26 February 2022 Corneal absorption spectra in the deep UV range
Dominik Inniger, Alessio Poretti, Manuel Ryser, Christoph Meier, Christian Rathjen, Thomas Feurer
Author Affiliations +
Abstract

Significance: Refractive surgery in ophthalmology uses pulsed lasers at 193, 210, or 213 nm. The reason is that most molecular constituents of cornea absorb strongly in this wavelength range. Precise refractive surgery via ablation requires an accurate knowledge of the absorption coefficient at the relevant wavelengths. Yet, the absorption coefficients of corneal tissue reported in literature vary by almost an order of magnitude; moreover, they were measured mostly at the wavelengths mentioned earlier.

Aim: By measuring the corneal absorption coefficient of intact eyeballs stored at different environmental conditions, prepared by following different procedures, and as a function of postmortem time, we determine the absorption coefficient for the entire wavelength range between 185 and 250 nm for as close as possible to in-vivo conditions.

Approach: We use a specially designed UV ellipsometer to measure refractive index and absorption coefficient. Specifically, we investigate the temporal evolution of refractive index and absorption coefficient after enucleation of the eyeballs under different environmental conditions and preparation procedures.

Results: Our measurements provide accurate values for refractive index as well as absorption coefficient of cornea in the wavelength range between 185 and 250 nm. We find that the absorption coefficient decreases with time and that neither storage conditions nor preparation procedures but a continuous degeneration of the cornea is responsible for the observed time evolution. We use the measured time evolution to extrapolate refractive index and absorption coefficient to in-vivo conditions.

Conclusion: Our measurements of the close to in-vivo absorption coefficient of cornea between 185 and 250 nm allow for a better understanding and modeling of refractive cornea surgery, also at other than the three commonly used wavelengths. In the future, this may be relevant when new pulsed laser sources with other wavelengths become available.

1.

Context

For ophthalmologic applications, specifically for refractive surgery, the wavelength range between 190 and 220 nm has been identified as particularly suitable, because of the high absorption coefficient in corneal tissue in this part of the spectrum.1 The most commonly used pulsed laser sources are the ArF excimer laser with a wavelength of 193 nm2 and upconverted solid-state lasers at 210 and 213 nm.3 Studies on corneal ablation with these lasers have shown a similar degree of precision and tissue damage.46 Moreover, no significant difference in corneal mutagenesis was reported.4,7,8 The main difference is the two- to threefold higher absorption coefficient at 193 nm and, as a consequence, the lower threshold fluence for ablation.911 Corneal ablation experiments were also performed with KrCl and KrF excimer lasers at 222 and 248 nm, respectively, but the severe thermal effects on the surrounding tissue12 made these lasers unsuitable for refractive surgery.

Due to their fundamental importance for refractive surgery, a number of research groups have investigated corneal absorption coefficients. Interestingly, the reported values vary quite dramatically, for instance, at 193 nm literature values range from 2410 to 39,900  cm1.13 A similar observation can be made for the absorption coefficient at 213 nm.1,14 The first measurement of the corneal absorption coefficient was in transmission mode using very thin slices to accommodate for the expected high absorption coefficients. The authors used a microtome to cut 32  μm (20  μm) thick slices from frozen dissected corneas15 (eyeballs1). Surprisingly, the results indicated an absorption coefficient as low as 2700  cm1 (2410  cm1) at 193 nm, which is an order of magnitude lower than the 20,000  cm1 estimated from the peptide bonds of the primary chromophores.14 Several subsequent indirect measurements, for instance ablation experiments,16 measurements on dissolved collagen,13 or interferometric photothermal spectroscopy17 suggested that the absorption coefficients should indeed be at least an order of magnitude higher. At 193 nm, they extracted values as high as 20,370, 16,000, and 19,000  cm1, respectively. From these results, it became evident that reflection-type experiments could shed more light on the discrepancy. Reflection-type measurement do not rely on thin slices, because we may use intact eyeballs, are minimally invasive, and reduce dehydration of the cornea. Pettit and Ediger reported the first reflection measurements.14 They determined the reflection coefficient of a glass–cornea interface for different angles of incidence and extracted absorption coefficients of 39,900 and 21,400  cm1 at 193 and 213 nm, respectively. Despite all advantages, reflection-type measurements can still yield fluctuating results, because sample storage conditions prior to the measurement or sample preparation can vary. For instance, samples have to be transported to the lab and hence are exposed to different ambient conditions (temperature, humidity, etc.) for up to several hours. The epithelium must be removed and superficial layers of water or physiological saline solutions are sometimes used to prevent degeneration of the cornea. For instance, at 193 nm, it has been shown that the ablation depth depends on the corneal hydration state18 and that saline solutions on the corneal surface may even act as a masking agent and hence may be used to treat corneal irregularities.19

