1.INTRODUCTIONSince 2007, the Diamond-NOM [1] has played a major role in the Optics Metrology Lab (OML) [2] at Diamond. Metrology feedback has helped to optimally clamp hundreds of X-ray mirrors and diffraction crystals into opto-mechanical holders for beamline operation [3]. With sufficient averaging, the sensitivity of the autocollimator (AC) of the Diamond-NOM can resolve ultra-small angle changes of < 20 nrad. Several NOM profilers now employ two or more AC’s to simultaneously measure the mirror under test and the parasitic angular errors of the scanning stage [4,5]. However, despite reducing the path-length dependency issues, this doesn’t solve all problems: calibration at a National Metrology Institute (NMI) is only valid for a specific curvature and reflectivity of the test mirror. To achieve fast and accurate scanning, we have developed the Diamond-VeNOM (velocity-NOM): a high-speed, slope profiler of X-ray optics which utilizes 10X faster, Elcomat5000 autocollimators (from Moller-Wedel) which are synchronized with movement of the multiple motion stages. The VeNOM concept combines the advantages of previous schemes to further minimize systematic errors. It also incorporates two, new distinct features: the scan head is programmed to execute an arbitrary, free-form speed profile during the scan; and the optic under test is simultaneously pitched in coordination. Arbitrary angular profiles can be purposefully chosen to null instrumental systematic errors of the autocollimators and environmental perturbances, thereby improving measurement accuracy. 2.EXPERIMENTAL2.1Instrument geometryThe classical Diamond-NOM uses a single Elcomat3000 electronic autocollimator (Möller-Wedel, Germany) and a high-grade pentaprism, mounted on an X-axis air-bearing stage (Q-Sys, Netherlands), to scan an aperture-defined region of an autocollimator beam along the length of the optic under test. A secondary, orthogonal, air-bearing translation stage (Y-axis) and motorized pitch stage Ry change the sagittal location and pitch of the mirror relative to the autocollimator beam. Figure 1 shows the upgraded hardware of the new Diamond-VeNOM system which enables high-speed, synchronized operation of all components. Elcomat autocollimators simultaneously monitor all moving components of the system. The first autocollimator, AC1, fixed on the left-hand granite column, observes a large aperture (> 10mm diameter), flat mirror attached to the X-axis stage to monitor small (< 5 μrad PV), parasitic angle errors θy and θz (rotations about the Y and Z axes) during translation. Path-length linearity errors of the AC are minimized for small angles [6]. The second autocollimator, AC2, is mounted on the X-axis scanning stage and directly measures a small region (~ 3.5 mm diameter) of the mirror under test at a nominally fixed distance. Nulling the measured angle using the pitch stage reduces the influence of angle dependent errors of AC2 originating from curvature and / or reflectivity of different optics. The third autocollimator, AC3, attached to the granite Y-stage, observes a large aperture, flat mirror mounted on the pitch stage at a nominally fixed distance. AC3 can be calibrated at an NMI over a wide range of angles for fixed experimental parameters, thereby further reducing systematic error and improving accuracy. 2.2Hardware upgrade and control logicDynamic scanning requires coordination and synchronization of multiple components at rates that surpass the capabilities of the classical NOM’s instrumentation hardware. Consequently, the following upgrades were made to the hardware and software infrastructure of the Diamond-VeNOM:
These upgrades collectively ensure that the VeNOM system is well-equipped to meet the demanding requirements of dynamic scanning, including higher-speed and more precise metrology data acquisition. The control logic of the Diamond VeNOM, as illustrated in Figure 2, is orchestrated within the Malcolm environment. Malcolm is a specialized data coordination framework developed at Diamond, which operates above the EPICS channel access layer, and offers advanced configuration capabilities, including the ability to define arbitrary, freeform motion trajectories, including variable speed and acceleration during each scan. Position encoder readback plays a central role in this control logic. It serves primarily to control the dynamic motion of the X-stage, but can also manage the behaviour of other components, such as the pitch stage and opening/closing of the mechanical shutters. 