|
1.IntroductionConventional light microscopy has been a vital tool in the life sciences by giving noninvasive visual access to micro-objects such as cells and microorganisms. Although significant improvements have been achieved to improve spatial resolution and contrast,1–12 one important restriction of light microscopy for many applications has been the limited field-of-view (FOV) and depth-of-focus (DOF). That is, for applications where large areas and/or volumes need to be screened, one typically needs to mechanically scan the sample under the microscope, demanding long scan times and relatively complex hardware. Lens-free on-chip holographic microscopy offers an alternative platform that overcomes this limitation.13–25 Unlike conventional light microscopy, lens-free on-chip microscopy does not rely on optical magnification. Instead, it uses the emerging large-format sensor architectures with small pixel sizes together with novel imaging algorithms to compensate for the lack of optical components such as lenses or objectives. Therefore, it enables imaging large areas (e.g., to ) at submicrometer spatial resolution in a compact and cost-effective set-up.18,19,21–25 In addition to the limited FOV, the short DOF associated with objective lenses poses further challenges, especially if the sample has a three-dimensional (3-D) structure, such as a tissue culture in a Petri dish where objects might be distributed at different depths. For such 3-D samples, in addition to lateral scanning, re-focusing by depth-scanning also becomes necessary. While high-end microscopes can perform both lateral and depth scanning automatically, these systems are cost-prohibitive and bulky. In contrast, lens-free on-chip holographic imaging provides particular advantages by offering the ability to perform postexposure digital auto-focusing (or depth-scanning) using holographic reconstruction over a depth-of-field of up to to 5 mm.21,22 Owing to the partially coherent holographic image acquisition scheme (unlike e.g., contact imaging techniques) the detector records the information regarding the entire 3-D structure of the sample, and lens-free images can be digitally reconstructed at any depth of interest over an extended depth-of-field.21,22 One important biological system that can be observed at the scale and resolution of lens-free imaging is a cultured microvascular capillary network. The study of capillary network formation, or capillary morphogenesis and angiogenesis, is vital to the fields of wound healing and tissue engineering of prevascularized implants,26 tumor mediated vascularization,27 and various diseases.28 Often times the growth and morphology of a complex capillary network must be monitored and quantified over many days in order to determine the effects of experimental conditions such as the addition of a putative drug.29 Capillary tubule formation is commonly quantified by the average tubule length, number of tubules, tubule area, and number of branch points.30 Since such measurements require micron-scale resolution, samples are traditionally imaged by conventional light microscopy, which due to its small FOV and shallow DOF precludes rapid quantification of the entire culture volume. As a compromise, results are averaged across a discrete number of regions using a time-intensive process.31–37 Numerous image analysis algorithms including packages for NIH ImageJ, Analyze Skeleton and AngioQuant38,39 have been developed for accurate hands-free quantification of capillary growth,31,33,40 leaving image acquisition as the primary rate-limiting step. While commercial high content imaging systems (e.g., BD Pathway 855) can scan the entire sample in 3-D, these systems are relatively expensive and slow because they rely on mechanical scanning. In contrast to existing approaches, our lens-free on-chip technique enables imaging of the entire sample with sufficient resolution to draw statistical conclusions regarding changing morphology