31 October 1996 X-ray CCD calibration for the AXAF CCD Imaging Spectrometer
Mark W. Bautz, Steven E. Kissel, Gregory Y. Prigozhin, Stephen E. Jones, Takashi Isobe, Herbert L. Manning, Michael Pivovaroff, George R. Ricker Jr., Jonathan W. Woo
Author Affiliations +
Abstract
We summarize the strategy and techniques used to calibrate x-ray CCD detectors for the AXAF CCD imaging spectrometer in the spectral range between 0.2 keV and 10 keV. The very demanding calibration requirements (energy scale knowledge error of order 0.1%; detection efficiency knowledge error of order 1%) are reviewed. The primary standards chosen for the calibration are discussed, with particular attention to the use of undispersed synchrotron radiation as a primary radiometric standard for the detection efficiency calibration. We review the basic models of the detector response which it is the objective of the calibration enterprise to constrain. The accuracy and reproducibility of the energy scale, spectral redistribution and detection efficiency calibration is discussed and illustrated with sample results from calibration of flight detectors.
© (1996) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Mark W. Bautz, Steven E. Kissel, Gregory Y. Prigozhin, Stephen E. Jones, Takashi Isobe, Herbert L. Manning, Michael Pivovaroff, George R. Ricker Jr., and Jonathan W. Woo "X-ray CCD calibration for the AXAF CCD Imaging Spectrometer", Proc. SPIE 2808, EUV, X-Ray, and Gamma-Ray Instrumentation for Astronomy VII, (31 October 1996); https://doi.org/10.1117/12.255992
Lens.org Logo
CITATIONS
Cited by 8 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Calibration

Charge-coupled devices

CCD image sensors

X-rays

Quantum efficiency

Data modeling

Back to Top