13 January 2020 Analysis of tissue microstructure with Mueller microscopy: logarithmic decomposition and Monte Carlo modeling
Author Affiliations +
Abstract

Significance: Definitive diagnostics of many diseases is based on the histological analysis of thin tissue cuts with optical white light microscopy. Extra information on tissue structural properties obtained with polarized light would help the pathologist to improve the accuracy of his diagnosis.

Aim: We report on using Mueller matrix microscopy data, logarithmic decomposition, and polarized Monte Carlo (MC) modeling for qualitative and quantitative analysis of thin tissue cuts to extract the information on tissue microstructure that is not available with a conventional white light microscopy.

Approach: Unstained cuts of human skin equivalents were measured with a custom-built liquid-crystal-based Mueller microscope in transmission configuration. To interpret experimental data, we performed the simulations with a polarized MC algorithm for scattering anisotropic media. Several optical models of tissue (spherical scatterers within birefringent host medium, and combination of spherical and cylindrical scatterers within either isotropic or birefringent host medium) were tested.

Results: A set of rotation invariants for the logarithmic decomposition of a Mueller matrix was derived to rule out the impact of sample orientation. These invariants were calculated for both simulated and measured Mueller matrices of the dermal layer of skin equivalents. We demonstrated that only the simulations with a model combining both spherical and cylindrical scatterers within birefringent host medium reproduced the experimental trends in optical properties of the dermal layer (linear retardance, linear dichroism, and anisotropic linear depolarization) with layer thickness.

Conclusions: Our studies prove that Mueller polarimetry provides relevant information not only on a size of dominant scatterers (e.g., cell nuclei versus subwavelength organelles) but also on its shape (e.g., cells versus collagen fibers). The latter is directly related to the state of extracellular collagen matrix, which is often affected by early pathology. Hence, using polarimetric data can help to increase the accuracy of diagnosis.

1.

Introduction

Probing the tissue with polarized light has been proven to be an efficient approach for fast noninvasive tissue diagnostics. Nevertheless, the interpretation of optical data and analytical prediction of the results of polarized light interaction with a biological tissue is not straightforward, as biological tissues are extremely complex objects containing a large number of randomly distributed multidisperse microscopic scatterers and anisotropic microstructures.1 It has been shown that Mueller matrix formalism is a powerful tool in the study of biological samples.24 This phenomenological approach describes the interaction of polarized light with a sample using a model of “equivalent optical circuit” built from the basic optical elements—diatenuators, retarders, and depolarizers. The extended toolkit of various Mueller matrix decompositions (data processing algorithms exploring nonlinear compression of a set of real values of 4×4 Mueller matrices) is available for data analysis and characterization of tissue optical properties.510 We have already proved the validity of the differential Mueller matrix formalism9,10 using Mueller transmission microscopy measurements of the dermal layer of histological cuts of skin equivalents,11,12 which were produced in vitro from human cells and accurately reflected the anatomy of human skin. Skin tissue models are widely employed as an alternative to animal models or human donor tissue. Since human skin equivalents can be produced with less variability compared with real human skin, these tissue models were chosen for our studies.

In this paper, we present the results of numerical modeling of optical properties of the dermal skin layer using various optical models and validate them with the experimental data of Lee et al.12 We use the Monte Carlo (MC) statistical algorithm for the solution of a vector radiative transfer equation and an appropriate optical model of tissue to simulate the propagation of polarized light through the biological sample. The polarized MC software for the modeling of scattering of polarized light on spherical and/or infinite-long cylindrical scatterers that are randomly distributed within either isotropic or uniaxial linear birefringent host medium was developed in prior studies.13,14 Each type of scatterer represents the different tissue components and microstructures (e.g., cell nuclei, cell organelles, extracellular collagen matrix, etc.). The optical model including both spheres and cylinders in birefringent medium [sphere-cylinder-birefringence (SCB) model] was previously tested on experimental data acquired from various biological samples.15,16 The SCB model predicted well the experimental trends in optical parameters extracted from the measured Mueller matrices.

It is known that values of some Mueller matrix elements and parameters extracted from Mueller matrix decomposition can be affected by the spatial orientation of a sample.6 Previously, a set of rotation invariants for Mueller matrix elements was proposed.17 In this study, we applied a logarithmic decomposition9 of both measured and simulated Mueller matrices and derived a set of invariant logarithmic Mueller matrix decomposition (LMMD) parameters to exclude the impact of sample orientation.