using existing image processing techniques. To demonstrate its proof of concept, here we present the use of lens-free on-chip microscopy to image capillaries grown in a 3-D in vitro system, without the use of lenses or mechanical scanning, and with sufficient spatial resolution to accurately determine vessel number, vessel length and vessel area. 2.Materials and Methods2.1.Capillary Morphogenesis ModelA model for capillary morphogenesis as first described by Nehls and Drenckhahn was implemented in Ref. 41. Primary human umbilical vein endothelial cells (HUVECs) were grown on microcarrier beads imbedded in a crosslinked fibrin hydrogel and given pro-angiogenic signaling factors naturally released by fibroblast cells cultured on top of the gel. HUVECs were first expanded in cell culture flasks to passage number three and then cultured on the surface of 150 to 210 μm diameter carrier beads (Cytodex). Beads were immersed in a solution of fibrinogen (bovine, Sigma) in EBM-2 media (Lonza) at approximately solution. The solution was then pipetted into a 35 mm diameter Petri dish with type 1 glass bottom (MatTek) containing 20 μl of thrombin (Sigma). A clotting cascade forms a fibrin gel thus imbedding the carrier beads within a 3-D extracellular matrix. Once the gel is formed, normal human lung fibroblast cells (NHLFs) grown out to passage five were seeded onto a Transwell container (Corning) that is placed on top of the gel allowing signaling factors released by the fibroblasts to diffuse down into the gel while keeping the fibroblasts mechanically separated from the gel (Fig. 1). Next, 2 ml of EGM-2 media (Lonza) was added to each sample with media exchanged every other day. Capillary sprouting from the carrier beads begins by day two with large interconnected capillary networks forming by day seven. Imaging was performed on day seven. 2.2.Lens-free On-Chip Imaging SetupThe lens-free on-chip imaging setup, shown in Fig. 1, is composed of a partially coherent light source and an opto-electronic sensor array (complementary metal-oxide-semiconductor (CMOS) Model #MT9P031, Micron Technology; pixel size: 2.2 μm, 5 mega pixels). The sample is placed directly on the top of the sensor chip such that the distance, , between the objects and the active area of the detector is to 2 mm. The sample is illuminated using a near infrared light emitting diode (LED, ) that is butt-coupled to a multimode optical fiber that has a core-diameter of 100 μm. The distance, , between the light source and the sample is typically , and its placement does not require sensitive alignment. An infrared LED was selected in our imaging experiments since it was found to reduce the background noise from multiple scattering within the fibrin gel, and this also minimizes the phase-wrapping problems in the reconstructed lens-free phase images as phase delay is inversely proportional to the wavelength. Since the light impinging on the sample is partially coherent both temporally and spatially,13,14 the unperturbed portion of the illumination interferes with the waves scattered by the objects. The sensor records this interference pattern, i.e., an in-line hologram of the objects placed on the chip. Owing to its unique geometrical configuration where , in-line holograms are recorded with unit magnification over a large imaging FOV that equals the active area of the sensor-array, in this case (Fig. 2), which can further increase to e.g., with a different choice of sensor-array.16 2.3.Digital Holographic ReconstructionLens-free images are reconstructed using digital beam propagation techniques based on the angular spectrum approach.42 In this approach, lens-free images are convolved with the impulse response of free space propagation. This operation is done in the Fourier domain, by multiplying the 2D Fourier transform of lens-free images with the transfer function of free space propagation, which can be expressed as:42 In Eq. (1), and denote spatial frequencies of the