7 August 2014 Design and characterization of a handheld multimodal imaging device for the assessment of oral epithelial lesions
Laura M. Higgins, Mark C. Pierce
Author Affiliations +
Abstract
A compact handpiece combining high resolution fluorescence (HRF) imaging with optical coherence tomography (OCT) was developed to provide real-time assessment of oral lesions. This multimodal imaging device simultaneously captures coregistered en face images with subcellular detail alongside cross-sectional images of tissue microstructure. The HRF imaging acquires a 712×594  μm2 field-of-view at the sample with a spatial resolution of 3.5  μm. The OCT images were acquired to a depth of 1.5 mm with axial and lateral resolutions of 9.3 and 8.0  μm, respectively. HRF and OCT images are simultaneously displayed at 25 fps. The handheld device was used to image a healthy volunteer, demonstrating the potential for in vivo assessment of the epithelial surface for dysplastic and neoplastic changes at the cellular level, while simultaneously evaluating submucosal involvement. We anticipate potential applications in real-time assessment of oral lesions for improved surveillance and surgical guidance.

1.

Introduction

Oral cancer is currently one of the top 10 most prevalent cancers worldwide and it is estimated that the incidence will reach 15 million new cases annually by the year 2020.1,2 Although treatment methods have improved in recent years, 5-year survival rates have remained fairly constant, at only 62% in the US, largely due to high rates of late stage diagnoses3 and local relapses.4,5 Current screening methods involve visual inspection and manual palpation by a trained clinician. However, the qualitative nature of these methods leads to poor diagnostic accuracy.6 Biopsy and histology of tissue from regions identified during screening is the gold standard for diagnosis, but this procedure is time consuming and painful for the patient.6 High rates of multiple primary lesions make adequate screening crucial so that all potential lesions are identified while also minimizing the number of unnecessary biopsies.6,7 With local recurrence estimated to account for up to 20% of cases, the ability to obtain negative surgical margins while minimizing the removal of surrounding normal tissue is also critical.4 Frozen section histopathology allows for intraoperative guidance on the margin status, but increases the cost and the duration of procedures and provides diagnostic information at only a few discrete locations. The long-term goal of this work is to develop an instrument for noninvasive, real-time assessment of the oral mucosa and to provide a more targeted approach for biopsy collection as well as real-time margin assessment during surgery, potentially decreasing rates of both missed diagnoses and local relapses.

Several optical imaging modalities, such as confocal microscopy,812 high resolution microendoscopy (HRME),1315 and optical coherence tomography (OCT),1622 have been proposed as methods for noninvasive “optical biopsy” to improve the accuracy of oral cancer screening. Reflectance and fluorescence confocal microscopy have demonstrated the ability to provide subcellular resolution of optically sectioned images within the epithelial layer due to native tissue contrast or when used with topically applied or intravenous contrast agents.812 However, these systems are expensive and typically involve custom-designed high numerical aperture optics, which are challenging to miniaturize. HRME is a simple, low-cost approach which can also provide subcellular resolution en face images of epithelial tissue, albeit without intrinsic optical sectioning. The distal end of a coherent fiber-optic bundle is placed in contact with the tissue in order to capture fluorescent emissions from an exogenous contrast agent.14 HRME then images the proximal face of the bundle onto a charge-coupled device (CCD) camera as in epi-fluorescence microscopy. When used with proflavine, a topically applied contrast agent that preferentially stains cell nuclei, morphologic parameters including the nuclear-to-cytoplasm ratio can be quantified and used to determine the disease state of oral epithelial tissues with good sensitivity and specificity.14,15 The limitations of HRME include a spatial resolution, which is determined by the size of the individual cores within the fiber bundle. Additionally, HRME can only image the first few cell layers within the epithelium and is, therefore, unable to assess abnormal lesions for any potential submucosal involvement.23,24

OCT, an optical imaging modality analogous to ultrasound, uses the interference of backscattered near-infrared (NIR) light to image tissue microstructures in cross section at depths of up to 1 to 2 mm.25,26 Contrast in OCT arises from the native optical properties of the tissue, providing the ability to distinguish tissue layers and potential disease states. By measuring factors such as the epithelial thickness1621 and vasculature,22 OCT has demonstrated potential as a useful aid in the diagnosis of oral cancer. However, while OCT can provide assessment of the entire epithelium and superficial stroma, spatial resolution at the 10 to 20-μm scale means that individual cells cannot be resolved.1622

