1 March 2006 Doppler flow imaging of cytoplasmic streaming using spectral domain phase microscopy
Author Affiliations +
Abstract
Spectral domain phase microscopy (SDPM) is a function extension of spectral domain optical coherence tomography. SDPM achieves exquisite levels of phase stability by employing common-path interferometry. We discuss the theory and limitations of Doppler flow imaging using SDPM, demonstrate monitoring the thermal contraction of a glass sample with nanometer per second velocity sensitivity, and apply this technique to measurement of cytoplasmic streaming in an Amoeba proteus pseudopod. We observe reversal of cytoplasmic flow induced by extracellular CaCl2, and report results that suggest parabolic flow of cytoplasm in the A. proteus pseudopod.

1.

Introduction

Doppler flow imaging was one of the first functional extensions developed for optical coherence tomography,1, 2 (OCT) There are two important aspects of OCT that enable phase imaging, in general, and Doppler flow imaging, in particular. First, since OCT is interferometric in nature, light that is Doppler shifted by the motion of scattering particles within the sample creates an electronic beat frequency proportional to the axial component of the sample particle velocity. Second, since OCT performs depth ranging with a broadband light source, this beat frequency can be recorded as a function of sample depth, which enables the generation of depth-indexed flow profiles.

Because Doppler flow imaging in OCT requires phase sensitive detection, interferometric phase stability is critical. Time domain (TD) and spectral domain (SD) systems have different considerations for phase stability. TD-OCT systems employ a broadband light source, a scanning delay line, and a single photodiode detector. As the group delay between the reference and sample arms is scanned, the interferometric signal generated is proportional to the sample reflectivity at the sample depth corresponding to the instantaneous group delay. As such, the phase stability in TD-OCT is limited by the linearity and repeatability of the reference arm group delay scan. SD-OCT systems employ a stationary reference arm and simultaneously collect light from all depths in the sample by collecting the interference signal as a function of optical wavelength. This spectrally indexed signal is Fourier transformed to calculate the depth-indexed sample reflectivity profile (i.e., A scan). The spectral interferometric signal is collected by either (1) using a broadband source and a spectrometer in the detector arm (Fourier domain OCT) or (2) measuring with a single photodiode the signal generated by sweeping a narrow-linewidth laser source through a broad bandwidth (swept source OCT). In SD-OCT systems employing a Michelson interferometer, the phase stability is limited by jitter in the relative path lengths between the sample and reference arms, and in the case of swept source OCT, the repeatability of the wavelength sweep. Indeed, it has been demonstrated that SD-OCT is more stable than its TD counterpart.3

Recently, several phase-sensitive extensions of OCT have been demonstrated. 1, 2, 4, 5, 6, 7, 8, 9 Recently, we demonstrated SD phase microscopy (SDPM), a functional extension of OCT that employs common path SD interferometry10, 11 to achieve phase stabilities in the milliradian regime.12 In this paper, we use SDPM to perform Doppler flow imaging of cytoplasmic streaming in an Amoeba proteus, which is very mobile unicellular protozoan that has been widely used for decades to study the mechanisms of cytoplasm physiology and cell motility.13, 14, 15, 16, 17, 18 Cytoplasmic streaming is an important cellular process that plays a role in events ranging from metastasis in cancer to cellular locomotion to cell division. We derive an expression for the shot-noise-limited flow velocity resolution of SDPM, and compare the performance of our system to this theoretical limit.

2.

Phase-Resolved SD Interference Signal

The SD photocurrent in SD-OCT and SDPM is given by19

Eq. 1

i(k)=12ρS(k)δk{RR+nRn+2RRnRncos[2k(Δxn+δxn)]}.
Here, S(k) is the source power spectral density, k is the wave number (radians per meter), δk is the spectral channel bandwidth, ρ is the detector responsivity, RR is the reference arm reflectivity, and Rn is the reflectivity of the nth sample reflector. The quantity Δxn+δxn is the position of the nth reflector; Δxn is an integer multiple of the discrete sampling interval in the x domain, given by mπΔk , where Δk is the total optical bandwidth interrogated and m is any positive or negative integer; and δxn accounts for subresolution deviations in reflector position away from mπΔk . Note that δxn is an important quantity because it primarily manifests in the phase of i(k) . Note also that n and m are distinct and separate variables: n indexes discrete reflectors in the sample, while m indexes elements in the 1-D x domain A scan array.