We have tested several optical models to reproduce the experimental dependence of the derived invariant parameters on tissue thickness. With our simulations, we ruled out the optical models of skin tissue that are based on (i) spheres and cylinders distributed in isotropic host medium [sphere-cylinder (SC) model] and (ii) spheres distributed in birefringent host medium model [sphere-birefringence (SB) model]. We found that only the optical model of the dermal layer, which includes both spheres and cylinders distributed in the birefringent host medium (SCB model), can reproduce the effects of anisotropic linear depolarization, linear dichroism, and linear retardance observed in our experiments. Moreover, we have confirmed that measured anisotropy of linear depolarization is a real effect that does not depend on sample orientation.

The trends in polarization and depolarization properties shown by the simulations qualitatively agree with the experiment results, thus, paving the way for use in optical biopsy, i.e., an understanding of the microstructure of biological samples from the polarimetric measurements.

2.

Model Description

2.1.

Monte Carlo Algorithm

The detailed description of the polarized MC algorithm and its implementation can be found in prior publications.13,14 Here, we recall the main steps of the statistical modeling approach while omitting some details for brevity. A point light source, placed at a fixed position in space, emits a given number (107 to 108) of monoenergetic photons with the preset states of polarization. Those photons impinge a top surface of the sample at a given incident angle. Every photon travels a certain distance within the sample before being scattered on a sphere or cylinder. For each collision event, this distance is determined statistically using a mean free path parameter calculated from the scattering cross-sections of scatterers and their number density. The sizes, refractive indices, and number densities of sphere and cylinder scatterers as well as refractive index of isotropic host medium or ordinary and extraordinary refractive indices and spatial orientation of the optical axis of linear birefringent host medium are the input parameters of the optical model and can be adjusted to mimic the conditions of a real biological sample.14 The photon changes its polarization state and direction of propagation after each scattering event. The angles of deflection and rotation of polarization plane are calculated using the rejection method.18 A transfer matrix for scattering is determined by Mie theory for spherical scatterers or scattering matrix theory for infinite-long cylinder.19 The host medium may also be absorbing (not implemented in this study). The random walk of a photon continues within a scattering medium until it is either absorbed within the sample or moved outside the sample volume, where it can be lost or hit a detector. The first version of the software was developed on C language and run on a CPU platform.13 The latest version of the software is accelerated using a GPU platform.20

2.2.

Optical Models of Skin Dermal Layer

To choose an appropriate optical model of a dermal layer of skin tissue cuts, one needs to account for both fibroblasts and well-aligned collagen fibers, which form the dermal equivalent of a skin model.21,22 While light scattering on cells and fibers produces depolarization, the optical anisotropy of the dermal layer results in retardance due to form birefringence.1 Thus, we used the monodisperse spherical scatterers in the optical model of the dermal layer to reproduce an isotropic scattering on cells. Infinitely long cylindrical scatterers were added to the optical model to simulate the effect of form birefringence due to the presence of aligned collagen fibers in dermis. The refractive indices of spherical and cylindrical scatterers (ns, nc) and isotropic medium (nm) were set to 1.45 and 1.33, respectively.

We also explored the validity of replacing a form birefringence by an intrinsic birefringence of a uniaxial linear anisotropic host medium with in-plane optical axis, ordinary index nmo=1.33, and extraordinary index nme=1.33+Δn, varying parameter Δn from 105 to 103. We took the values of refractive indices as for bulk fresh tissue,1 while noting that those values may be somewhat different for the studied fixed unstained tissue cuts. This is most probably not so important for our consideration because the refractive index of a scatterer and its size are highly correlated parameters in the Mie electromagnetic scattering problem. With our choice of refractive index values for both scatterers and host medium, the optical contrast ns/nm (or nc/nm) is more than 1. Keeping constant the value of optical contrast, we varied the size of scatterers to reproduce the general trends in polarization and depolarization parameters in our simulations. We believe this is a reasonable assumption for performing the parametric numerical studies to reproduce the experimental trends.

