1.IntroductionPhotoacoustic tomography (PAT), also called optoacoustic tomography, is a three-dimensional (3-D) imaging modality based on the photoacoustic (PA) effect. Although the PA effect was discovered more than a century ago by Alexander Graham Bell, it has found applications in biomedical imaging only in the last decade or so. Nowadays, PAT is one of the largest research areas in biomedical optics and is still growing rapidly.1 Harnessing both rich optical absorption contrast and high ultrasonic resolution, PAT is a hybrid imaging modality that can image deep tissues. While pure optical imaging modalities can also detect optical absorption by monitoring intensity variations in transmitted or reflected light, their sensitivities are usually two orders of magnitude lower than that of PAT.2 In addition, because acoustic waves scatter much more weakly than light in biological tissues, they can propagate a greater distance than photons without losing their original propagation directions, providing PAT with high spatial resolution at depths. While pure ultrasonic imaging can also achieve high spatial resolution in deep tissues, its mechanical contrast is incapable of providing certain physiological parameters, such as the oxygen saturation of hemoglobin and the metabolic rate of oxygen. Several comprehensive reviews of PAT can be found in the literature;1,3–21 some are general,1,3–5,11,21 and some focus on specific areas, such as PAT’s application in imaging molecules,9 microvasculature,6 tumors,17 the brain,16 and small animals.15 Here, we will review the fundaments of PAT, including its principles, major implementations, system characteristics, main contrast agents, and recent applications. 2.Principles2.1.Initial Pressure RiseIn PAT, a short-pulsed light source is typically used to irradiate the tissue, resulting in broadband PA waves with a frequency content extending to several tens or even hundreds of megahertz for acoustic detection. Following absorption of the light, an initial temperature rise induces a pressure rise, which propagates as a photoacoustic wave and finally is detected by a single-element ultrasonic transducer or a transducer array. There are two important time scales in the generation of PA waves,2 the thermal relaxation time () and the stress relaxation time (). denotes the thermal relaxation (thermal diffusion) time of the desired voxel, and is given by where is the desired spatial resolution and is the thermal diffusivity (). , which characterizes the stress relaxation time of the desired voxel, is given by where is the speed of sound (m/s).Upon laser excitation, the fractional volume expansion of the heated region can be expressed as where denotes the isothermal compressibility (), denotes the change in pressure (Pa), denotes the thermal coefficient of volume expansion (), and denotes the change in temperature (K).If the laser pulse duration is shorter than and , the excitation satisfies both thermal and stress confinements. In this situation, the fractional volume change is negligible. Thus, the initial pressure rise can be derived from Further, the local temperature rise can be expressed as where is the percentage of absorbed light converted into heat, and is the specific optical energy deposition (). Substituting Eq. (5) into Eq. (4), we haveBy defining the Gruneisen parameter (dimensionless) as Equation (6) becomes For single-photon optical absorption, is proportional to the local optical fluence (). In this case, Eq. (8) becomes where is the optical absorption coefficient (). Based on Eq. (9), the initial pressure rise is proportional to and . and are usually approximated as constants, although they have been found to depend on the tissue type;22,23 thus, if can be measured and is known, can be recovered. After the generation of the initial pressure , an acoustic wave starts to propagate at the speed of sound in the material. The propagation in an inviscid medium can be described by general photoacoustic equations in the time-domain, as discussed in Sec. 2.2.2.2.Photoacoustic Wave PropagationThe propagation and generation of acoustic pressure at position and time t is governed by the following wave equation: Note that represents the temperature rise instead of the temperature, i.e., the temperature rise above its initial value. Under the condition of thermal confinement, where heat conduction is negligible, the heat diffusion equation becomes where is the heating function, defined as the thermal energy deposited per unit volume and per unit time. Note that the heating function is related to the specific optical energy deposition by the following equation:Substituting Eq. (11) into Eq. (10), we have where is the specific heat capacity at constant pressure.Solving Eq. (13) with the Green function approach, we have the following delta heating response: where is the initial pressure rise at location . If the heating pulse has a finite duration, the response can be computed by convolution where is the temporal profile of the excitation pulse.3.Photoacoustic Image FormationThe goal of PA imaging is to retrieve the local pressure rise inside the tissue. Based on Eq. (9), if we know the local optical fluence , the absorption coefficient can then be calculated. In practice, the adjacent fluence in the tissue is usually comparable, but the absorption coefficient differs considerably. For example, blood in the visible light region has much stronger absorption than other components in tissue.14 Thus, if is assumed to be regionally homogeneous in anatomic PA imaging, then can be used to directly map the relative absorption coefficient . There are two basic methods to recover the original distribution inside the target once the pressures at the observation points are measured:11 reconstruction-based image formation and focused-scanning image formation. While the former is the basis of photoacoustic computed tomography (PACT), the latter is commonly used for photoacoustic microscopy (PAM) and occasionally for photoacoustic macroscopy (PAMac). Because PAM and PAMac mainly differ in their spatial resolution, i.e., PAM has a spatial resolution less than while PAMac does not, we will only discuss PAM in this review paper. 3.1.Image Formation in Photoacoustic Computed TomographyFor PACT, the light is expanded to illuminate the whole object to be imaged. PA signals are acquired at multiple locations around the object, either by using a transducer array or by scanning a single-element transducer to simulate an array. Next, back-projecting all the PA data, similar to traditional computed tomography or positron emission tomography imaging, generates PA images of the object. Note that in order to detect PA signals from the same object at multiple locations, a transducer or transducer array with a large acceptance angle is desirable. Several methods are widely used for PA image formation,24,25–30 such as universal back-projection (UBP)25 and time reversal.26 Demonstrated in spherical, cylindrical, and planar detection geometries, the UBP algorithm has the following formula: where is the solid angle of the entire detection surface with respect to a source point at . The factor weighs the contribution from each element on the detection surface . The term is the direct back-projection of the detected PA signals onto a spherical surface centered at . The first derivative with respect to time represents a ramp filter, which suppresses low frequency signals. In the UBP algorithm, the medium is assumed to be acoustically lossless and homogeneous.The time-reversal method has recently been recognized as the least restrictive reconstruction algorithm: it can work for any closed geometry and can incorporate acoustic heterogeneities. In the time-reversal method, the measured acoustic pressure is retransmitted into the medium in time reversed order. The same wave equation is solved from to , where is the maximum time for the acoustic wave to traverse the detection region. The measured pressure data are treated either as boundary conditions or as a source. To solve the wave equation, numerical methods are employed, such as time-domain finite-difference techniques or k-space pseudospectral methods. Thus, the time-reversal method is more computationally intensive than the UBP method, especially for 3-D image reconstruction. 