In swept source OCT i(k) is directly measured, whereas in spectrometer-based Fourier domain OCT, i(k) is integrated over the A scan acquisition time in a charge-coupled device (CCD) or similar charge-accumulation detector. In either case, after Fourier transformation of the k domain signal, the complex-valued x domain signal and shot noise are given by, respectively,

Eq. 2

Isignal(2Δxn)=ρS0(RRRn)122efascanE(2δxn)exp(jk02δxn),

Eq. 3

Inoise(2Δxn)=(ρS0RRefascan)12exp(jϕrand).
Here, E() is the unity-amplitude x domain autocorrelation function, S0 is the total sample illumination power [i.e., S(k)dk ], k0 is the source center wavenumber, ϕrand is the random phase of the shot noise, and fascan is the rate in hertz at which A scans were acquired. In situations where the detector integration time is less than the interval between acquired A scans, fascan is the numerical inverse of the integration time. It is assumed that RRRn . The amplitude signal to noise ratio of the nth reflector is given by the square of the ratio of the amplitudes of Eqs. 2, 3. The shot-noise-limited phase stability of the nth reflector signal is limited by the phase angle δϕsens between Isignal(2Δxn) and I(2Δxn)=Isignal(2Δxn)+Inoise(2Δxn) . In our previous work,12 we derived an expression for δϕsens by considering its upper limit when Isignal and Inoise are orthogonal (i.e., ϕrand=2k0δxn±π2 ). The issue can be more generally approached by considering the average value of the phase angle between Isignal(2Δxn) and I(2Δxn) over all values of ϕrand . This is given by (Fig. 1 ):

Eq. 4

δϕsens=2π0π2tan1(InoiseIsignalsinϕrand)dϕrand.

Fig. 1

Phase stability of spectral domain phase microscopy. In the x domain the signal and noise are complex-valued signals that add in a vectoral manner. If the phase of the noise is defined to be zero when Inoise is parallel to Isignal , then the error in the phase of Isignal is given by the component of Inoise that is perpendicular to Isignal (i.e., Inoisesinϕrand ). The average rotation of Isignal caused by Inoisesinϕrand taken over all ϕrand defines the phase stability.

024014_1_035602jbo1.jpg

Equation 4 is derived in part from the representation of signal and noise in Fig. 1. At any particular instant, the signal vector (Isignal) and the noise vector (Inoise) have a random angular orientation with respect to each other. Since the phase of Isignal is not random, the phase of Inoise (i.e., ϕrand ) can be conveniently defined with respect to Isignal , and Inoise can be decomposed into components that are parallel (Inoisecosϕrand) and orthogonal (Inoisesinϕrand) to the signal vector. The parallel component contributes to amplitude sensitivity, while the orthogonal component contributes to the phase sensitivity. The phase noise of Isignal is defined by the magnitude of the rotation of Isignal by Inoisesinϕrand . The phase noise also defines the phase sensitivity (δϕsens) since the smallest observable change in the phase of the signal vector is determined by the phase noise. In other words, an observable change in the signal phase must be larger than the phase noise. If InoiseIsignal , which is the usual case, then Isignal , Inoisesinϕrand , and Isignal+Inoisesinϕrand form a right triangle, and δϕsens is the angle opposite Inoisesinϕrand . Here, δϕsens is defined by the argument of the integral in Eq. 4, while the integral itself takes the average value of δϕsens over all possible random values of ϕrand . This integral assumes that ϕrand has a uniform distribution. If it is assumed that InoiseIsignal , then the arctangent function can be approximated by the value of its argument. The integral in Eq. 4 then simplifies to the mean value of the sine function over a quarter period, yielding:

Eq. 5

δϕsens=2π[1SNR(S0,Rn,fascan)]12,
where SNR (S0,Rn,fascan) is the SNR of the nth reflector.

Because the phase of I(2Δxn) is proportional to δxn , displacements in a sample reflector can be tracked over time by tracking the phase over time. The instantaneous velocity of a reflector is given by the difference of δxn on two successive A scans divided by the temporal sampling interval, which is equivalent to defining the instantaneous Doppler shift as the derivative of the phase with respect to time. This gives

Eq. 6

v(t)=λ02cosθfdopp=λ04πcosθfascan[I(2Δxn,t)I(2Δxn,tfascan1)].
Here, ∠ is the phase operator, λ0 is the source center wavelength, fdopp is Doppler frequency shift, and θ is the Doppler angle between the optical axis and the direction of motion. The phase error in v(t) is 2 times higher than the phase error in I(2Δxn,t0) . The factor of 2 arises because velocity is proportional to the numerical difference between two successive phase measurements. As such, the uncertainty in difference (i.e., velocity) must be larger than the individual data points in the difference (i.e., phase). We assume that the summation of N random data points with identical standard deviations (or errors) has an error that is N12 times larger that the error of each data point. The velocity sensitivity is thus (Fig. 2 ):

Eq. 7

vsens=λ022πcosθδϕsensfascan=λ0fascan2π2cosθ(SNR(S0,Rn,fascan))12.

