Paper
30 May 1995 Radiometric calibration of infrared detectors and thermal imaging systems
Nigel P. Fox, Tim R. Prior, Evangelos Theocharous
Author Affiliations +
Abstract
The paper describes the establishment of an infrared spectral responsivity scale at NPL with an uncertainty of 1.6% and the development of facilities to characterize the optical properties of infrared detector, arrays and cameras. The paper will illustrate the use of the calibration facilities with examples of uniformity, linearity and spectral responsivity measurements on a range of detectors.
© (1995) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Nigel P. Fox, Tim R. Prior, and Evangelos Theocharous "Radiometric calibration of infrared detectors and thermal imaging systems", Proc. SPIE 2474, Smart Focal Plane Arrays and Focal Plane Array Testing, (30 May 1995); https://doi.org/10.1117/12.210558
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Calibration

Thermography

Imaging systems

Infrared detectors

Infrared radiation

Nanolithography