3.2.Image Formation in Photoacoustic MicroscopyDifferent from reconstruction-based methods, the focused-scanning scheme of PAM usually focuses both the optical excitation and acoustic detection. If the optical focus is tighter than the acoustic focus, the technique is called optical resolution photoacoustic microscopy (OR-PAM);31,32 otherwise, it is called acoustic resolution photoacoustic microscopy (AR-PAM).33–35 In both cases, each laser pulse generates a one-dimensional (1-D) photoacoustic image (A-line) along the axial direction. Raster-scanning laterally and then piecing together all the A-lines provides a 3-D PA image. Because each signal acquired by the transducer directly represents a 1-D image of a single line inside the object after minimal signal processing, there is no need for image reconstruction. Although raster-scanning is used in most cases, there are alternative scanning methods, such as circumferential-section-scanning for endoscopic imaging,36,37–40 3-D arbitrary scanning for blood vessel monitoring,41 and random access scanning for cell tracking.42 4.Photoacoustic Systems and Their CharacteristicsAs mentioned earlier, based on their different image formation mechanism, PA systems can be classified as either reconstruction-based PACT or focused-scanning-based PAM. In this section, we discuss typical PA systems and their characteristics, including spatial resolution, imaging speed, and penetration depth. 4.1.Photoacoustic Computed TomographyCurrent PACT systems use spherical,43,44–46 cylindrical,15,47–52 or planar detection geometry.53–60 Each geometry has several implementations. For spherical-view systems, either an arc-shaped transducer array44 or a hemispherical array with a spiral pattern45,46 is used. In both cases, mechanical scanning is required for dense spatial sampling, and 3-D reconstruction is performed. Figure 1(a) shows a hemispherical array based PACT system,46 with 512 ultrasonic transducer elements distributed in a hemispherical shell with a radius of 127 mm. The diameter of each transducer element is 3 mm, and the center frequency is 2 MHz, with a 70% bandwidth. The transducer array is scanned spirally to achieve dense spatial sampling for image reconstruction. As shown in Fig. 1(a), the bottom of the hemispherical shell contains an aperture for light delivery. In this case, pulses from an Alexandrite laser at 756 nm are used to excite the target. Depending on the spiral scanning pattern, total data acquisition time varies from 12 s for the smallest spiral (24-mm radius) to 3.2 min for the largest spiral (96-mm radius). Using a thin graphite fiber phantom ( diameter), the resolution of this system was quantified to be 0.42 mm. Because spherical-view detection was implemented, this resolution was constant at different graphite fiber orientations. A 5.3-cm penetration was achieved in the phantom. As shown in Fig. 1(b), blood vessels in the breasts of two healthy volunteers were clearly imaged although their depths are not shown. A representative cylindrical-view PACT system48 is shown in Fig. 2. The system contains a 512-element full ring transducer array with a ring diameter of 5 cm. Each transducer element has a center frequency of 5 MHz and an 80% (one-way) detection bandwidth.61 To improve the cross-sectional imaging ability, each element is cylindrically focused to reject out-of-plane signals. The combined foci from all elements provide a central imaging region with a diameter and thickness. Strictly speaking, such a system is a circular-view system, since only a ring is used to reconstruct the image instead of a cylinder. However, by taking advantage of its cylindrical focusing capability, high-quality two-dimensional (2-D) cross-sectional images are attainable. In addition, by scanning the sample or the array along the elevational direction, 3-D images can be acquired. Within the imaging region, the system provides 100- to transverse resolution in the circumferential direction and axial resolution in the radial direction. Limited by its 64-channel data acquisition and 10-Hz laser, the system acquires one . However, by employing a 512-channel real-time data acquisition system and a faster laser, higher rate imaging can be realized. Because of the fixed ring diameter, only 1-cm penetration depth was reported. As shown in Figs. 2(b) and 2(c), blood-rich organs, such as the liver, kidneys, spleen, spine, and GI tracts can be clearly visualized. In addition, blood-poor organs, such as the bladder, can also be imaged with the help of a near-infrared contrast agent (IRDye800, LI-COR, Inc.), as shown in Fig. 2(d). There are different implementations for planar-view PACT, using either a 2-D ultrasound transducer array44,53 or a Fabry–Perot interferometer (FPI).55,60 Generally, the 2-D transducer array based PACT system has a higher frame rate, while the FPI-based system has higher sensitivity and a larger receiving angle. Figure 3(a) is a schematic of the FPI-based PACT system.62 In this type of PA imaging, the deformation of pressure-sensitive materials (e.g., polymer) is measured by optical resonance. The PA excitation beam was at 640 nm, and the PA probing beam was at 1550 nm. By raster scanning the probe beam across the FPI surface, photoacoustic waves can be mapped in 2-D. Depending on the detector bandwidth (22 MHz in this work), the axial resolution of this system was . The lateral resolution was about , which was determined by primarily the detection bandwidth and angular range. To scan an area of , the image acquisition time was about 8 min, which was limited by the 50-Hz pulse repetition rate of the excitation laser. The in vivo penetration depth of this system was demonstrated to be more than 10 mm. As shown in Fig. 3(b), two embryos (shaded in red) with detailed structures, such as the liver, ribs, pulmonary vein, and right atrium, can be clearly imaged. 4.2.Photoacoustic MicroscopyAs mentioned earlier, there are two types of PAM: OR-PAM and AR-PAM. In OR-PAM, although the optical focus determines its lateral resolution, the detection transducer is also placed confocally with the optical focus to maximize the system’s detection sensitivity. Similarly, in AR-PAM, the excitation beam fills the entire acoustic focus to maximize the system’s sensitivity. The axial resolution in both OR-PAM and AR-PAM is determined by the detection bandwidth of the transducer. Due to the frequency dependence of acoustic attenuation, the bandwidth is chosen according to the desired imaging depth. As shown in Fig. 4(a), a typical OR-PAM system employs an optical lens to focus light into the sample. A light-sound combiner transmits the light and reflects the sound. The combiner is composed of a right-angled prism, a thin layer of silicone oil, and a rhomboid prism for acoustic-optical coaxial alignment. Usually the laser beam is tightly focused, whose diameter can range from several hundred nanometers to several micrometers, depending on the numerical aperture (NA) of the optical focusing lens, the wavelength of the excitation beam, and the desired imaging depth. Relying on the tight optical focus, the penetration of an OR-PAM system is limited to about one transport mean free path in tissue ().12,63,64 However, by using longer wavelength laser pulses, which have longer transport mean free paths, the penetration limit can be increased.65 OR-PAM can image vasculature in a mouse ear,31 eye,66–69 and brain31,70,71 clearly. Figure 4(b) shows a representative mouse ear image acquired with OR-PAM, where both a capillary bed and flowing red blood cells can be clearly visualized. As shown in Fig. 5(a), in a typical AR-PAM system, the laser beam passes through a conical lens to form a ring-shaped illumination pattern.34,72,73–75 The beam is then focused into the target by custom-made mirrors. The optical illumination on the skin surface has a donut shape with a dark center to minimize strong surface signals. Since acoustic scattering is much weaker than optical scattering in tissue, tight acoustic focusing can be maintained at depths. For example, using a 50-MHz center frequency transducer with an NA of 0.44, a lateral resolution of was achieved with an imaging depth of more than 3 mm. By choosing transducers with different center frequencies and NAs, the lateral resolutions can be scaled. Note that although acoustic scattering is weak in tissue, high frequency ultrasound signals suffer strong attenuation. In the end, the attenuation becomes the limiting factor for deep high-resolution AR-PAM imaging. Figure 5(b) shows a representative AR-PAM image of a human forearm, where detailed vasculatures can be clearly discerned. Figure 5(c) is a photo of the forearm, where the red box indicates the image area of . The lateral and axial resolutions as well as imaging depth are summarized in Table 1. Table 1Summary of typical PA system characteristics.
