Abstract
Use of pure diffusion theory produced errors in the measurement of absorbance and scattering using a time-of- flight system. In an attempt to produce quantitative values for absorbance and scattering, we performed both software simulations and empirical collections. For the empiric correction, we first developed a calibration system based upon a homogeneous liquid test phantom. Next, we performed multiwavelength (785, 808, 850, 904 nm) time-resolved spectroscopy upon known standards using this model, and then developed an algorithm to correct for the differences between our system and the known values. We suspect that part of the difference between theory and actual measured values, in the absence of correction, may be due to the response characteristics of our diode-based laser and detection system. If so, then our calibration represents a deconvolution of the response characteristics of the system, and suggests that `less clean' components can be used, reducing cost of time-resolved systems.
© (1998) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Joshua L. Duckworth, David A. Benaron M.D., Susan R. Hintz, Frank W.H. Liu, David K. Stevenson, and Wai-Fung Cheong "Calibration of time-of-flight optical spectroscopy", Proc. SPIE 3194, Photon Propagation in Tissues III, (1 January 1998); https://doi.org/10.1117/12.301037
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Optical calibration

Optical spectroscopy

Absorbance

Algorithm development

Scattering

Spectral calibration

Back to Top