The current state-of-the-art has motivated us to measure the corneal absorption coefficient in the range between 185 to 250 nm with a specially designed UV ellipsometer. We laid out the measurement protocols to assess the influence of storage conditions prior to the actual measurement as well as preparation procedures to, hopefully, extrapolate the absorption coefficient to in-vivo conditions. On the other hand, we aimed to cover the entire wavelength range between 185 and 250 nm to provide the absorption coefficient also for other than the three standard wavelengths. In the future, this may be relevant when new efficient pulsed laser sources with wavelengths different from the three standard sources at 193, 210, and 213 nm become available.

However, under ablation conditions additional dynamic effects may arise which modify light absorption. For instance, it has been reported that during ablative ArF laser irradiation light transmission decreases,20 tissue scattering properties change,21,22 and water absorption may increase by several orders of magnitude.23 Moreover, the ablation plume or plasma formation may decrease the fluence incident on the cornea.24 This effect is known to also depend on repetition rate and spot size. Hence, the small signal absorption coefficient is one of several but a fundamental ingredient to understand, model, and optimize refractive surgery.

We organized the paper as follows. We first outline the experimental setup, sample storage conditions, and preparation procedures. Then, we present the measurements of the absorption coefficient, compare the results to previous measurements, and discuss the influence of different environmental parameters. We use the results to extrapolate the absorption coefficient of cornea to in-vivo conditions.

2.

Experiment

Ellipsometry is a standard technology to measure complex refractive indices or thin film thicknesses.25

In ellipsometry, a linearly polarized plane wave illuminates a sample so the incident light has an s- as well as a p-polarized electric field component with respect to the sample surface. The self-referencing detection scheme measures the difference between the complex-valued reflection coefficients for s- and p-polarized light, from which the optical properties of interest are extracted. Ellipsometry comes in different variants and the best-suited setup for our purpose is the “rotating compensator ellipsometry” (RCE), see Table 4.3 in Ref. 25. Figure 1 shows a photographic image and a schematic of the RCE setup. Knowing that the cornea’s index of refraction is of order 1.5 in the wavelength range of interest,14 we set the ellipsometer angle to 75 deg. The emitted broadband UV radiation from a 30 W deuterium lamp (BWTEK, BDS130A) is coupled in a 200 mm long solarization-resistant multimode patch fiber (Thorlabs FG200AEA) and collimated by a CaF2 lens (Thorlabs LA5315) to produce a broadband, nearly plane-wave illumination. A UV polarizer (Bernhard Halle Nachfl., PUM 1.08) with an azimuth angle of 45 deg defines the polarization state relative to the sample surface and an aperture after the polarizer limits the beam diameter to 2 mm. The sample holder is an inverted UV prism (Bernhard Halle Nachfl., Heraeus Suprasil 1, angles 30/75/75 deg, base size 14  mm×8  mm, and surface quality P4). A prism angle of 75 deg allows the incident and the reflected light to enter and to exit the prism normal to its side surfaces. After exiting the prism, the beam passes the compensator (Bernhard Halle Nachfl., RZM 4.10), which is mounted in a motorized rotation state (PI, RS-40 DC) and the analyzer (same type as polarizer). To maximize the dynamic range, the azimuth angle of the analyzer is set to 45  deg. The transmitted polarization component is focused (with the same lens as used for collimation) to a second 200 mm long multimode patch fiber (Thorlabs, FG105ACA), which is connected to the UV spectrometer (Ocean Insight, Maya2000 Pro-Deep-UV). The antireflection coatings of all optical elements as well as the operational wavelengths of polarizer, compensator, analyzer, and spectrometer allowed the ellipsometer to cover the wavelength range between 185 and 250 nm. The entire ellipsometer resided in a flow-box at room temperature.

Fig. 1

(a) Photographic image and (b) schematic of the RCE setup. A deuterium lamp coupled to a multimode fiber with collimator lens produces a broadband, nearly plane wave illumination. A downstream polarizer sets its polarization state. The light reflected from the sample is analyzed by a rotating compensator, which is followed by an analyzer and a UV spectrometer. The eyeballs are mounted on the base of an inverted UV prism.