5.Photoacoustic Contrast AgentsTheoretically speaking, any material with sufficiently high optical absorption can be detected by PAT. Thus, by choosing the right wavelengths, PAT potentially can be used to image all materials. Specifically, in biological applications, absorbers are usually divided into endogenous and exogenous categories. We will discuss these two categories in Secs. 5.1 and 5.2. 5.1.Endogenous Contrast AgentsThe primary advantage of endogenous contrast agents is that they allow label-free imaging, so they do not affect the original biological environment. In biological tissue, there are varieties of optical absorbers,14 such as DNA/RNA,79,80 cytochromes,81 bilirubin,82 myoglobin,83,84 hemoglobin,85,86 methemoglobin,87 carboxyhemoglobin,88 melanin,54,89,90 lipid,91–93 water,94,95 and glucose.96,97 Among all these agents, DNA/RNA is commonly used for cell nuclear imaging in the ultraviolet region [Fig. 6(a)],79 hemoglobin is widely used for vascular imaging in the visible and the near-infrared spectral regions [Fig. 6(b)],19 and melanin is employed for melanoma tumor imaging in the near-infrared region [Fig. 6(c)].54 In addition, in the near-infrared region, lipids and water are used for atherosclerotic plaque98,99 and injury94 imaging, respectively. As shown in Fig. 7, different endogenous contrast agents have different absorption spectra. Thus, PAT can separate them with spectral measurement, when the local optical fluence is known. A representative application of spectral decomposition is to quantify the oxygen saturation of hemoglobin () in blood vessels by differentiating signals from oxy-hemoglobin () and deoxy-hemoglobin (HbR). As derived above, the measured PA amplitude of blood is proportional to the local fluence and the absorption coefficient of blood where is the measured PA amplitude at wavelength , is a system constant, is the blood absorption coefficient, is the optical fluence. and are the molar extinction coefficients of HbR and , respectively, and and are the concentrations of HbR and , respectively. For a given system, is a constant. In OR-PAM, which works within the optical ballistic regime, can be corrected for by measuring the surface optical fluence. Thus, by performing measurements at two wavelengths, the relative concentrations of HbR and can be quantified and where stands for the second wavelength. By definition, the value can be calculated as However, in AR-PAM and PACT, which work in the optical (quasi)diffusive regime, it is challenging to correct for . So far, several methods have been proposed to address this issue and achieve more accurate measurement in the optical (quasi)diffusive regime,100–103 such as directly fitting the temporal profiles of PA signals,100,103 analyzing their acoustic spectra,101 and measuring their dynamics.1025.2.Exogenous Contrast AgentsCompared with endogenous contrast agents, exogenous ones can be engineered to absorb at specific wavelengths to maximize their detection sensitivities. In addition, exogenous agents can be made to bind to only certain molecules; thus, these molecules can be selectively imaged by PAT. So far, a variety of exogenous agents have been developed,9,104,105 including nanoparticles,106–109 organic dyes,110–115 and proteins.116,117 Recently, for PA imaging, agents with specially designed functions have been developed, such as photosensitizing,118,119 activatable,120 and switchable agents.121–123 Note that since hemoglobin and water have strong absorption in the visible and midinfrared regions, respectively, most of these exogenous agents are designed to work in the near-infrared window, i.e., from 700 to 1350 nm. Usually, to enable more accurate spectral decomposition, two or more wavelengths are chosen to image blood and exogenous contrasts separately. Figure 8 shows two examples of using exogenous contrast agents in PA imaging: methylene blue115 and gold nanoparticles.108 Before the contrasts were injected, only blood vessels could be detected, as shown in Figs. 8(a) and 8(c). However, after injecting the contrasts, a sentinel lymph node and tumor appeared with high contrast, as seen in Figs. 8(b) and 8(d), respectively. 6.Recent ApplicationsWith its multiparameter and multiscale imaging capability, PAT has been applied in many different disciplines, including cardiology,77,124,125 dermatology,126–129 oncology,130–133 ophthalmology,67,69,134–136 gastroenterology,36–38,40,137–140 hematology,141–143 and neurology.144–147 In terms of imaging locations, PAT can be used for human breast imaging148–150 and small-animal imaging of the brain,70,151,152 ear,153–155 eye,134–136 liver,15 intestine,156 and skin.54,90 In terms of functionality, PAT has been widely used for anatomical,157,158 functional,159–161 molecular,9,82 and metabolic imaging.162,163 As for object size, PAT can detect objects ranging from organelles to human organs or whole-body small animals.11 PAT has been used for both small animal imaging (such as zebrafish,164 mice,165 rats,151 and rabbits166) and human imaging.72 In this review paper, rather than cover details of all the applications, we will simply list the most recent advances and significant applications of PAT. 6.1.High Speed Imaging of Mouse Brain FunctionsPAT has been extensively applied for brain studies. Wang et al.47 reported the first in vivo mouse brain function study with PAT. Using a circular-view PACT system, they imaged rat brain responses to vibrational stimulations to whiskers. Because a single-element ultrasonic transducer with a center frequency of 3.5 MHz was used, both the frame rate and spatial resolution were limited. With newer techniques, both the frame rate and spatial resolution have been significantly improved. For example, using an ultrasonic transducer array with 512 elements, mouse brain imaging at one frame per 1.25 s has been achieved.48 Using OR-PAM, optical resolution mouse brain images, i.e., at micrometer or submicrometer level resolution, were obtained.167 Recently, Yao et al.70 reported fast functional PAM (ffPAM) for mouse brain function imaging in action. Working as OR-PAM, this system has a lateral resolution of and an axial resolution of . Using a water-immersed microelectromechanical system (MEMS) scanning mirror along with a 500 kHz repetition rate laser, ffPAM has a 2-D frame rate of 400 Hz over a 3-mm scanning range. With a field of view, a 3-D volumetric rate of 1 Hz can be achieved. By using a single-wavelength pulse-width-based method (PW-), ffPAM can perform high speed imaging of up to a 1-D rate of 100 kHz. As shown in Fig. 9(a), to implement the PW- method, lasers with different pulse widths are used, i.e., 3 ps and 3 ns. Figure 9(b) shows a representative PA image of the mouse brain with the skull intact, where the cortex vasculature can be visualized in detail. With the PW- method, only 40 s were required to acquire the map shown in Fig. 9(c). To further demonstrate the fast functional imaging capability, the authors studied mouse cortical responses to electrical stimulations of the hindlimbs. As shown in Fig. 10(a), upon stimulations, PA signals in the corresponding regions increased. In addition, levels increased in veins and deep capillary beds upon stimulation, as shown in Fig. 10(b). However, there was no arterial response, which is consistent with the fact that arterial blood had not reached the capillaries for oxygen exchange and thus maintained a high oxygenation level. 6.2.Sentinel Lymph Node Biopsy Guidance in Patients with Breast CancerSentinel lymph node (SLN) biopsy is a standard of care in diagnosing cancer, including breast and melanoma cancers. Because the SLN is the first node in the lymphatic system that drains a tumor site, metastasis can be diagnosed by SLN biopsy. A positive biopsy result suggests that cancer has spread to the node and probably to distant organs. Recently, a handheld dual-modality ultrasound (US) and PAT system was used to accurately identify an SLN and thus to precisely guide SLN biopsy.168 The dual-modality system was modified from a clinical US scanner (iU22, Philips Healthcare). To highlight SLNs in patients, a very common clinical contrast agent, methylene blue, was used. As shown in Fig. 11, laser pulses at two wavelengths of 665 and 1064 nm spectrally differentiated methylene blue from other major absorbers, such as blood. The pulse width was around 6.5 ns, and the pulse repetition rate was 10 Hz. With a custom-built data acquisition computer, this system could achieve a frame rate of 5 Hz for coregistered US and PA imaging. To improve the detection sensitivity and operation convenience, the light delivery fiber bundles and ultrasonic transducer array were integrated into a single probe, as shown in Fig. 11. Because the probe is handheld, it is convenient for physicians to operate for SLN biopsy guidance. As shown in Fig. 12, by combining both US and PA, the SLN in a patient with breast cancer can be clearly detected. In addition, taking advantage of its high frame rate, this system can provide guidance for fine needle aspiration biopsy (FNAB) with minimal invasiveness. Because both the SLN and the needle can be imaged with high contrast, FNAB can be performed with high accuracy. 