17 January 2024 Improved intraoperative identification of close margins in oral squamous cell carcinoma resections using a dual aperture fluorescence ratio approach: first in-human results
Author Affiliations +
Abstract

Significance

Surgical excision is the main treatment for solid tumors in oral squamous cell carcinomas, where wide local excision (achieving a healthy tissue margin of >5 mm around the excised tumor) is the goal as it results in reduced local recurrence rates and improved overall survival.

Aim

No clinical methods are available to assess the complete surgical margin intraoperatively while the patient is still on the operating table; and while recent intraoperative back-bench fluorescence-guided surgery approaches have shown promise for detecting “positive” inadequate margins (<1 mm), they have had limited success in the detection of “close” inadequate margins (1 to 5 mm). Here, a dual aperture fluorescence ratio (dAFR) approach was evaluated as a means of improving detection of close margins.

Approach

The approach was evaluated on surgical specimens from patients who were administered a tumor-specific fluorescent imaging agent (cetuximab-800CW) prior to surgery. The dAFR approach was compared directly against standard wide-field fluorescence imaging and pathology measurements of margin thickness in specimens from three patients and a total of 12 margin locations (1 positive, 5 close, and 6 clear margins).

Results

The area under the receiver operating characteristic curve, representing the ability to detect close compared to clear margins (>5 mm) was found to be 1.0 and 0.57 for dAFR and sAF, respectively. Improvements in dAFR were found to be statistically significant (p<0.02).

Conclusions

These results provide evidence that the dAFR approach potentially improves detection of close surgical margins.

1.

Introduction

Over 370,000 new cases of oral squamous cell carcinomas (OSCC) were reported in 2020.1 The primary treatment for these tumors is surgical resection. 2 Yet, owing to the complex anatomy of the head and neck, complete resection of primary tumors is challenging and inadequate margins are observed in up to 40% of cases.35 These include both positive margins (tumor within 1 mm of the edge of the resected specimen) and close margins (tumor between 1 and 5 mm of the edge of the resected specimen).6,7 In cases where inadequate margins are found (any <5  mm thick), patients often require intensive adjuvant therapy8 because a secondary surgery is typically not possible due to altered anatomy or the proximity of vital structures. In either scenario, patient outcomes are generally worse than if clear margins were obtained on the initial surgery.9

Currently, the surgical margin status is determined by histopathology, which takes 3 to 7 days following surgery. This prevents the ability to directly act on results while the patient is still on the operating table. Frozen-section histopathological analysis of at-risk areas can be performed intraoperatively, as results are typically obtained in around 30 min.10,11 However, the staining quality of frozen sections is variable and, more importantly, it requires a subjective decision on part of the surgical team to decide the locations to investigate, leading to sampling error that can result in discrepancies between final pathology in up to 13% of cases.12 Such limitations have spurred efforts to develop faster, more accurate methods of assessing the entire margin of the resected specimen intraoperatively.

Over the last decade, significant efforts have been made to establish intraoperative margin assessment using fluorescence molecular guided surgery approaches.13,14 In OSCC, these efforts have predominantly employed imaging agents developed from epidermal growth factor receptor (EGFR)-targeted antibodies. Recently, a phase II clinical trial in 65 OSCC patients showed that EGFR-targeted ex vivo fluorescence imaging—referred to in this work as single-aperture fluorescence (sAF)—of entire surgical margins could identify positive margins with a sensitivity of 100%, yet close margin detection yielded a sensitivity of only 70%.15 All evidence suggested that the lower detection rates of close margins were caused by the effects of heterogeneous tissue optical properties in the surgical margins.