In our experiments, a dermal layer of all skin model cuts demonstrated higher circular depolarization compared with the linear one (|α44|>|α22|, |α44|>|α33|). This suggests the dominance of the Rayleigh scattering regime over the Mie scattering regime23 and justifies the use of subwavelength spherical and cylindrical scatterers in an optical model of the dermal layer. The wavelength of probing light was fixed at 0.533  μm, so we tested spherical and cylindrical scatterers whose diameter ranged from 0.01 to 0.5  μm. Their concentrations cs and cc were described by the scattering coefficients μs and μc [μs,c=1/(cs,cσs,c), where σs,c is the scattering cross-sections of sphere and cylinder, respectively), which were varied from 5 to 5000  cm1. The parameter Δn for uniaxial birefringent host medium was adjusted to fit the experimental results12 for total linear retardance parameter RT (see Sec. 3.1). The optical axis of linear birefringent host medium was always oriented parallel to the sample surface, reflecting the arrangement of collagen fibers in a dermal layer of histological cuts. The GPU acceleration allowed us to carry out the simulations in a wide range of parameters to find the best-fit values.

Histological cuts of skin tissue models of varying thicknesses (nominal values 3 to 30  μm) were mounted on 1-mm-thick microscopy glass slides in our experiments.12 During the calibration of the Mueller polarimetric microscope with the eigenvalue calibration method,24 the measurements of air (one of the reference samples) were performed through a bare microscopy glass slide. Hence, the contribution of glass was excluded from Mueller matrices of all measured histological cuts. To model our experimental setup, we performed MC simulations in transmission configuration for the range of histological cut thicknesses defined by profilometer measurements11 without adding a 1-mm-thick glass layer to our optical model. A spatially uniform light beam was normally incident onto the flat front surface of a sample. No back surface roughness was taken into account in our optical model. The simulated images of forward scattering Mueller matrix elements were spatially averaged over a central circle of 600  μm in diameter to reproduce the experimental conditions, and the resulting Mueller matrices were decomposed using the LMMD method.9

3.

Results and Discussion

3.1.

Rotation Invariants of Logarithmic Decomposition

The set of polarization and depolarization parameters obtained from the LMMD includes the values of linear (LB,LB) and circular (CB) retardance, linear (LD,LD) and circular (CD) dichroism, and linear (α22, α33) and circular (α44) depolarization coefficients.9 The parameters LB, CD, and α22 are defined with respect to the framework of 0 deg to 90 deg (linear polarization); the parameters LB, LD, and α33 are defined with respect to the framework of ±45  deg (linear polarization).

Neither optical activity nor circular dichroism was detected in the polarimetric measurement data for skin model histological cuts (CB=0, CD=0). The presence of well-aligned collagen fibers in a dermal layer of skin model cuts indicated the direction of the optical axis of uniaxial linear birefringent medium. In our experiments,11,12 the orientation of histological cuts in a sample holder plane was performed manually, thus, producing an uncertainty in azimuth of the optical axis from sample to sample. Therefore, the measured values of polarization parameters LB, LB for skin model cuts of different thicknesses were affected not only by the different optical paths of the detected photons but also by the in-plane rotation of the samples. The experimental results for histological cuts of skin models have also demonstrated the existence of anisotropy of linear depolarization (α22α33).11,12 Both parameters α22 and α33 are not invariant under the in-plane rotation of an anisotropic sample. Therefore, we also derived a rotation-invariant parameter for the linear retardance and linear depolarization anisotropy parameters to eliminate the impact of sample orientation with respect to the laboratory coordinate frame.

The logarithm of Mueller matrix M is calculated from the eigenvalue decomposition as

Eq. (1)

L=lnM=ln(UΛU1)=Uln(Λ)U1,
where Λ is a diagonal matrix of eigenvalues of M and U is a matrix with the columns—eigenvectors of matrix M. The rotational transformation of a Mueller matrix in transmission configuration is described by M=R(α)MR(α), where

Eq. (2)

R(α)=[10000cos(2α)sin(2α)00sin(2α)cos(2α)00001].
Rotational transformation does not affect the eigenvalues of the matrix; therefore, we have

Eq. (3)

L=lnM=ln(R(α)UΛU1R(α))=R(α)Uln(Λ)U1R(α)=R(α)LR(α),
which means that the rotation transformation of matrix L is the same as for Mueller matrix M. As a result, the rotation invariants of matrix L should take the same form as the invariants for Mueller matrix M.17 If we denote sn=sin(nα) and cn=cos(nα) and decompose the matrix R(α)LR(α) into the sum of polarization and depolarization matrices Lm and Lu (G-antisymmetric and G-symmetric components9), we get the following expressions:

Eq. (4)

