Paper
2 March 2010 Improved safety of retinal photocoagulation with a shaped beam and modulated pulse
Christopher Sramek, Jefferson Brown, Yannis M. Paulus, Hiroyuki Nomoto, Daniel Palanker
Author Affiliations +
Proceedings Volume 7550, Ophthalmic Technologies XX; 75500V (2010) https://doi.org/10.1117/12.841079
Event: SPIE BiOS, 2010, San Francisco, California, United States
Abstract
Shorter pulse durations help confine thermal damage during retinal photocoagulation, decrease treatment time and minimize pain. However, safe therapeutic window (the ratio of threshold powers for rupture and mild coagulation) decreases with shorter exposures. A ring-shaped beam enables safer photocoagulation than conventional beams by reducing the maximum temperature in the center of the spot. Similarly, a temporal pulse modulation decreasing its power over time improves safety by maintaining constant temperature for a significant portion of the pulse. Optimization of the beam and pulse shapes was performed using a computational model. In vivo experiments were performed to verify the predicted improvement. With each of these approaches, the pulse duration can be decreased by a factor of two, from 20 ms down to 10 ms while maintaining the same therapeutic window.
© (2010) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Christopher Sramek, Jefferson Brown, Yannis M. Paulus, Hiroyuki Nomoto, and Daniel Palanker "Improved safety of retinal photocoagulation with a shaped beam and modulated pulse", Proc. SPIE 7550, Ophthalmic Technologies XX, 75500V (2 March 2010); https://doi.org/10.1117/12.841079
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication and 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Beam shaping

Laser coagulation

Modulation

Eye

Pulsed laser operation

Safety

Retina

RELATED CONTENT

Modelling of heat transfer in a laser irradiated eye retina
Proceedings of SPIE (December 30 2019)
UV collagen cross linking of the cornea safety aspects...
Proceedings of SPIE (February 25 2008)
Patterned retinal coagulation with a scanning laser
Proceedings of SPIE (March 05 2007)

Back to Top