input field along and , respectively, is the illumination wavelength in free space, and is the distance to which the original field is propagated. After multiplication by this transfer function, an inverse Fourier transform provides the output at the desired plane.When the above-mentioned digital beam-propagation is performed on the measured lens-free holograms, the reconstructed images exhibit artifacts due to the “twin-image” noise. Stated differently, since imaging sensors are only sensitive to the intensity of an optical field, the initial phase of the complex field at the hologram plane, which is unknown, is assumed to be zero before digital propagation, giving rise to artifacts in the reconstructed image. In order to obtain a refined reconstructed image that does not suffer from this twin-image noise, phase recovery algorithms should be used to retrieve the unknown phase at the hologram plane. One way of achieving this is to use a size-constrained iterative phase recovery algorithm,14 where the object size is used as additional information (i.e., size-constraint) in the reconstruction process. The object size can generally be estimated from the initial reconstruction despite the contamination due to the twin-image noise, as the boundaries of the objects can still be determined. For lens-free imaging of capillaries, however, the complex morphology of the sample makes it challenging to estimate the capillary shape and size, hampering the use of size-constrained phase recovery algorithms. As a result, here we utilized an alternative phase-recovery algorithm, multi-height phase recovery (MHPR),24,25,43 which does “not” require the knowledge of the object size or shape as additional information, but instead uses multiple intensity measurements. 2.4.Multi-Height Phase RecoveryIn this iterative phase-retrieval technique, multiple in-line holograms (i.e., intensity measurements) are recorded for the same object, where each measurement is performed at a different distance. For lens-free imaging of capillaries, two intensity measurements were sufficient to effectively retrieve the phase of the optical field at the hologram plane. The change in distance is achieved by placing or replacing a coverslip with a thickness of, for example, as a spacer between the sample and the sensor. This operation may also cause a slight translation and rotation of the sample with respect to the sensor array between two successive lens-free image acquisitions. Therefore, the two measured lens-free holograms are first registered to each other in order to compensate for any translational or rotational motion. The exact distances do not need to be known a priori, since an auto-focus algorithm17 as described below is utilized to estimate this parameter. After image registration, the MHPR algorithm is invoked to retrieve the missing phase of the sample holograms. The algorithm works by propagating the measured fields back and forth between the two planes of measurement. At each iteration, the algorithm enforces the recorded (i.e., measured) amplitude at the corresponding height (i.e., ), while keeping the updated phase for the next iteration. This way, the missing phase is retrieved in iterations without modifying the measured amplitude values, resulting in a refined lens-free image where the twin-image noise is significantly suppressed. 2.5.Digital Autofocusing AlgorithmAn important attribute of lens-free holography is that it enables imaging of samples over a long depth-of-field, as long as the holograms of objects at large distances above the sensor (e.g., 1 to 5 mm) have sufficient signal-to-noise ratio.21,22 This ability is particularly useful for imaging capillary morphogenesis due to the inherent 3-D nature of the sample. Imaging an extended depth-of-field is achieved by reconstructing the holograms at different depths-of-interest, which is equivalent to focusing a conventional microscope objective lens at different depths. However, digitally