22 April 2020 Reliability assessment for blood oxygen saturation levels measured with optoacoustic imaging
Leonie Ulrich, Kai Gerrit Held, Michael Jaeger, Martin Frenz, Hidayet Günhan Akarçay
Author Affiliations +
Abstract

Significance: Quantitative optoacoustic (OA) imaging has the potential to provide blood oxygen saturation (SO2) estimates due to the proportionality between the measured signal and the blood’s absorption coefficient. However, due to the wavelength-dependent attenuation of light in tissue, a spectral correction of the OA signals is required, and a prime challenge is the validation of both the optical characterization of the tissue and the SO2.

Aim: We propose to assess the reliability of SO2 levels retrieved from spectral fitting by measuring the similarity of OA spectra to the fitted blood absorption spectra.

Approach: We introduce a metric that quantifies the trends of blood spectra by assigning a pair of spectral slopes to each spectrum. The applicability of the metric is illustrated with in vivo measurements on a human forearm.

Results: We show that physiologically sound SO2 values do not necessarily imply a successful spectral correction and demonstrate how the metric can be used to distinguish SO2 values that are trustworthy from unreliable ones.

Conclusions: The metric is independent of the methods used for the OA data acquisition, image reconstruction, and spectral correction, thus it can be readily combined with existing approaches, in order to monitor the accuracy of quantitative OA imaging.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Leonie Ulrich, Kai Gerrit Held, Michael Jaeger, Martin Frenz, and Hidayet Günhan Akarçay "Reliability assessment for blood oxygen saturation levels measured with optoacoustic imaging," Journal of Biomedical Optics 25(4), 046005 (22 April 2020). https://doi.org/10.1117/1.JBO.25.4.046005
Received: 8 January 2020; Accepted: 25 March 2020; Published: 22 April 2020
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
KEYWORDS
Blood

Veins

Arteries

Reliability

Tissue optics

Optoacoustics

Blood oxygen saturation

Back to Top