Paper
18 September 2007 Calibration of gated MCP responses in the x-ray region: spatial gain variation
George A. Kyrala, Scott Evans, Tom Archuleta, Joseph Cowan, John Oertel, Philip Sanchez
Author Affiliations +
Abstract
We will discuss our attempts to measure of the absolute gain and its variation across the face of fast gated multichannel plate [MCP] detectors for 4.75 keV x-rays. We found that some of the gated strips had variations in the gain along and perpendicular to the direction of travel, and significant variation along the time axis that requires these calibrations to obtain the correct time history of gated events. We will also present some of the results on the linearity of such gain with input x-ray signal amplitude.
© (2007) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
George A. Kyrala, Scott Evans, Tom Archuleta, Joseph Cowan, John Oertel, and Philip Sanchez "Calibration of gated MCP responses in the x-ray region: spatial gain variation", Proc. SPIE 6703, Ultrafast X-Ray Sources and Detectors, 67030M (18 September 2007); https://doi.org/10.1117/12.740663
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
X-rays

Cameras

Optical filters

Titanium

Calibration

Microchannel plates

Laser energy

RELATED CONTENT

Performance of CID camera x ray imagers at NIF in...
Proceedings of SPIE (September 26 2013)
Ultrafast X-Ray Diagnostics For Laser Fusion Experiments
Proceedings of SPIE (January 18 1977)
A Soft X Ray Streak Camera Using A Microchannel Plate...
Proceedings of SPIE (February 04 1988)
Energy-resolved soft x-ray framing camera
Proceedings of SPIE (August 01 2003)
Picosecond X-Ray Spectral Studies
Proceedings of SPIE (September 14 1977)

Back to Top