Paper
22 February 2010 Calibration protocol for Fourier-domain OCT using optical fiber gratings
Author Affiliations +
Abstract
We present a calibration protocol to get the alignment factors of a custom-made spectrometer and the nonlinear fitting function between the measured CCD pixel domain and the wavelength domain to apply the Fourier-domain optical coherence tomography (FD-OCT) using optical fiber gratings. We have used 5 different center wavelength gratings covered the broadband source spectral range with a narrow spectral bandwidth (<0.05 nm) and the same reflectivity (>92 %) to calibrate and align the custom-made spectrometer. The implemented SD-OCT system following the proposed protocol showed the alignment factors as 44.37o incident angle, 53.11o diffraction angle, and 70.0 mm focal length. The spectral resolution of 0.187 nm was recalculated from the alignment factors.
© (2010) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Tae Joong Eom, Bong-Ahn Yu, Woojin Shin, Yeung Lak Lee, Chang-Seok Kim, Yeh-Chan Ahn, and Zhongping Chen "Calibration protocol for Fourier-domain OCT using optical fiber gratings", Proc. SPIE 7555, Advanced Biomedical and Clinical Diagnostic Systems VIII, 75551M (22 February 2010); https://doi.org/10.1117/12.842905
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Spectroscopy

Fiber Bragg gratings

Optical coherence tomography

Calibration

Charge-coupled devices

Diffraction gratings

Spectral resolution