Lm=12[0(L12+L21)c2(L13+L31)c2L14+L41(L13+L31)s2+(L12+L21)s2(L12+L21)c20L23L32(L24L42)c2(L13+L31)s2+(L43L34)s2(L13+L31)c2L32L230(L34L43)c2+(L12+L21)s2+(L24L42)s2L14+L41(L42L24)c2(L43L34)c20+(L34L43)s2+(L42L24)s2],

Eq. (5)

Lu=12[2L11(L12L21)c2(L13L31)c2L14L41+(L31L13)s2+(L12L21)s2(L21L12)c2L22+L33+(L22L33)c4(L23+L32)c4(L24+L42)c2+(L13L31)s2(L23+L32)s4+(L22L33)s4(L43+L34)s2(L31L13)c2(L23+L32)c4L22+L33+(L33L22)c4(L34+L43)c2+(L21L12)s2+(L22L33)s4+(L23+L32)s4+(L24+L42)s2L41L14(L42+L24)c2(L43+L34)c22L44(L34+L43)s2+(L42+L24)s2]
The rotation invariants of the matrix Lm are total linear birefringence

Eq. (6)

RT=Lm422+Lm432=Lm242+Lm342=12[(L24L42)2+(L34L43)2],
total linear dichroism

Eq. (7)

DT=Lm122+Lm132=Lm212+Lm312=12[(L12+L21)2+(L13+L31)2],
circular birefringence

Eq. (8)

RC=Lm23=Lm32=(L23L32)/2,
and circular dichroism

Eq. (9)

DC=Lm14=Lm41=(L14+L41)/2.
Using the notation αii=Luii, (i=2,3,4) for the diagonal elements of matrix Lu, the rotation invariants of the matrix Lu can be written in terms of linear (isotropic) depolarization

Eq. (10)

αL=(α22+α33)/2=(L22+L33)/2,
and circular depolarization

Eq. (11)

α44=L44.
The four elements at the corners of the matrix L are also invariant under rotation, as well as the squared sum of the matrix elements from the first and last columns and the first and last rows. To find the rotation-invariant parameter for a linear anisotropic depolarization, we applied to matrix L the same procedure that was applied to matrix M in Mueller matrix transformation theory17 for defining a degree of anisotropy parameter t1=12(m22m33)2+(m23+m32)2 and obtained

Eq. (12)

αLA=12(α22α33)2+(Lu23+Lu32)2.
In our optical model of tissue, the orientation of the optical axis of birefringence and the orientation of the axis of cylindrical scatterers were always along the X axis. In the experiments, the orientation of the aligned collagen fibers with respect to the edge of the microscope glass slide (i.e., laboratory X axis) depends on a sample preparation and may slightly vary from one histological cut to another. Therefore, we used the derived set of rotation invariants with nonzero values, namely, RT, DT αL, α44, and αLA for the comparison of the results of measurements and simulations.

3.2.

Choice of an Appropriate Optical Model of Dermis

Several optical models were tested to reproduce the optical effects observed in a dermal zone of samples. Some models (e.g., the SC model) were ruled out for the reasons discussed in Sec. 3.2.1. Finally, the SCB model was adopted in these studies, and the results are presented in Sec. 3.2.2.

3.2.1.

Ruled out optical models

Experimental studies of tissue phantoms25 and tissue models12 with transmission Mueller polarimetry had confirmed that retardance values calculated using LMMD depend linearly on thickness, while the depolarization parameters α22, α33, and α44, calculated using LMMD show quadratic dependence on layer thickness. To reproduce the experimental values of polarization and depolarization parameters of a dermal layer of histological cuts measured with the Mueller microscope in transmission configuration,12 we tested the SC, SB, and SCB optical models.

It was already demonstrated that at normal incidence an isotropic medium with spherical scatterers does not exhibit any retardance effect.26,27 A phase shift in the detected signal can be induced by scattering of polarized light by cylindrical scatterers as well as by light passing through a birefringent medium. Our modeling results demonstrated that the SC model with isotropic host medium fell short of reproducing the experimental values of retardance for low values of scattering coefficient μc. The volume density (or concentration) of cylindrical scatterers had to increase significantly to fit the experimental trends in retardance, but with increase of parameter μc this model produced very high values of dichroism and depolarization which by far exceed the corresponding experimental values. Hence, we concluded that a uniaxial linear birefringent host medium is a necessary component of our optical model. It will increase the simulated retardance values without pushing up the dichroism and depolarization parameters of a simulated medium. Therefore, we ruled out the SC model for further consideration.