Fig. 2

Velocity sensitivity of Doppler flow imaging. Here, fdopp is the Doppler frequency shift, and dϕdt is experimentally calculated by taking the numerical derivative of the phase with respect to time.

024014_1_035602jbo2.jpg

Equation 7 is consistent with the Cramer-Rao lower bound for a model-based velocity estimator,20 which has been previously verified21 in TD OCT. In Doppler OCT imaging it has been suggested that the minimum observable Doppler shift is related to the inverse of the observation period, which yields a velocity sensitivity22, 23 of λ0fascan2 . The basis for this Fourier-limited assumption is that at least one cycle of the Doppler-induced electronic beat frequency must be sampled to detect the motion of a reflector. If we define the instantaneous Doppler shift (or velocity) as being the derivative of the phase (or position), it is clear the Fourier limit is overly restrictive. For example, if a reflector is moving at a constant velocity, the phase of the interoferometric signal increases linearly over time. The accuracy of the phase difference between two sequential points, which is directly proportional to the instantaneous Doppler shift, is limited only by the phase noise on each individual point (Fig. 2). In this light, there is no obvious requirement to sample an entire fringe of the Doppler-induced electronic beat frequency in the detector photocurrent.

3.

Verification of Doppler Sensitivity Prediction

The major result in the previous section is that the velocity sensitivity in SD Doppler imaging is limited by the phase stability of the x domain signal, according to Eq. 7. To test this prediction, we used SDPM to measure the velocity of thermal expansion of an uncoated glass coverslip transiently heated by a butane flame. A spectral domain interferometer was constructed per Fig. 3a . The source was a 5-mW superluminescent diode (SLD) with a center wavelength and a 3-dB bandwidth of 830 and 45nm , respectively. The spectrometer (Spec) had a 5-ms integration time and a 40-Hz maximum readout rate. The coverslip surface reflection most proximal to the interferometer was used as the reference reflection. Reflectors distal to this surface were designated as having positive displacement with respect to the reference reflector (i.e., Δx> 0 ). Moreover, by design there are no reflectors with negative displacement (Δx<0) . While this configuration does not resolve the complex conjugate ambiguity of spectral domain24, 25 OCT, it sidesteps the issue by forcing all sample displacements to have the same sign. The change in the optical path length (OPL) of the coverslip during heating and cooling was tracked by recording the phase of the x domain interference signal at a depth corresponding to the thickness of the coverslip. This phase was converted to OPL by multiplication by the quantity λ0(4πn) , where n was assumed as the index of the coverslip.

Fig. 3

(a) SDPM interferometer used for measuring thermal expansion of boroslilcate glass. The coverslip surface proximal to the interferometer generates the reference reflection. All reflections distal to this reference surface have positive displacements (i.e., Δx> 0 ). (b) Thermal expansion and contraction of a glass coverslip. The upswing corresponds to the application of a flame to the coverslip, and the downswing corresponds to cooling down to room temperature. The inset highlights the expansion of the coverslip.

024014_1_035602jbo3.jpg

The rapid expansion and slow contraction of the coverslip is shown in Fig. 3b. The baseline phase stability of the interference signal immediately before placing the flame near the coverslip was 0.4mrad (18pm) . The phase stability is defined as the standard deviation of the x domain interference phase at the depth corresponding to the coverslip thickness. The instantaneous velocity of expansion and contraction was calculated by numerically differentiating the OPL on sequential successive A scans and multiplying that quantity by the line rate [Eq. 6].

Figure 4a shows the instantaneously calculated velocity while the coverslip cooled off after flame removal. The yellow curve is a smoothed estimate of the actual velocity generated by low pass filtering the phase data before calculation of the Doppler shift. The red curves represent the estimated velocity plus/minus half of the velocity sensitivity calculated using Eq. 7. The black vertical line at t=20s represents the approximate time at which the magnitude of the velocity fell below the sensitivity of 1nms , as calculated using Eq. 7. The inset to Fig. 4a shows a histogram of the measured velocity values for t> 20s . This data distribution, which is approximately Gaussian, has a standard deviation of 1.1nms , consistent with the predicted velocity sensitivity of 1nms .