6.3.Multiscale Photoacoustic Tomography with Photo-Switchable Protein ContrastA reversibly switchable bacterial phytochrome, Rhodopseudomonas palustris (BphP1), has been recently combined with PAT for deep molecular imaging with improved detection sensitivity and spatial resolution.121 BphP1 has two states: Pfr and Pr. Upon 730- to 790-nm light illumination, it undergoes a Pfr→Pr photoconversion; while upon 630- to 690-nm light illumination, Pr→Pfr photoconversion happens. For simplicity, the Pfr state of BphP1 was denoted as the ON state and the Pr state as the OFF state. In the reported work, 780 nm light was used for Pfr→Pr photoconversion, and 630-nm light was used for Pr→Pfr photoconversion. Because the background absorbers, primarily blood, did not have the same switchable properties as BphP1, taking differential images largely suppressed the background signals and thus increased the detection sensitivity for BphP1-expressing tumors in deep tissue. First implementing BphP1 with a circular-view PACT system, the authors observed a noise-equivalent detection sensitivity of U87 human glioblastoma cells expressing BphP1, as shown in Fig. 13(a). With the single-wavelength differential method, the CNR was about 34-fold higher than the two-wavelength spectral unmixing method. In the in vivo experiment, a mouse was imaged 1 week after injection of BphP1-expressing U87 cells into its left kidney. As shown in Fig. 13(b), although major organs, such as the skin, kidneys, spleen, and bladder, can all be clearly imaged, the tumor in the left kidney could not be detected because of the overwhelming blood signals. However, after 20 cycles of photoswitching, the differential image showed the tumor at depths up to 8 mm with high contrast, as seen in Figs. 13(b) and 13(c). The line profiles in Fig. 13(b) show that while the background blood signals remained unchanged, the photoswitchable tumor had clearly different signals in the ON- and OFF-state images [Fig. 13(d)]. A histological examination taken after PA imaging confirmed the tumor, as shown in Fig. 13(e). Another tumor detection experiment, in a mouse brain, also showed the superior sensitivity of the BphP1-enhanced PACT system, as shown in Fig. 13(f). Combined with a high NA (1.4 in this work) OR-PAM system [Fig. 14(a)], this reversely switchable protein can also be used for super-resolution PA imaging (RS-SPAM). As shown in Fig. 14(b), because the switching-off rate is proportional to the local excitation intensity, PA signals from the center of the excitation spot will decay faster than those from the periphery. By fitting the nonlinear signal-decay process, a high-order coefficient can be extracted and thus subdiffraction resolution can be achieved. As shown in Figs. 14(c)–14(g), RS-SPAM showed much finer lateral and axial resolutions. The lateral resolution was quantified to be , which is about twofold better than that of conventional PAM, and the axial resolution was around 400 nm in RS-SPAM, which was around 75 times better than that of conventional PAM. 7.SummaryIn summary, PAT is a highly scalable imaging modality with major implementations of PAM and PACT. In PAM, light is focused into the target and a focused transducer is typically used for the signal detection. Thus, its lateral resolution is determined by either the optical focus (OR-PAM) or acoustic focus (AR-PAM), depending on which one is tighter. In OR-PAM, the lateral resolution is given by , where is the light wavelength, and is the NA of the optical focusing lens. In AR-PAM, the lateral resolution is given by , where is the central acoustic wavelength of the ultrasonic transducer, and is the NA of the acoustic lens. The difference in the scaling factors arises because optical excitation is based on light intensity, while ultrasonic detection is based on acoustic amplitude. In both cases, the axial resolution is given by , where is the speed of sound in soft tissue, and is the bandwidth of the ultrasonic transducer. This axial resolution formula also applies to PACT systems. However, unlike PAM, lateral resolutions in PACT are usually not a constant. Thus, we can see that by choosing different optical focusing lenses or ultrasonic transducers, both lateral and axial resolution can be changed. Super optical resolution has been achieved in OR-PAM.169 In addition, because acoustic attenuation in tissues increases with the acoustic frequency, ultrasonic transducers with different center frequencies should be chosen according to the desired imaging penetration limit. The maximum penetration achieved in PA images is with a lateral resolution about .11 PAT is also a multiparameter imaging modality. In most of the cases, because hemoglobin provides the highest contrast in biological imaging, most studies focus on extracting and studying blood-related parameters, such as blood vessel diameter,31 blood flow speed,170,171 hemoglobin oxygen concentration,19,31 blood pulse wave velocity,172 and the metabolic rate of oxygen.162 PAT has provided valuable information for studying vasculature-related diseases, such as stroke,173 diabetes,174 and atherosclerosis.98,175,176 In addition, because neural activities are closely related to hemodynamics, measuring these hemodynamic parameters is also useful for neurological studies, such as studies on epilepsy,177 resting state functional connectivity,152 and stimulation responses.147 Because of its high contrast in comparison to other absorbers, such as melanin54 and DNA/RNA,79 PAT can also image other important biological parameters. Imaging melanin can provide the depth of melanoma, as well as its rate of growth and metastatic rate,54 which are all very important parameters in diagnosing and treating melanoma patients. Imaging DNA/RNA provides a tool for label-free measurement of cell nuclear density, which potentially can be used for tumor demarcation. PAT has become one of the fastest growing fields in biomedical imaging. So far, PAT has translated several important applications into clinics, which may help solve existing problems in health care. For example, noninvasive detection of SLNs168 in patients can potentially provide minimally invasive cancer staging, and quantification of melanoma depth54 can potentially guide more accurate surgeries, both reducing morbidity and costs. With advances in new techniques, we anticipate that PAT will provide valuable information for disease diagnosis as well as treatment. AcknowledgmentsThe authors would like to thank Professor James Ballard for manuscript editing. This work was sponsored in part by National Institutes of Health Grant Nos. DP1 EB016986 (NIH Director’s Pioneer Award), R01 CA186567 (NIH Director’s Transformative Research Award), and S10 RR026922. L. V. W. has a financial interest in Microphotoacoustics, Inc., which, however, did not support this work. ReferencesL. V. Wang and L. Gao,
“Photoacoustic microscopy and computed tomography: from bench to bedside,”
Annu. Rev. Biomed. Eng., 16 155
–185
(2014). http://dx.doi.org/10.1146/annurev-bioeng-071813-104553 Google Scholar
L. V. Wang,
“Tutorial on photoacoustic microscopy and computed tomography,”
IEEE J. Sel. Top. Quantum Electron., 14
(1), 171
–179
(2008). http://dx.doi.org/10.1109/JSTQE.2007.913398 Google Scholar
P. Beard,
“Biomedical photoacoustic imaging,”
Interface Focus, 1
(4), 602
–631
(2011). http://dx.doi.org/10.1098/rsfs.2011.0028 Google Scholar
J. Yao and L. V. Wang,
“Photoacoustic tomography: fundamentals, advances and prospects,”
Contrast Media Mol. Imaging, 6
(5), 332
–345
(2011). http://dx.doi.org/10.1002/cmmi.443 Google Scholar
V. Ntziachristos,
“Going deeper than microscopy: the optical imaging frontier in biology,”
Nat. Methods, 7
(8), 603
–614
(2010). http://dx.doi.org/10.1038/nmeth.1483 1548-7091 Google Scholar
S. Hu and L. V. Wang,
“Photoacoustic imaging and characterization of the microvasculature,”
J. Biomed. Opt., 15
(1), 011101
(2010). http://dx.doi.org/10.1117/1.3281673 JBOPFO 1083-3668 Google Scholar
C. Li and L. V. Wang,
“Photoacoustic tomography and sensing in biomedicine,”
Phys. Med. Biol., 54
(19), R59
–R97
(2009). http://dx.doi.org/10.1088/0031-9155/54/19/R01 Google Scholar
L. V. Wang,
“Multiscale photoacoustic microscopy and computed tomography,”
Nat. Photonics, 3
(9), 503
–509
(2009). http://dx.doi.org/10.1038/nphoton.2009.157 Google Scholar
C. Kim, C. Favazza and L. V. Wang,
“In vivo photoacoustic tomography of chemicals: high-resolution functional and molecular optical imaging at new depths,”
Chem. Rev., 110
(5), 2756
–2782
(2010). http://dx.doi.org/10.1021/cr900266s Google Scholar
X. Cai et al.,
“Photoacoustic microscopy in tissue engineering,”
Mater. Today, 16
(3), 67
–77
(2013). http://dx.doi.org/10.1016/j.mattod.2013.03.007 Google Scholar
L. V. Wang and S. Hu,
“Photoacoustic tomography: in vivo imaging from organelles to organs,”
Science, 335
(6075), 1458
–1462
(2012). http://dx.doi.org/10.1126/science.1216210 SCIEAS 0036-8075 Google Scholar
J. Yao and L. V. Wang,
“Photoacoustic microscopy,”
Laser Photon. Rev., 7
(5), 758
–778
(2013). http://dx.doi.org/10.1002/lpor.2013.7.issue-5 Google Scholar
J. Yao, L. Song and L. V. Wang,
“Photoacoustic microscopy superdepth, superresolution, and superb contrast,”
IEEE Pulse, 6
(3), 34
–37
(2015). http://dx.doi.org/10.1109/MPUL.2015.2409100 Google Scholar
J. Yao and L. V. Wang,
“Sensitivity of photoacoustic microscopy,”
Photoacoustics, 2
(2), 87
–101
(2014). http://dx.doi.org/10.1016/j.pacs.2014.04.002 Google Scholar
J. Xia and L. V. Wang,
“Small-animal whole-body photoacoustic tomography: a review,”
IEEE Trans. Bio-Med. Eng., 61
(5), 1380
–1389
(2014). http://dx.doi.org/10.1109/TBME.2013.2283507 Google Scholar
