheld optical coherence tomography (OCT) systems facilitate imaging of young children, bedridden subjects, and those with less stable fixation. Smaller and lighter OCT probes allow for more efficient imaging and reduced operator fatigue, which is critical for prolonged use in either the operating room or neonatal intensive care unit. In addition to size and weight, the imaging speed, image quality, field of view, resolution, and focus correction capability are critical parameters that determine the clinical utility of a handheld probe. Here, we describe an ultra-compact swept source (SS) OCT handheld probe weighing only 211 g (half the weight of the next lightest handheld SSOCT probe in the literature) with 20.1 µm lateral resolution, 7 µm axial resolution, 102 dB peak sensitivity, a 27° x 23° field of view, and motorized focus adjustment for refraction correction between -10 to +16 D. A 2D microelectromechanical systems (MEMS) scanner, a converging beam-at-scanner telescope configuration, and an optical design employing 6 different custom optics were used to minimize device size and weight while achieving diffraction limited performance throughout the system’s field of view. Custom graphics processing unit (GPU)-accelerated software was used to provide real-time display of OCT B-scans and volumes. Retinal images were acquired from adult volunteers to demonstrate imaging performance.
|