1 July 2003 Embedded progressive-three-layered fiber long-period gratings for respiratory monitoring
Thomas D. P. Allsop, T. Earthrowl-Gould, David J. Webb, Ian Bennion
Author Affiliations +

1.

Introduction

A long-period grating (LPG) is a UV light induced periodic modulation of the refractive index of the core of a single-mode optical fiber with a modulation depth of about 10−4 and a period typically between 100 and 600×10−6 m. The effect of the LPG is to introduce a number of attenuation bands into the fiber’s transmissivity, occurring at wavelengths where the LPG enables phase-matched coupling between the guided core mode and cladding modes.1 The study of the LPG’s attenuation bands has yielded many potential applications. One important area is in fiber-optic telecommunications, where, for example, they have found use as wavelength-selective filters for flattening the gain spectrum of an erbium-doped fiber amplifier.2 Over the past few years, the LPG has also found numerous applications in the field of sensing through its sensitivities to strain, bending and shape, temperature, and the refractive index of the surrounding medium.3 4 5 6 7 The LPG sensitivity can manifest itself in two ways; first through the spectral shift in the attenuation band, which we refer to here as spectral sensitivity, and second through a change in the strength of the attenuation band, which we refer to as a spectral profile variation.

In general, problems can arise when using an LPG that is written into a conventional step-index single-mode fiber, owing to its spectral sensitivity to external influences. For telecommunications and structural monitoring applications, problems occur when the fiber is recoated after the inscription process or embedded in another material. In a step-index fiber, the spectral position and strength of the LPG attenuation bands are dependent upon the surrounding medium’s refractive index (ns) and in the process of recoating the fiber, ns changes, thereby altering the spectral profile and position of the attenuation bands. In some cases the attenuation bands may even disappear. To overcome such problems, there has been some interest in the spectral behavior of LPGs fabricated in multilayered single mode optical fibers. Examples include the use of air-clad8 or depressed-ladding9 fibers to desensitize the structure to changes in the refractive index of the surrounding medium.

Recent reports on the fabrication of LPGs in a single-mode progressive three-layered (PTL) fiber have shown that the lower order cladding modes are insensitive to ns. 10 In this paper, for the first time to the authors’ knowledge, experimental data relating to temperature and bending are presented for an embedded LPG written in the PTL fiber. The intended application for this device is as a shape sensor in monitoring respiratory function, where the movement of the torso during respiration must be monitored. The measurement of lung function is generally based upon the use of spirometric devices that either measure the volume of gases expired (or inspired) at the mouth, or the corresponding flow rate. The measured output requires that an individual blow forcibly into a mouthpiece attached to a recording system and as such provides no more than an averaged representation of lung function behavior and the condition of the respiratory system as a whole.11 12 It is desirable to obtain detailed information on the contribution to breathing of different regions of the chest and abdomen,12 and spirometric devices do not provide this information.

The sensor investigated in this work is intended to be a component of a multiplexed system capable of assessing lung function that avoids the need for flow measurement at the mouth and provides more detailed information on the dynamics of chest motion during breathing. Such a system would facilitate further studies of respiratory physiology since no product is currently available that can provide a completely noninvasive and quantitative appreciation of respiratory function.12

2.

Fabrication and Transmission Spectrum of the LPG Sensor

The PTL optical fiber was supplied by Fibercore Ltd. and was originally designed for a cladding-pumped, erbium-doped fiber laser. This fiber has a core of GeO 2 /SiO 2 with a radius of 2.6×10−6 m, an inner cladding of SiO 2 with a radius of 45.2×10−6 m, and an outer cladding of SiO 2 /F/P 2 O 5 with a radius of 62.5×10−6 m. The fiber is not specifically designed to be photosensitive and so its photosensitivity was increased by hydrogenation at a pressure of 120 bar for a period of two weeks at room temperature. A number of LPGs of various lengths and periods were fabricated in the fiber using the point-by-point writing technique13; a programmable shutter provided this facility. The UV writing beam was generated from a frequency-doubled argon ion laser and its power of 70 mW was focused to a size of approximately 10×10−6 m.

The periods of the LPGs were chosen so that the attenuation bands associated with cladding modes ncl(1,1) to ncl(1,10), which are insensitive to ns, 14 were spectrally located in the range 1400–1550 nm to facilitate interrogation. Two LPGs with periods of 350×10−6 m and 240×10−6 m were embedded to produce a sensor.

