20 April 2005 Physical evaluation of a high-frame-rate extended dynamic range flat panel detector for real-time cone beam computed tomography applications
Author Affiliations +
Abstract
The use of flat panel detectors in computed tomography (CT) systems can improve resolution, reduce system cost, and add operational flexibility by combining fluoroscopy and radiography applications within CT systems. However, some prior studies have suggested that flat panel detectors would not perform well in CT applications due to their lack of high dynamic range, lag artifacts, and inadequate frame rate. The purpose of this study was to perform a physical evaluation of a prototype flat panel detector capable of high frame rates and extended dynamic range. The flat panel detector used had a pixel size of 194 microns and a matrix size of 2048x1536. The detector could be configured for several combinations of frame rate and matrix size up to 750 frames per second for a 512x16 matrix size with 4x4 binning. The evaluation was performed in terms of the MTF and DQE as a function of frame rate and exposure at the IEC RQA5 (~75 kVp, 21 mm Al) beam quality. The image lag was evaluated in terms of temporal-frequency dependent transfer function. Offset shift were also evaluated. Preliminary results indicate 0.1 MTF at 0.92 cycles/mm and DQE(0) of approximately 0.8, 0.6, 0.4, and 0.22 at 0.144, 0.065, 0.035, and 0.008 mR per frame exposures. The temporal MTF exhibited a low-frequency drop and a value of 0.5 at the Nyquist frequency. Offset shift was negligible. Considering high frame rate capabilities of the new detector, the results suggest that the detector has potential for use in real-time CT applications including CT angiography.
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Sarah J. Boyce, Amarpreet Chawla, and Ehsan Samei "Physical evaluation of a high-frame-rate extended dynamic range flat panel detector for real-time cone beam computed tomography applications", Proc. SPIE 5745, Medical Imaging 2005: Physics of Medical Imaging, (20 April 2005); https://doi.org/10.1117/12.593881
Lens.org Logo
CITATIONS
Cited by 9 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Modulation transfer functions

Sensors

Computed tomography

Computing systems

Prototyping

X-rays

Imaging systems

Back to Top