Although both HRME and OCT have been individually used to distinguish dysplastic from healthy oral mucosa,1322 the combination of these complementary imaging modalities may overcome the practical limitations of each system alone. The use of these complementary imaging modalities may aid the clinician in the evaluation of oral lesions during both diagnosis and resection. Potential lesions initially identified through visual examination can be evaluated by HRF imaging to determine the degree of dysplasia. If classified as abnormal by HRF, submucosal involvement can then be assessed using OCT. A method combining OCT with HRME, a simple high resolution en face imaging modality, was first described by Wall and Barton27 for the diagnosis of colon cancer. This system uses multiple optical fibers evenly spaced around a central fiber bundle in order to capture sequential OCT and HRME images. This design requires the use of complex distal end optics and limits HRME resolution to the size of the fiber bundle cores.27 The use of similar systems, combining OCT with complementary high-resolution imaging modalities such as confocal, two-photon, fluorescence lifetime, and photoacoustic imaging, has recently shown potential for minimally enhanced invasive assessment of epithelial and endothelial lesions.2837

In this paper, we report the optical design and the experimental performance of a compact, handheld instrument for multimodal optical imaging which combines cellular resolution en face imaging at the epithelial surface with cross-sectional imaging of the tissue microstructure to the submucosal layer. The high resolution fluorescence (HRF) imaging component described in this paper builds on the previously described HRME system.13 Here, we image the tissue in free-space rather than through a fiber bundle, which (i) allows coregistered, simultaneous imaging with OCT, and (ii) eliminates the fiber bundle pixelation effect, increasing spatial resolution without sacrificing field-of-view. The design uses only six off-the-shelf lenses in a compact and cost-effective approach that made it ideal for clinical imaging.

2.

Methods

2.1.

Optical Design

A schematic of the multimodal imaging system can be seen in Fig. 1(a). This handpiece is located within the sample arm of a spectral domain OCT (SD-OCT) system. The SD-OCT component was based on a system previously described by Yun et al.38 Briefly, the OCT system uses a 100-nm bandwidth (FWHM) superluminescent diode source centered at 1325 nm (Thorlabs, Newton, New Jersey). Light from the source is split into the sample and the reference arms by a fiber-optic beamsplitter, so that 90% of the light travels to the sample arm, as described in detail below, illuminating the sample with 1.8 mW. The remaining 10% of the light is directed to a stationary reference arm. Recombined light is directed to a custom built spectrometer where it is dispersed by a 1200-l/mm transmission diffraction grating (Wasatch, Logan, Utah) and focused onto a 1024 element InGaAs line scan camera (SUI Goodrich, Princeton, New Jersey). The parameters of the OCT light source and the spectrometer provide a theoretical axial resolution and an imaging depth of 8.0μm and 3.0 mm, respectively. This system acquires data at a rate of 25,500A-lines/s, which corresponds to 25.3 fps with a sample overlap of 6.0μm.

Fig. 1

Combination optical coherence tomography high resolution fluorescence (OCT/HRF) system. (a) Schematic of HRF imaging integrated within the sample arm of a spectral domain OCT (SD-OCT) system. SLD: superluminescent diode source, C: fiber-optic circulator, PC: polarization controller, L: lens, RM: reference mirror, DG: diffraction grating, LSC: line scan camera, LED: LED excitation source, ex: excitation filter, D: dichroic, em: emission filter, CCD: charge-coupled device camera. (b) Rendering of the OCT/HRF handpiece, drawn to scale. Scale bar is 25 mm.