Details of the sensor construction are shown in Fig. 1. Assembly was carried out on a base pad of natural latex rubber 250 by 120 mm and with a 2-mm depth. This provided both a flexible stage as well as a thermally insulating layer for monitoring respiratory function (reducing the effects of rapid ambient temperature fluctuations). The two LPGs were fixed using a standard cyanoacrylate adhesive to strips of carbon steel 200×12×0.254 mm. The LPGs were placed in straight grooves along the length of the steel strips to ensure that there was no bending of the fiber during the gluing process. This would ensure that the curvatures and shape changes experienced by the LPGs were the same as the steel strips; it was also thought that the steel would help to stabilize the temperature. Using the same adhesive, these strips were fixed parallel to each other on the rubber base 75 mm apart. Two additional steel strips 80 mm long provided a perpendicular link between the sensors at either end. This framework lent longitudinal rigidity to the system, the steel having an elastic modulus of 210 GPa,15 and this was designed to prevent the fiber (elastic modulus ∼70 GPa) from experiencing significant axial strain. The whole setup was then sealed on the sensor side of the rubber base with a room temperature vulcanizing clear silicone rubber to a depth of approximately 3 mm. The refractive index of the silicone encapsulation was known to be greater than 1.5 as a result of previous experiments involving embedded LPGs written in standard fiber, where the attenuation bands were observed to disappear.

Figure 1

The construction of the long-period grating sensor.

021303j.1.jpg

The PTL fiber with the LPG was fusion spliced at both ends to a standard single-mode telecom optical fiber. The spectral characteristics of the attenuation bands before and after embedding were obtained by illuminating the LPG using a broadband light source and observing the transmission spectrum with an optical spectrum analyzer (OSA) with an accuracy of ±0.04 nm (see Fig. 2).

Figure 2

The transmission spectrum of an LPG (period of 350×10−6 m) written in the PTL fiber before and after embedding; an offset of 6 dBm is included on the transmission spectrum of the LPG in air for clarity.

021303j.2.jpg

The two transmission spectra in Fig. 2 show that the central wavelengths of the observed attenuation bands (1,2), (1,3), and (1,4) of the LPG have not visibly had their locations changed as a result of being embedded. The spectral profiles have altered, but not significantly. Changing the shape or position of the standard single-mode telecom optical fiber input lead alters the spectral profile (although not the position) of the LPG’s attenuation bands. This effect may be due to the setting up of an interferometric cavity formed as a result of Fresnel reflections from the splices connecting the standard and PTL fibers. Further refinement of the fusion splicing of these two fibers should reduce this effect; alternatively, the PTL fiber could be used for the whole of the sensing fiber.

3.

Spectral Sensitivity of an LPG Sensor

The LPG sensor was subjected to three measurands—temperature, compression, and bending—the sensitivities to all of which are important for the intended application. The sensitivity of the bare-fiber LPG to temperature had already been investigated14 by placing the fiber on a hot plate monitored using a probe with an accuracy of ±0.1°C. The spectral location of the central wavelength of the attenuation band associated with the fourth cladding mode was monitored and its sensitivity determined to be 0.04±0.005 nm °C−1.

The temperature sensitivity of the sensor unit was measured by immersing the sensor in a bath of hot water and monitoring the wavelength as the water cooled. This approach gave a sensitivity of (dλ/dT)=0.0952±0.5×10−3 nm °C−1, roughly twice that of the bare-fiber LPG. This increase in sensitivity (which for our application is of course undesirable) is due to the bonding of the fiber onto the sensor, for although the variation of refractive index with temperature will not be affected by the embedding, the axial thermal expansion of the LPG will be closer to that of the steel (typical value: 5×10−6 °C−1),16 than glass (typical value: 4×10−7 °C−1).14

It should be noted, however, that this way of determining the temperature sensitivity may not necessarily correspond all that well to the way in which the sensor will be used in practice. The sensing fiber is embedded inside an insulating material and in the intended application, one side of this material may well be close to body temperature while the other side is exposed to the vagaries of the environment. To illustrate the effect of this, we measured the temperature sensitivity of the sensor lying on a hot plate (as was done with the bare fiber) with the other surface exposed to room temperature. In this configuration, the temperature of the fiber will be somewhere between that of the hot plate and that of the ambient air. Measured in this way the spectral temperature sensitivity was only 0.02 nm °C−1 with the natural rubber side on the hot plate and 0.04 nm °C−1 with the sensor inverted. The results are illustrated in Fig. 3, along with the normal temperature variation of the skin.17 The sensor is intended to be worn close to the skin and be insulated from the surrounding air, and so this is the temperature range that the sensor must normally accommodate.