JBO_27_2_025004_f001.png

For all measurements reported hereafter, we used pig eyes, which were enucleated within 3 min after death. Immediately after enucleation, the eyeballs were stored in a sealable container at a temperature of 4°C either on paper soaked with a 0.9% NaCl water solution or fully immersed in 0.9% NaCl water solution. Just before the measurement, the epithelium was removed with a hockey epithelium remover. Then, the eyeballs were mounted on a xyz translation stage and pressed against the prism base by moving only the z axis (perpendicular to the prism base) to avert shear forces. Moreover, we carefully prevented air entrapment and bubble formation at the interface. The time span between removing the eyeball from the cooled container to mounting it was typically 2 min. The mount was equipped with a thermocouple (0.1 K resolution) and a scale (0.1 g resolution) to monitor temperature and contact force before, during, and after the measurements. The initial contact force was below 100 mN. Then, we let the system relax for 5 to 15 min (stabilization time) while continuously monitoring the spectrally resolved ellipsometer signal and the measurement was started when the signal change was below 5‰ per minute. At this point, the contact force was below 70 mN and the temperature at the bottom of the eyeball was between 14 to 17°C.

Since the polished prism surface (P4, ISO 10110-8) defines the interface and cornea is highly absorbing, we assume a perfectly flat sample with infinite thickness to extract the cornea’s pseudodielectric function. Note that the microscopic structure of the stroma consists of collagen fibril bundles called lamellae, which per se are birefringent.26 However, these bundles are randomly oriented within the central part of the cornea,27 and we assume a zero net birefringence when probing an area of 12  mm2. For a single measurement, we rotated the compensator angle C by 180 deg in steps of 10 deg. At every angle twelve spectra, each with an integration time of 80 ms, were recorded and averaged. The total time for an entire scan was 2.5 min and we define the postmortem time as the time elapsed between enucleation and end of measurement. For the RCE configuration, the signal as a function of the compensator angle C is28,29

Eq. (1)

ID(C)=a0+a2ccos2C+a2ssin2C+a4ccos4C+a4ssin4C.

The wavelength-dependent Fourier coefficients ai are defined via

Eq. (2)

a0=12I0(1+cosδc)(sin2Asin2ξcosΔcos2Acos2ξ)+I0a2c=I0sin2Asin2ξsinΔsinδca2s=I0cos2Asin2ξsinΔsinδca4c=12I0(cosδc1)(sin2Asin2ξcosΔ+cos2Acos2ξ)a4s=12I0(cosδc1)(cos2Asin2ξcosΔsin2Acos2ξ),
with the compensator phase retardance δc, the analyzer angle A, the amplitude ratio of p- and s-polarized light after reflection from the sample ξ=atan(tanΨ/tanP), the polarizer angle P, and the three sample parameters I0, Ψ, and Δ. Fitting Eq. (1) for every wavelength to the measured data reveals the ellipsometry parameters Ψ and Δ from which we extract the wavelength-dependent pseudodielectric function25

Eq. (3)

εt(Ψ,Δ)=εitanθi[1+tanθi(1tanΨeiΔ1+tanΨeiΔ)2],
with the dielectric function of the prism material εi and the incident angle θi. From the definition n+ik=εt, the real and the imaginary parts of the refraction index are calculated. The absorption coefficient is obtained via

Eq. (4)

α=4πλk.

By taking into account the alignment precision of all relevant optical elements, we find that the absolute error decreases from ±1400  cm1 at 185 to ±600  cm1 at 200 nm and remains constant at ±600  cm1 above.

3.

Results and Discussion

3.1.

Ellipsometer Calibration

To calibrate the ellipsometer system, we performed two measurements, one at a prism–air and the other at a prism–distilled water interface, with the goal to correct for imperfect optical elements.

Figure 2 shows theoretical (dashed green curve) and measured (solid red curve) phase difference Δ (a) and amplitude ratio Ψ (b) as a function of wavelength for the prism-air interface. For Δ, we find a deviation from the expected values between 2 and 4 deg, which we attribute to a stress induced birefringence of the prism. This has been confirmed by exerting additional pressure on the prism. The average measurement of Ψ agrees well with the expected value of 45 deg and the oscillations with an amplitude of 0.1 deg result from an imperfect compensator. Both deviations δΔair  and δΨair are used to correct further measurements.

