Paper
28 June 2001 Antialiasing weighting function for helical CT
Author Affiliations +
Abstract
Longitudinal aliasing effects plague most volumes reconstructed from single-slice helical computed tomography data, and its presence can degrade resolution and distort image structures. We have recently developed a Fourier-based approach to longitudinal interpolation in helical CT that can, under certain conditions, essentially eliminate this longitudinal aliasing by exploiting a generalization of the sampling theorem whose conditions are satisfied by the interlaced pairs of direct and complementary longitudinal samples. However, the algorithm is computationally intensive and cannot be pipelined. In the present work, we address this shortcoming by deriving a spatial-domain, projection- data weighting function that approximates the application of the Fourier-based approach, and preserves its aliasing suppression properties to some degree, while allowing for a pipelined implementation. We call the approach 180AA, for anti-aliasing.
© (2001) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Patrick J. La Riviere and Xiaochuan Pan "Antialiasing weighting function for helical CT", Proc. SPIE 4320, Medical Imaging 2001: Physics of Medical Imaging, (28 June 2001); https://doi.org/10.1117/12.430863
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Optical spheres

Computed tomography

Modulation transfer functions

Sensors

Reconstruction algorithms

Computer simulations

Collimation

Back to Top