selecting the object-to-detector distance (), which maximizes the contrast and the signal-to-noise ratio of the object, can be a tedious task when imaging objects over a large volume. Towards this end, we implemented an autofocus algorithm to automatically determine the best plane of focus for imaging different capillaries across the sample volume.17 This autofocus algorithm is based on the fact that the edges of the object should be the sharpest at the plane of best focus. To estimate the sharpness of the edges, first a Sobel operator is used to detect edges in the vertical and horizontal directions. Second, the edge images are combined by using the two-norm (i.e., the square root of the sum of the squares). Third, the variance of the resultant image is calculated, where a high variance indicates existence of sharp edges. Therefore our autofocus algorithm scans several distances spaced by e.g., 1 μm distances in order to find the distance with the maximum edge variance, i.e., the best focus. It should be noted that this auto-focus algorithm is invoked twice in order to estimate the distances for both measurement planes. For a typical sample, the difference in the distances of the two lens-free holograms does not exceed 50 μm. Consequently, once the distance for one of the lens-free holograms is determined, the search space for the distance of the second hologram is rather small, achieving fast convergence. 3.Results and DiscussionTo obtain microscopic 3-D images of the sample, the recorded lens-free in-line holograms are digitally reconstructed14 where the recorded amplitude is initially multiplied by a reference plane wave and the resulting optical field is back-propagated42 to the sample/object plane. Due to the in-line holographic recording scheme, the reconstructed images exhibit twin image noise [e.g., Fig. 3(a2) and 3(b2)], which degrades the lens-free image quality by concealing an object’s fine features and giving rise to artificial intensity modulations that may lead to a biological misinterpretation. The arrows in Fig. 3 point to exemplary regions where the twin image hampers accurate interpretation of the reconstructed images, either by concealing the actual structure, or sometimes creating artifacts that misleadingly appear like capillaries. To digitally eliminate this twin image noise, as detailed in Sec. 2, a MHPR algorithm is invoked to enable accurate interpretation of our lens-free reconstructed images. A visual comparison of Fig. 3(a3) and 3(b3) to 3(a2) and 3(b2), particularly in the regions indicated by the arrows, clearly demonstrates the significance of our MHPR approach to lens-free 3-D imaging of capillaries. Figure 3(a4) and 3(b4) further demonstrate this improvement by showing profiles along the white lines in 3(a2), 3(b2), 3(a3), and 3(b3). These plots illustrate that using multi-height phase retrieval the modulations due to capillaries get stronger, increasing the image contrast, while false modulations due to the twin-image artifacts are now significantly suppressed. It should be noted that pixel super-resolution techniques that digitally increase the spatial resolution of lens-free on-chip imaging were not implemented here.18,19,21–23 As a result, the lens-free imaging platform utilized here provides a modest spatial resolution of to 2.0 μm (still subpixel considering that the CMOS chip has a pixel size of 2.2 μm),14 which was more than sufficient for the quantitative analysis of vessel length in our application. We compared the average capillary length of a 3-D sample measured by both the MHPR-based lens-free microscopy [Fig. 4(a)] and an inverted microscope with a objective lens (Nikon, ) under bright field illumination [Fig. 4(b)]. In these measurements, capillary length was estimated as the linear distance from the center of the bead to the tip of each capillary. Based on these measurements, vessel length differences between the two methods were found to be less than 2% indicating that