Paper
23 May 2005 A sampling wattmeter for noise power measurements
Author Affiliations +
Proceedings Volume 5844, Noise in Devices and Circuits III; (2005) https://doi.org/10.1117/12.608787
Event: SPIE Third International Symposium on Fluctuations and Noise, 2005, Austin, Texas, United States
Abstract
Low-frequency noise measurements are usually performed by measuring the voltage across a dipole, or the current through a wire. Here we demonstrate the feasibility of the measurement of the noise power flowing through a line connecting two dipoles. A sampling wattmeter, with spectral display capabilities, having a bandwidth in the range 1 Hz - 10 kHz and sensitivity better than 10-22 W/Hz is here described, and employed to measure the noise power flowing between two resistors at different temperatures. Possible applications of the device include noise thermometry, noise measurement of active devices under load condition, investigations of excess noise below thermal threshold.
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Luca Callegaro and Marco Pisani "A sampling wattmeter for noise power measurements", Proc. SPIE 5844, Noise in Devices and Circuits III, (23 May 2005); https://doi.org/10.1117/12.608787
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Amplifiers

Wattmeters

Resistors

Interference (communication)

Temperature metrology

Thermometry

Capacitance

RELATED CONTENT

Low-noise preamplifier based on PMT and its signal processing
Proceedings of SPIE (September 19 2013)
Self-optimising piezoelectric damping
Proceedings of SPIE (April 21 2008)
Properties of liquid-nitrogen-cooled electronic elements
Proceedings of SPIE (August 01 1991)

Back to Top