J. Yao and L. V. Wang,
“Photoacoustic brain imaging: from microscopic to macroscopic scales,”
Neurophotonics, 1
(1), 011003
(2014). http://dx.doi.org/10.1117/1.NPh.1.1.011003 Google Scholar
S. Mallidi, G. P. Luke and S. Emelianov,
“Photoacoustic imaging in cancer detection, diagnosis, and treatment guidance,”
Trends Biotechnol., 29
(5), 213
–221
(2011). http://dx.doi.org/10.1016/j.tibtech.2011.01.006 Google Scholar
B. Cox et al.,
“Quantitative spectroscopic photoacoustic imaging: a review,”
J. Biomed. Opt., 17
(6), 061202
(2012). http://dx.doi.org/10.1117/1.JBO.17.6.061202 JBOPFO 1083-3668 Google Scholar
S. Hu and L. V. Wang,
“Optical-resolution photoacoustic microscopy: auscultation of biological systems at the cellular level,”
Biophys. J., 105
(4), 841
–847
(2013). http://dx.doi.org/10.1016/j.bpj.2013.07.017 Google Scholar
L. V. Wang,
“Prospects of photoacoustic tomography,”
Med. Phys., 35
(12), 5758
–5767
(2008). http://dx.doi.org/10.1118/1.3013698 Google Scholar
L. V. Wang,
“Multiscale photoacoustic microscopy and computed tomography,”
Nat. Photonics, 3
(9), 503
–509
(2009). http://dx.doi.org/10.1038/nphoton.2009.157 NPAHBY 1749-4885 Google Scholar
D. K. Yao et al.,
“Photoacoustic measurement of the Gruneisen parameter of tissue,”
J. Biomed. Opt., 19
(1), 017007
(2014). http://dx.doi.org/10.1117/1.JBO.19.1.017007 JBOPFO 1083-3668 Google Scholar
L. Gao et al.,
“Intracellular temperature mapping with fluorescence-assisted photoacoustic-thermometry,”
Appl. Phys. Lett., 102
(19),
(2013). http://dx.doi.org/10.1063/1.4807140 Google Scholar
M. H. Xu and L. H. V. Wang,
“Photoacoustic imaging in biomedicine,”
Rev. Sci. Instrum., 77
(4), 041101
(2006). http://dx.doi.org/10.1063/1.2195024 Google Scholar
M. H. Xu and L. H. V. Wang,
“Universal back-projection algorithm for photoacoustic computed tomography,”
Phys. Rev. E, 71
(1), 016706
(2005). http://dx.doi.org/10.1117/12.589146 Google Scholar
B. E. Treeby and B. T. Cox,
“k-Wave: MATLAB toolbox for the simulation and reconstruction of photoacoustic wave fields,”
J. Biomed. Opt., 15
(2), 021314
(2010). http://dx.doi.org/10.1117/1.3360308 Google Scholar
M. Xu and L. V. Wang,
“Analytic explanation of spatial resolution related to bandwidth and detector aperture size in thermoacoustic or photoacoustic reconstruction,”
Phys. Rev. E, 67
(5 Pt 2), 056605
(2003). http://dx.doi.org/10.1103/PhysRevE.67.056605 Google Scholar
K. Wang et al.,
“Investigation of iterative image reconstruction in three-dimensional optoacoustic tomography,”
Phys. Med. Biol., 57
(17), 5399
–5423
(2012). http://dx.doi.org/10.1088/0031-9155/57/17/5399 Google Scholar
K. Wang et al.,
“Discrete imaging models for three-dimensional optoacoustic tomography using radially symmetric expansion functions,”
IEEE Trans. Med. Imaging, 33
(5), 1180
–1193
(2014). http://dx.doi.org/10.1109/TMI.2014.2308478 Google Scholar
K. Wang et al.,
“An imaging model incorporating ultrasonic transducer properties for three-dimensional optoacoustic tomography,”
IEEE Trans. Med. Imaging, 30
(2), 203
–214
(2011). http://dx.doi.org/10.1109/TMI.2010.2072514 Google Scholar
S. Hu, K. Maslov and L. V. Wang,
“Second-generation optical-resolution photoacoustic microscopy with improved sensitivity and speed,”
Opt. Lett., 36
(7), 1134
–1136
(2011). http://dx.doi.org/10.1364/OL.36.001134 OPLEDP 0146-9592 Google Scholar
K. Maslov et al.,
“Optical-resolution photoacoustic microscopy for in vivo imaging of single capillaries,”
Opt. Lett., 33
(9), 929
–931
(2008). http://dx.doi.org/10.1364/OL.33.000929 OPLEDP 0146-9592 Google Scholar
H. F. Zhang, K. Maslov and L. H. V. Wang,
“In vivo imaging of subcutaneous structures using functional photoacoustic microscopy,”
Nat. Protoc., 2
(4), 797
–804
(2007). http://dx.doi.org/10.1038/nprot.2007.108 Google Scholar
H. F. Zhang et al.,
“Functional photoacoustic microscopy for high-resolution and noninvasive in vivo imaging,”
Nat. Biotechnol., 24
(7), 848
–851
(2006). http://dx.doi.org/10.1038/nbt1220 Google Scholar
K. Maslov, G. Stoica and L. H. V. Wang,
“In vivo dark-field reflection-mode photoacoustic microscopy,”
Opt. Lett., 30
(6), 625
–627
(2005). http://dx.doi.org/10.1364/OL.30.000625 OPLEDP 0146-9592 Google Scholar
J. M. Yang et al.,
“Photoacoustic endoscopy,”
Opt. Lett., 34
(10), 1591
–1593
(2009). http://dx.doi.org/10.1364/OL.34.001591 OPLEDP 0146-9592 Google Scholar
J. M. Yang et al.,
“Catheter-based photoacoustic endoscope,”
J. Biomed. Opt., 19
(6), 066001
(2014). http://dx.doi.org/10.1117/1.JBO.19.6.066001 JBOPFO 1083-3668 Google Scholar
J. M. Yang et al.,
“Optical-resolution photoacoustic endomicroscopy in vivo,”
Biomed. Opt. Express., 6
(3), 918
–932
(2015). http://dx.doi.org/10.1364/BOE.6.000918 Google Scholar
J. M. Yang et al.,
“Simultaneous functional photoacoustic and ultrasonic endoscopy of internal organs in vivo,”
Nat. Med., 18
(8), 1297
–1302
(2012). http://dx.doi.org/10.1038/nm.2823 1078-8956 Google Scholar
C. Y. Li et al.,
“Urogenital photoacoustic endoscope,”
Opt. Lett., 39
(6), 1473
–1476
(2014). http://dx.doi.org/10.1364/OL.39.001473 OPLEDP 0146-9592 Google Scholar
C. Yeh et al.,
“Three-dimensional arbitrary trajectory scanning photoacoustic microscopy,”
J. Biophotonics, 8
(4), 303
–308
(2015). http://dx.doi.org/10.1002/jbio.v8.4 Google Scholar
J. Y. Liang et al.,
“Random-access optical-resolution photoacoustic microscopy using a digital micromirror device,”
Opt. Lett., 38
(15), 2683
–2686
(2013). http://dx.doi.org/10.1364/OL.38.002683 OPLEDP 0146-9592 Google Scholar
R. A. Kruger et al.,
“Thermoacoustic molecular imaging of small animals,”
Mol. Imaging, 2
(2), 113
–123
(2003). http://dx.doi.org/10.1162/153535003322331993 Google Scholar
H. P. Brecht et al.,
“Whole-body three-dimensional optoacoustic tomography system for small animals,”
J. Biomed. Opt., 14
(6), 064007
(2009). http://dx.doi.org/10.1117/1.3259361 JBOPFO 1083-3668 Google Scholar
R. A. Kruger et al.,
“Photoacoustic angiography of the breast,”
Med. Phys., 37
(11), 6096
–6100
(2010). http://dx.doi.org/10.1118/1.3497677 Google Scholar
R. A. Kruger et al.,
“Dedicated 3D photoacoustic breast imaging,”
Med. Phys., 40
(11), 113301
(2013). http://dx.doi.org/10.1118/1.4824317 Google Scholar
X. D. Wang et al.,
“Noninvasive laser-induced photoacoustic tomography for structural and functional in vivo imaging of the brain,”
Nat. Biotechnol., 21
(7), 803
–806
(2003). http://dx.doi.org/10.1038/nbt839 Google Scholar
J. Xia et al.,
“Whole-body ring-shaped confocal photoacoustic computed tomography of small animals in vivo,”
J. Biomed. Opt., 17
(5), 050506
(2012). http://dx.doi.org/10.1117/1.JBO.17.5.050506 JBOPFO 1083-3668 Google Scholar
C. H. Li and L. H. V. Wang,
“Photoacoustic tomography of the mouse cerebral cortex with a high-numerical-aperture-based virtual point detector,”
J. Biomed. Opt., 14
(2), 024047
(2009). http://dx.doi.org/10.1117/1.3122365 JBOPFO 1083-3668 Google Scholar
C. H. Li et al.,
“Real-time photoacoustic tomography of cortical hemodynamics in small animals,”
J. Biomed. Opt., 15
(1), 010509
(2010). http://dx.doi.org/10.1117/1.3302807 JBOPFO 1083-3668 Google Scholar
A. Buehler et al.,
“Video rate optoacoustic tomography of mouse kidney perfusion,”
Opt. Lett., 35
(14), 2475
–2477
(2010). http://dx.doi.org/10.1364/OL.35.002475 OPLEDP 0146-9592 Google Scholar
L. M. Nie et al.,
“Photoacoustic tomography through a whole adult human skull with a photon recycler,”
J. Biomed. Opt., 17
(11), 110506
(2012). http://dx.doi.org/10.1117/1.JBO.17.11.110506 JBOPFO 1083-3668 Google Scholar
Y. Wang et al.,
“In vivo three-dimensional photoacoustic imaging based on a clinical matrix array ultrasound probe,”
J. Biomed. Opt., 17
(6), 061208
(2012). http://dx.doi.org/10.1117/1.JBO.17.6.061208 JBOPFO 1083-3668 Google Scholar
Y. Zhou et al.,
“Handheld photoacoustic probe to detect both melanoma depth and volume at high speed in vivo,”
J. Biophotonics, 1
(7),
(2015). http://dx.doi.org/10.1002/jbio.201400143 Google Scholar
E. Zhang, J. Laufer and P. Beard,
“Backward-mode multiwavelength photoacoustic scanner using a planar Fabry-Perot polymer film ultrasound sensor for high-resolution three-dimensional imaging of biological tissues,”
Appl. Opt., 47
(4), 561
–577
(2008). http://dx.doi.org/10.1364/AO.47.000561 Google Scholar
D. Piras et al.,
“Photoacoustic imaging of the breast using the twente photoacoustic mammoscope: present status and future perspectives,”
IEEE J. Sel. Top. Quantum Electronics, 16
(4), 730
–739
(2010). http://dx.doi.org/10.1109/JSTQE.2009.2034870 Google Scholar
M. Heijblom et al.,
“Visualizing breast cancer using the Twente photoacoustic mammoscope: What do we learn from twelve new patient measurements?,”
Opt. Express., 20
(11), 11582
–11597
(2012). http://dx.doi.org/10.1364/OE.20.011582 OPEXFF 1094-4087 Google Scholar
M. Heijblom et al.,
“Appearance of breast cysts in planar geometry photoacoustic mammography using 1064-nm excitation,”
J. Biomed. Opt., 18
(12), 126009
(2013). http://dx.doi.org/10.1117/1.JBO.18.12.126009 JBOPFO 1083-3668 Google Scholar
J. Laufer et al.,