We also examined the SB optical model of the skin dermal layer. The values of Δn parameter, radius of spherical scatterer Rs, and scattering coefficient μs were varied to find the best-fit to the experimental data. The SB optical model fits well the experimental values of retardance11 with optimal values of Δn=0.009 and Rs=0.05  μm [see Fig. 1(a)]. It is worth mentioning that the optimal value of Δn for the fixed tissue cuts was found to be about 2 orders of magnitude larger compared with the values reported for the fresh biological tissue.1 The simulation results with the SB optical model confirmed the linear dependence of total linear retardance RT (calculated from LMMD) on layer thickness. The simulated values of depolarization parameters αL, α44 have demonstrated a quadratic dependence on thickness within the experimental range of errors [see Figs. 1(b) and 1(c)]. No anisotropy of linear depolarization was observed with the SB model, as simulated values of αLA=0 for all layer thickness [Fig. 1(d)].

Fig. 1

Results of MC simulations with the SB optical model for different layer thicknesses: (a) total linear retardance RT (radians) and dimensionless depolarization parameters (b) αL, (c) α44, and (d) αLA. Simulated data are shown by open symbols corresponding to different scattering coefficients (i.e., different concentrations of spherical scatterers). Open boxes with error bars represent the experimental data. Solid lines show the results of (a), (d) linear and (b), (c) quadratic fit of the simulated data.

JBO_25_1_015002_f001.png

The experimental values of depolarization coefficients α22, α33, and α44 for a dermal layer of histological cuts obey the relation |α22|<|α33|<|α44|.12 However, the simulations with the SB model could not reproduce the effect of anisotropy of linear depolarization (|αLA|0) observed experimentally [see Fig. 1(d)]. Moreover, no linear dichroism can be simulated with the SB model, while the nonzero values of the linear dichroism were measured in our experiments.12 Therefore, we conclude that the optical SB model of the dermal layer of skin model histological cuts has to be modified to reproduce experimental trends.

3.2.2.

SCB model

We then added cylindrical scatterers to the optical model of dermis to simulate the effects of anisotropy of linear depolarization and linear dichroism. First, we used the same set of parameter values as for the SB optical model, but added a group of cylindrical scatterers of radius Rc=0.05  μm=Rs, aligned along the X axis. The scattering coefficient for spherical scatterers μs was fixed at 1500  cm1 and scattering coefficient for cylindrical scatterers μc was varied from 500 to 1500  cm1.

The results obtained with the SCB optical model after applying LMMD to the simulated Mueller matrices of layers of varying thicknesses are shown in Fig. 2. The presence of cylindrical scatterers had very limited influence on the values of linear retardance, but it had a significant impact on the values of linear dichroism and anisotropic depolarization effect. The SCB model yields the values of total linear retardance that also match well the experimental data [see Fig. 2(a)] The linear increase of total linear retardance RT and linear dichroism values on thickness are shown in Figs. 2(b) and 2(a), respectively. The nonzero intercept of linear regression curve for the experimental linear dichroism values with the Y axis [Fig. 2(b)] was explained by scattering of transmitted light on the rough surface of tissue.12 It was shown that for an anisotropic media a surface scattering does not affect the retardance values but has an impact on linear dichroism values.28 The quadratic dependence of αL, α44, and αLA on thickness is also confirmed within the experimental range of errors. Moreover, an anisotropic depolarization effect (|αLA|0) is well reproduced with the SCB optical model [see Figs. 2(c)2(e)].

Fig. 2

Results of MC simulations using the SCB optical model. Dependence on layer thickness of (a) total linear dichroism DT (dashed line is a linear regression curve for the experimental data), (b) total linear retardance RT, (c)–(e) depolarization parameters αL, α44, and αLA, respectively. Simulated data are shown by open symbols corresponding to different scattering coefficients μc (i.e., different concentrations of cylindrical scatterers), μs=1500  cm1, Rs=Rc=0.05  μm. Open boxes with error bars represent the experimental data. Solid lines show the results of (a), (b) linear and (c)–(e) quadratic fit of the simulated data.

JBO_25_1_015002_f002.png

The impact of the radius of a cylindrical scatterer Rc on anisotropic linear depolarization parameter αLA was also studied (see Fig. 3). Our simulations show that cylinders with a smaller radius produce stronger anisotropy in linear depolarization. Hence, the presence of anisotropic linear depolarization is an indication of the scattering on small-size fibroid scatterers in the studied medium (so-called form birefringence). The values of parameter αLA can be used for the estimation of the characteristic size of nonspherical scatterers.