2.3Data synchronization with mechanical shuttersElcomat autocollimators lack high-speed triggering capabilities. To achieve synchronization between multiple autocollimators and positional encoders, we have incorporated electro-mechanical shutters into the optical path. These shutters are open by default, letting the autocollimator beam pass through. Once triggered by TTL signals from the PandA box, shutters simultaneously block the optical path of all autocollimators for a brief interval (typically 20 milliseconds). The trigger action is programmed to occur when the X-stage reaches a series of user-defined positions. Shutters introduce null values at specific points within each autocollimator data stream. These null values serve as synchronization markers to align all autocollimator data with readings from the position encoders and measured location on the optical surface. This synchronization process is essential for precise correlation and analysis during dynamic scanning operations. Figure 3 demonstrates a representative application of synchronization logic. The upper left image (purple lines) shows all shutters are triggered to close (logical value = 1) when the positional encoder readback confirms that the X-stage has reached specific, user-defined locations of X = 10, 30, 50, 70, and 90 mm. In this example, the X-stage is programmed to execute a non-linear velocity profile (constant acceleration), as shown in the black curve in the lower left chart. Note that although the shutter pulses are evenly distributed in space, they are not evenly distributed in time. As seen in the right column of images in Figure 3, null values are incorporated into each of the autocollimator data streams. 3.RESULTS3.1Noise levels of the Elcomat5000 autocollimatorStability tests were conducted for the new Elcomat5000 autocollimator mounted on a passively damped, pneumatic optical table, situated within a protective enclosure. The autocollimator observed a flat mirror (25 mm diameter) located ~ 200 mm away. A beam-defining iris was placed a few millimeters away from the mirror. As shown in Figure 4, the optical path was optionally shielded to reduce air turbulence and external influences. In Figure 5, the power spectral density (PSD) of the angle measured by the Elcomat5000 is shown for various iris diameters (between 4 and 6 mm), with and without shielding of the air path. For comparison, data from an Elcomat3000 mounted on the Diamond-NOM is included. To improve statistics, the long-duration data stream was batched into smaller temporal blocks. Calculating the average FFT for all blocks (within a specific experimental configuration) reduced random noise to aid visualization of the frequency response of the autocollimators. As expected, path shielding leads to a significant reduction in the PSD amplitude of autocollimator data for all iris diameters, especially below 8 Hz. For the shielded path, the Elcomat5000 data is less noisy than the Elcomat3000 above ~ 0.5 Hz. Two resonance frequencies were detected in the ‘shielded’ datasets at ~44.2Hz and ~79.3Hz. These are induced by floor vibrations (transmitted via the optical table) and the ventilation system (transmitted via the roof of the thermal enclosure). As seen in Figure 5 and Figure 6, measurement noise is asymptotically reduced towards ~25 nrad rms as the aperture diameter was increased. The smallest diameter was in this study was 3 mm, which increased noise levels to ~70 nrad rms. 3.2Mechanical shuttersInitial testing of the mechanical shutters consisted of varying the pulse duration of the shutters to block each autocollimator beam. Ideally, the shortest duration signal would be chosen to minimize loss of information about the optical surface. As seen in Figure 7, the shortest pulse signal from the shutter was just over 20 ms, corresponding to 6 readings from the Elcomat5000 (which records angles at 4 ms intervals). However, variation in the duration of the pulse signal could be desirable in a future scenario to uniquely distinguish between specific locations on the optical surface, or to encode experimental parameters into the data. For example, longer pulses to encode the ends of the optical surface, or variations in the pulse width as a function of the