JBO_19_8_086004_f001.png

Using ZEMAX optical system design software, the handheld OCT sample arm was designed for clinical imaging specifically within the oral cavity. The HRF system was incorporated within the handheld OCT sample arm, providing inherently coregistered and simultaneous imaging. The HRF component of the system is designed for use with proflavine, a fluorescent dye with peak excitation and emission at 445 and 515 nm, respectively.13 Excitation wavelengths are isolated from a 455 nm LED (Thorlabs) using an excitation filter with a 45-nm bandwidth centered at 445 nm (Semrock, Rochester, New York). This light is collected using a 10-mm focal length aspheric condenser lens (L2, Thorlabs). A long-pass dichroic mirror (D2) with a 495-nm cut-on (Chroma, Bellows Falls, Vermont) reflects the excitation beam toward the sample. This beam passes through a 950-nm cut-off short-pass dichroic (D1, Semrock), where it is combined with the collimated OCT beam. This short-pass dichroic, mounted on a galvanometer, scans the reflected NIR OCT beam while transmitting the unaffected visible HRF excitation and emission light. The coregistered OCT and HRF beams then travel to the sample along a common path through a pair of 12.7-mm diameter, 30-mm focal length achromatic relay lenses (L3 and L4, Thorlabs). The final lens in the sequence is a 5-mm diameter, 10-mm focal length achromatic doublet (L5, Thorlabs). The lens sequence L3–L5 is designed to simultaneously focus the OCT beam onto the sample, while providing uniform (Köhler) illumination of the region with HRF excitation light. A glass window is fixed at the sample plane to ensure the tissue is flat and to increase imaging stability. The backscattered OCT signal and fluorescent emissions from the proflavine travel back along their original paths and are separated by the short-pass dichroic mirror (D1), which de-scans the OCT beam. Proflavine emissions are isolated from the HRF excitation light by the long-pass dichroic (D2) and a separate emission filter with an 88-nm bandwidth centered at 550 nm (Semrock). The fluorescent emission is focused onto a compact (29×29×30mm3), 1288×728 pixel CCD camera (Point Grey, BFLY-PGE-09S2M-CS, Richmond, British Columbia, Canada) using a 50-mm focal length achromatic doublet (L6, Thorlabs). The CCD camera’s exposure time was set to match the OCT frame rate for simultaneous image acquisition in order to minimize HRF image flicker due to the scanning OCT dichroic mirror (D1).

A rendering of the OCT-HRF handpiece, drawn to scale, can be seen in Fig. 1(b). This model shows the HRF component of the system integrated within the sample arm of the OCT system. The pair of relay lenses previously mentioned (L3 and L4), acts to extend the imaging arm in order to improve ease of clinical imaging. The segment containing L3–L5 was designed to be positioned within the oral cavity. It is 12.6-cm long and tapers to a 5-mm tip for access to confined tissue sites. The maximum dimensions of the handpiece are 29-cm long ×7-cm wide ×6-cm high and the instrument weighs 320 g.

2.2.

Performance Characterization

Standard targets were imaged in order to evaluate the imaging performance of both OCT and HRF imaging with the handpiece. The axial resolution of the OCT system was determined by measuring the intensity profile for a silver mirror scanned through the focus. The OCT lateral resolution was measured by scanning across the reflective elements of a high resolution USAF target (Edmund Optics, Barrington, New Jersey). The HRF imaging parameters were evaluated at the center (on-axis) and at the edge of the field (full-field). For these measurements, “full-field” was defined as the point on the x-axis at the edge of the field-of-view (356 μm from the center of the object). A high resolution USAF target (Edmund Optics) was imaged to determine the experimental resolution of the HRF imaging system across the entire field-of-view. Distortion was evaluated using a 100-μm-square grid target (Thorlabs). A target with Ronchi rulings at varying spatial frequencies (Thorlabs) was used to measure the experimental modulation transfer function (MTF) of the HRF system.

One healthy volunteer was imaged as a proof-of-concept experiment for obtaining simultaneous, coregistered OCT and HRF images of oral mucosa in vivo. Imaging was performed under an Institutional Review Board-approved protocol. As used in the previous studies,14,15 a 0.01% w/v solution of proflavine in sterile phosphate buffered saline was topically applied to a small region of tissue within the oral cavity. Immediately following application, the distal tip of the handpiece was placed on the tissue for imaging. Simultaneous, coregistered OCT and HRF images were acquired, processed, and displayed in real time using a custom LabView program.

3.

Results

In comparison to the theoretical OCT axial resolution of 8.0μm (based on source spectral properties), an axial resolution of 9.3μm was measured. Assuming a refractive index of 1.4 in tissue, this corresponds to an axial resolution of 6.6μm. The sample arm optics provide an OCT numerical aperture of 0.105, which results in a theoretical lateral resolution of 8.0μm and a Rayleigh parameter of 76μm. The lateral resolution was measured to be 12.4μm.

