Paper
1 August 1991 Radiometric calibration of SPOT 2 HRV: a comparison of three methods
Stuart F. Biggar, Magdeleine C. Dinguirard, David I. Gellman, Patrice J. Henry, Ray D. Jackson, M. Susan Moran, Philip N. Slater
Author Affiliations +
Abstract
Three methods for determining an absolute radiometric calibration of a spacecraft optical sensor are compared. They are the well-known reflectance-based and radiance-based methods and a new method based on measurements of the ratio of diffuse-to-global irradiance at the ground. The latter will be described in detail and the comparison of the three approaches will be made with reference to the SPOT-2 HRV cameras for a field campaign 1990-06-19 through 1990-06-24 at the White Sands Missile Range in New Mexico.
© (1991) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Stuart F. Biggar, Magdeleine C. Dinguirard, David I. Gellman, Patrice J. Henry, Ray D. Jackson, M. Susan Moran, and Philip N. Slater "Radiometric calibration of SPOT 2 HRV: a comparison of three methods", Proc. SPIE 1493, Calibration of Passive Remote Observing Optical and Microwave Instrumentation, (1 August 1991); https://doi.org/10.1117/12.46693