pitch angle. The shutters provided a reliably method to block each autocollimator beam. However, in certain instances, the action of the mechanical shutters influences other aspects of the experimental apparatus. For example, damped oscillations are observed in Figure 7 in the autocollimator data immediately after the shutter opens. Such oscillations decay within a few tenths of a second, but this still causes loss of data about the optical surface. Mechanical vibrations from the activation of one shutter were also observed to influence the performance of other autocollimators. For example, in Figure 8, when only the shutter for AC2 was triggered, an effect was induced in AC1. This is attributed to the fact that the shutter for AC2 and the reflector for AC1 are both mounted on the X-axis air-bearing stage. Conversely, AC3 is impervious to the shutter activity of AC2, since AC3 and its reflector are mounted on the heavy granite of the Y-axis, which is decoupled from the X-axis. Time delays from all system components were calibrated by varying the experimental parameters and observing the effects on the autocollimator data. Contributions to the time-lag include: the X-axis encoder readback rate; mechanical shutter latency; and processing time to receive inputs and trigger outputs. A shutter test was performed to compare the pulse delays relative to the target position. Shutter pulses of 100 ms duration were applied to fly-scans with various velocities of the X-axis. Angle data was aligned along the X-direction by comparison with an earlier, step scan of the same mirror. “Roll-off” slope features, at each end of the optic, were used as alignment fiducials, as seen in Figure 9. Third-order polynomials were removed to aid visualisation. A variable, spatial offset was observed for null values from the target position, which increased with scan velocity. When plotted against time, these spatial delays convert into a temporal delay of 20 – 30 ms between the moment the X-axis passes the target position and the shutter pulse registering in the autocollimator data. Further investigation and optimization of such factors is required to further improve the synchronisation of pulses. 3.3X-stage parasitic errorAnother systematic error to be calibrated in the parasitic angle error of the X-stage. Such errors are well studied for quasi-static step scans, but the motion performance of the X-axis could change for variable speed and accelerations. To investigate, a comprehensive series of tests were performed. The geometry for this investigation is shown in Figure 10. Two autocollimators (Elcomat3000 at AC1, and Elcomat5000 at a modified position for AC3) both monitor the same reflective cube mounted on the X-axis. This setup provides enhanced time / space resolution data for parasitic angle readings, which is particularly useful when monitoring higher speed scans. Figure 11 shows the parasitic angles during translation of the X-axis translation, as measured by AC3, for different types of motion. Each curve is the average of 20 repeated scans. The parasitic angle error of the X-axis has a peak to valley of ~1.5 µrad. The fly-scans at 20 mm s-1 (maximum speed of the X-stage) have a slight difference in the low frequency component compared to the lower speeds, but there is very good repeatability between all other speeds. The 1 mm s-1 fly-scan is most comparable with the pseudo-static step-scan. There is also the expected loss of high frequency features for the faster scans. To assess whether the Elcomat3000, with its lower acquisition rate, is well-suited for dynamic scanning, a comparison is shown in Figure 12. For a fly-scan speed of 1 mm s-1 there is excellent agreement, despite the significant difference in acquisition rates. These preliminary results provide confidence about the suitability of the Elcomat3000 for dynamic scanning. 