Figures 2(a) and 2(b) show the HRF geometrical spot diagrams at the image (camera) plane (IMA) for the on-axis and full-field points, respectively, displayed with the diffraction-limited spot size shown as a black circle. The on-axis imaging point is diffraction limited (8.7μm at the camera) while the rays at the edge of the field focus to a root mean square (RMS) spot size of 27.8μm. Accounting for the 5× magnification from object to image, this corresponds to a resolution at the sample of 1.7μm on-axis and 5.6μm at the edge of the field. These theoretical predictions are supported by the experimental results seen in Fig. 2(c). The image of the high resolution USAF resolution target [Fig. 2(c)] shows that the system is able to clearly resolve group 8 element 2 (with 26.4% contrast), corresponding to an on-axis resolution of 3.5μm.

Fig. 2

HRF imaging resolution. Root mean square (RMS) spot diagrams calculated in ZEMAX at image plane (IMA) of (a) on-axis and (b) field-edge imaging points. The on-axis spot diagram is diffraction limited (8.7μm), while rays at the edge of the field focus to a 27.8-μm spot (RMS). (c) HRF image of a high resolution USAF target. The system is capable of resolving group 8, element 2 with 26.4% contrast, consistent with the Rayleigh criterion. This corresponds to an imaging resolution of 3.5μm. The inset shows groups 7 to 9 in greater detail.

JBO_19_8_086004_f002.png

Figure 3 shows plots of the theoretical (a) field curvature and (b) distortion from the center of the image to the edge of the field. The maximum field curvature is 1.1 mm. The distortion of the image, which increases on moving toward the edge of the field, does not exceed 0.20%. Corresponding experimental results are demonstrated in an image of a 100-μm-square grid target [Fig. 3(c)]. The image is well focused at the center, but slightly out of focus toward the edges. The system distortion calculated from this image is 0.29% at the edge of the field, again in good agreement with theoretical predictions.

Fig. 3

HRF field curvature and distortion. Theoretical (a) field curvature and (b) distortion plotted as a function of distance from the center of the image. Field curvature does not exceed 1.1 mm and the maximum distortion is 0.20%. (c) Experimental distortion was measured by imaging a 100-μm-square grid target, yielding a maximum measured distortion of 0.29% at the edge of the image.

JBO_19_8_086004_f003.png

Figure 4 compares the theoretical and the experimental MTF curves for points (a) on-axis and (b) at the edge of the field (356μm from the center of the object). The 26.4% MTF contrast value, indicating a limiting resolution based on the Rayleigh criterion, occurs at 462.4cycles/mm. Loss of resolution was calculated for the experimental results by interpolating between the measured spatial frequencies. Experimentally, the on-axis tangential and the sagittal components differ slightly. The tangential component shows a resolution of 317.1cycles/mm and the sagittal component shows a resolution of 308.0cycles/mm, both at the 26.4% contrast level. At the edge of the field [Fig. 4(b)], the theoretical tangential and the sagittal components are distinct. The theoretical resolution at Rayleigh’s criterion occurs at 258.4cycles/mm for the tangential component and 75.4cycles/mm for the sagittal component. Experimentally, resolution figures at 26.4% contrast for the tangential and the sagittal components were measured to be 231.8 and 157.4cycles/mm, respectively. The Strehl ratio, calculated by taking the ratio of the areas under the MTF and the diffraction limited MTF, is a single metric used to evaluate imaging performance. In comparison to the theoretical on-axis Strehl ratio (0.81), a Strehl ratio of 0.63 was calculated for the on-axis experimental data.

Fig. 4

HRF system modulation transfer functions (MTFs) for points on axis (a) and at the field edge (b). Theoretical system MTF plots, calculated in ZEMAX, and experimental measurements shown.

JBO_19_8_086004_f004.png

Figure 5 shows a video capture of real-time in vivo (a) OCT and (b) HRF imaging of a volunteer’s oral mucosa, in the XZ and XY planes, respectively. The epithelium, the basement membrane, and the lamina propria are visible in the OCT image. In the HRF image, the fluorescently labeled nuclei can be distinguished from the unstained cytoplasm. The dashed line in (b) shows the intersection of the OCT scan within the HRF image.

Fig. 5

Coregistered, cross-sectional, and en face images of healthy oral mucosa (Video 1). (a) OCT and (b) HRF images taken simultaneously using the multimodal handpiece. (a) The primary tissue layers can be distinguished in the cross-sectional OCT image (EP: epithelium, BM: basement membrane, LP: lamina propria). (b) Epithelial cell nuclei appear as discrete bright dots in the en face HRF image. Scale bars are 250 μm (Video 1, QuickTime, 2.5 MB) [URL: http://dx.doi.org/10.1117/1.JBO.19.8.086004.1].