Figure 3

The spectral sensitivity of an LPG (period of 350×10−6 m) written in the PTL fiber and embedded in the sensor. (a) The sensor’s silicone rubber in contact with a hot plate. (b) The sensor’s natural rubber in contact with a hot plate. The error bars indicate the accuracy of the OSA (±0.04nm).

021303j.3.jpg

The second investigation was to determine the influence of transverse compression on the sensor. This was achieved by placing flat weights on the top surface (the silicone encapsulation layer) of the sensor, which was itself placed on an optical bench (see Fig. 4). For the loads applied (up to 3.6 kg, much higher than any load likely in the intended application), there was no observable wavelength shift or change in the spectral profile of the LPG’s attenuation band. The error bars shown in Fig. 4 again indicate the accuracy of the OSA used.

Figure 4

Spectral stability of embedded LPG (period 350×10−6 m) under compression.

021303j.4.jpg

The third investigation was to obtain the spectral sensitivity of the sensor to bending, and this was achieved using the experimental setup shown in Fig. 5. The sensor platform was clamped between two towers; one of the clamps was mounted on a translation stage, which was moved inward to induce a bend in the optical fiber. For this arrangement where the LPG is midway between the clamps, the sensor’s curvature, R, is given by18

Eq. (1)

R=2×d(d2+L2),
where L is the half-distance between the edges of the two towers and d is the bending displacement at the center of the LPG.

Figure 5

Schematic of bending apparatus.

021303j.5.jpg

The results for this experiment, along with a theoretically predicted wavelength shift, are shown in Fig. 6. The theoretical wavelength shift was obtained using the same approach as in Ref. 19 and is given by

Eq. (2)

Δλ=[λ(δneffδng)×dδneffdR+(δneff)3δng ddR]ΔR,
where δn eff =n coeff −n cleff ν is the differential effective index between the cladding and the core mode and δng=n cog −n clg ν is the differential group index. The effective refractive indices of the core and fourth cladding mode were calculated as a function of curvature using a 2-D curvilinear hybrid mode eigenvalue equation20 21 using the optical constants given in Ref. 14 and assuming that the amount of birefringence induced in a typical single-mode fiber is negligibly small for curvatures of <2 m−1 (no splitting of the attenuation bend). Figure 6 demonstrates reasonable agreement between the theoretical wavelength shift calculated at a number of different curvatures and the experimental results. Note that the small amount of “noise” visible on the theoretical data is a result of computational truncation errors, owing to the limited precision of the Mathcad software. Mathcad performs all computations in double precision (floating point: 64 bits), which causes errors to occur when using the built-in Bessel functions and their cross products with small arguments. Over the range of curvature investigated, the spectral sensitivity of the sensor was found to be linear, with (dλ/dR)=3.747±0.006 nm m.

Figure 6

The central wavelength of the attenuation band (fourth mode) of the LPG with period 350×10−6 m embedded in the sensor as a function of curvature.

021303j.6.jpg

4.

Respiratory Monitoring

In order to validate this sensor design for the intended application, a commercial manikin used as a resuscitation training aid was employed.22 This manikin consists of a rigid underframe over which is stretched a polymer skin. An air bag is placed between the frame and skin and when inflated or deflated is used to simulate expansion and contraction of the surface of the torso in volumetric proportions similar to those of breathing. This provides a test platform for changes in dimension and shape similar to those that are expected in humans.

The response of the sensor was recorded at various states of inflation at five different orientations on the manikin. Each location is represented by a set of dimensions as indicated in Fig. 7 and Table 1. The shift in central wavelength was also recorded with reference to the peripheral expansion of the dummy’s skin in Fig. 8 and Fig. 9. The variation in response apparent between the upper and lower chest regions mimics that which might be expected in a real subject since the expansion of the rib cage has a more significant contribution at higher levels of ventilation.23 In our experiment, however, this phenomenon can be attributed to the profile and position of the internal airbag.

Figure 7

The positioning of the LPG sensor platform on the manikin.

021303j.7.jpg

Figure 8

Spectral functionality of the LPG sensor positioned on the lower chest.

021303j.8.jpg

Figure 9

Spectral functionality of the LPG sensor positioned on the upper chest.