lens-free imaging can faithfully replace traditional microscopy in our model system. Importantly, the lens-free system has the advantage of its extraordinarily large FOV and DOF. The microscopic image shown in Fig. 4(b) acquired with a objective lens has a FOV of allowing only one carrier bead to be imaged at a time, while the full FOV of the lens-free image (see Fig. 5) is (i.e., fold larger than the FOV of the microscope objective) allowing all the carrier beads to be imaged simultaneously without the need for mechanical scanning. Note that this FOV of the lens-free imaging technique is only limited by the active area of the detector chip, and sample areas larger than e.g., can also be imaged by using wide-field CCD sensor-arrays.16,44 In summary, we have demonstrated that imaging of capillary morphogenesis is possible without the use of any lenses or mechanical scanning over a large FOV in 3-D culture. The resolution, image contrast and signal-to-noise-ratio achieved in the reconstructed holographic images were sufficient to extract useful structural information such as vessel length, diameter, area, number of vessels, vessel shape, vessel interactions and the volume fraction of tissue that had been vascularized. The large FOV, depth-of-field, cost-effectiveness, and ease-of-use of the lens-free on-chip imaging platform make it attractive to researchers studying capillary growth under the influence of, for example, chemical or mechanical cues. A high throughput system for measuring growth over large volumes in multiple samples may be possible with the use of lens-free imaging, thus enabling high content screening of 3-D tissue cultures. AcknowledgmentsWe wish to acknowledge Dr. Ekaterina Kniazeva for assisting with the capillary culture, Dr. Samir Shreim for assisting with the optical setup and Prof. Christopher Hughes for the supply of HUVECs. E. Botvinick acknowledges the Laser Microbeam and Medical Program (LAMMP), an NIH/NIBIB Biomedical Technology Center (P41EB05890) and the NSF Physical and Engineering Sciences in Engineering program (CMMI-1233697). A. Ozcan also acknowledges the support of the Presidential Early Career Award for Scientists and Engineers (PECASE), ARO Young Investigator Award, NSF CAREER Award, the ONR Young Investigator Award 2009 and the NIH Director’s New Innovator Award DP2OD006427 from the Office of The Director, NIH. ReferencesS. A. Joneset al.,
“Fast, three-dimensional super-resolution imaging of live cells,”
Nat. Methods, 8
(6), 499
–508
(2011). http://dx.doi.org/10.1038/nmeth.1605 1548-7091 Google Scholar
S. W. Hell,
“Toward fluorescence nanoscopy,”
Nat. Biotechnol., 21
(11), 1347
–1355
(2003). http://dx.doi.org/10.1038/nbt895 NABIF9 1087-0156 Google Scholar
M. G. L. Gustafsson,
“Nonlinear structured-illumination microscopy: wide-field fluorescence imaging with theoretically unlimited resolution,”
Proc. Natl. Acad. Sci. U. S. A., 102
(37), 13081
–13086
(2005). http://dx.doi.org/10.1073/pnas.0406877102 PNASA6 0027-8424 Google Scholar
E. Betziget al.,
“Imaging intracellular fluorescent proteins at nanometer resolution,”
Science, 313
(5793), 1642
–1645
(2006). http://dx.doi.org/10.1126/science.1127344 SCIEAS 0036-8075 Google Scholar
M. J. RustM. BatesX. Zhuang,
“Sub-diffraction-limit imaging by stochastic optical reconstruction microscopy (STORM),”
Nat. Methods, 3
(10), 793
–796
(2006). http://dx.doi.org/10.1038/nmeth929 1548-7091 Google Scholar
S. T. HessT. P. GirirajanM. D. Mason,
“Ultra-high resolution imaging by fluorescence photoactivation localization microscopy,”
Biophys. J., 91
(11), 4258
–4272
(2006). http://dx.doi.org/10.1529/biophysj.106.091116 BIOJAU 0006-3495 Google Scholar
Z. Maet al.,
“Fluorescence near-field microscopy of DNA at sub-10 nm resolution,”
Phys. Rev. Lett., 97
(26), 260801
(2006). http://dx.doi.org/10.1103/PhysRevLett.97.260801 PRLTAO 0031-9007 Google Scholar
E. Chunget al.,
“Two-dimensional standing wave total internal reflection fluorescence microscopy: superresolution imaging of single molecular and biological specimens,”
