21 April 2020 Development and validation of the Lesion Synthesis Toolbox and the Perception Study Tool for quantifying observer limits of detection of lesions in positron emission tomography
Hanif Gabrani-Juma, Zamzam Al Bimani, Lionel S. Zuckier, Ran Klein
Author Affiliations +
Abstract

Purpose: Accurate detection of cancer lesions in positron emission tomography (PET) is fundamental to achieving favorable clinical outcomes. Therefore, image reconstruction, processing, visualization, and interpretation techniques must be optimized for this task. The objective of this work was to (1) develop and validate an efficient method to generate well-characterized synthetic lesions in real patient data and (2) to apply these lesions in a human perception experiment to establish baseline measurements of the limits of lesion detection as a function of lesion size and contrast using current imaging technologies.

Approach: A fully integrated software package for synthesizing well-characterized lesions in real patient PET was developed using a vendor provided PET image reconstruction toolbox (REGRECON5, General Electric Healthcare, Waukesha, Wisconsin). Lesion characteristics were validated experimentally for geometric accuracy, activity accuracy, and absence of artifacts. The Lesion Synthesis Toolbox was used to generate a library of 133 synthetic lesions of varying sizes (n  =  7) and contrast levels (n  =  19) in manually defined locations in the livers of 37 patient studies. A lesion-localization perception study was performed with seven observers to determine the limits of detection with regard to lesion size and contrast using our web-based perception study tool.

Results: The Lesion Synthesis Toolbox was validated for accurate lesion placement and size. Lesion intensities were deemed accurate with slightly elevated activities (5% at 2:1 lesion-to-background contrast) in small lesions (Ø  =  15  mm spheres), and no bias in large lesions (Ø  =  22.5  mm). Bed-stitching artifacts were not observed, and lesion attenuation correction bias was small (−1.6  ±  1.2  %  ). The 133 liver lesions were synthesized in ∼50  h, and readers were able to complete the perception study of these lesions in 12  ±  3  min with consistent limits of detection amongst all readers.

Conclusions: Our open-source utilities can be employed by nonexperts to generate well-characterized synthetic lesions in real patient PET images and for administering perception studies on clinical workstations without the need to install proprietary software.

© 2020 Society of Photo-Optical Instrumentation Engineers (SPIE) 2329-4302/2020/$28.00 © 2020 SPIE
Hanif Gabrani-Juma, Zamzam Al Bimani, Lionel S. Zuckier, and Ran Klein "Development and validation of the Lesion Synthesis Toolbox and the Perception Study Tool for quantifying observer limits of detection of lesions in positron emission tomography," Journal of Medical Imaging 7(2), 022412 (21 April 2020). https://doi.org/10.1117/1.JMI.7.2.022412
Received: 24 September 2019; Accepted: 23 March 2020; Published: 21 April 2020
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Positron emission tomography

Image restoration

Optical spheres

Scanners

Computer simulations

Liver

Data modeling

Back to Top