3.4Motorized pitch stageThe motorized pitch stage in this study is a Kohzu goniometer with a stepper motor operating in open-loop. With AC3 repositioned in the geometry shown in Figure 1, the motion characteristics of the pitch stage were investigated. Angle stability of the pitch stage was assessed after moving to a demand angle. Angular drift of ~ 120 nrad (peak-to-valley) was observed over a period of 50 minutes. The process of returning the pitch stage to a specific angle (always from the same starting point) was repeated multiple times, and good repeatability was achieved. The average of such scans is shown in Figure 13. This demonstrates the necessity to continuously monitor angular drifts of the pitch stage, even for constant demand angles. To assess its dynamic performance, the pitch stage was rotated at a rate of 5 µrad/s, whilst monitoring the actual angle using AC3. Figure 14 shows the linearity errors of the pitch stage, which is the difference between the demand and measured angles. This error needs to be compensated when calculating the motion profile of the X-axis to match to the slope range of the mirror under test. In conclusion, the simple, motorized pitch stage has some limitations, both in static and dynamic operational modes, but these issues can easily be corrected using the real-time measurement from AC3. Such corrections confirm the basic suitability of the hardware for dynamic scanning via the new VeNOM concept. A new, in-house goniometer is being built. Once commissioned, closed-loop operation will provide superior stability and resolution of pitch, roll and yaw adjustment of the mirror under test. 4.CONCLUSIONSWe have developed a new instrument: the Diamond-VeNOM which utilises multiple autocollimators, synchronised with motion stage movements, to rapidly measure the optical surface and parasitic motion errors. The optic under test is placed on a motorized tilt stage, and synchronization of motion trajectories and autocollimator data is achieved by temporarily blocking the beam paths of the autocollimators using high-speed electronic shutters, based on triggering signals from positional encoders. Enhanced motion control capabilities enable freeform velocity profiles of the scanning stage, alongside coordinated pitch of the optic under test during each scan. This new concept enables innovative dynamic scanning strategies, including on-the-fly, automated nulling of the optical surface throughout the scan to reduce systematic errors. Preliminary testing of the new system shows that the hardware, software and automation upgrades are capable of delivering the desired performance of dynamic scanning. Overall, using 10X faster, Elcomat5000 autocollimators, and synchronized motion stages, enables 20X faster scanning and innovative dynamic strategies including automated nulling of the optical surface to reduce systematic errors. ACKOWLEDGEMENTSThis work was carried out with the support of Diamond Light Source. The Authors would like to thank numerous people at Diamond, including: Emilio Perez-Juarez & Adam Howell (Diamond EPICS & Python scripts); Brian Nutter & Nico Rubies (Diamond Motion controls); and Andy Malandain (Diamond Mech. Tech.). We are also indebted to Henry Over & staff at Q-Sys, Netherlands (installation of new controller for NOM), and Ralf Geckeler at PTB and Carsten Schlewitt at Moeller-Wedel for valuable discussions about autocollimators. REFERENCESS. G. Alcock, K. J. S. Sawhney, S. Scott, U. Pedersen, R. Walton, F. Siewert, T. Zeschke, F. Senf, T. Noll, and H. Lammert,
“The Diamond-NOM: A non-contact profiler capable of characterizing optical figure error with sub-nanometre repeatability,”
Nucl Instrum Methods Phys Res A, 616 224
–228
(2010). https://doi.org/10.1016/j.nima.2009.10.137 Google Scholar
I.-T. Nistea, S. G. Alcock, M. Bazan da Silva, and K. J. S. Sawhney,
“The Optical Metrology Laboratory at Diamond: pushing the limits of nano-metrology,”
Advances in Metrology for X-Ray and EUV Optics VIII, 1110906 5
(2019). https://doi.org/10.1117/12.2529401 Google Scholar
S. G. Alcock, I. Nistea, and K. Sawhney,
“Nano-metrology: The art of measuring X-ray mirrors with slope errors <100 nrad,”
Review of Scientific Instruments, 87 1
–8
(2016). https://doi.org/10.1063/1.4949272 Google Scholar
S. Qian and M. Idir,
“Innovative nano-accuracy surface profiler for sub-50 nrad rms mirror test,”
9687 81
–90
(2016). https://doi.org/10.1117/12.2247575 Google Scholar
I. Lacey, J. Adam, G. P. Centers, G. S. Gevorkyan, S. M. Nikitin, B. V. Smith, and V. V. Yashchuk,
“Development of a high performance surface slope measuring system for two-dimensional mapping of x-ray optics,”
10385 103
–115
(2017). https://doi.org/10.1117/12.2273029 Google Scholar
R. Geckeler and A. Just,
“Distance-dependent influences on angle metrology with autocollimators in deflectometry,”
7077 85
–96
(2008). https://doi.org/10.1117/12.793742 Google Scholar
|