Paper
27 July 1976 Image Intensifier Scintillation Cameras For Nuclear Medicine Applications
Gerd Muehllehner
Author Affiliations +
Proceedings Volume 0078, Low Light Level Devices for Science and Technolgy; (1976) https://doi.org/10.1117/12.954789
Event: 1976 SPIE/SPSE Technical Symposium East, 1976, Reston, United States
Abstract
Various scintillation cameras using image intensifiers are described and compared in their important imaging characteristics with existing instrumentation. The main advantages of image intensifier cameras are in the areas of positional resolution, count-rate capability and system simplicity. The main problem in the past has been the general lack of good gamma-ray energy discrimination which results in low contrast images in clinical situations. Good energy discrimination is in principle possible with several of the described systems.
© (1976) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Gerd Muehllehner "Image Intensifier Scintillation Cameras For Nuclear Medicine Applications", Proc. SPIE 0078, Low Light Level Devices for Science and Technolgy, (27 July 1976); https://doi.org/10.1117/12.954789