Open Access
6 March 2021 Improving laser standards for three-photon microscopy
Deano M. Farinella, Arani Roy, Chao J. Liu, Prakash Kara
Author Affiliations +
Abstract

Significance: Three-photon excitation microscopy has double-to-triple the penetration depth in biological tissue over two-photon imaging and thus has the potential to revolutionize the visualization of biological processes in vivo. However, unlike the plug-and-play operation and performance of lasers used in two-photon imaging, three-photon microscopy presents new technological challenges that require a closer look at the fidelity of laser pulses.

Aim: We implemented state-of-the-art pulse measurements and developed innovative techniques for examining the performance of lasers used in three-photon microscopy. We then demonstrated how these techniques can be used to provide precise measurements of pulse shape, pulse energy, and pulse-to-pulse intensity variability, all of which ultimately impact imaging.

Approach: We built inexpensive tools, e.g., a second harmonic generation frequency-resolved optical gating (SHG-FROG) device and a deep-memory diode imaging (DMDI) apparatus to examine laser pulse fidelity.

Results: First, SHG-FROG revealed very large third-order dispersion (TOD). This extent of phase distortion prevents the efficient temporal compression of laser pulses to their theoretical limit. Furthermore, TOD cannot be quantified when using a conventional method of obtaining the laser pulse duration, e.g., when using an autocorrelator. Finally, DMDI showed the effectiveness of detecting pulse-to-pulse intensity fluctuations on timescales relevant to three-photon imaging, which were otherwise not captured using conventional instruments and statistics.

Conclusions: The distortion of individual laser pulses caused by TOD poses significant challenges to three-photon imaging by preventing effective compression of laser pulses and decreasing the efficiency of nonlinear excitation. Moreover, an acceptably low pulse-to-pulse amplitude variability should not be assumed. Particularly for low repetition rate laser sources used in three-photon microscopy, pulse-to-pulse variability also degrades image quality. If three-photon imaging is to become mainstream, our diagnostics may be used by laser manufacturers to improve system design and by end-users to validate the performance of their current and future imaging systems.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Deano M. Farinella, Arani Roy, Chao J. Liu, and Prakash Kara "Improving laser standards for three-photon microscopy," Neurophotonics 8(1), 015009 (6 March 2021). https://doi.org/10.1117/1.NPh.8.1.015009
Received: 9 September 2020; Accepted: 5 February 2021; Published: 6 March 2021