Here, we introduce a new approach that involves dividing wide numerical-aperture (NA) fluorescence images with narrow NA fluorescence images taken from the same point-of-view, pixel-by-pixel. The resulting ratio image can be shown to provide an estimate of fluorescence depth that is insensitive to tissue optical properties as long as the depth of the fluorescence is considered to be in the sub-diffuse regime.16,17 While previous work employed photon energy bands in the ultraviolet range, limiting the depth of its penetration to <1  mm, here near-infrared (NIR) light is utilized, which contains a much deeper depth of penetration (where the reduced mean free path length is much larger, between 5 mm and 1 cm).18

This study compares dual aperture fluorescence ratio (dAFR) against conventional sAF and histopathological assessment of margin thickness in a clinical pilot study, evaluating a total of 12 margins from three patients. We show that dAFR provides accurate information on the surgical margin status in minutes and has the potential to improve the detection of close margins.

2.

Materials and Methods

2.1.

Dual Aperture Fluorescence Ratio Imaging System

A dAFR system, capable of capturing fluorescence at a range of numerical apertures, was constructed in house (Fig. 1). The system was fitted with an 800-mW, 780-nm wavelength light-emitting diode (LED) for illumination (Thorlabs, Newton, New Jersey), from which light was passed through a collimation adaptor (Thorlabs, Newton, New Jersey) and a 10-nm bandpass excitation filter centered at 780 nm (Chroma, Bellows Falls, Vermont) to achieve a widefield illumination on the surface of surgical specimens. The power density of the excitation light at the surface of the specimens was 5  mW/cm2, on average, over a field-of-view of 10.5  cm×10.5  cm. Emitted fluorescence light was collected in epi-illumination mode with a 24-mm-focal-length high-resolution visible-near infrared (VIS-NIR) lens (Edmund Optics, Barrington, New Jersey), installed 20 cm from the plane of imaging. The lens was connected to a 4.2-megapixel sCMOS camera (PCO Panda, Kelheim, Germany) and an 800-nm long pass fluorescence filter (Chroma, Bellows Falls, Vermont) was fit between the opening of the lens and the camera sensor. The spatial resolution of the system was set to 205  μm/pixel, isotropic, at a final binned-pixel image size of 512×512, for a field of view of 10.5×10.5  cm.

Fig. 1

Detailed schematic of the custom widefield imaging system capable of a wide range of numerical apertures at the detector, side view (TOP-left) and top view (TOP-right), and a picture of the true in-house system (BOTTOM)—components: (a) heavy-duty rotation motor, (b) mounted 780 nm LED, (c) 4.2 MP sCMOS fluorescence camera, (d) color camera, (e) 800 nm long pass emission filter, (f) widefield zoom lens, (g) f/2.0-f/16.0 aperture 24 mm widefield lens, (h) LED collimation adaptor, (i) 780±10  nm bandpass excitation filter, (j) height adjustable rail and stage. Note that the clinical adaptation of the system featured the LED in the position detailed in the top schematic.

JBO_29_1_016003_f001.png

2.2.

Clinical Study

2.2.1.

Fluorescent antibody

Clinical grade cetuximab-800CW was produced under good manufacturing practice in a dedicated facility at the University Medical Center Groningen (UMCG). A detailed protocol for its synthesis and purification has been described previously.19 Briefly, commercially available cetuximab (Erbitux®, Eli Lilly and Company, Indianapolis, Indiana) was conjugated with the NIR fluorescent dye IRDye 800CW (LI-COR Biosciences, Lincoln, Nebraska), and the resulting conjugate was purified using PD-10 buffer exchange columns (GE Healthcare, Chicago, Illinois). The purified solution was diluted to 1  mg/mL in sodium-phosphate buffer solution and sterile loaded into vials for later use.

2.2.2.

Patients

A subset of six patients with biopsy-confirmed OSCC included in a clinical trial for margin assessment using fluorescence molecular imaging (NCT03134846) were enrolled in the study. Approval for the clinical trial was obtained at the Institutional Review Board of the UMCG (METc 2016/395). The study was performed following the Dutch Act on Medical Research involving Medical Subjects and the Helsinki Declaration (adapted version 2013, Fortaleza, Brazil). Written informed consent was obtained from all patients prior to any study-related procedure.