Fig. 3

Dependence of parameter αLA on layer thickness for different radii of cylindrical scatterers. The parameters of the SBC optical model are: μs=μc=1500  cm1, Rs=0.05  μm. Open symbols correspond to the different radii of cylindrical scatterers: Rc=0.05, 0.5, and 2  μm, respectively. The concentration of cylindrical scatterers cc was adjusted to keep a constant value of the scattering coefficient μc. Open boxes with error bars represent the experimental data. Solid lines show the results of a parabolic fit of the simulated data.

JBO_25_1_015002_f003.png

To summarize, an appropriate optical model for a dermal layer of unstained fixed histological cuts of skin model tissue should include the subwavelength spherical scatterers and well-aligned cylindrical scatterers both distributed in a uniaxial linear birefringent medium. This model can qualitatively reproduce the thickness dependence of polarization and depolarization properties obtained from LMMD of the experimental Mueller matrices of a dermal layer of skin model histological cuts.12

4.

Conclusions

We used MC software to model scattering of polarized light within the uniaxial birefringent scattering media and applied the logarithmic decomposition of simulated Mueller matrices to study a dependence of linear retardance, linear dichroism, and depolarization parameters on thickness in transmission configuration. Previously, it was confirmed that linear retardance and linear dichroism of the dermal layer of skin histological cuts depend linearly on thickness, while the depolarization varies quadratically with thickness. In these studies, we tested several different optical models to explain the results of transmission Mueller microscopy measurements of the histological cuts of full-thickness human skin equivalents. We found that the linear birefringence of the host medium is a necessary parameter of the optical model for reproducing the total linear retardance values, and anisotropic scatterers are the essential component of the optical model of dermis for reproducing both linear dichroism and anisotropic depolarization effects. We also derived the rotation-invariant parameters for LMMD and proposed using parameter αLA as a marker for anisotropy of linear depolarization.

It was shown that measured values of linear retardance for a dermal layer of histological cuts can be simulated with either the SB or SCB model. Despite the fact that depolarization of transmitted light was reproduced with both optical models of dermis, the experimentally observed effect of anisotropy of linear depolarization (αLA0) required using the SCB optical model. Both skin tissue cuts measurements and simulations with the SCB optical model confirmed the presence of the nonzero linear dichroism calculated with LMMD for both measured and simulated Mueller matrices. The offset between experimental and simulated values of linear dichroism can be explained by the effect of light surface scattering, which was not included in our optical model. Extending the optical model of tissue by taking into account a rough interface between two media as well as testing different types of anisotropic scatterers (e.g., ellipsoids) will be the subject of future work.

With MC simulations, we have shown that applying the logarithmic decomposition of transmission Mueller matrix of tissue may provide the relevant information not only on average size of dominant scatterers but also on their shape. For example, the presence of the anisotropy of linear depolarization may point to the dominant scattering by nonspherical scatterers, thus, providing the information on tissue microstructure for optical diagnostics.

Disclosures

The authors have no relevant financial interests in the paper and no other potential conflicts of interest to disclose.

Acknowledgments

Pengcheng Li acknowledges the funding from the China Scholarship Council for his internship at LPICM, Ecole Polytechnique, France. Hee Ryung Lee gratefully acknowledges the funding from the Doctoral School “Interfaces” of Ecole Polytechnique, France. Shubham Chandel acknowledges the bilateral Franco-Indian Raman–Charpak program scholarship for his internship at LPICM, Ecole Polytechnique, France.

References

1. 

V. V. Tuchin, L. V. Wang and D. A. Zimnyakov, Optical Polarization in Biomedical Applications, Springer, Berlin, Heidelberg (2006). Google Scholar

2. 

A. Vitkin, N. Ghosh, A. De Martino, “Tissue polarimetry,” Photonics: Biomedical Photonics, Spectroscopy, and Microscopy, IV 239 –321 1st ed.John Wiley & Sons, Inc., Hoboken, New Jersey (2015). Google Scholar

3. 

J. Qi and D. S. Elson, “Mueller polarimetric imaging for surgical and diagnostic applications: a review,” J. Biophotonics, 10 (8), 950 –982 (2017). https://doi.org/10.1002/jbio.v10.8 Google Scholar

4. 