021303j.9.jpg

Table 1

The measured central wavelength shift of the LPG sensor at various positions on the manikin.
Position on Dummy
Fig. 7(a)
d1
(mm)
d2
(mm)
d3
(mm)
d4
(mm)
Maximum detected wavelength shift
(nm)
1 90 50 50 250 1.05
2 185 50 50 185 3.18
3 287 50 50 75 0.95
Position on Dummy
Fig. 7(b)
p1
(mm)
p2
(mm)
p3
(mm)
p4
(mm)
Maximum detected wavelength shift
(nm)
4 100 25 165 165 1.98
5 205 128 165 165 1.70

An inspection of Fig. 8 and Fig. 9 shows that the spectral response of the LPG sensor as a function of peripheral expansion of the manikin’s skin varies from position to position. The errors illustrated in these figures correspond to the spectral accuracy of the OSA (±0.04 nm) and a circumference error of ±1 mm. The circumference error is an estimate of the variation of the manikin’s skin deformation between each set of results. The curves shown in Fig. 8 and Fig. 9 are presented simply as guides for the eye to better illustrate the different responses at the various locations.

As noted earlier, the sensor is temperature sensitive. In normal use it would be worn next to the skin and insulated from the environment, so that it would be subjected to a likely temperature variation of only 32 to 35 °C.17 The maximum wavelength uncertainty associated with this temperature range is about ±0.14 nm (using the temperature sensitivity of 0.095 nm °C−1). Combining this with the curvature resolution of ±1.1×10−2 m−1 for the sensing scheme leads to a relative error in the circumference that varies between ±0.3 and ±2.3, corresponding, respectively, to positions 1 and 3 on the manikin.

To improve on the accuracy, we plan to investigate two approaches to dealing with the temperature cross-sensitivity. First the steel supports are to be replaced by carbon-fiber sheets in which the thermal expansion coefficient is typically in the range of ±4×10−7 °C−1, depending on the material’s composition. This should in principle allow the temperature sensitivity of the sensor platform to be significantly reduced. Second, a fiber Bragg grating may be used to discriminate between bending and temperature.24

A conservative estimate of the number of sensors required for the intended application was made using chest profile data obtained from a computed tomography (CT) imager. A single-line profile from the upper chest of a male subject was chosen because it constituted the most complex geometry from the image data. This profile was then modeled using a seventh-order polynomial, and the profile length of the curve was evaluated symbolically as a line integral to give the cross-sectional area and then squared for the volumetric error. This value was then compared with the same line evaluated numerically (simple trapezoidal integration) at varying discrete intervals along the profile. Assuming that an identical longitudinal and lateral resolution is needed to represent the overall profile of the torso by squaring the results, Fig. 10 depicts the total number of sensors required versus volumetric error. This error may be reduced further with the application of cubic-spline interpolation. Linking this to generally accepted volumetric performance standards for spirometric devices,25 about twenty such sensors would be required for this application using cubic-spline interpolation. Of course the resuscitation manikin provides only a simple approximation to the behavior of the human torso during respiration. The location and number of sensors required to monitor respiratory function within acceptable standards25 will need to be investigated experimentally using volunteers instrumented with an array of improved sensor platforms working in conjunction with a suitable multiplexing and interrogation scheme.

Figure 10

Density of sensing positions versus volumetric error for reconstructed CT scan chest profile data. Solid and broken curves represent the errors from the trapezoidal approximation and 4-point cubic spline interpolation, respectively.

021303j.10.jpg

There are several possible multiplexing and interrogation techniques that could be used. We have considered the use of a broadband light source illuminating a series of in-line LPGs combined with a miniature fiber-optic spectrometer (e.g., those from Ocean Optics26) to record the transmission spectra of the LPGs. Our concern is that the limited spectral resolution of these spectrometers would seriously limit the curvature resolution of the interrogation scheme. The sensing scheme that is being investigated is based upon a derivative spectroscopy technique in which an in-line series of sensing LPGs are being interrogated by spectrally matched fiber Bragg gratings.27

For the intended application, a number of alternative (nonoptical) sensing elements could also in principle be used. Most schemes are based on strain measurement (e.g., capacitance28 and resistance29 gauges or bistable switches30) and as such are not directly applicable to surface profile reconstruction because a dynamic model of the motion of the thorax and abdomen during breathing would be required on which localized strain measurement could be used to infer changes in geometry. It is a considerable task to create a model that would accurately describe respiratory interactions over a broad range of subjects. The solution detailed here sidesteps the need for such a model because geometry is the fundamental measurand and the chest surface profile can be obtained directly from sensor signals.