Fig. 2

Calibration of the ellipsometer. Measured (solid red curve) and theoretical (dashed green curve) (a) phase difference Δ and (b) amplitude ratio Ψ for a prism–air interface. Measured (solid red curve) and theoretical (dashed green curve: Segelstein;30 dotted pink curve: Diamon31 assuming zero absorption) (c) phase difference Δ and (d) amplitude ratio Ψ for a prism–water interface. The simulations (dashed blue curve) take into account a rotation by 11 deg of the prism birefringence with respect to the beam coordinate system. (e) Measured refractive index and (f) absorption coefficient compared with literature values.30,31

JBO_27_2_025004_f002.png

Next, we perform a measurement on distilled water with the aim to determine whether the birefringence of the prism is rotated with respect to the beam coordinate system. We use water because its refractive index is closer to that of cornea. The measured ellipsometry parameters Δ and Ψ (solid red curves) are shown in Figs. 2(c) and 2(d) and the resulting index of refraction and absorption coefficient in Figs. 2(e) and 2(f), respectively. While the measured refractive index is between the two literature values, the absorption coefficient is significantly higher than the literature values especially below 225 nm. If we rotate the coordinate system of the prism birefringence by 11 deg, we find perfect agreement between the measured and the simulated results. Note that the rotation affects only the amplitude ratio Ψ, and hence the absorption coefficient, but does not alter the phase difference Δ. Hence, we find δΔwater=0 and a nonzero δΨwater. As a result of the calibration measurements, all subsequent measurements on pig eyes are corrected by (δΔwater+δΔair) and (δΨwater+δΨair) before the index of refraction and the absorption coefficient are calculated.

3.2.

Cornea

Figure 3 shows the real part of the refractive index and the absorption coefficient of porcine cornea as a function of wavelength between 185 and 250 nm. In detail, 10 eyeballs, fully immersed in 0.9% NaCl solution, were prepared one after the other as described above. The time span between two subsequent measurements was 45 min and all 10 eyeballs were measured between 2 and 9 h postmortem. While the solid red curves indicate mean values, the light red shaded regions indicate the range between the minimum and the maximum values. We would like to stress that the maximum curve corresponds tendentially to the first and the minimum curve to the last measurement, in other words, during the measurement time of 7 h, the refractive index and the absorption coefficient were slowly decreasing.

Fig. 3

(a) Refractive index and (b) absorption coefficient of porcine cornea between 185 and 250 nm. The red curves represent the average values and the light red-shaded regions indicate the range between the minimum and maximum values. Literature values are included as indicated in the legend.

JBO_27_2_025004_f003.png

The real part of the refractive index shows a continuous increase toward shorter wavelengths. Below 195 nm, we find an almost constant value of 1.51. The refractive index in the entire wavelengths range is significantly higher than that of water and exhibits a slightly more complex behavior. The values published by Pettit et al.14 (eyes stored in a plastic bag on ice) for 193 and 213 nm deviate by 0.01 at 193 nm and 0.02 at 213 nm. The absorption coefficient also increases with decreasing wavelength but shows two peaks around 187 nm with a peak absorption coefficient of 30,000  cm1 and around 213 nm with a peak absorption coefficient of 7200  cm1. Between 204 and 216 nm, the absorption coefficient is roughly constant on a level of 7000  cm1. For wavelengths larger than 225 nm, the absorption coefficient drops below 2000  cm1, which agrees well with the transmission measurement of Lembares et al.;1 below 225 nm, however, we observe a significant deviation from these transmission measurements. A better agreement is found when comparing our data to the results of Pettit et al.,14 who reported (39,900±9800)  cm1 at 193 nm and (21,400±4900)  cm1 at 213 nm. The two values deviate from our results by 15,900 and 14,200  cm1, respectively. At 193 nm, we find more literature values,13,14,16,17 but all of them are lower by 4000 to 8000  cm1 when compared with our values. As discussed in Sec. 1, we believe that these deviations are mostly due to different sample preparation, measurement technique, or degree of sample degeneration. To further investigate these issues, we carried out additional measurements under different environmental conditions.

We started by investigating the influence of the contact force with which the eyeballs were pressed against the prism. When mounted, we apply a contact force of 100  mN, which relaxes to <70  mN after the stabilization time. When applying contact force in excess of 100 mN, we found a stress-induced birefringence in the prism that corrupts the measurements.