Two days before primary tumor resection, patients (n=6) were administered an i.v. bolus injection of 75 mg of unlabeled cetuximab to minimize off-target receptor binding of labeled cetuximab-800CW,2022 followed by 15 mg of cetuximab-800CW conjugate 1 h later. Patient vital signs were monitored prior to and after administration of the study drugs for 1 h and if no complications occurred, the patients were discharged. Surgical resection was performed according to the standard of care. Previous work of the UMCG group showed that fluorescence imaging of surgical specimens does not disturb the standard of care.2325 Short exposure times in the narrow aperture images for the specimens examined from the first three patients resulted in a low signal-to-noise ratio (<5, where the corresponding exposure times were 2, 5, and 10 s), and consequently those samples were removed from analyses, leaving a total of three surgical specimens, which were each analyzed at four distinct locations.

2.2.3.

Patient exclusion criteria

Patients were excluded from the study if they presented with a life expectancy of <12 weeks, had a Karnofsky performance status <70%%, had a history of infusion reactions to monoclonal antibody therapies, had a QT prolongation on screening electrocardiogram, were pregnant, had abnormal electrolyte status, were using class IA or III antiarrhythmic drugs, were administered any investigational drugs within 30 days prior to the infusion of cetuximab-800CW, had any episodes within 6 months prior to enrollment (such as myocardial infarction and/or cerebrovascular accident), or presented with any uncontrolled medical conditions (including uncontrolled hypertension, significant cardiopulmonary issues, and/or liver disease).

2.2.4.

Specimen basal margin imaging

Immediately following resection, basal surgical margins were imaged using the dAFR system. The height of the stage was adjusted manually to ensure that the basal margin was at the focal plane of the dAFR system (20 cm). Fluorescence signal was first collected at a wide aperture (NA=0.3) followed by a narrow aperture (NA=0.03) at full illumination power. The last three specimens were exposed for 5- and 45-s durations at the wide and narrow apertures, respectively. To maintain specimen moisture, 0.5 ml of 0.9% NaCl solution was dropped onto the specimen at 2-min intervals throughout handling and the excess liquid was carefully blotted off so as to not introduce imaging and motion artifacts in later analysis.

2.2.5.

Pathology

Following imaging, the surgical specimens were submitted to the Department of Pathology for routine histopathological assessment. Specimens were formalin-fixed for at least 24 h (depending on specimen size). The margins of the formalin-fixed surgical specimens were inked for orientation purposes. The specimens were then serially sliced into 3 to 4 mm thick “bread-loaf” slices. To register the sAF and dAFR images with the corresponding pathology data, white light images of the specimens were taken. The locations and orientations of each bread-loaf slice were noted on the white light image and the image was then aligned with the sAF and dAFR images.26 A synopsis of this process is presented in Fig. 2. Each bread-loaf slice was then embedded in paraffin and a 4-μm tissue section was cut from the edge and stained with hematoxylin and eosin (H&E). A pathologist blinded to the fluorescence imaging results inspected the sections to verify the true margin thickness, as per standard-of-care protocol. These measurements were correlated with the sAF and dAFR image values at each corresponding location. For three of the specimens, margin thickness was measured in four locations at the discretion of the pathologist, including the location of the closest margin, yielding a total of 12 margin thickness values that were compared to co-registered regions from the sAF and dAFR images. Owing to the complex and heterogeneous anatomy of the imaged specimens, each margin measurement was considered an independent sample with regard to the statistical analyses.

Fig. 2

A depiction of the co-registration process employed to evaluate true margin thickness in locations of interest in a whole excised specimen imaged on the deep margin side. (a) The widefield fluorescent image is aligned with the white light image that contains the bread-loaf slice number. (b) Bread-loaf slice number is noted on the widefield image and locations of interest are chosen. Here the dashed red line corresponds to the boundaries of the basal margin. (c) The histology slices of interest are pulled from records and the margin value each location is assessed. Given the bread-loaf slices were 3 to 4 mm thick and the fluorescence signal is being compared with a 4-μm section, and further given the uncertainty in the position of the slice within each noted bread loaf, here each ROI has been drawn an 1.5  mm in diameter, with pixel values chosen within the 2  mm of uncertainty in true margin location.