JBO_19_8_086004_f005.png

4.

Discussion

A multimodal system capable of simultaneously capturing images of cellular level detail and subsurface tissue microstructure has been developed. The combination of OCT and HRF beams at a scanning dichroic mirror provides a common optical path and inherent spatial coregistration. The dual beam-shaping requirements for high-resolution wide field imaging for HRF in tandem with focused point scanning for OCT were satisfied by the use of only six off-the-shelf lenses, allowing a compact, low-cost instrument to be assembled.

The complementary nature of these modalities makes them well suited for combination in clinical applications. The HRF imaging provides en face images of cellular level detail at the epithelial surface, but can only assess the most superficial cell layers.23,24 In contrast, OCT allows for cross-sectional imaging of tissue microstructure at depths up to a few millimeters.38 Isolation of each signal for measurement is feasible because the visible HRF wavelengths are spectrally distinct from the NIR light used for OCT.

The HRF component described here builds upon recent work on HRME in the oral cavity, which showed that the dysplastic lesions can be accurately identified in images of proflavine-stained tissue acquired through a fiber-optic bundle.14,15 Here, by eliminating the pixelating fiber bundle, the HRF system achieves higher spatial resolution than HRME (and most confocal platforms) while maintaining the field-of-view. Wall and Barton27 report a resolution of 8μm for their HRME component. In contrast, the free-space HRF system described here can resolve 3.5-μm structures. Images of standard targets indicated there is some falloff of resolution toward the edge of the HRF field [Fig. 3(c)]. Since distortion in the system is minimal, the falloff is likely due to field curvature, a result of illuminating the sample using a single lens rather than a multielement objective. As noted by other developers of in vivo high resolution imaging systems, a slightly curved field is not a major concern when imaging a thick tissue sample.39 Even with this falloff, a 5.7-μm resolution was achieved at the field edge.

With a 5× HRF system magnification, the 5.26×2.97-mm2 CCD sensor views a 1051×594-μm2 field at the sample. However, the use of a single-achromatic doublet (Fig. 1, L5) instead of a well-corrected objective lens affects illumination uniformity and image quality, especially at the field edges. Because the illumination across the entire field-of-view was not perfectly uniform, the imaged field was cropped on the left and the right sides at the location where the illumination was halfway between the maximum and the minimum pixel values, corresponding to a 712×594-μm2 field-of-view.

Although high numerical aperture is desirable for HRF imaging, a low numerical aperture is required for OCT, where it is important to match the optical depth of focus to the imaging depth range provided by the spectrometer. By filling the clear aperture of the focusing lens (Fig. 1, L5) with the HRF excitation beam and underfilling with the OCT beam, we were able to achieve a HRF numerical aperture of 0.200 and an OCT numerical aperture of 0.105. Compared to traditional OCT systems, a numerical aperture of 0.105 is fairly high and results in a shorter depth of focus. In air, this produces a useful imaging depth of about 1.5 mm instead of the theoretical 3.0 mm. A smaller numerical aperture and larger depth of focus could be achieved by increasing the focal length of the last lens (Fig. 1, L5) in the sample arm. This, however, would decrease the numerical aperture available for HRF imaging, resulting in lower magnification and reduced resolution across the HRF image. The lens selected (L5) provides a compromise between HRF imaging resolution and OCT imaging depth. The epithelium in oral mucosa is generally 90 to 300-μm thick.40 Therefore, a 1.5-mm OCT imaging depth remains capable of imaging the entire epithelium, basement membrane, and into the lamina propria, sufficient for initial assessment of submucosal involvement.