Next, we investigated sample degeneration possibly caused by UV radiation. We prepared two identical series of samples and recorded reference spectra for both. Then, we continuously irradiated one of them while shielding the other from the UV radiation. After the stabilization time, we again recorded spectra and found no evidence of UV-related sample degeneration.

The next four series of measurements were designed to better understand the influence of the hydration state of the corneal surface. Prior to the measurement, all eyeballs were stored in a sealable container at a temperature of 4°C. While the eyeballs of the first series were immersed in 0.9% NaCl water solution, the eyeballs of the second, third, and fourth series were stored on paper soaked with 0.9% NaCl water solution. We removed the epithelium of the first and second series and mounted and measured them immediately after. Similarly, the epithelium of the third series was removed but the stroma was moistened with a 0.9% NaCl water solution before mounting them. Finally, the eyeballs of the fourth series were treated as those of the second series, but they were stored about 25 h longer. The conditions are summarized in Table 1.

Table 1

Measurement series and corresponding measurement conditions.

SeriesStoring conditionPreparation# of eyeballs in seriesMeasurement time
1Sealable container at 4°C, eyeballs immersed in 0.9% NaCl water solutionMeasured immediately after epithelium removal10First eyeball: 2 h
Last eyeball: 9 h
2Sealable container at 4°C, eyeballs on paper soaked with 0.9% NaCl water solutionMeasured immediately after epithelium removal5First eyeball: 5 h
Last eyeball: 11 h
3Sealable container at 4°C, eyeballs on paper soaked with 0.9% NaCl water solutionAfter epithelium removal, stroma was moistened with 0.9% NaCl water solution4First eyeball: 4 h
Last eyeball: 10 h
4Sealable container at 4°C, eyeballs on paper soaked with 0.9% NaCl water solutionMeasured immediately after epithelium removal5First eyeball: 30 h
Last eyeball: 35 h

Figure 4 compares the refractive indices as well as the absorption coefficients of the four series. The results suggest that neither the storage conditions (immersed in 0.9% NaCl water solution or stored on paper soaked with a 0.9% NaCl water solution) nor moisturizing the corneal surface with a 0.9% NaCl water solution prior to mounting has any major influence on the results. The only significant difference we observe when eyeballs are stored for longer times. In this case, we find a decrease of the refractive index as well as of the absorption coefficient for the most part of the spectrum.

Fig. 4

(a) Refractive index and (b) absorption coefficient of porcine corneas between 185 and 250 nm for the four different series. For details, see Table 1 and text.

JBO_27_2_025004_f004.png

These findings led us to the conclusion that a continuous slow degeneration of the cornea is the main reason for the decreasing absorption coefficient. This becomes even more obvious when we plot the absorption coefficient at 193 and 213 nm as a function of postmortem time for all series of eyeballs investigated above. Figures 5(a) and 5(b) show that, irrespective of storage condition or preparation procedure, the absorption coefficient always decreases with time, and the effect is somewhat more pronounced at 213 nm than at 193 nm. Similar observations are also reported elsewhere in the literature. For instance, it has been found that the thickness of the cornea changes within 30 min postmortem32 and that the corneal hydration33 and the collagen diameter of the corneal matrix34 increase with postmortem time. Because collagen fibers dominate the corneal UV absorption below 220 nm, we believe that the decreasing absorption coefficient is related to an increased hydration level or the associated structural changes. Needless to say, that the observed degeneration is irrelevant in refractive surgery, since the procedure takes place within a few minutes and the endothelium is not affected (responsible for the active dehydration process).26

Fig. 5

Absorption coefficient at different postmortem measurement times (a) 193 nm and (b) 213 nm.

JBO_27_2_025004_f005.png

Zooming in on the first 9 h postmortem of the first series strongly suggests that the absorption coefficient decreases more or less linearly within this time span. This allows us to extrapolate the wavelength-dependent absorption coefficient to time zero, i.e., to the time of death. Exemplary, we show the linear fit in Figs. 6(a) and 6(b) for the two important wavelengths of 193 and 213 nm.

Fig. 6

Absorption coefficient of the first series at (a) 193 nm and (b) 213 nm as a function of postmortem time. Applying a linear regression (dashed black lines) to the measured data (red stars) allows us to extrapolate the absorption coefficient to the time of death. (c) Absorption coefficient measured 2.3 h (pink curve), 6 h (green curve), and 8.7 h (blue curve) postmortem and the absorption coefficient extrapolated to the time of death (black curve). The two vertical red lines indicate the two wavelengths shown in (a) and (b).

