Paper
12 May 1995 Pulsed infrared laser ablation rates and characteristics in otic capsule
Brian Jet-Fei Wong M.D., Joseph Neev, Vivian Sung, Belinda Simonian, Eric Mun-Wight Cheung, Michael W. Berns
Author Affiliations +
Abstract
Holmium:YAG (2.1 micrometers ) and Erbium:YAG (2.9 micrometers ) lasers have been suggested for use in neurotologic operations. The relatively high absorption in biologic tissues and minimal thermal diffusion associated with pulsed lasers make them attractive alternatives to CO2 and visible wavelength lasers, which currently enjoy widespread clinical use. There has been no systematic investigation of the properties of these lasers in a tissue model relevant to temporal bone operations, though there are numerous clinically oriented studies using these wavelengths. The microarchitecture of otic capsule (bone tissue of the inner ear) differs dramatically from bone found elsewhere in the body. Porcine otic capsule was machined into 0.8 mm thick specimens. The ablation rates of Ho:YAG and Erbium:YAG lasers were determined with incident flux varying from 6 - 20 J/cm2/pulse at 4 Hz PPR. Otic capsule specimens were ablated under both wet (physiologic) and dry (chemical dehydration) conditions. With Ho:YAG irradiation, dry otic capsule ablated with multiple episodes of stalling out. The relative number of specimens where stalling out occurred decreased with increasing laser flux. Wet tissue demonstrated similar characteristics, but with a lower ablation rate for a given fluence. In Erbium:YAG treated tissue, similar observations were noted, but the dependence on fluence is less clear in both wet and dry tissues. Water is shown to play a key role in the ablation process. The threshold of ablation is lower in wet tissue than in dehydrated tissue. The homogeneous microarchitecture of otic tissue also strongly influences the ablation process. Possible mechanisms are discussed.
© (1995) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Brian Jet-Fei Wong M.D., Joseph Neev, Vivian Sung, Belinda Simonian, Eric Mun-Wight Cheung, and Michael W. Berns "Pulsed infrared laser ablation rates and characteristics in otic capsule", Proc. SPIE 2395, Lasers in Surgery: Advanced Characterization, Therapeutics, and Systems V, (12 May 1995); https://doi.org/10.1117/12.209112
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Laser ablation

Tissues

Bone

Pulsed laser operation

Er:YAG lasers

Ear

Absorption

RELATED CONTENT

Compact infrared laser for stapedotomy
Proceedings of SPIE (February 08 2008)
Pulsed Tm:YAG laser ablation of knee joint tissues
Proceedings of SPIE (June 01 1992)
Laser ablation of otic capsule tissue
Proceedings of SPIE (February 20 2008)
Erbium laser in middle ear surgery: laboratory study
Proceedings of SPIE (September 07 1994)
Effects of pulsed lasers on hard biological tissue
Proceedings of SPIE (July 07 1993)

Back to Top