JBO_29_1_016003_f002.png

2.3.

Image Analyses

2.3.1.

Single aperture fluorescence analysis

Single “wide” aperture (NA=0.3) fluorescence images first underwent a flat-field correction by a pixel-by-pixel division of a background image.27 This was followed by segmentation of the basal margin from the whole image by the surgical team, where other tissue types (e.g., bone, mucosa) were masked out of the fluorescence images to ensure the image analyses included only the basal margins. Within each basal margin, a background region of interest (ROI) was drawn that included adjacent tissue from the same origin (e.g., connective tissue or muscle). sAF measurements were carried out by dividing the mean pixel value at the locations interest (those delineated by the pathologist) by the background intensity, yielding a signal-to-background ratio (SBR), as described in detail by Ref. 15.

2.3.2.

Dual aperture fluorescence ratio analysis

In addition to the wide aperture images used in the sAF analyses, a second image was taken with a “narrow” aperture (NA=0.03), which also underwent a flat-field correction. A 3×3-pixel median filter was then applied to narrow aperture images during post-processing to reduce the contribution of noise in the final analyses. A third “dAFR” image was then created by taking a pixel-by-pixel ratio of the wide aperture divided by the narrow aperture. Measurements from this image were taken directly from the test ROIs with no background normalization.

2.3.3.

Statistics

Statistical analyses of the results were done by constructing receiver operating characteristic curves (ROCs) for sAF and dAFR measures of margin status versus true margin depth provided by pathological assessment and areas under the curve (AUCs) were calculated. Statistical comparison between sAF and dAFR AUCs was carried out using the pre-validated iMRMC software,28 with which a Student’s t-test was conducted, yielding confidence intervals and p-values to evaluate the existence of statistically significant differences between AUC of the ROC of different datasets. Pearson’s correlation coefficients were calculated to evaluate the correlation between sAF or dAFR values and the true margin thicknesses as determined by histopathology.

3.

Results

Patient demographics, clinical data, and histopathological data are provided in the Supplemental Material for the three patient specimens and 12 margins assessed (four margins per specimen). Representative sAF and dAFR images of a specimen alongside its corresponding pathology data are shown in Fig. 3. Figure 4 illustrates the depth versus intensity for both sAF and dAFR measurements. Statistically significant linear correlations were observed between dAFR values and actual margin thickness [Fig. 4(b); r=0.72, p=0.0052], as well as sAF values and actual margin thickness [Fig. 4(a); r=0.61, p=0.022], and the difference in the strength of the correlations was not found to be statistically significant.

Fig. 3

In a whole excised specimen, widefield images of (a) sAF, (b) dAFR, and (c) the associated pathology for each of the noted locations. Here, we note positive margins in red, close margins in blue, and clear margins in green. The boundary of the basal margin is noted with a dashed red line.

JBO_29_1_016003_f003.png

Fig. 4

SBR versus true depth of margin for (a) sAF and (b) dAFR.

JBO_29_1_016003_f004.png

The dAFR ROC [Fig. 5(b)] curve yielded an AUC of 1.0 for each “positive,” “close,” and “inadequate” (any <5  mm thick) margins. For sAF [Fig. 5(a)], the AUCs of the ROCs were 1.0 for “positive” margins, 0.64 for “inadequate,” and 0.57 for “close” margins. The iMRMC software indicated that the improved detection of “close” margins with dAFR compared to sAF was statistically significant (p<0.02), yielding an estimated 95% confidence interval of the AUC of ROC difference to be between 0.10 and 0.63 if extended to a larger population.

Fig. 5

Corresponding ROC analysis for (a) sAF and (b) dAFR. sAF measurements yield AUCs of 1.0 for positive margins (<1  mm), 0.64 for inadequate margins (0 to 5 mm), and 0.57 for close margins (1 to 5 mm). dAFR measurements yield AUCs of 1.0 for all margin classifications.