JBO_27_2_025004_f006.png

Figure 6(c) shows the absorption coefficient measured 2.3 h (pink curve), 6 h (green curve), and 8.7 h (blue curve) postmortem and the absorption coefficient extrapolated to time zero, i.e., the time of death (black curve). The result of the fitting procedure is the “in-vivo” corneal absorption coefficient between 185 and 250 nm.

4.

Conclusion

A specially designed UV ellipsometer was used to determine the absorption coefficient of porcine cornea in the wavelength range from 185 to 250 nm. Measurements up to 35 h postmortem reveal that the absorption coefficient continuously decreases with time for the most part of the spectrum. The observed decrease does not correlate with the UV irradiation dose, the applied force, nor the storage condition, but relates to a continuous slow degeneration of the cornea. Using the measured time series allows for extrapolating the absorption coefficient to the time of death and to report, for the first time, the “in-vivo” absorption coefficient in the wavelength range between 185 and 250 nm.

Disclosures

The authors have no financial interests in this article and no potential conflicts of interest to disclose.

Acknowledgments

This research was funded by Innosuisse – Swiss Innovation Agency (project no. 46956.1 IP).

Code, Data, and Materials Availability

Theory, original data, and data processing supporting the plots in this paper are available via DOI number 10.48350/161507.

References

1. 

A. Lembares, A. X. H. Hu and G. W. Kalmus, “Absorption spectra of corneas in the far ultraviolet region,” Invest. Ophthal. Vis. Sci., 38 (6), 1283 –1287 (1997). IOVSDA 0146-0404 Google Scholar

2. 

A. Pidro et al., “Excimer lasers in refractive surgery,” Acta Inf. Med., 27 (4), 278 –283 (2019). https://doi.org/10.5455/aim.2019.27.278-283 Google Scholar

3. 

S. Shah and M. Piovella, “Solid-state laser platforms: two reviews,” 26 –30 (2013). Google Scholar

4. 

N. S. Tsiklis et al., “Photorefractive keratectomy using solid state laser 213 nm and excimer laser 193 nm: a randomized, contralateral, comparative, experimental study,” Invest. Ophthal. Vis. Sci., 49 (4), 1415 –1420 (2008). https://doi.org/10.1167/iovs.07-1280 IOVSDA 0146-0404 Google Scholar

5. 

Q. Ren et al., “Ablation of the cornea and synthetic polymers using a UV (213 nm) solid-state laser,” IEEE J. Quantum Electron., 26 (12), 2284 –2288 (1990). https://doi.org/10.1109/3.64366 IEJQA7 0018-9197 Google Scholar

6. 

A. M. Roszkowska et al., “Experimental and clinical investigation of efficiency and ablation profiles of new solid-state deep-ultraviolet laser for vision correction,” J. Refract. Surg., 30 (12), 2536 –2542 https://doi.org/10.1016/j.jcrs.2004.04.071 Google Scholar

7. 

P. P. Van Saarloos and J. Rodger, “Histological changes and unscheduled DNA synthesis in the rabbit cornea following 213·-nm, 193-nm, and 266-nm irradiation,” J. Refract. Surg., 23 477 –481 (2007). https://doi.org/10.3928/1081-597X-20070501-10 JRSUEY 0883-0444 Google Scholar

8. 

T. Kaido et al., “Cytotoxic and mutagenic action of 193-nm and 213-nm laser radiation,” J. Refract. Surg., 18 529 –534 (2002). https://doi.org/10.3928/1081-597X-20020901-07 JRSUEY 0883-0444 Google Scholar

9. 

J. H. Shen et al., “Ablation rate of PMMA and human cornea with a frequency-quintupled Nd:YAG laser (213 nm),” Lasers Surg. Med., 21 (2), 179 –185 (1997). https://doi.org/10.1002/(SICI)1096-9101(1997)21:2<179::AID-LSM9>3.0.CO;2-N LSMEDI 0196-8092 Google Scholar

10. 

G. T. Dair et al., “Investigation of corneal ablation efficiency using ultraviolet 213-nm solid state laser pulses,” Invest. Ophthal. Vis. Sci., 40 (11), 2752 –2756 (1999). IOVSDA 0146-0404 Google Scholar

11. 