M. Borovkova et al., “Complementary analysis of Muller-matrix images of optically anisotropic highly scattering biological tissues,” J. Eur. Opt. Soc. Rapid Publ., 14 20 (2018). https://doi.org/10.1186/s41476-018-0085-9 Google Scholar

5. 

S. R. Cloude, “Group theory and polarisation algebra,” Optik, 75 26 –36 (1986). OTIKAJ 0030-4026 Google Scholar

6. 

S. Y. Lu and R. A. Chipman, “Interpretation of Mueller matrices based on polar decomposition,” J. Opt. Soc. Am. A, 13 (5), 1106 –1113 (1996). https://doi.org/10.1364/JOSAA.13.001106 JOAOD6 0740-3232 Google Scholar

7. 

J. Morio and F. Goudail, “Influence of the order of diattenuator, retarder, and polarizer in polar decomposition of Mueller matrices,” Opt. Lett., 29 (19), 2234 –2236 (2004). https://doi.org/10.1364/OL.29.002234 OPLEDP 0146-9592 Google Scholar

8. 

R. Ossikovski, “Analysis of depolarizing Mueller matrices through a symmetric decomposition,” J. Opt. Soc. Am. A, 26 (5), 1109 –1118 (2009). https://doi.org/10.1364/JOSAA.26.001109 Google Scholar

9. 

R. Ossikovski, “Differential matrix formalism for depolarizing anisotropic media,” Opt. Lett., 36 (12), 2330 –2332 (2011). https://doi.org/10.1364/OL.36.002330 OPLEDP 0146-9592 Google Scholar

10. 

N. Ortega-Quijano and J. L. Arce-Diego, “Depolarizing differential Mueller matrices,” Opt. Lett., 36 (13), 2429 –2431 (2011). https://doi.org/10.1364/OL.36.002429 OPLEDP 0146-9592 Google Scholar

11. 

H. R. Lee et al., “Mueller microscopy of anisotropic scattering media: theory and experiments,” Proc. SPIE, 10677 1067718 (2018). https://doi.org/10.1117/12.2306943 PSISDG 0277-786X Google Scholar

12. 

H. R. Lee et al., “Digital histology of tissue with Mueller microscopy: how to mitigate an impact of tissue cut thickness fluctuations,” J. Biomed. Opt., 24 (7), 076004 (2019). https://doi.org/10.1117/1.JBO.24.7.076004 JBOPFO 1083-3668 Google Scholar

13. 

T. Yun et al., “Monte Carlo simulation of polarized photon scattering in anisotropic media,” Opt. Express, 17 (19), 16590 –16602 (2009). https://doi.org/10.1364/OE.17.016590 OPEXFF 1094-4087 Google Scholar

14. 

E. Du et al., “Two-dimensional backscattering Mueller matrix of sphere-cylinder birefringence media,” J. Biomed. Opt., 17 (12), 126016 (2012). https://doi.org/10.1117/1.JBO.17.12.126016 JBOPFO 1083-3668 Google Scholar

15. 

H. He et al., “Application of sphere-cylinder scattering model to skeletal muscle,” Opt. Express, 18 (14), 15104 –15112 (2010). https://doi.org/10.1364/OE.18.015104 OPEXFF 1094-4087 Google Scholar

16. 

Y. Guo et al., “Study on retardance due to well-ordered birefringent cylinders in anisotropic scattering media,” J. Biomed. Opt., 19 (6), 065001 (2014). https://doi.org/10.1117/1.JBO.19.6.065001 JBOPFO 1083-3668 Google Scholar

17. 

P. Li et al., “Separating azimuthal orientation dependence in polarization measurements of anisotropic media,” Opt. Express, 26 (4), 3791 –3800 (2018). https://doi.org/10.1364/OE.26.003791 OPEXFF 1094-4087 Google Scholar

18. 

J. C. Ramella-Roman, S. A. Prahl and S. L. Jacques, “Three Monte Carlo programs of polarized light transport into scattering media: part I,” Opt. Express, 13 (12), 4420 –4438 (2005). https://doi.org/10.1364/OPEX.13.004420 OPEXFF 1094-4087 Google Scholar

19. 

C. F. Bohren and D. R. Huffman, Absorption and Scattering of Light by Small Particles, 83194 –113208 John Wiley and Sons, New York (1983). Google Scholar

20. 