JBO_29_1_016003_f005.png

An optimal threshold value was found using Youden’s index for both sAF and dAFR measures.29 An SBR of 1.3 was calculated for sAF, while a dAFR value of 0.63 were found to be ideal for discrimination between “close” and “clear” margins. These thresholds yielded a sensitivity of 100% and 100%, a specificity of 50% and 100%, a positive-predictive value of 63% and 100%, and a negative-predictive value of 100% and 100%, for sAF and dAFR metrics, respectively.

4.

Discussion

This study shows that a dAFR approach provides significant improvements in the detection of close margins compared to standard sAF imaging in OSCC patients injected with cetuximab-800CW prior to surgery (p<0.02) in <1  min of imaging time. This is the first time dAFR has been used in a clinical trial to directly analyze surgical margins, based on these statistically significant improvements in the detection of close margins, could have major implications for intra-operative decision making.

Another key advantage of the dAFR technique over sAF is that no subjective selection of a “background” tissue is required, as is need for the sAF measurements. Rather than normalizing a suspicious signal to a background, dAFR includes a normalization of the wide aperture image with the narrow aperture image, providing an intrinsic point-by-point correction for patient-to-patient or location-to-location differences in exposure time, tracer uptake, and/or heterogeneities in specimen illumination and optical properties. Conversely, sAF analyses of SBR (which help to minimize variability among specimens) require clinical experts to select ROIs in a suspicious region and an appropriate “background” region (one suspected to have a clear margin). This selection process requires a trained and experienced operator, ultimately limiting standardization and reproducibility of results, which is key to ensuring clinical translation of the approach.30,31

While an AUC of the ROC of 1.0 for close margin detection with dAFR is impressive, this is a small sample size and so expansion of this work is required for a more definitive estimation of detection accuracy.32 Yet, although the study was limited to three patient specimens, a more representative sample size of 12 was approximated by evaluating each specimen at four distinct locations. While these are technically “within-subjects” variables, the optical property heterogeneity within any single sample is expected to be representative of the variability across multiple samples owing to the complex anatomy of the specimens imaged. Furthermore, given the statistical significance of the results presented here from a head-to-head comparison with the state-of-the-art SBR of a single wide aperture image, these results are encouraging and merit expansion and further investigation in future work.

In terms of evaluating positive margin detection, both sAF and dAFR were able to distinguish the single identified positive margin in the data set with an AUC of the ROC of 1.0; however, since there was only a single case, it was not possible to evaluate these results statistically. It should be noted that in the larger sAF study,15 the AUC of the ROC for positive margin detection accuracy was found to be 0.91, and it is not expected that dAFR need to outperform sAF for positive margin detection for it to be clinically useful, as long as it provides a better detection of close margins, as observed in this pilot study.

While the potential of dAFR for improving intraoperative margin assessment of close margins is evident, this pilot study has some limitations, including the potential for low photon budget when imaging at the narrow aperture and the ability to adapt dAFR to specimens with highly heterogeneous surface topographies. In the context of a low photon budget, the PCO camera used here, while capable of imaging with a relatively low read-noise of 2.3  rmse-, was only capable of generating sufficient contrast in the narrow numerical aperture images with long exposure times (at least 45 s per image). To combat this, one could use a more advanced qCMOS detector, which can be capable of “single photon counting” sensitivity and could yield a greater than four time improvement in SNR and allow for shorter imaging times. Second, a more powerful light source would also improve the system used in this work, which employed a light source two to three orders of magnitude lower than the ANSI safety limit. With improved photon budgets at narrow apertures, it is conceivable that the relationship between dAFR and margin thickness can be improved substantially either through greater SNR or enhanced narrowing of the narrow-aperture image, which is also expected to improve discrimination/correction, as it can reject more highly scattered photons. A full optimization of what is possible in clinically relevant imaging times and without damaging samples is ongoing using Monte Carlo simulations.33

To partially account for heterogeneous margin surface topographies, the designed system was fitted with a heavy-duty rotation motor to which a height-adjustable stage was attached. The height-adjustable stage allowed for imaging of specimens that vary from 1 to 10 cm thickness while keeping the surface of the specimen in the focal plane of the lens (where depth of field was 1 and 3  cm, respectively, for the narrow and wide aperture images). It should also be noted that the heavy-duty rotation mount presents the ability to image the surface of the specimen at multiple angles, opening the possibility to correct for surface topography34 or for limited-angle tomography35 that could be combined with dAFR to further improve depth estimation in future.

