7 April 2023 A comparison of digital-based CNT and analogue-based filament x-ray source in regard to intraoperative specimen x-ray system’s development
Author Affiliations +
Conference Poster
Abstract
In this paper, we demonstrated the comparison of digital-based CNT and analogue-based filament X-ray sources by taking the X-ray images of ACR mammo phantom for developing the intraoperative specimen X-ray system. The X-ray image of ACR mammo phantom taken at 25 kV 1 mA by filament shows overall the better image quality. However, when X-ray images of phantom were compared at 40 kV 1 mA by both sources, it showed that CNT X-ray source showed better image quality because of Aluminum window.
© (2023) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Amar Prasad Gupta, Taewon Kim, Mrinal Bhusal Sharma, Jinho Choi, Jaekyu Jang, Seung Jun Yeo, Jaeik Jung, Jeung Sun Ahn, Beom-Seok Ko, and Jehwang Ryu "A comparison of digital-based CNT and analogue-based filament x-ray source in regard to intraoperative specimen x-ray system’s development", Proc. SPIE 12463, Medical Imaging 2023: Physics of Medical Imaging, 124634F (7 April 2023); https://doi.org/10.1117/12.2655782