While other optical-based bend-sensing elements have been developed (e.g., the Shape Tape developed by the Canadian Space Agency31), these do not possess the ability to be serially multiplexed, relying as they do on curvature-dependent attenuation, and commercially available configurations of these sensors do not lend themselves to this application because they are arranged linearly on one strip of steel, and that platform cannot be stretched. The curvature resolution of the Shape Tape is about 10−3 m−1, which is comparable to the interrogation scheme we have under investigation.27 Our approach allows the simple multiplexing of the sensors, providing the ability to scale up the approach to multiple sensors without a dramatic increase in the size of the peripheral electronics.

6.

Conclusion

The spectral characteristics of an embedded long-period grating fabricated in a PTL fiber were investigated. The spectral location and profile of the attenuation bands were found to be unaffected by embedding the LPG, although the temperature sensitivity was increased. A curvature spectral sensitivity of 3.747 nm m with a curvature resolution of ±1.1×10−2 m−1 was found for this sensor. Axial strain for this sensor in the intended application can be assumed to be negligible because the sensor is fixed to an axially rigid steel support.

Normal skin temperature variations limit the accuracy of the sensor to a worst-case equivalent error in a circumference measurement of ±2.3. Where this is insufficient or where it cannot be guaranteed that the sensor will be in contact with the skin, we have identified two avenues toward improving or controlling the temperature sensitivity—the use of a carbon fiber support frame or the incorporation of an additional Bragg grating temperature sensor.

It has also been demonstrated that such a sensor can be used to distinguish among the geometric variations associated with different locations on a simulated torso during respiratory movement. We have shown that, in theory, an array of approximately twenty such sensors should be sufficient to generate a geometric profile of the chest and abdomen in three dimensions with the necessary accuracy.

Acknowledgment

The authors would like to thank Tim Hart of Fibercore Ltd. for fruitful discussions and supplying the material constants and specification of the PTL fiber. We also thank M. R. Miller M.D., Consultant in respiratory medicine, Selly Oak Hospital, Birmingham, UK, for the computed tomography image data. This work was funded by the UK Engineering and Physical Sciences Research Council.

REFERENCES

1. 

S. Vasilev and O. Medvedkov , “Long-period refractive index fiber gratings: properties, applications and fabrication techniques,” Proc. SPIE , 4083 212 –223 (2000). Google Scholar

2. 

Y. Ruan, X. Huang, Y. Jiang, and J. Wang, “The design and simulation of cascade long-period fiber gratings for flattening EDFA’s gain,” Proc. SPIE 3801, 240–245 (1999).

3. 

A. Vengsarkar , P. Lemaire , J. Judkins , V. Bhatia , and T. Erdogan , “Long-period gratings as band-rejection filters,” J. Lightwave Technol. , 14 (1), 58 –64 (1996). Google Scholar

4. 

V. Bhatia and A. M. Vengsarkar , “Optical fibre long-period grating sensors,” Opt. Lett. , 21 (9), 692 –694 (1996). Google Scholar

5. 

H. J. Patrick , A. D. Kersey , and F. Bucholtz , “Analysis of the response of long period fiber gratings to external index of refraction,” J. Lightwave Technol. , 16 (9), 1606 –1612 (1998). Google Scholar

6. 

V. Grubsky and J. Feinberg , “Long-period gratings with variable coupling for real-time sensing applications,” Opt. Lett. , 25 (4), 203 –205 (2000). Google Scholar

7. 

V. Bhatia , T. D’Alberto , N. Zabaronick , and R. Claus , “Temperature-insensitive and strain-insensitive long-period gratings sensor for smart structures,” Opt. Eng. , 36 1872 (1997). Google Scholar

8. 

R. P. Espindola , R. Windeler , A. Abramov , B. Eggleton , T. Strasser , and D. DiGiovanni , “External refractive index insensitive air-clad long period fibre grating,” Electron. Lett. , 35 (4), 327 –328 (1999). Google Scholar

9. 

L. Dong , L. Reekie , and J. L. Cruz , “Long period gratings formed in depressed cladding fibers,” Electron. Lett. , 33 (22), 1897 –1898 (1997). Google Scholar

10. 

J. Jang , Se Yoon Kim , Sun-Wook Kim , and Min-Sung Kim , “Temperature insensitive long-period fibre gratings,” Electron. Lett. , 35 (24), 2134 –2136 (1999). Google Scholar

11. 

T. A. Wilson , K. Rehder , S. Krayer , E. Hoffman , C. Whitney , and J. Rodarte , “Geometry and displacement of human rib,” J. Appl. Physiol. , 62 (5), 1872 –1877 (1987). Google Scholar

12. 