“Three-dimensional noninvasive imaging of the vasculature in the mouse brain using a high resolution photoacoustic scanner,”
Appl. Optics, 48
(10), D299
–D306
(2009). http://dx.doi.org/10.1364/AO.48.00D299 Google Scholar
J. Laufer et al.,
“In vivo preclinical photoacoustic imaging of tumor vasculature development and therapy,”
J. Biomed. Opt., 17
(5), 056016
(2012). http://dx.doi.org/10.1117/1.JBO.17.5.056016 JBOPFO 1083-3668 Google Scholar
J. Xia et al.,
“Wide-field two-dimensional multifocal optical-resolution photoacoustic-computed microscopy,”
Opt. Lett., 38
(24), 5236
–5239
(2013). http://dx.doi.org/10.1364/OL.38.005236 OPLEDP 0146-9592 Google Scholar
J. Laufer et al.,
“In vivo photoacoustic imaging of mouse embryos,”
J. Biomed. Opt., 17
(6), 061220
(2012). http://dx.doi.org/10.1117/1.JBO.17.6.061220 JBOPFO 1083-3668 Google Scholar
Y. Zhou, J. J. Yao and L. H. V. Wang,
“Optical clearing-aided photoacoustic microscopy with enhanced resolution and imaging depth,”
Opt. Lett., 38
(14), 2592
–2595
(2013). http://dx.doi.org/10.1364/OL.38.002592 OPLEDP 0146-9592 Google Scholar
Y. Liu, C. Zhang and L. H. V. Wang,
“Effects of light scattering on optical-resolution photoacoustic microscopy,”
J. Biomed. Opt., 17
(12), 126014
(2012). http://dx.doi.org/10.1117/1.JBO.17.12.126014 JBOPFO 1083-3668 Google Scholar
P. F. Hai et al.,
“Near-infrared optical-resolution photoacoustic microscopy,”
Opt. Lett., 39
(17), 5192
–5195
(2014). http://dx.doi.org/10.1364/OL.39.005192 OPLEDP 0146-9592 Google Scholar
S. Hu et al.,
“Label-free photoacoustic ophthalmic angiography,”
Opt. Lett., 35
(1), 1
–3
(2010). http://dx.doi.org/10.1364/OL.35.000001 OPLEDP 0146-9592 Google Scholar
N. Wu et al.,
“High-resolution dual-modality photoacoustic ocular imaging,”
Opt. Lett., 39
(8), 2451
–2454
(2014). http://dx.doi.org/10.1364/OL.39.002451 OPLEDP 0146-9592 Google Scholar
W. Song et al.,
“Integrating photoacoustic ophthalmoscopy with scanning laser ophthalmoscopy, optical coherence tomography, and fluorescein angiography for a multimodal retinal imaging platform,”
J. Biomed. Opt., 17
(6), 061206
(2012). http://dx.doi.org/10.1117/1.JBO.17.6.061206 JBOPFO 1083-3668 Google Scholar
X. J. Liu et al.,
“Optical coherence photoacoustic microscopy for in vivo multimodal retinal imaging,”
Opt. Lett., 40
(7), 1370
–1373
(2015). http://dx.doi.org/10.1364/OL.40.001370 OPLEDP 0146-9592 Google Scholar
J. J. Yao et al.,
“High-speed label-free functional photoacoustic microscopy of mouse brain in action,”
Nat. Methods, 12
(5), 407
–410
(2015). http://dx.doi.org/10.1038/nmeth.3336 Google Scholar
S. Hu et al.,
“Functional transcranial brain imaging by optical-resolution photoacoustic microscopy,”
J. Biomed. Opt., 14
(4), 040503
(2009). http://dx.doi.org/10.1117/1.3194136 JBOPFO 1083-3668 Google Scholar
Y. Zhou et al.,
“Microcirculatory changes identified by photoacoustic microscopy in patients with complex regional pain syndrome type I after stellate ganglion blocks,”
J. Biomed. Opt., 19
(8), 086017
(2014). http://dx.doi.org/10.1117/1.JBO.19.8.086017 JBOPFO 1083-3668 Google Scholar
H. F. Zhang et al.,
“In vivo volumetric imaging of subcutaneous microvasculature by photoacoustic microscopy,”
Opt. Express., 14
(20), 9317
–9323
(2006). http://dx.doi.org/10.1364/OE.14.009317 OPEXFF 1094-4087 Google Scholar
C. P. Favazza, L. A. Cornelius and L. H. V. Wang,
“In vivo functional photoacoustic microscopy of cutaneous microvasculature in human skin,”
J. Biomed. Opt., 16
(2), 026004
(2011). http://dx.doi.org/10.1117/1.3536522 JBOPFO 1083-3668 Google Scholar
C. P. Favazza et al.,
“In vivo photoacoustic microscopy of human cutaneous microvasculature and a nevus,”
J. Biomed. Opt., 16
(1), 016015
(2011). http://dx.doi.org/10.1117/1.3528661 JBOPFO 1083-3668 Google Scholar
C. Zhang, K. Maslov and L. H. V. Wang,
“Subwavelength-resolution label-free photoacoustic microscopy of optical absorption in vivo,”
Opt. Lett., 35
(19), 3195
–3197
(2010). http://dx.doi.org/10.1364/OL.35.003195 OPLEDP 0146-9592 Google Scholar
L. D. Wang et al.,
“Video-rate functional photoacoustic microscopy at depths,”
J. Biomed. Opt., 17
(10), 106007
(2012). http://dx.doi.org/10.1117/1.JBO.17.10.106007 JBOPFO 1083-3668 Google Scholar
H. X. Ke et al.,
“Performance characterization of an integrated ultrasound, photoacoustic, and thermoacoustic imaging system,”
J. Biomed. Opt., 17
(5), 056010
(2012). http://dx.doi.org/10.1117/1.JBO.17.5.056010 JBOPFO 1083-3668 Google Scholar
D. K. Yao et al.,
“In vivo label-free photoacoustic microscopy of cell nuclei by excitation of DNA and RNA,”
Opt. Lett., 35
(24), 4139
–4141
(2010). http://dx.doi.org/10.1364/OL.35.004139 OPLEDP 0146-9592 Google Scholar
D. K. Yao et al.,
“Optimal ultraviolet wavelength for in vivo photoacoustic imaging of cell nuclei,”
J. Biomed. Opt., 17
(5), 056004
(2012). http://dx.doi.org/10.1117/1.JBO.17.5.056004 JBOPFO 1083-3668 Google Scholar
C. Zhang et al.,
“Label-free photoacoustic microscopy of cytochromes,”
J. Biomed. Opt., 18
(2), 020504
(2013). http://dx.doi.org/10.1117/1.JBO.18.2.020504 JBOPFO 1083-3668 Google Scholar
Y. Zhou et al.,
“Photoacoustic microscopy of bilirubin in tissue phantoms,”
J. Biomed. Opt., 17
(12), 126019
(2012). http://dx.doi.org/10.1117/1.JBO.17.12.126019 JBOPFO 1083-3668 Google Scholar
C. Zhang et al.,
“Label-free photoacoustic microscopy of myocardial sheet architecture,”
J. Biomed. Opt., 17
(6), 060506
(2012). http://dx.doi.org/10.1117/1.JBO.17.6.060506 JBOPFO 1083-3668 Google Scholar
B. S. Goldschmidt et al.,
“Photoacoustic measurement of refractive index of dye solutions and myoglobin for biosensing applications,”
Biomed. Opt. Express., 4
(11), 2463
–2476
(2013). http://dx.doi.org/10.1364/BOE.4.002463 Google Scholar
H. F. Zhang et al.,
“Imaging of hemoglobin oxygen saturation variations in single vessels in vivo using photoacoustic microscopy,”
Appl. Phys. Lett., 90
(5), 053901
(2007). http://dx.doi.org/10.1063/1.2435697 Google Scholar
Y. Zhou et al.,
“Calibration-free absolute quantification of particle concentration by statistical analyses of photoacoustic signals in vivo,”
J. Biomed. Opt., 19
(3), 037001
(2014). http://dx.doi.org/10.1117/1.JBO.19.3.037001 JBOPFO 1083-3668 Google Scholar
M. Tang et al.,
“Noninvasive photoacoustic microscopy of methemoglobin in vivo,”
J. Biomed. Opt., 20
(3), 036007
(2015). http://dx.doi.org/10.1117/1.JBO.20.3.036007 JBOPFO 1083-3668 Google Scholar
Z. J. Chen, S. H. Yang and D. Xing,
“In vivo detection of hemoglobin oxygen saturation and carboxyhemoglobin saturation with multiwavelength photoacoustic microscopy,”
Opt. Lett., 37
(16), 3414
–3416
(2012). http://dx.doi.org/10.1364/OL.37.003414 OPLEDP 0146-9592 Google Scholar
J. Staley et al.,
“Growth of melanoma brain tumors monitored by photoacoustic microscopy,”
J. Biomed. Opt., 15
(4), 040510
(2010). http://dx.doi.org/10.1117/1.3478309 JBOPFO 1083-3668 Google Scholar
Y. Zhou et al.,
“Handheld photoacoustic microscopy to detect melanoma depth in vivo,”
Opt. Lett., 39
(16), 4731
–4734
(2014). http://dx.doi.org/10.1364/OL.39.004731 OPLEDP 0146-9592 Google Scholar
H. W. Wang et al.,
“Label-free bond-selective imaging by listening to vibrationally excited molecules,”
Phys. Rev. Lett., 106
(23), 238106
(2011). http://dx.doi.org/10.1103/PhysRevLett.106.238106 Google Scholar
V. V. Yakovlev et al.,
“Stimulated Raman photoacoustic imaging,”
Proc. Natl. Acad. Sci. U. S. A., 107
(47), 20335
–20339
(2010). http://dx.doi.org/10.1073/pnas.1012432107 Google Scholar
V. V. Yakovlev et al.,
“Monitoring stimulated Raman scattering with photoacoustic detection,”
Opt. Lett., 36
(7), 1233
–1235
(2011). http://dx.doi.org/10.1364/OL.36.001233 OPLEDP 0146-9592 Google Scholar
Z. Xu, Q. I. Zhu and L. H. V. Wang,
“In vivo photoacoustic tomography of mouse cerebral edema induced by cold injury,”
J. Biomed. Opt., 16
(6), 066020
(2011). http://dx.doi.org/10.1117/1.3584847 JBOPFO 1083-3668 Google Scholar
Z. Xu, C. H. Li and L. V. Wang,
“Photoacoustic tomography of water in phantoms and tissue,”
J. Biomed. Opt., 15
(3), 036019
(2010). http://dx.doi.org/10.1117/1.3443793 JBOPFO 1083-3668 Google Scholar
J. Kottmann et al.,
“Glucose sensing in human epidermis using mid-infrared photoacoustic detection,”
Biomed. Opt. Express, 3
(4), 667
–680
(2012). http://dx.doi.org/10.1364/BOE.3.000667 Google Scholar
M. A. Pleitez et al.,
“In vivo noninvasive monitoring of glucose concentration in human epidermis by mid-infrared pulsed photoacoustic spectroscopy,”
Anal. Chem., 85
(2), 1013
–1020
(2013). http://dx.doi.org/10.1021/ac302841f Google Scholar
P. Wang et al.,
“High-speed intravascular photoacoustic imaging of lipid-laden atherosclerotic plaque enabled by a 2-kHz barium nitrite Raman laser,”
Sci. Rep., 4
(2014). http://dx.doi.org/10.1038/srep06889 Google Scholar
B. Wang et al.,
“Detection of lipid in atherosclerotic vessels using ultrasound-guided spectroscopic intravascular photoacoustic imaging,”
Opt. Express., 18
(5), 4889
–4897
(2010). http://dx.doi.org/10.1364/OE.18.004889 OPEXFF 1094-4087 Google Scholar
Y. Wang and R. K. Wang,