The HRF imaging component described here was designed for use with the contrast agent proflavine because of its ability to rapidly stain cell nuclei and, therefore, allow assessment of nuclear morphology, which is a well-established marker of dysplasia and neoplasia. Proflavine has also been previously used in imaging epithelial tissues in the oral cavity and GI tract.1215,41 The use of 1325-nm OCT and a short-pass dichroic with a 950-nm cut-off allows the system to be easily modified for use with different dyes within the visible and the NIR spectrum for HRF imaging. By changing only the excitation source, and the excitation and the emission filters, the HRF system has the potential to image with other fluorescent contrast agents, including fluorescein, indocyanine green, or 5-aminolevulinic acid.11,12,41 These dyes, which stain different tissue features, could potentially be used to assess the disease state based on indicators other than nuclear morphology.41

In summary, we describe a multimodal optical imaging instrument capable of simultaneously capturing en face images of subcellular level detail within the epithelium, alongside cross-sectional images of submucosal tissue microstructure. This work reports the design principles, evaluates the imaging performance of the instrument, and demonstrates that real-time, high resolution imaging can be achieved in vivo in a compact design, using only off-the-shelf lenses. Future work will evaluate the ability of the instrument to distinguish dysplastic from healthy oral mucosa in the clinical setting.

Acknowledgments

This work was supported by NIH grant R21EB011598.

References

1. 

P. BoyleB. Levin, (2008). Google Scholar

2. 

P. E. Petersen, “Oral cancer prevention and control–the approach of the World Health Organization,” Oral Oncol., 45 (4–5), 454 –460 (2009). http://dx.doi.org/10.1016/j.oraloncology.2008.05.023 EJCCER 1368-8375 Google Scholar

3. 

(2014). Google Scholar

4. 

P. J. Slootweget al., “Treatment failure and margin status in head and neck cancer. A critical view on the potential value of molecular pathology,” Oral Oncol., 38 (5), 500 –503 (2002). http://dx.doi.org/10.1016/S1368-8375(01)00092-6 EJCCER 1368-8375 Google Scholar

5. 

B. J. M. Braakhuiset al., “Molecular analysis of surgical margins in head and neck cancer: more than a marginal issue,” Oral Oncol., 46 (7), 485 –491 (2010). http://dx.doi.org/10.1016/j.oraloncology.2010.01.019 EJCCER 1368-8375 Google Scholar

6. 

M. W. Lingenet al., “Critical evaluation of diagnostic aids for the detection of oral cancer,” Oral Oncol., 44 (1), 10 –22 (2008). http://dx.doi.org/10.1016/j.oraloncology.2007.06.011 EJCCER 1368-8375 Google Scholar

7. 

B. W. NevilleT. A. Day, “Oral cancer and precancerous lesions,” CA Cancer J. Clin., 52 (4), 195 –215 (2002). http://dx.doi.org/10.3322/canjclin.52.4.195 CAMCAM 0007-9235 Google Scholar

8. 

A. L. Clarket al., “Confocal microscopy for real-time detection of oral cavity neoplasia,” Clin. Cancer Res., 9 (11), 4714 –4721 (2003). http://dx.doi.org/10.1117/1.1805558 CCREF4 1078-0432 Google Scholar

9. 

K. C. Maitlandet al., “In vivo imaging of oral neoplasia using a miniaturized fiber optic confocal reflectance microscope,” Oral Oncol., 44 (6), 1059 –1066 (2008). http://dx.doi.org/10.1016/j.oraloncology.2008.02.002 EJCCER 1368-8375 Google Scholar

10. 

Y. Wanget al., “Portable oral cancer detection using a miniature confocal imaging probe with a large field of view,” J. Micromech. Microeng., 22 (1), 065001 (2012). http://dx.doi.org/10.1088/0960-1317/22/6/065001 JMMIEZ 0960-1317 Google Scholar

11. 

P. S. Thonget al., “Laser confocal endomicroscopy as a novel technique for fluorescence diagnostic imaging of the oral cavity,” J. Biomed. Opt., 12 (1), 014007 (2007). http://dx.doi.org/10.1117/1.2710193 JBOPFO 1083-3668 Google Scholar

12. 

B. R. Haxelet al., “Confocal endomicroscopy: a novel application for imaging of oral and oropharyngeal mucosa in human,” Eur. Arch. Otorhinolaryngol., 267 (3), 443 –448 (2010). http://dx.doi.org/10.1007/s00405-009-1035-3 0937-4477 Google Scholar

13. 

M. C. PierceD. YuR. Richards-Kortum, “High-resolution fiber-optic microendoscopy for in situ cellular imaging,” JOVE, (47), e2306 (2011). http://dx.doi.org/10.3791/2306 JVEOA4 1940-087X Google Scholar

14. 