Disclosures

The authors have no conflicts of interest to disclose.

Code and Data Availability

All code, data, and materials used in this project and analysis can be obtained by direct email to the corresponding author, Dr. Kenneth M. Tichauer at ktichaue@iit.edu.

Acknowledgments

Funding for this project was received from the Pritzker Institute for Biomedical Science, as well as NIH R37CA212187.

References

1. 

H. Sung et al., “Global cancer statistics 2020: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries,” CA Cancer J. Clin., 71 (3), 209 –249 https://doi.org/10.3322/caac.21660 CAMCAM 0007-9235 (2021). Google Scholar

2. 

H. M. Baddour, K. R. Magliocca and A. Y. Chen, “The importance of margins in head and neck cancer: margins in head and neck cancer,” J. Surg. Oncol., 113 (3), 248 –255 https://doi.org/10.1002/jso.24134 JSONAU 0022-4790 (2016). Google Scholar

3. 

J. A. Woolgar and A. Triantafyllou, “A histopathological appraisal of surgical margins in oral and oropharyngeal cancer resection specimens,” Oral Oncol., 41 (10), 1034 –1043 https://doi.org/10.1016/j.oraloncology.2005.06.008 EJCCER 1368-8375 (2005). Google Scholar

4. 

D. N. Sutton et al., “The prognostic implications of the surgical margin in oral squamous cell carcinoma,” Int. J. Oral Maxillofacial Surg., 32 (1), 30 –34 https://doi.org/10.1054/ijom.2002.0313 (2003). Google Scholar

5. 

A. S. Pierik, C. R. Leemans and R. H. Brakenhoff, “Resection margins in head and neck cancer surgery: an update of residual disease and field cancerization,” Cancers, 13 (11), 2635 https://doi.org/10.3390/cancers13112635 (2021). Google Scholar

6. 

M. J. Bullock et al., “Data set for the reporting of nodal excisions and neck dissection specimens for head and neck tumors: explanations and recommendations of the guidelines from the international collaboration on cancer reporting,” Arch. Pathol. Lab. Med., 143 (4), 452 –462 https://doi.org/10.5858/arpa.2018-0421-SA APLMAS 0003-9985 (2019). Google Scholar

7. 

T. Helliwell et al., “Data set for the reporting of carcinomas of the hypopharynx, larynx, and trachea: explanations and recommendations of the guidelines from the international collaboration on cancer reporting,” Arch. Pathol. Lab. Med., 143 (4), 432 –438 https://doi.org/10.5858/arpa.2018-0419-SA APLMAS 0003-9985 (2019). Google Scholar

8. 

M. L. Hinni et al., “Surgical margins in head and neck cancer: a contemporary review,” Head Neck, 35 (9), 1362 –1370 https://doi.org/10.1002/hed.23110 (2013). Google Scholar

9. 

M. Ooms et al., “Impact of secondary tumor‐free resection margins in soft tissue on local, regional, and distant recurrence in R0‐resected oral squamous cell carcinoma,” Head Neck, 44 (5), 1172 –1181 https://doi.org/10.1002/hed.27014 (2022). Google Scholar

10. 

P. Patil et al., “Accuracy of frozen section analysis in correlation with surgical pathology diagnosis,” Int. J. Res. Med. Sci., 3 (2), 399 https://doi.org/10.5455/2320-6012.ijrms20150203 (2015). Google Scholar

11. 