T. Earthrowl-Gould and B. Jones “Chest and abdominal surface motion measurement for continuous monitoring of respiratory function,” in Proceedings of ImechE, Part H, Engineering in Medicine (2001).

13. 

Y. Liu , L. Zhang , and I. Bennion , “Design and realization of long-period grating devices in conventional and high birefringence fibers and their novel applications as fiber-optic load sensors,” IEEE J. Sel. Top. Quantum Electron. , 5 (5), 1373 –1378 (1999). Google Scholar

14. 

T. Allsop , D. J. Webb , and I. Bennion , “Investigations of the spectral sensitivity of long period gratings fabricated in 3-layered optical fiber,” J. Lightwave Technol. , 21 (1), 264 –268 (2003). Google Scholar

15. 

A. Wolfenden , G. M. Ludtka , and L. S. Cook , “An evaluation of three methods to measure the dynamic elastic modulus of steel,” J. Test. Eval. , 23 (3), 176 –179 (1995). Google Scholar

16. 

R. D. Harrison, Book of Data: Chemistry, Physical, Science Physics. Penguin, Bristol, UK (1978).

17. 

T. Miyanaga , W. Urabe , and Y. Nakano , “Simplified human body-model for evaluating thermal radiant environment in a radiant cooled space,” Build Env. , 36 (7), 801 –808 (2001). Google Scholar

18. 

W. Du, H. Tam, M. Liu, and X. Tao, “Long-period fiber grating bending sensors in laminated composite structures,” in Sensory Phenomena and Measurement Instrumentation for Smart Structures and Materials, Proc. SPIE 3330, 284–292 (1998).

19. 

V. Bhatia , “Applications of long-period gratings to single and multi-parameter sensing,” Opt. Express , 4 (11), 457 –466 (1999). Google Scholar

20. 

C. Tsao, Optical Fibre Waveguide Analysis, Oxford, New York (1992).

21. 

J. Ma , W. Tang , and W. Zhou , “Optical-fibre sensor for simultaneous measurement of pressure and temperature: analysis of cross sensitivity,” Appl. Opt. , 35 (25), 5206 –5210 (1996). Google Scholar

22. 

Little Anne CPR Manikin, Laerdal Medical, http://www.laerdal.co.uk

23. 

K. Konno and J. Mead , “Measurement of the separate volume changes of the rib cage and abdomen during breathing,” J. Appl. Physiol. , 22 (3), 407 –422 (1967). Google Scholar

24. 

H. Patrick G. Williams , A. Kersey , J. Pedrazzani , and A. Vengsarkar , “Hybrid fiber Bragg grating/long period fiber grating sensor for strain/temperature discrimination,” Photon. Technol. Lett. , 8 (9), 1223 –1225 (1998). Google Scholar

25. 

American Thoracic Society , Am. Rev. Respir. Dis. , 119 831 –838 (1979). Google Scholar

26. 

USB2000 Fiber Optic Spectrometer, Ocean Optics, Inc., http://www.oceanoptics.com

27. 

T. Allsop, T. Earthrowl-Gould, R. Reeves, D. J. Webb, and I. Bennion, “The interrogation/multiplexing of long period grating curvature sensors using derivative spectroscopy technique based upon fiber Bragg gratings” (in preparation).

28. 

R. P. Kampen, “Bulk-micromachined capacitive servo-accelerometer,” Ph.D. Thesis, Laboratory of Electronic Instrumenta-tion, Delft Univer-sity of Technology, Delft University Press, Delft, The Netherlands (1995).

29. 

Strain gauges, Minebea Co. Ltd., http://www.minebea.co.jp

30. 

U.S. Patent #5563458, Apparatus and method for sensing surface flexure sensing shape and motion with bistable switches that digitize material strain caused by flexure.

31. 

Shape Tape, Measurand Shape Advantage, http://www.measurand.com
©(2003) Society of Photo-Optical Instrumentation Engineers (SPIE)
Thomas D. P. Allsop, T. Earthrowl-Gould, David J. Webb, and Ian Bennion "Embedded progressive-three-layered fiber long-period gratings for respiratory monitoring," Journal of Biomedical Optics 8(3), (1 July 2003). https://doi.org/10.1117/1.1578497
Published: 1 July 2003
Lens.org Logo
CITATIONS
Cited by 33 scholarly publications and 6 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Signal attenuation

Chest

Skin

Cladding

Temperature metrology

Refractive index

Back to Top