T. J. Muldoonet al., “Noninvasive imaging of oral neoplasia with a high-resolution fiber-optic microendoscope,” Head Neck, 34 (3), 305 –312 (2012). http://dx.doi.org/10.1002/hed.v34.3 HEANEE 1097-0347 Google Scholar

15. 

M. C. Pierceet al., “Accuracy of in vivo multimodal optical imaging for detection of oral neoplasia,” Cancer Prev. Res., 5 (6), 801 –809 (2012). http://dx.doi.org/10.1158/1940-6207.CAPR-11-0555 CPRACC 1940-6207 Google Scholar

16. 

P. Wilder-Smithet al., “Noninvasive imaging of oral premalignancy and malignancy,” J. Biomed. Opt., 10 (5), 051601 (2005). http://dx.doi.org/10.1117/1.2098930 JBOPFO 1083-3668 Google Scholar

17. 

J. M. Ridgwayet al., “In vivo optical coherence tomography of the human oral cavity and oropharynx,” Arch. Otolaryngol. Head Neck Surg., 132 (10), 1074 –1081 (2006). http://dx.doi.org/10.1001/archotol.132.10.1074 AONSEJ 0886-4470 Google Scholar

18. 

M. T. Tsaiet al., “Effective indicators for diagnosis of oral cancer using optical coherence tomography,” Opt. Express, 16 (20), 15847 –15862 (2008). http://dx.doi.org/10.1364/OE.16.007778 OPEXFF 1094-4087 Google Scholar

19. 

M. T. Tsaiet al., “Differentiating oral lesions in different carcinogenesis stages with optical coherence tomography,” J. Biomed. Opt., 14 (4), 044028 (2009). http://dx.doi.org/10.1117/1.3200936 JBOPFO 1083-3668 Google Scholar

20. 

P. Wilder-Smithet al., “In vivo diagnosis of oral dysplasia and malignancy using optical coherence tomography: preliminary studies in 50 patients,” Lasers Surg. Med., 41 (5), 353 –357 (2009). http://dx.doi.org/10.1002/(ISSN)1096-9101 LSMEDI 0196-8092 Google Scholar

21. 

W. Jerjeset al., “In vitro examination of suspicious oral lesions using optical coherence tomography,” Br. J. Oral Maxillofac. Surg., 48 (1), 18 –25 (2010). http://dx.doi.org/10.1016/j.bjoms.2009.04.019 0266-4356 Google Scholar

22. 

C. K. Leeet al., “Diagnosis of oral precancer with optical coherence tomography,” Biomed. Opt. Express, 3 (7), 1632 –1646 (2012). http://dx.doi.org/10.1364/BOE.3.001632 BOEICL 2156-7085 Google Scholar

23. 

M. H. KouckyM. C. Pierce, “Axial response of high-resolution microendoscopy in scattering media,” Biomed. Opt. Express, 4 (10), 2247 –2256 (2013). http://dx.doi.org/10.1364/BOE.4.002247 BOEICL 2156-7085 Google Scholar

24. 

S. El Hallaniet al., “Ex vivo confocal imaging with contrast agents for the detection of oral potentially malignant lesions,” Oral Oncol., 49 (6), 582 –590 (2013). http://dx.doi.org/10.1016/j.oraloncology.2013.01.009 EJCCER 1368-8375 Google Scholar

25. 

D. Huanget al., “Optical coherence tomography,” Science, 254 (5035), 1178 –1181 (1991). http://dx.doi.org/10.1126/science.1957169 SCIEAS 0036-8075 Google Scholar

26. 

B. J. Vakocet al., “Cancer imaging by optical coherence tomography: preclinical progress and clinical potential,” Nat. Rev. Cancer, 12 (5), 363 –368 (2012). http://dx.doi.org/10.1038/nrc3235 NRCAC4 1474-175X Google Scholar

27. 

R. A. WallJ. K. Barton, “Fluorescence-based surface magnifying chromoendoscopy and optical coherence tomography endoscope,” J. Biomed. Opt., 17 (8), 086003 (2012). http://dx.doi.org/10.1117/1.JBO.17.8.086003 JBOPFO 1083-3668 Google Scholar

28. 

H. MakhloufA. R. RouseA. F. Gmitro, “Dual modality fluorescence confocal and spectral-domain optical coherence tomography microendoscope,” Biomed. Opt. Express, 2 (3), 634 –644 (2011). http://dx.doi.org/10.1364/BOE.2.000634 BOEICL 2156-7085 Google Scholar

29. 

