20 April 2005 Potential of software-based scatter corrections in cone-beam volume CT
Author Affiliations +
Abstract
This study deals with a systematic assessment of the potential of different schemes for computerized scatter correction in flat detector based cone-beam X-ray computed tomography. The analysis is based on simulated scatter of a CT image of a human head. Using a Monte-Carlo cone-beam CT simulator, the spatial distribution of scattered radiation produced by this object has been calculated with high accuracy for the different projected views of a circular tomographic scan. Using this data and, as a reference, a scatter-free forward projection of the phantom, the potential of different schemes for scatter correction has been evaluated. In particular, the ideally achievable degree of accuracy of schemes based on estimating a constant scatter level in each projection was compared to approaches aiming at estimation of a more complex spatial shape of the scatter distribution. For each scheme, remaining cupping artifacts in the reconstructed volumetric image were quantified and analyzed. It was found that already accurate estimation of a constant scatter level for each projection allows for comparatively accurate compensation of scatter-caused artifacts.
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Matthias Bertram, Jens Wiegert, and Georg Rose "Potential of software-based scatter corrections in cone-beam volume CT", Proc. SPIE 5745, Medical Imaging 2005: Physics of Medical Imaging, (20 April 2005); https://doi.org/10.1117/12.595032
Lens.org Logo
CITATIONS
Cited by 26 scholarly publications and 3 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Monte Carlo methods

Sensors

Head

Computer simulations

X-rays

Signal attenuation

X-ray computed tomography

RELATED CONTENT


Back to Top