K. T. Schomacker et al., “Co:MgF2 laser ablation of tissue: effect of wavelength on ablation threshold and thermal damage,” Lasers Surg. Med., 11 (2), 141 –151 (1991). https://doi.org/10.1002/lsm.1900110208 LSMEDI 0196-8092 Google Scholar

12. 

S. L. Trokel et al., “Excimer laser surgery of the cornea,” Am. J. Ophthalmol., 96 (6), 710 –715 (1983). https://doi.org/10.1016/S0002-9394(14)71911-7 AJOPAA 0002-9394 Google Scholar

13. 

B. T. Fisher and D. W. Hahn, “Measurement of small-signal absorption coefficient and absorption cross section of collagen for 193-nm excimer laser light and the role of collagen in tissue ablation,” Appl. Opt., 43 (29), 5443 –5451 (2004). https://doi.org/10.1364/AO.43.005443 APOPAI 0003-6935 Google Scholar

14. 

G. H. Pettit and M. N. Ediger, “Corneal-tissue absorption coefficients for 193- and 213-nm ultraviolet radiation,” Appl. Opt., 35 (19), 3386 –3391 (1996). https://doi.org/10.1364/AO.35.003386 APOPAI 0003-6935 Google Scholar

15. 

C. A. Puliafito et al., “Excimer laser ablation of the cornea and lens: experimental studies,” Ophthalm., 92 (6), 741 –748 (1985). https://doi.org/10.1016/S0161-6420(85)33962-3 Google Scholar

16. 

Z. Bor et al., “Physical problems of excimer laser cornea ablation,” Opt. Eng., 32 (10), 2481 –2486 (1993). https://doi.org/10.1117/12.145357 Google Scholar

17. 

A. D. Yablon et al., “Measurement of tissue absorption coefficients by use of interferometric photothermal spectroscopy,” Appl. Opt., 38 (7), 1259 –1272 (1999). https://doi.org/10.1364/AO.38.001259 APOPAI 0003-6935 Google Scholar

18. 

P. J. Dougherty et al., “Excimer laser ablation rate and corneal hydration,” Am. J. Ophthalmol., 118 (2), 169 –176 (1994). https://doi.org/10.1016/S0002-9394(14)72896-X AJOPAA 0002-9394 Google Scholar

19. 

E.W. Kornmehl et al., “A comparative study of masking fluids for excimer laser phototherapeutic keratectomy,” Arch. Ophthal., 109 (6), 860 –863 (1991). https://doi.org/10.1001/archopht.1991.01080060124039 AROPAW 0003-9950 Google Scholar

20. 

G. H. Pettit et al., “Excimer laser ablation of the cornea,” Opt. Eng., 34 (3), 661 –667 (1995). https://doi.org/10.1117/12.190408 Google Scholar

21. 

G. H. Pettit and M. N. Ediger, “Pump/probe transmission measurements of corneal tissue during excimer laser ablation,” Lasers Surg. Med., 13 (3), 363 –367 (1993). https://doi.org/10.1002/lsm.1900130313 LSMEDI 0196-8092 Google Scholar

22. 

G. H. Pettit et al., “Dynamic optical properties of collagen-based tissue during ArF excimer laser ablation,” Appl. Opt., 32 (4), 488 –493 (1993). https://doi.org/10.1364/AO.32.000488 APOPAI 0003-6935 Google Scholar

23. 

P. T. Staveteig and J. T. Walsh, “Dynamic 193-nm optical properties of water,” Appl. Opt., 35 (19), 3392 –3403 (1996). https://doi.org/10.1364/AO.35.003392 APOPAI 0003-6935 Google Scholar

24. 

A. Vogel and V. Venugopalan, “Mechanisms of pulsed laser ablation of biological tissues,” Chem. Rev., 103 (2), 577 –644 (2003). https://doi.org/10.1021/cr010379n CHREAY 0009-2665 Google Scholar

25. 

H. Fujiwara, Spectroscopic Ellipsometry: Principles and Applications, John Wiley & Sons Ltd., Chichester (2007). Google Scholar

26. 

P. Fratzl, Collagen Structure and Mechanics, Springer, New York (2008). Google Scholar

27. 

P. M. Pinsky et al., “Computational modeling of mechanical anisotropy in the cornea and sclera,” J. Cataract Refract. Surg., 31 (1), 136 –145 (2005). https://doi.org/10.1016/j.jcrs.2004.10.048 Google Scholar

28. 