Fig. 4

(a) Velocity of thermal contraction of glass coverslip (t=4.8to48s) . The black curve is generated by taking the numerical derivative of the data in (b). The yellow (inner) line is the velocity estimated by lowpass filtering the raw data in black. The red (outer) lines are the velocity estimate plus/minus one-half of the predicted velocity sensitivity of 1nms . The inset shows a histogram [yellow (gray) bars] of the velocity values measured for t> 20s . The mean (μ) and standard deviation (σ) of the measured data are shown. The black curve is a Gaussian distribution with the measured μ and σ . (b) Absolute value of the velocity of expansion and contraction calculated from (a) and plotted on a log scale.

024014_1_035602jbo4.jpg

Figure 4a illustrates two points. First, the experimental velocity data were bound by a range defined by the actual velocity (estimated by lowpass filtering the velocity data) and the predicted velocity sensitivity using Eq. 7. This supports Eq. 7 as a valid expression for the noise and uncertainty in a Doppler calculation given a level of phase stability. Second, it demonstrates that the magnitude of the velocity must be greater than the velocity sensitivity to be resolved from zero velocity. In other words, when the velocity magnitude is equal to the velocity sensitivity, the “velocity SNR” is unity, rendering the velocity measurement indistinguishable from zero velocity. Figure 4b shows the absolute value of the expansion and contraction velocity on a log scale. The predicted velocity sensitivity is shown as a horizontal red line. This figure draws an analogy with amplitude sensitivity for OCT in that the level of the “height” of the noise floor on a log plot is determined by the measurement sensitivity.

4.

Measurement of Cytoplasmic Flow Using SDPM

For in vivo cellular measurements, an SDPM setup was relayed into a Zeiss Axiovert 200 using a documentation port, as illustrated in Fig. 5 . This setup employed the source and spectrometer described in Sec. 3. Simultaneous acquisition of SDPM and visible light microscopy video was achieved by placement of an 8020 beamsplitter in the microscope optical path. There was thus real-time display of video and A scan data. The SDPM spot size ( 1e diameter) on the sample was estimated from the magnification factor of the coupler fiber core being imaged onto the sample. This factor was (L2L1)×(TLOBJ) , giving a calculated spot size of 12μm and a calculated depth of focus of 1mm ( Ln , nth lens; TL, tube lens; OBJ, objective lens; see Fig. 5 for more detail). The ratio OBJ/TL is specified by the manufacturer as the effective or net magnification of a sample object onto the documentation port. The reflection from an uncoated CS surface proximal to the SDPM interferometer acted as the reference reflection.

Fig. 5

SDPM setup adapted to Zeiss Axiovert 200 inverted microscope with simultaneous acquisition of SDPM and video light microscopy. A 635-nm aiming beam (Aim) is combined with an 840-nm SLD ( 50-nm FWHM bandwidth) with a wavelength-division multiplexing (WDM) fiber coupler. The combined light enters a 2×1 5050 fiber coupler whose output fiber core is imaged via lenses (L) 1 and 2 onto a documentation port (DP) of the microscope with a magnification of L2L1=22 . The image formed at the documentation port is then relayed onto the sample (SAMP) with magnification of 110 . The sample also is imaged onto a documentation port CCD. CS, coverslip; OBJ, microscope objective; REF, reference reflection; Spec, spectrometer; TL, tube lens.

024014_1_035602jbo5.jpg

Several amoebas (species A. proteus) were placed on the other CS surface in a springwater solution. The amoeba A. proteus moves by extruding a pseudopod along the CS surface, and cytoplasm flows within the pseudopod along the direction of motion. Since this cytoplasmic streaming is nominally parallel to the coverslip surface, lens 1 (L1) was tilted to make a Doppler angle of θ=87.7deg between the SDPM light and the streaming. This Doppler angle represents a compromise between recoupling efficiency of the reference beam (highest at 90deg ) and optimal Doppler angle (optimal at 0deg ). The angle was verified through image analysis of the position of the aiming beam on the video image taken at calibrated displacements of the objective lens along the optical axis. With respect to recoupling efficiency of the reference beam, note that standard OCT sensitivity expressions here and in the literature are typically independent of reference arm power provided that (1) reference power is much greater than sample power and (2) the system operates in the shot noise limit.