“Photoacoustic recovery of an absolute optical absorption coefficient with an exact solution of a wave equation,”
Phys. Med. Biol., 53
(21), 6167
–6177
(2008). http://dx.doi.org/10.1088/0031-9155/53/21/018 Google Scholar
Z. J. Guo et al.,
“Quantitative photoacoustic microscopy of optical absorption coefficients from acoustic spectra in the optical diffusive regime,”
J. Biomed. Opt., 17
(6), 066011
(2012). http://dx.doi.org/10.1117/1.JBO.17.6.066011 JBOPFO 1083-3668 Google Scholar
J. Xia et al.,
“Calibration-free quantification of absolute oxygen saturation based on the dynamics of photoacoustic signals,”
Opt. Lett., 38
(15), 2800
–2803
(2013). http://dx.doi.org/10.1364/OL.38.002800 OPLEDP 0146-9592 Google Scholar
Y. Y. Petrov et al.,
“Multiwavelength optoacoustic system for noninvasive monitoring of cerebral venous oxygenation: a pilot clinical test in the internal jugular vein,”
Opt. Lett., 31
(12), 1827
–1829
(2006). http://dx.doi.org/10.1364/OL.31.001827 OPLEDP 0146-9592 Google Scholar
S. Zackrisson, S. M. W. Y. van de Ven and S. S. Gambhir,
“Light in and sound out: emerging translational strategies for photoacoustic imaging,”
Cancer Res., 74
(4), 979
–1004
(2014). http://dx.doi.org/10.1158/0008-5472.CAN-13-2387 Google Scholar
G. P. Luke, D. Yeager and S. Y. Emelianov,
“Biomedical applications of photoacoustic imaging with exogenous contrast agents,”
Ann. Biomed. Eng., 40
(2), 422
–437
(2012). http://dx.doi.org/10.1007/s10439-011-0449-4 Google Scholar
A. De La Zerda et al.,
“Carbon nanotubes as photoacoustic molecular imaging agents in living mice,”
Nat. Nanotechnol., 3
(9), 557
–562
(2008). http://dx.doi.org/10.1038/nnano.2008.231 Google Scholar
C. Kim et al.,
“In vivo photoacoustic mapping of lymphatic systems with plasmon-resonant nanostars,”
J. Mater. Chem., 21
(9), 2841
–2844
(2011). http://dx.doi.org/10.1039/c0jm04194g Google Scholar
C. Kim et al.,
“In vivo molecular photoacoustic tomography of melanomas targeted by bioconjugated gold nanocages,”
ACS Nano, 4
(8), 4559
–4564
(2010). http://dx.doi.org/10.1021/nn100736c ANCAC3 1936-0851 Google Scholar
D. P. J. Pan et al.,
“Photoacoustic sentinel lymph node imaging with self-assembled copper neodecanoate nanoparticles,”
ACS Nano, 6
(2), 1260
–1267
(2012). http://dx.doi.org/10.1021/nn203895n ANCAC3 1936-0851 Google Scholar
J. J. Yao et al.,
“Evans blue dye-enhanced capillary-resolution photoacoustic microscopy in vivo,”
J. Biomed. Opt., 14
(5), 054049
(2009). http://dx.doi.org/10.1117/1.3251044 JBOPFO 1083-3668 Google Scholar
S. Ashkenazi,
“Photoacoustic lifetime imaging of dissolved oxygen using methylene blue,”
J. Biomed. Opt., 15
(4), 040501
(2010). http://dx.doi.org/10.1117/1.3465548 JBOPFO 1083-3668 Google Scholar
X. D. Wang et al.,
“Noninvasive photoacoustic angiography of animal brains in vivo with near-infrared light and an optical contrast agent,”
Opt. Lett., 29
(7), 730
–732
(2004). http://dx.doi.org/10.1364/OL.29.000730 OPLEDP 0146-9592 Google Scholar
G. Ku and L. H. V. Wang,
“Deeply penetrating photoacoustic tomography in biological tissues enhanced with an optical contrast agent,”
Opt. Lett., 30
(5), 507
–509
(2005). http://dx.doi.org/10.1364/OL.30.000507 OPLEDP 0146-9592 Google Scholar
M. R. Chatni et al.,
“Functional photoacoustic microscopy of pH,”
J. Biomed. Opt., 16
(10), 100503
(2011). http://dx.doi.org/10.1117/1.3644495 JBOPFO 1083-3668 Google Scholar
K. H. Song et al.,
“Noninvasive photoacoustic identification of sentinel lymph nodes containing methylene blue in vivo in a rat model,”
J. Biomed. Opt., 13
(5), 054033
(2008). http://dx.doi.org/10.1117/1.2976427 JBOPFO 1083-3668 Google Scholar
L. Li et al.,
“Photoacoustic imaging of lacZ gene expression in vivo,”
J. Biomed. Opt., 12
(2), 020504
(2007). http://dx.doi.org/10.1117/1.2717531 JBOPFO 1083-3668 Google Scholar
X. Cai et al.,
“Multi-scale molecular photoacoustic tomography of gene expression,”
PLoS One, 7
(8), e43999
(2012). http://dx.doi.org/10.1371/journal.pone.0043999 POLNCL 1932-6203 Google Scholar
C. J. H. Ho et al.,
“Multifunctional photosensitizer-based contrast agents for photoacoustic imaging,”
Sci. Rep., 4
(2014). http://dx.doi.org/10.1038/srep05342 Google Scholar
A. B. Attia et al.,
“Phthalocyanine photosensitizers as contrast agents for in vivo photoacoustic tumor imaging,”
Biomed. Opt. Express, 6
(2), 591
–598
(2015). http://dx.doi.org/10.1364/BOE.6.000591 Google Scholar
E. Morgounova et al.,
“Photoacoustic lifetime contrast between methylene blue monomers and self-quenched dimers as a model for dual-labeled activatable probes,”
J. Biomed. Opt., 18
(5), 056004
(2013). http://dx.doi.org/10.1117/1.JBO.18.5.056004 JBOPFO 1083-3668 Google Scholar
J. Yao et al.,
“Multiscale photoacoustic tomography using reversibly switchable bacterial phytochrome as a near-infrared photochromic probe,”
Nat Methods,
(2015). http://dx.doi.org/10.1038/nmeth.3656 Google Scholar
A. C. Stiel et al.,
“High-contrast imaging of reversibly switchable fluorescent proteins via temporally unmixed multispectral optoacoustic tomography,”
Opt. Lett., 40
(3), 367
–370
(2015). http://dx.doi.org/10.1364/OL.40.000367 OPLEDP 0146-9592 Google Scholar
E. I. Galanzha et al.,
“Photoacoustic and photothermal cytometry using photoswitchable proteins and nanoparticles with ultrasharp resonances,”
J. Biophotonics, 8
(1–2), 81
–93
(2015). http://dx.doi.org/10.1002/jbio.201300140 Google Scholar
R. J. Zemp et al.,
“Realtime photoacoustic microscopy of murine cardiovascular dynamics,”
Opt. Express., 16
(22), 18551
–18556
(2008). http://dx.doi.org/10.1364/OE.16.018551 OPEXFF 1094-4087 Google Scholar
A. Taruttis et al.,
“Real-time imaging of cardiovascular dynamics and circulating gold nanorods with multispectral optoacoustic tomography,”
Opt. Express., 18
(19), 19592
–19602
(2010). http://dx.doi.org/10.1364/OE.18.019592 OPEXFF 1094-4087 Google Scholar
L. A. Song et al.,
“Ultrasound-array-based real-time photoacoustic microscopy of human pulsatile dynamics in vivo,”
J. Biomed. Opt., 15
(2), 021303
(2010). http://dx.doi.org/10.1117/1.3333545 JBOPFO 1083-3668 Google Scholar
K. Aizawa et al.,
“Photoacoustic monitoring of burn healing process in rats,”
J. Biomed. Opt., 13
(6), 064020
(2008). http://dx.doi.org/10.1117/1.3028005 JBOPFO 1083-3668 Google Scholar
S. Sato et al.,
“Photoacoustic diagnosis of burns in rats,”
J. Trauma, 59
(6), 1450
–1455
(2005). http://dx.doi.org/10.1097/01.ta.0000197389.94466.04 Google Scholar
H. F. Zhang et al.,
“Imaging acute thermal burns by photoacoustic microscopy,”
J. Biomed. Opt., 11
(5), 054033
(2006). http://dx.doi.org/10.1117/1.2355667 JBOPFO 1083-3668 Google Scholar
A. de la Zerda et al.,
“Ultrahigh sensitivity carbon nanotube agents for photoacoustic molecular imaging in living mice,”
Nano Lett., 10
(6), 2168
–2172
(2010). http://dx.doi.org/10.1021/nl100890d NALEFD 1530-6984 Google Scholar
M. L. Li et al.,
“In-vivo photoacoustic microscopy of nanoshell extravasation from solid tumor vasculature,”
J. Biomed. Opt., 14
(1), 010507
(2009). http://dx.doi.org/10.1117/1.3081556 JBOPFO 1083-3668 Google Scholar
J. Kang et al.,
“Photoacoustic imaging of breast microcalcifications: a validation study with 3-dimensional ex vivo data and spectrophotometric measurement,”
J. Biophotonics, 8
(1–2), 71
–80
(2015). http://dx.doi.org/10.1002/jbio.201300100 Google Scholar
C. Sim et al.,
“Photoacoustic-based nanomedicine for cancer diagnosis and therapy,”
J. Control Release, 203 118
–125
(2015). http://dx.doi.org/10.1016/j.jconrel.2015.02.020 Google Scholar
S. Y. Nam and S. Y. Emelianov,
“Array-based real-time ultrasound and photoacoustic ocular imaging,”
J. Opt. Soc. Korea, 18
(2), 151
–155
(2014). http://dx.doi.org/10.3807/JOSK.2014.18.2.151 Google Scholar
W. Song et al.,
“Multimodal photoacoustic ophthalmoscopy in mouse,”
J. Biophotonics, 6
(6–7), 505
–512
(2013). http://dx.doi.org/10.1002/jbio.v6.6/7 Google Scholar
T. Liu et al.,
“Fundus camera guided photoacoustic ophthalmoscopy,”
Curr. Eye Res., 38
(12), 1229
–1234
(2013). http://dx.doi.org/10.3109/02713683.2013.815219 Google Scholar
J. M. Yang et al.,
“Three-dimensional photoacoustic endoscopic imaging of the rabbit esophagus,”
PLoS One, 10
(4),
(2015). http://dx.doi.org/10.1371/journal.pone.0120269 POLNCL 1932-6203 Google Scholar
P. Hajireza, W. Shi and R. Zemp,
“Label-free in vivo GRIN-lens optical resolution photoacoustic micro-endoscopy,”
Laser Phys. Lett., 10
(5), 055603
(2013). http://dx.doi.org/10.1088/1612-2011/10/5/055603 Google Scholar
P. Hajireza et al.,
“Optical resolution photoacoustic microendoscopy with ultrasound-guided insertion and array system detection,”
J. Biomed. Opt., 18
(9), 090502
(2013). http://dx.doi.org/10.1117/1.JBO.18.9.090502 JBOPFO 1083-3668 Google Scholar
J. M. Yang et al.,
“A 2.5-mm diameter probe for photoacoustic and ultrasonic endoscopy,”
Opt. Express, 20
(21), 23944
–23953
(2012). http://dx.doi.org/10.1364/OE.20.023944 OPEXFF 1094-4087 Google Scholar
M. Kneipp et al.,
“Functional real-time optoacoustic imaging of middle cerebral artery occlusion in mice,”
PLoS One, 9
(4),
(2014). http://dx.doi.org/10.1371/journal.pone.0096118 POLNCL 1932-6203 Google Scholar
