10 March 2009 Optimization of 4D cone-beam CT: evaluation of streaking artifacts and noise with various simulated gantry rotation speeds
Moiz Ahmad, Peter Balter, Peter Munro, Jun Lu, Tinsu Pan
Author Affiliations +
Proceedings Volume 7258, Medical Imaging 2009: Physics of Medical Imaging; 725808 (2009) https://doi.org/10.1117/12.813739
Event: SPIE Medical Imaging, 2009, Lake Buena Vista (Orlando Area), Florida, United States
Abstract
We have investigated the relationship between scan parameters and image quality in fourdimensional cone-beam computed tomography (4D-CBCT) performed with a flat panel imager in image-guided radiotherapy. We have determined upper bounds on scan time while achieving objective thresholds of image quality, namely in noise performance and minimization of view aliasing artifacts. A slow-gantry design for 4D-CBCT was used, in which we slow down clinical linear accelerator gantry speed from the typical 1.0 rpm speed to 0.1 - 0.125 rpm, to ensure the projection angle spacing between two consecutive respiratory cycles is less than 3 degrees. A respiratory monitoring device was used to record the respiratory signal for temporal correlation of the projection data for 4D-CBCT image reconstruction. Four patient data sets were acquired. Reference images were reconstructed with all projection data and were compared with images reconstructed with 50%, 33% and 20% of the projection data. These three partial data reconstructions are simulations of scans with shorter acquisition times. The main image degradations in the short scan simulation image sets are streaking artifacts and poor signal to noise ratio, both caused by sparse projection sampling. The amount of streaking artifacts and SNR in each image set is quantified. By allowing some streaking artifacts and not compromising the assessment of tumor motion, we produce images that suggest that a reduction in scan time from 3 to 6 min to approximately 2 min may be possible, making 4D-CBCT feasible in a clinical setting.
© (2009) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Moiz Ahmad, Peter Balter, Peter Munro, Jun Lu, and Tinsu Pan "Optimization of 4D cone-beam CT: evaluation of streaking artifacts and noise with various simulated gantry rotation speeds", Proc. SPIE 7258, Medical Imaging 2009: Physics of Medical Imaging, 725808 (10 March 2009); https://doi.org/10.1117/12.813739
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Signal to noise ratio

Image quality

Imaging systems

Radiotherapy

Computer simulations

Tumors

Computed tomography

Back to Top