Paper
27 September 2013 Assessing consistency of a Mueller matrix measurement by rotation of the sample under test
Author Affiliations +
Abstract
We present a method for checking the consistency of a Mueller matrix measurement. The method is based upon the rotational properties of a Mueller matrix. An anisotropic sample is placed in the polarimeter in a precision rotation stage. The Mueller matrix is then measured as a function of sample rotation. Each Mueller matrix is then transformed by matrix rotation back into the Mueller matrix at zero angle. The consistency between these resulting matrices is then taken as an indication of measurement uncertainties. The method is demonstrated with two different samples measured on two different polarimeters.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Thomas A. Germer "Assessing consistency of a Mueller matrix measurement by rotation of the sample under test", Proc. SPIE 8873, Polarization Science and Remote Sensing VI, 887303 (27 September 2013); https://doi.org/10.1117/12.2024853
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Polarimetry

Calibration

Wave plates

Sensors

Mueller matrices

Transmittance

Modulation