R. A. Ord and S. Aisner, “Accuracy of frozen sections in assessing margins in oral cancer resection,” J. Oral Maxillofacial Surg., 55 (7), 663 –669 https://doi.org/10.1016/S0278-2391(97)90570-X (1997). Google Scholar

12. 

E. Mahe et al., “Intraoperative pathology consultation: error, cause and impact,” Can. J. Surg., 56 (3), E13 –E18 https://doi.org/10.1503/cjs.011112 CJSUAX 0008-428X (2013). Google Scholar

13. 

S. van Keulen et al., “Rapid, non-invasive fluorescence margin assessment: optical specimen mapping in oral squamous cell carcinoma,” Oral Oncol., 88 58 –65 https://doi.org/10.1016/j.oraloncology.2018.11.012 EJCCER 1368-8375 (2019). Google Scholar

14. 

J. Vonk et al., “Improving oral cavity cancer diagnosis and treatment with fluorescence molecular imaging,” Oral Dis., 27 (1), 21 –26 https://doi.org/10.1111/odi.13308 (2021). Google Scholar

15. 

J. G. De Wit et al., “EGFR-targeted fluorescence molecular imaging for intraoperative margin assessment in oral cancer patients: a phase II trial,” Nat. Commun., 14 (1), 4952 https://doi.org/10.1038/s41467-023-40324-8 NCAOBW 2041-1723 (2023). Google Scholar

16. 

L. Quan and N. Ramanujam, “Relationship between depth of a target in a turbid medium and fluorescence measured by a variable-aperture method,” Opt. Lett., 27 (2), 104 https://doi.org/10.1364/OL.27.000104 OPLEDP 0146-9592 (2002). Google Scholar

17. 

C. Rounds et al., “Rapid, variable aperture approach to quantify depth of fluorescence in a heterogenous medium,” Proc. SPIE, 12361 1236105 https://doi.org/10.1117/12.2650919 PSISDG 0277-786X (2023). Google Scholar

18. 

S. L. Jacques, “Optical properties of biological tissues: a review,” Phys. Med. Biol., 58 (11), R37 –R61 https://doi.org/10.1088/0031-9155/58/11/R37 PHMBA7 0031-9155 (2013). Google Scholar

19. 

M. D. Linssen et al., “Roadmap for the development and clinical translation of optical tracers Cetuximab-800CW and Trastuzumab-800CW,” J. Nucl. Med., 60 (3), 418 –423 https://doi.org/10.2967/jnumed.118.216556 JNMEAQ 0161-5505 (2019). Google Scholar

20. 

F. J. Voskuil et al., “Fluorescence-guided imaging for resection margin evaluation in head and neck cancer patients using cetuximab-800CW: a quantitative dose-escalation study,” Theranostics, 10 (9), 3994 –4005 https://doi.org/10.7150/thno.43227 (2020). Google Scholar

21. 

L. S. Moore et al., “Effects of an unlabeled loading dose on tumor-specific uptake of a fluorescently labeled antibody for optical surgical navigation,” Mol. Imaging Biol., 19 (4), 610 –616 https://doi.org/10.1007/s11307-016-1022-1 (2017). Google Scholar

22. 

G. Lu et al., “Co-administered antibody improves penetration of antibody–dye conjugate into human cancers with implications for antibody–drug conjugates,” Nat. Commun., 11 (1), 5667 https://doi.org/10.1038/s41467-020-19498-y NCAOBW 2041-1723 (2020). Google Scholar

23. 

G. M. van Dam et al., “Intraoperative tumor-specific fluorescence imaging in ovarian cancer by folate receptor-α targeting: first in-human results,” Nat. Med., 17 (10), 1315 –1319 https://doi.org/10.1038/nm.2472 1078-8956 (2011). Google Scholar

24. 

N. J. Harlaar et al., “Molecular fluorescence-guided surgery of peritoneal carcinomatosis of colorectal origin: a single-centre feasibility study,” Lancet Gastroenterol. Hepatol., 1 (4), 283 –290 https://doi.org/10.1016/S2468-1253(16)30082-6 (2016). Google Scholar

25. 