H. Yooet al., “Intra-arterial catheter for simultaneous microstructural and molecular imaging in vivo,” Nat. Med., 17 (12), 1680 –1684 (2011). http://dx.doi.org/10.1038/nm.2555 1078-8956 Google Scholar

30. 

J. Xiet al., “Integrated multimodal endomicroscopy platform for simultaneous en face optical coherence and two-photon fluorescence imaging,” Opt. Lett., 37 (3), 362 –364 (2012). http://dx.doi.org/10.1364/OL.37.000362 OPLEDP 0146-9592 Google Scholar

31. 

E. Auksoriuset al., “Dual-modality fluorescence and full-field optical coherence microscopy for biomedical imaging applications,” Biomed. Opt. Express, 3 (3), 661 –666 (2012). http://dx.doi.org/10.1364/BOE.3.000661 BOEICL 2156-7085 Google Scholar

32. 

Y. Zhaoet al., “Integrated multimodal optical microscopy for structural and functional imaging of engineered and natural skin,” J. Biophotonics, 5 (5–6), 437 –448 (2012). http://dx.doi.org/10.1002/jbio.v5.5/6 JBOIBX 1864-063X Google Scholar

33. 

S. Lianget al., “Intravascular atherosclerotic imaging with combined fluorescence and optical coherence tomography probe based on a double-clad fiber combiner,” J. Biomed. Opt., 17 (7), 070501 (2012). http://dx.doi.org/10.1117/1.JBO.17.7.070501 JBOPFO 1083-3668 Google Scholar

34. 

J.M. Jabbouret al., “Fluorescence lifetime imaging and reflectance confocal microscopy for multiscale imaging of oral precancer,” J. Biomed. Opt., 18 (4), 046012 (2013). http://dx.doi.org/10.1117/1.JBO.18.4.046012 JBOPFO 1083-3668 Google Scholar

35. 

N. Iftimiaet al., “Combined reflectance confocal microscopy/optical coherence tomography imaging for skin burn assessment,” Biomed. Opt. Express, 4 (5), 680 –695 (2013). http://dx.doi.org/10.1364/BOE.4.000680 BOEICL 2156-7085 Google Scholar

36. 

H. Fatakdawalaet al., “Multimodal in vivo imaging of oral cancer using fluorescence lifetime, photoacoustic and ultrasound techniques,” Biomed. Opt. Express, 4 (9), 1721 –1741 (2013). http://dx.doi.org/10.1364/BOE.4.001724 BOEICL 2156-7085 Google Scholar

37. 

T. Wanget al., “Gradient index lens based combined two-photon microscopy and optical coherence tomography,” Opt. Express, 22 (11), 12962 –12970 (2014). http://dx.doi.org/10.1364/OE.22.012962 OPEXFF 1094-4087 Google Scholar

38. 

S. H. Yunet al., “High-speed spectral-domain optical coherence tomography at 1.3 μm wavelength,” Opt. Express, 11 (26), 3598 –3604 (2003). http://dx.doi.org/10.1364/OE.11.003598 OPEXFF 1094-4087 Google Scholar

39. 

M. Kyrishet al., “Ultra-slim plastic endomicroscope objective for non-linear microscopy,” Opt. Express, 19 (8), 7603 –7615 (2011). http://dx.doi.org/10.1364/OE.19.007603 OPEXFF 1094-4087 Google Scholar

40. 

S. Prestinet al., “Measurement of epithelial thickness within the oral cavity using optical coherence tomography,” Head Neck, 34 (12), 1777 –1781 (2012). http://dx.doi.org/10.1002/hed.v34.12 HEANEE 1097-0347 Google Scholar

41. 

A. L. Polglaseet al., “A fluorescence confocal endomicroscope for in vivo microscopy of the upper- and the lower-GI tract,” Gastrointest. Endosc., 62 (5), 686 –695 (2005). http://dx.doi.org/10.1016/j.gie.2005.05.021 GAENBQ 0016-5107 Google Scholar

Biographies of the authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Laura M. Higgins and Mark C. Pierce "Design and characterization of a handheld multimodal imaging device for the assessment of oral epithelial lesions," Journal of Biomedical Optics 19(8), 086004 (7 August 2014). https://doi.org/10.1117/1.JBO.19.8.086004