R. Kleim et al., “Systematic errors in rotating-compensator ellipsometry,” J. Opt. Soc. Am. A, 11 (9), 2550 –2559 (1994). https://doi.org/10.1364/JOSAA.11.002550 JOAOD6 0740-3232 Google Scholar

29. 

D. E. Aspnes, “Optimizing precision of rotating-analyzer and rotating-compensator ellipsometers,” J. Opt. Soc. Am. A, 21 (3), 403 –410 (2004). https://doi.org/10.1364/JOSAA.21.000403 JOAOD6 0740-3232 Google Scholar

30. 

D. J. Segelstein, “The complex refractive index of water,” (1975). Google Scholar

31. 

M. Daimon and A. Masumura, “Measurement of the refractive index of distilled water from the near-infrared region to the ultraviolet region,” Appl. Opt., 46 (18), 3811 –3820 (2007). https://doi.org/10.1364/AO.46.003811 APOPAI 0003-6935 Google Scholar

32. 

P. E. Napoli et al., “Post-mortem corneal thickness measurements with a portable optical coherence tomography system: a reliability study,” Sci. Rep., 6 1 –7 (2016). https://doi.org/10.1038/srep30428 SRCEC3 2045-2322 Google Scholar

33. 

L. Zhou et al., “Image analysis on corneal opacity: a novel method to estimate postmortem interval in rabbits,” J. Huazhong Univ. Sci. Technol. Med. Sci., 30 (2), 235 –239 (2010). https://doi.org/10.1007/s11596-010-0221-2 Google Scholar

34. 

S. H. Zhu et al., “An ultrastructural observation of corneal collagen matrix of rabbit in the different time after death,” Chin. J. For. Med., 23 (2), 99 –101 (2008). Google Scholar

Biography

Dominik Inniger received his BSc degree in microtechnology and his MSE degree in industrial technologies from Bern University of Applied Science in 2013 and 2016, respectively. He is currently pursuing his PhD at the Graduate School for Cellular and Biomedical Sciences at Bern University. His research is focused on UV ablation lasers and applications in medicine.

Alessio Poretti received his BSc degree in microtechnology at Bern University of Applied Science in 2019 and is currently pursuing his master’s degree in biomedical engineering at Bern University. During his master’s studies, he works in the optics research group at the Bern University of Applied Science.

Manuel Ryser received his PhD in experimental physics from the University of Bern in 2009. He is currently cohead of the Optical Fibers and Fiber Lasers group at the same university, cofounder and CEO of the university spinoff 8photonics GmbH, and high school teacher for physics and mathematics. His research interests include fiber lasers, fiber optic sensors, and fabrication and application of specialty optical fibers.

Christoph Meier received his engineering diploma degree from the School of Engineering, HTL Biel, and his master’s degree in physics from the University of Neuchâtel. Since 1991, he has been working at Bern University of Applied Science, as a professor of physics and electrical technology. In 2001, he was elected professor of optics and head of the optics research group. His main interests are R&D projects in the field of optical coherence tomography since 2008.

Christian Rathjen received his MS degree in mechanical engineering and his PhD in metrology from the University of Hannover, Germany. After working at the same university in the field of student education and research, he held a R&D post at CERN, Geneva, Switzerland. Currently, employed as vice president technology and strategy at Ziemer Ophthalmic Systems Port, Switzerland. His strong focus on innovation and technology is reflected in numerous patents.

Thomas Feurer received his PhD in experimental physics from the University of Würzburg in 1994 and his habilitation from the University of Jena in 2001. Beginning of 2001, he started working as a research associate in the Department of Chemistry of MIT until 2004 when he became full professor at the University of Bern. His research interests are optical fibers, fiber lasers, quantum optics, THz science, and ultrafast spectroscopy of molecules and solids.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Dominik Inniger, Alessio Poretti, Manuel Ryser, Christoph Meier, Christian Rathjen, and Thomas Feurer "Corneal absorption spectra in the deep UV range," Journal of Biomedical Optics 27(2), 025004 (26 February 2022). https://doi.org/10.1117/1.JBO.27.2.025004
Received: 29 November 2021; Accepted: 31 January 2022; Published: 26 February 2022
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
KEYWORDS
Absorption

Cornea

Refractive index

Water

Prisms

Ultraviolet radiation

Deep ultraviolet

Back to Top