Biophys. J., 93
(5), 1747
–1757
(2007). http://dx.doi.org/10.1529/biophysj.106.097907 BIOJAU 0006-3495 Google Scholar
W. R. ZipfelR. M. WilliamsW. W. Webb,
“Nonlinear magic: multiphoton microscopy in the biosciences,”
Nat. Biotechnol., 21
(11), 1369
–1377
(2003). http://dx.doi.org/10.1038/nbt899 NABIF9 1087-0156 Google Scholar
W. Choiet al.,
“Tomographic phase microscopy,”
Nat. Methods, 4
(9), 717
–719
(2007). http://dx.doi.org/10.1038/nmeth1078 1548-7091 Google Scholar
R. P. J. BarrettoB. MesserschmidtM. J. Schnitzer,
“In vivo fluorescence imaging with high-resolution microlenses,”
Nat. Methods, 6
(7), 511
–512
(2009). http://dx.doi.org/10.1038/nmeth.1339 1548-7091 Google Scholar
J. RosenG. Brooker,
“Non-scanning motionless fluorescence three-dimensional holographic microscopy,”
Nat. Photon., 2
(3), 190
–195
(2008). http://dx.doi.org/10.1038/nphoton.2007.300 1749-4885 Google Scholar
C. Ohet al.,
“On-chip differential interference contrast microscopy using lensless digital holography,”
Opt. Express, 18
(5), 4717
–4726
(2010). http://dx.doi.org/10.1364/OE.18.004717 OPEXFF 1094-4087 Google Scholar
O. Mudanyaliet al.,
“Compact, light-weight and cost-effective microscope based on lensless incoherent holography for telemedicine applications,”
Lab Chip, 10
(11), 1417
–1428
(2010). http://dx.doi.org/10.1039/c000453g LCAHAM 1473-0197 Google Scholar
D. Tsenget al.,
“Lensfree microscopy on a cell-phone,”
Lab Chip, 10
(14), 1787
–1792
(2010). http://dx.doi.org/10.1039/c003477k LCAHAM 1473-0197 Google Scholar
S. Seoet al.,
“High-throughput lensfree blood analysis on a chip,”
Anal. Chem., 82
(11), 4621
–4627
(2010). http://dx.doi.org/10.1021/ac1007915 ANCHAM 0003-2700 Google Scholar
O. Mudanyaliet al.,
“Detection of waterborne parasites using field-portable and cost-effective lensfree microscopy,”
Lab Chip, 10
(18), 2419
–2423
(2010). http://dx.doi.org/10.1039/c004829a LCAHAM 1473-0197 Google Scholar
W. Bisharaet al.,
“Lensfree on-chip microscopy over a wide field-of-view using pixel super-resolution,”
Opt. Express, 18
(11), 11181
–11191
(2010). http://dx.doi.org/10.1364/OE.18.011181 OPEXFF 1094-4087 Google Scholar
W. Bisharaet al.,
“Holographic pixel super-resolution in portable lensless on-chip microscopy using a fiber-optic array,”
Lab Chip, 11
(7), 1276
–1279
(2011). http://dx.doi.org/10.1039/c0lc00684j LCAHAM 1473-0197 Google Scholar
T. Suet al.,
“A compact and light-weight automated semen analysis platform using lensfree on-chip microscopy,”
Anal. Chem., 82
(19), 8307
–8312
(2010). http://dx.doi.org/10.1021/ac101845q ANCHAM 0003-2700 Google Scholar
S. O. Isikmanet al.,
“Field-portable lensfree tomographic microscope,”
Lab Chip, 11
(13), 2222
–2230
(2011). http://dx.doi.org/10.1039/c1lc20127a LCAHAM 1473-0197 Google Scholar
S. O. Isikmanet al.,
“Lensfree optical tomographic microscope with a large imaging volume on a chip,”
Proc. Nat. Acad. Sci. U. S. A., 108
(18), 7296
–7301
(2011). http://dx.doi.org/10.1073/pnas.1015638108 PNASA6 0027-8424 Google Scholar
S. O. Isikmanet al.,
“Optofluidic tomography on a chip,”
Appl. Phys. Lett., 98
(16), 161109
(2011). http://dx.doi.org/10.1063/1.3548564 APPLAB 0003-6951 Google Scholar
A. GreenbaumA. Ozcan,
“Maskless imaging of dense samples using pixel super-resolution based multi-height lensfree on-chip microscopy,”
Opt. Express, 20
(3), 3129
–3143
(2012). http://dx.doi.org/10.1364/OE.20.003129 OPEXFF 1094-4087 Google Scholar
A. GreenbaumU. SikoraA. Ozcan,
“Field-portable wide-field microscopy of dense samples using multi-height pixel super-resolution based lensfree imaging,”
Lab Chip, 12
(7), 1242
–1245
(2012). http://dx.doi.org/10.1039/c2lc21072j LCAHAM 1473-0197 Google Scholar
P. S. Sahotaet al.,
“Approaches to improve angiogenesis in tissue-engineered skin,”
Wound Repair Regen., 12
(6), 635
–642
(2004). http://dx.doi.org/10.1111/wrr.2004.12.issue-6 WREREU 1067-1927 Google Scholar
J. Folkman,
“Tumor angiogenesis: therapeutic implications,”