X. S. Bai et al.,
“Intravascular optical-resolution photoacoustic tomography with a 1.1 mm diameter catheter,”
PLoS One, 9
(3),
(2014). http://dx.doi.org/10.1371/journal.pone.0092463 POLNCL 1932-6203 Google Scholar
E. I. Galanzha et al.,
“In vivo flow cytometry of circulating clots using negative photothermal and photoacoustic contrasts,”
Cytom Part A, 79A
(10), 814
–824
(2011). http://dx.doi.org/10.1002/cyto.a.v79a.10 Google Scholar
L. D. Liao et al.,
“Transcranial imaging of functional cerebral hemodynamic changes in single blood vessels using in vivo photoacoustic microscopy,”
J. Cereb. Blood Flow Metab., 32
(6), 938
–951
(2012). http://dx.doi.org/10.1038/jcbfm.2012.42 Google Scholar
L. D. Liao et al.,
“Imaging brain hemodynamic changes during rat forepaw electrical stimulation using functional photoacoustic microscopy,”
NeuroImage, 52
(2), 562
–570
(2010). http://dx.doi.org/10.1016/j.neuroimage.2010.03.065 NEIMEF 1053-8119 Google Scholar
J. J. Yao et al.,
“Noninvasive photoacoustic computed tomography of mouse brain metabolism in vivo,”
NeuroImage, 64 257
–266
(2013). http://dx.doi.org/10.1016/j.neuroimage.2012.08.054 NEIMEF 1053-8119 Google Scholar
V. Tsytsarev et al.,
“Photoacoustic microscopy of microvascular responses to cortical electrical stimulation,”
J. Biomed. Opt., 16
(7), 076002
(2011). http://dx.doi.org/10.1117/1.3594785 JBOPFO 1083-3668 Google Scholar
M. Heijblom, W. Steenbergen and S. Manohar,
“Clinical photoacoustic breast imaging,”
IEEE Pulse, 6
(3), 42
–46
(2015). http://dx.doi.org/10.1109/MPUL.2015.2409102 Google Scholar
T. Gould, Q. Z. Wang and T. J. Pfefer,
“Optical-thermal light-tissue interactions during photoacoustic breast imaging,”
Biomed. Opt. Express, 5
(3), 832
–847
(2014). http://dx.doi.org/10.1364/BOE.5.000832 Google Scholar
W. F. Xia et al.,
“An optimized ultrasound detector for photoacoustic breast tomography,”
Med. Phys., 40
(3), 032901
(2013). http://dx.doi.org/10.1118/1.4792462 Google Scholar
L. Lin et al.,
“In vivo deep brain imaging of rats using oral-cavity illuminated photoacoustic computed tomography,”
J. Biomed. Opt., 20
(1), 016019
(2015). http://dx.doi.org/10.1117/1.JBO.20.1.016019 JBOPFO 1083-3668 Google Scholar
M. Nasiriavanaki et al.,
“High-resolution photoacoustic tomography of resting-state functional connectivity in the mouse brain,”
Proc. Natl. Acad. Sci. U. S. A., 111
(1), 21
–26
(2014). http://dx.doi.org/10.1073/pnas.1311868111 Google Scholar
R. Y. Zhang et al.,
“In vivo optically encoded photoacoustic flowgraphy,”
Opt. Lett., 39
(13), 3814
–3817
(2014). http://dx.doi.org/10.1364/OL.39.003814 OPLEDP 0146-9592 Google Scholar
C. Zhang et al.,
“Slow-sound photoacoustic microscopy,”
Appl. Phys. Lett., 102
(16),
(2013). http://dx.doi.org/10.1063/1.4803444 Google Scholar
C. Zhang et al.,
“Reflection-mode submicron-resolution in vivo photoacoustic microscopy,”
J. Biomed. Opt., 17
(2), 020501
(2012). http://dx.doi.org/10.1117/1.JBO.17.2.020501 JBOPFO 1083-3668 Google Scholar
J. J. Yao et al.,
“Double-illumination photoacoustic microscopy,”
Opt. Lett., 37
(4), 659
–661
(2012). http://dx.doi.org/10.1364/OL.37.000659 OPLEDP 0146-9592 Google Scholar
X. T. Zou et al.,
“Polydimethylsiloxane thin film characterization using all-optical photoacoustic mechanism,”
Appl. Opt., 52
(25), 6239
–6244
(2013). http://dx.doi.org/10.1364/AO.52.006239 Google Scholar
E. Z. Zhang et al.,
“In vivo high-resolution 3D photoacoustic imaging of superficial vascular anatomy,”
Phys. Med. Biol., 54
(4), 1035
–1046
(2009). http://dx.doi.org/10.1088/0031-9155/54/4/014 Google Scholar
Y. Zhou et al.,
“Calibration-free in vivo transverse blood flowmetry based on cross correlation of slow time profiles from photoacoustic microscopy,”
Opt. Lett., 38
(19), 3882
–3885
(2013). http://dx.doi.org/10.1364/OL.38.003882 OPLEDP 0146-9592 Google Scholar
J. J. Yao et al.,
“Absolute photoacoustic thermometry in deep tissue,”
Opt. Lett., 38
(24), 5228
–5231
(2013). http://dx.doi.org/10.1364/OL.38.005228 OPLEDP 0146-9592 Google Scholar
J. Y. Liang et al.,
“Cross-correlation-based transverse flow measurements using optical resolution photoacoustic microscopy with a digital micromirror device,”
J. Biomed. Opt., 18
(9), 096004
(2013). http://dx.doi.org/10.1117/1.JBO.18.9.096004 JBOPFO 1083-3668 Google Scholar
J. J. Yao et al.,
“Label-free oxygen-metabolic photoacoustic microscopy in vivo,”
J. Biomed. Opt., 16
(7), 076003
(2011). http://dx.doi.org/10.1117/1.3594786 JBOPFO 1083-3668 Google Scholar
B. Ning et al.,
“Ultrasound-aided multi-parametric photoacoustic microscopy of the mouse brain,”
Sci. Rep., 5 18775
(2015). http://dx.doi.org/10.1038/srep18775 Google Scholar
G. Li et al.,
“Multiview Hilbert transformation for full-view photoacoustic computed tomography using a linear array,”
J. Biomed. Opt., 20
(6), 066010
(2015). http://dx.doi.org/10.1117/1.JBO.20.6.066010 JBOPFO 1083-3668 Google Scholar
J. Xia et al.,
“Retrospective respiration-gated whole-body photoacoustic computed tomography of mice,”
J. Biomed. Opt., 19
(1), 016003
(2014). http://dx.doi.org/10.1117/1.JBO.19.1.016003 JBOPFO 1083-3668 Google Scholar
S. N. Hennen et al.,
“Photoacoustic tomography imaging and estimation of oxygen saturation of hemoglobin in ocular tissue of rabbits,”
Exp. Eye Res., 138 153
–158
(2015). http://dx.doi.org/10.1016/j.exer.2015.05.022 EXERA6 0014-4835 Google Scholar
L. D. Wang, K. Maslov and L. H. V. Wang,
“Single-cell label-free photoacoustic flowoxigraphy in vivo,”
Proc. Natl. Acad. Sci. U. S. A., 110
(15), 5759
–5764
(2013). http://dx.doi.org/10.1073/pnas.1215578110 Google Scholar
A. Garcia-Uribe et al.,
“Dual-modality photoacoustic and ultrasound imaging system for noninvasive sentinel lymph node detection in patients with breast cancer,”
Sci. Rep., 5 15748
(2015). http://dx.doi.org/10.1038/srep15748 Google Scholar
A. Danielli et al.,
“Label-free photoacoustic nanoscopy,”
J. Biomed. Opt., 19
(8), 086006
(2014). http://dx.doi.org/10.1117/1.JBO.19.8.086006 JBOPFO 1083-3668 Google Scholar
Y. Zhou, J. Liang and L. V. Wang,
“Cuffing-based photoacoustic flowmetry in humans in the optical diffusive regime,”
J. Biophotonics,
(2015). http://dx.doi.org/10.1002/jbio.201500181 Google Scholar
Y. Zhou et al.,
“In vivo photoacoustic flowmetry at depths of the diffusive regime based on saline injection,”
J. Biomed. Opt., 20
(8), 087001
(2015). http://dx.doi.org/10.1117/1.JBO.20.8.087001 Google Scholar
C. H. Yeh et al.,
“Photoacoustic microscopy of blood pulse wave,”
J. Biomed. Opt., 17
(7), 070504
(2012). http://dx.doi.org/10.1117/1.JBO.17.7.070504 JBOPFO 1083-3668 Google Scholar
X. H. Zhu et al.,
“Simultaneous and noninvasive imaging of cerebral oxygen metabolic rate, blood flow and oxygen extraction fraction in stroke mice,”
NeuroImage, 64 437
–447
(2013). http://dx.doi.org/10.1016/j.neuroimage.2012.09.028 NEIMEF 1053-8119 Google Scholar
A. Krumholz et al.,
“Functional photoacoustic microscopy of diabetic vasculature,”
J. Biomed. Opt., 17
(6), 060502
(2012). http://dx.doi.org/10.1117/1.JBO.17.6.060502 JBOPFO 1083-3668 Google Scholar
D. Yeager et al.,
“Intravascular photoacoustic imaging of exogenously labeled atherosclerotic plaque through luminal blood,”
J. Biomed. Opt., 17
(10), 106016
(2012). http://dx.doi.org/10.1117/1.JBO.17.10.106016 JBOPFO 1083-3668 Google Scholar
M. Wu et al.,
“Impact of device geometry on the imaging characteristics of an intravascular photoacoustic catheter,”
Appl. Opt., 53
(34), 8131
–8139
(2014). http://dx.doi.org/10.1364/AO.53.008131 Google Scholar
B. Wang et al.,
“Photoacoustic tomography system for noninvasive real-time three-dimensional imaging of epilepsy,”
Biomed. Opt. Express, 3
(6), 1427
–1432
(2012). http://dx.doi.org/10.1364/BOE.3.001427 Google Scholar
BiographyYong Zhou is currently a graduate student in biomedical engineering at Washington University in St. Louis, under the supervision of Dr. Lihong V. Wang, Gene K. Beare Distinguished Professor. His research focuses on the development of photoacoustic imaging systems. Junjie Yao received his BE and ME degrees in biomedical engineering from Tsinghua University, Beijing, in 2006 and 2008, respectively, under the tutelage of Prof. Jing Bai. He received his PhD in biomedical engineering at Washington University in St. Louis (WUSTL), in 2013, under the tutelage of Prof. Lihong V. Wang. He is currently a postdoctoral research associate at WUSTL. His research interest is in photoacoustic, optical, and ultrasound imaging technologies in biomedicine. Lihong V. Wang is the Beare distinguished professor at Washington University, has published 425 journal articles (h-index = 96, citations >36,000) and delivered 420 keynote/plenary/invited talks. His laboratory published the first functional photoacoustic CT and 3-D photoacoustic microscopy. He received the Goodman Award for his Biomedical Optics textbook, NIH Director’s Pioneer Award, OSA Mees Medal, IEEE Technical Achievement and Biomedical Engineering Awards, SPIE Britton Chance Biomedical Optics Award, and an honorary doctorate from Lund University, Sweden. |