F. J. Voskuil et al., “Exploiting metabolic acidosis in solid cancers using a tumor-agnostic pH-activatable nanoprobe for fluorescence-guided surgery,” Nat. Commun., 11 (1), 3257 https://doi.org/10.1038/s41467-020-16814-4 NCAOBW 2041-1723 (2020). Google Scholar

26. 

M. Koller et al., “Implementation and benchmarking of a novel analytical framework to clinically evaluate tumor-specific fluorescent tracers,” Nat. Commun., 9 (1), 3739 https://doi.org/10.1038/s41467-018-05727-y NCAOBW 2041-1723 (2018). Google Scholar

27. 

D. E. Wolf, C. Samarasekera and J. R. Swedlow, “Quantitative analysis of digital microscope images,” Methods Cell Biol., 81 365 –396 https://doi.org/10.1016/S0091-679X(06)81017-4 MCBLAG 0091-679X (2007). Google Scholar

28. 

RST Catalog, “iMRMC: software for the statistical analysis of multi-reader multi-case studies,” (2022). Google Scholar

29. 

W. J. Youden, “Index for rating diagnostic tests,” Cancer, 3 (1), 32 –35 https://doi.org/10.1002/1097-0142(1950)3:1<32::AID-CNCR2820030106>3.0.CO;2-3 CANCAR 0008-543X (1950). Google Scholar

30. 

W. Heeman et al., “A guideline for clinicians performing clinical studies with fluorescence imaging,” J. Nucl. Med., 63 (5), 640 –645 https://doi.org/10.2967/jnumed.121.262975 JNMEAQ 0161-5505 (2022). Google Scholar

31. 

W. S. Tummers et al., “Recommendations for reporting on emerging optical imaging agents to promote clinical approval,” Theranostics, 8 (19), 5336 –5347 https://doi.org/10.7150/thno.27384 (2018). Google Scholar

32. 

J. M. Lobo, A. Jiménez-Valverde and R. Real, “AUC: a misleading measure of the performance of predictive distribution models,” Global Ecol. Biogeogr., 17 (2), 145 –151 https://doi.org/10.1111/j.1466-8238.2007.00358.x GEBIFS 1466-8238 (2008). Google Scholar

33. 

L. Sinha et al., “Comparison of time- and angular-domain scatter rejection in mesoscopic optical projection tomography: a simulation study,” Biomed. Opt. Express, 10 (2), 747 https://doi.org/10.1364/BOE.10.000747 BOEICL 2156-7085 (2019). Google Scholar

34. 

H. Dehghani et al., “Quantitative bioluminescence tomography using spectral derivative data,” Biomed. Opt. Express, 9 (9), 4163 https://doi.org/10.1364/BOE.9.004163 BOEICL 2156-7085 (2018). Google Scholar

35. 

M. Jermyn et al., “Macroscopic-imaging technique for subsurface quantification of near-infrared markers during surgery,” J. Biomed. Opt., 20 (3), 036014 https://doi.org/10.1117/1.JBO.20.3.036014 JBOPFO 1083-3668 (2015). Google Scholar

Biographies of the authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Cody C. Rounds, Jaron G. de Wit, Jasper Vonk, Jennifer Vorjohan, Sophia Nelson, Allyson Trang, Brooke Villinski, Kimberley S. Samkoe, Jovan G. Brankov, Floris J. Voskuil, Max J. H. Witjes, and Kenneth M. Tichauer "Improved intraoperative identification of close margins in oral squamous cell carcinoma resections using a dual aperture fluorescence ratio approach: first in-human results," Journal of Biomedical Optics 29(1), 016003 (17 January 2024). https://doi.org/10.1117/1.JBO.29.1.016003
Received: 19 October 2023; Accepted: 21 December 2023; Published: 17 January 2024
Advertisement
Advertisement
KEYWORDS
Fluorescence

Resection

Tumors

Imaging systems

Surgery

Cancer detection

Pathology

Back to Top