P. Li et al., “GPU acceleration of Monte Carlo simulations for polarized photon scattering in anisotropic turbid media,” Appl. Opt., 55 (27), 7468 –7476 (2016). https://doi.org/10.1364/AO.55.007468 APOPAI 0003-6935 Google Scholar

21. 

A. Rossi et al., “Generation of a three-dimensional full thickness skin equivalent and automated wounding,” J. Visualized Exp., (96), e52576 (2015). https://doi.org/10.3791/52576 Google Scholar

22. 

C. Reuter, H. Walles and F. Groeber, “Preparation of a three-dimensional full thickness skin equivalent,” Methods Mol. Biol., 1612 191 –198 (2017). https://doi.org/10.1007/978-1-4939-7021-6 Google Scholar

23. 

H. C. van de Hulst, Light Scattering by Small Particles, John Wiley and Sons, New York (1957). Google Scholar

24. 

E. Compain, S. Poirier and B. Drévillon, “General and self-consistent method for the calibration of polarization modulators, polarimeters and Mueller matrix ellipsometers,” Appl. Opt., 38 3490 –3502 (1999). https://doi.org/10.1364/AO.38.003490 APOPAI 0003-6935 Google Scholar

25. 

N. Agarwal et al., “Spatial evolution of depolarization in homogeneous turbid media within the differential Mueller matrix formalism,” Opt. Lett., 40 5634 –5637 (2015). https://doi.org/10.1364/OL.40.005634 OPLEDP 0146-9592 Google Scholar

26. 

S. Bartel and A. H. Hielscher, “Monte Carlo simulations of the diffuse backscattering Mueller matrix for highly scattering media,” Appl. Opt., 39 (10), 1580 –1588 (2000). https://doi.org/10.1364/AO.39.001580 APOPAI 0003-6935 Google Scholar

27. 

M.-R. Antonelli et al., “Impact of model parameters on Monte Carlo simulations of backscattering Mueller matrix images of colon tissue,” Biomed. Opt. Express, 2 1836 –1851 (2011). https://doi.org/10.1364/BOE.2.001836 BOEICL 2156-7085 Google Scholar

28. 

G. J. Simpson and K. L. Rowlen, “Molecular orientation at surfaces: surface roughness contributions to measurements based on linear dichroism,” J. Phys. Chem. B, 103 3800 –3811 (1999). https://doi.org/10.1021/jp983912k JPCBFK 1520-6106 Google Scholar

Biography

Pengcheng Li is a postdoctoral fellow at the Laboratory of Optical Imaging and Sensing, Department of Physics, Tsinghua University, China. His research topics are Monte Carlo (MC) simulation of polarized photon scattering in turbid media, Mueller matrix transformation theory (orientation invariant parameters and mirror symmetry breaking of Mueller matrix), and its application in distinguishing different sources of anisotropies.

Hee Ryung Lee is a PhD student in the Applied Optics and Polarimetry Group of the Laboratory of Physics of Interfaces and Thin Films, Ecole Polytechnique, France. He received his MS degree in physics from Kyung Hee University (Seoul, Korea) and Ecole Polytechnique (Palaiseau, France) in 2017. His current research interests include biophotonics, Mueller polarimetry, and polarized MC modeling. He is a member of the SPIE Student Chapter at Ecole Polytechnique, France.

Tatiana Novikova is a head of the Applied Optics and Polarimetry Group at the Laboratory of Physics of Interfaces and Thin Films, Ecole Polytechnique, France. She received her PhD in applied mathematics from the Institute of Mathematical Modeling (Moscow, Russia) in 1999 and HDR (habilitation) in physics from the University Paris XI (Orsay, France) in 2015. Her research activities are focused on Mueller polarimetry and its applications in biomedical engineering and microelectronics, computational optics of polarized light, interaction of electromagnetic waves with matter, and physics of weakly ionized plasmas for technological applications. She is a senior OSA member and SPIE member.

Biographies of the other authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Pengcheng Li, Hee Ryung Lee, Shubham Chandel, Christian Lotz, Florian Kai Groeber-Becker, Sofia Dembski, Razvigor Ossikovski, Hui Ma, and Tatiana Novikova "Analysis of tissue microstructure with Mueller microscopy: logarithmic decomposition and Monte Carlo modeling," Journal of Biomedical Optics 25(1), 015002 (13 January 2020). https://doi.org/10.1117/1.JBO.25.1.015002
Received: 30 October 2019; Accepted: 23 December 2019; Published: 13 January 2020