Paper
28 May 2001 Flat-panel cone-beam CT: a novel imaging technology for image-guided procedures
Jeffrey H. Siewerdsen, David A. Jaffray, Gregory K. Edmundson, W. P. Sanders, John W. Wong, Alvaro A. Martinez
Author Affiliations +
Abstract
The use of flat-panel imagers for cone-beam CT signals the emergence of an attractive technology for volumetric imaging. Recent investigations demonstrate volume images with high spatial resolution and soft-tissue visibility and point to a number of logistical characteristics (e.g., open geometry, volume acquisition in a single rotation about the patient, and separation of the imaging and patient support structures) that are attractive to a broad spectrum of applications. Considering application to image-guided (IG) procedures - specifically IG therapies - this paper examines the performance of flat-panel cone-beam CT in relation to numerous constraints and requirements, including time (i.e., speed of image acquisition), dose, and field-of-view. The imaging and guidance performance of a prototype flat panel cone-beam CT system is investigated through the construction of procedure-specific tasks that test the influence of image artifacts (e.g., x-ray scatter and beam-hardening) and volumetric imaging performance (e.g., 3D spatial resolution, noise, and contrast) - taking two specific examples in IG brachytherapy and IG vertebroplasty. For IG brachytherapy, a procedure-specific task is constructed which tests the performance of flat-panel cone-beam CT in measuring the volumetric distribution of Pd-103 permanent implant seeds in relation to neighboring bone and soft-tissue structures in a pelvis phantom. For IG interventional procedures, a procedure-specific task is constructed in the context of vertebroplasty performed on a cadaverized ovine spine, demonstrating the volumetric image quality in pre-, intra-, and post-therapeutic images of the region of interest and testing the performance of the system in measuring the volumetric distribution of bone cement (PMMA) relative to surrounding spinal anatomy. Each of these tasks highlights numerous promising and challenging aspects of flat-panel cone-beam CT applied to IG procedures.
© (2001) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jeffrey H. Siewerdsen, David A. Jaffray, Gregory K. Edmundson, W. P. Sanders, John W. Wong, and Alvaro A. Martinez "Flat-panel cone-beam CT: a novel imaging technology for image-guided procedures", Proc. SPIE 4319, Medical Imaging 2001: Visualization, Display, and Image-Guided Procedures, (28 May 2001); https://doi.org/10.1117/12.428085
Lens.org Logo
CITATIONS
Cited by 34 scholarly publications and 2 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-ray computed tomography

Therapeutics

Imaging systems

Computed tomography

Visualization

X-rays

Image processing

Back to Top