N. Engl. J. Med., 285
(7), 1182
–1186
(1971). http://dx.doi.org/10.1056/NEJM197108122850711 NEJMAG 0028-4793 Google Scholar
P. Carmeliet,
“Angiogenesis in health and disease,”
Nat. Med., 9
(6), 653
–660
(2003). http://dx.doi.org/10.1038/nm0603-653 1078-8956 Google Scholar
L. Evensenet al.,
“A novel imaging-based high-throughput screening approach to anti-angiogenic drug discovery,”
Cytom. Part A, 77
(1), 41
–51
(2010). http://dx.doi.org/10.1002/cyto.a.20808 1552-4922 Google Scholar
C. A. StatonM. W. R. ReedN. J. Brown,
“A critical analysis of current in vitro and in vivo angiogenesis assays,”
Int. J. Exp. Pathol., 90
(3), 195
–221
(2009). http://dx.doi.org/10.1111/j.1365-2613.2008.00633.x IJEPEI 1365-2613 Google Scholar
E. KniazevaA. J. Putnam,
“Endothelial cell traction and ECM density influence both capillary morphogenesis and maintenance in 3D,”
Am. J. Physiol. Cell Physiol., 297
(1), 179
–187
(2009). http://dx.doi.org/10.1152/ajpcell.00018.2009 0363-6143 Google Scholar
T. KorffH. G. Augustin,
“Tensional forces in fibrillar extracellular matrices control directional capillary sprouting,”
J. Cell Sci., 112
(19), 3249
–3258
(1999). JNCSAI 0021-9533 Google Scholar
M. N. Nakatsuet al.,
“Angiogenic sprouting and capillary lumen formation modeled by human umbilical vein endothelial cells (HUVEC) in fibrin gels: the role of fibroblasts and angiopoietin-1,”
Microvasc. Res., 66
(2), 102
–112
(2003). http://dx.doi.org/10.1016/S0026-2862(03)00045-1 MIVRA6 0026-2862 Google Scholar
V. NehlsD. Drenckhahn,
“A novel, microcarrier-based in vitro assay for rapid and reliable quantification of three-dimensional cell migration and angiogenesis,”
Microvasc. Res., 50
(3), 311
–322
(1995). http://dx.doi.org/10.1006/mvre.1995.1061 MIVRA6 0026-2862 Google Scholar
A. Collenet al.,
“Influence of fibrin structure on the formation and maintenance of capillary-like tubules by human microvascular endothelial cells,”
Angiogenesis, 2
(2), 153
–165
(1998). http://dx.doi.org/10.1023/A:1009240522808 0969-6970 Google Scholar
T. I. Koblizeket al.,
“Angiopoietin-1 induces sprouting angiogenesis in vitro,”
Curr. Biol., 8
(9), 529
–532
(1998). http://dx.doi.org/10.1016/S0960-9822(98)70205-2 CUBLE2 0960-9822 Google Scholar
A. Wengeret al.,
“Modulation of in vitro angiogenesis in a three-dimensional spheroidal coculture model for bone tissue engineering,”
Tissue Eng., 10
(9–10), 1536
–1547
(2004). http://dx.doi.org/10.1089/ten.2004.10.1536 1937-3341 Google Scholar
I. Arganda-Carreraset al.,
“3D reconstruction of histological sections: application to mammary gland tissue,”
Microsc. Res. Techniq., 73
(11), 1019
–1029
(2010). http://dx.doi.org/10.1002/jemt.v73:11 MRTEEO 1059-910X Google Scholar
A. Niemistoet al.,
“Robust quantification of in vitro angiogenesis through image analysis,”
IEEE Trans. Med. Imaging, 24
(4), 549
–553
(2005). http://dx.doi.org/10.1109/TMI.2004.837339 ITMID4 0278-0062 Google Scholar
J. Plendlet al.,
“Primitive endothelial cell lines from the porcine embryonic yolk sac,”
In Vitro Cell Dev. Biol. Animal, 38
(6), 334
–342
(2002). http://dx.doi.org/10.1290/1071-2690(2002)038<0334:PECLFT>2.0.CO;2 1071-2690 Google Scholar
V. NehlsD. Drenckhahn,
“A microcarrier-based cocultivation system for the investigation of factors and cells involved in angiogenesis in three-dimensional fibrin matrices in vitro,”
Histochem. Cell Biol., 104
(6), 459
–466
(1995). Google Scholar
J. W. Goodman, Introduction to Fourier Optics, 3rd ed.Roberts & Company Publishers, Greenwood Village, Colo
(2005). Google Scholar
L. J. AllenM. P. Oxley,
“Phase retrieval from series of images obtained by defocus variation,”
Opt. Commun., 199
(1–4), 65
–75
(2001). http://dx.doi.org/10.1016/S0030-4018(01)01556-5 OPCOB8 0030-4018 Google Scholar
A. F. Coskunet al.,
“Lensfree fluorescent on-chip imaging of transgenic caenorhabditis elegans over an ultra-wide field-of-view,”
PLoS One, 6
(1), e15955
(2011). http://dx.doi.org/10.1371/journal.pone.0015955 1932-6203 Google Scholar
|