A visible light microscopic image selected from a video recording of an extruding A. proteus is in Fig. 6 . SDPM data were recorded from the location marked with the white triangle. This location was identified with a 635-nm aiming beam that was turned off after the acquisition window was marked with the triangle. The aiming beam was turned off to avoid contamination of the data by photophobic reflex in A. proteus. M-mode recordings (repeated recordings at a given spatial location) of magnitude, phase, and derived Doppler images are shown in Fig. 6b. M-mode images have a vertical axis with units of depth, a horizontal axis with units of time, and image intensity proportional to the measurement of interest (e.g., reflectivity, velocity). The Doppler map was calculated using Eq. 6 after lowpass filtering of the phase data. Between t=18 and 20s , several drops of a 50-mM CaCl2 solution were added to the springwater solution. This triggered a slowdown and subsequent reversal in the cytoplasmic flow. The flow reversal is manifest as a decrease in accumulated phase [Fig. 6c] and as a change in the sign of the Doppler shift [Figs. 6d and 6e]. The flow reversed again at t35s . Overall, measured flow rates were consistent with previously reported13 values for A. proteus.

Fig. 6

(a) Photomicrograph of A. proteus pseudopod (p). The white triangle marks location of data collection, and black box in lower left is 10×10μm . The inset to (a) shows an A scan of the pseudopod (abscissa has units of depth in micrometers, ordinate has units of reflectivity in decibels). The coverslip (cs) on which the pseudopod sits is located at zero displacement. The pseudopod/water (p/w) interface is clearly identified near 80μm by a reflectivity peak followed by the decreased reflectivity of water. (b) M-mode magnitude image of A. proteus pseudopod. (c) and (d) are M-mode phase and Doppler images, respectively. The arrows marked with “f” in (b) demarcate the flowing portion of the cytoplasm as determined from the phase and Doppler data in (c) and (d), respectively. (e) Doppler shift versus time at a depth of 36μm . (f) Doppler shift versus depth at t=20s . The green line is a least-squares parabolic fit of the flow profile (R2=80%) .

024014_1_035602jbo6.jpg

Lowpass filtering of the Doppler data was performed to mitigate the influence of SNR variations due to speckle on the calculation of Doppler shift. Since speckle is multiplicative noise imposed on I(2Δx) [i.e., I=Ispeckle×(Isignal+Inoise) ], SNR (S0,Rn2,fascan) is modulated by this multiplicative noise as well. Additionally, the “nulls” of the speckle pattern have a phase SNR of zero since there is zero signal. These nulls give the false impression that there is zero flow in an otherwise flowing sample. Likewise, SNR is maximum at the “peaks” of the speckle pattern, and the Doppler shift at these peaks presumably will most accurately represent the sample flow velocity. The Doppler data in Figs. 6 and 7 were lowpass filtered with a moving average filter with a time constant (or width) of 3459ms . This relatively longer time constant was chosen to emphasize changes in flow over the course of a few seconds. Figures 8 and 9 show the Doppler shift recorded at a depth of 36μm that was lowpass filterd with time constant ranging from 0to3459ms . The Doppler data is clearly interpretable with little to no filtering. Figure 8 shows the entire time trace, while Fig. 9 focuses on t=20to30s when the initial flow reversal occurred.

Fig. 7

Surface plot of flow-induced Doppler shift as a function of time and depth. Laminar (parabolic) cytoplasmic flow is suggested.

024014_1_035602jbo7.jpg

Fig. 8

Doppler shift versus time at a depth of 36μm after processing with moving-average filters of various time constants (tau). All data points in the recording are shown. Abscissa has units of time in seconds; ordinate, Doppler shift in hertz.

024014_1_035602jbo8.jpg

Fig. 9

Doppler shift versus time at a depth of 36μm after processing with moving-average filters of various time constants (tau). Data points corresponding to flow reversal that occurs between 20 and 30s are shown. Abscissa has units of time in seconds; ordinate, Doppler shift in hertz.

024014_1_035602jbo9.jpg

Visual inspection of the extruding A. proteus pseudopod on light microscopy indicates that the cytoplasm flows within a channel delinated on either side by nonflowing cytoplasm (the so-called “gel,” in contrast to the flowing “sol”). The gel has high viscosity and acts as a stationary conduit, while the sol, which has much lower viscosity, flows within that conduit. Flow is generated by active (i.e., adenosine triphosphate (ATP)-dependent) cytoskeletal and cytoplasmic processes. In the absence of turbulence, which would be difficult to generate owing to the viscosity of cytoplasm, the amplitude of the flow is expected to follow a parabolic profile as a function of depth. Figure 6f, which shows the Doppler frequency as a function of depth at t=20s , supports this laminar flow hypothesis. This measured flow profile matches that of an ideal parabolic flow profile with an R2 value of 80%. The laminar flow profile also is suggested from the plot of the flow-induced Doppler shift against both depth and time in a 3-D surface plot (Fig. 7).

5.

Discussion and Conclusions

Common-path interferometry has been a subject of recent interest in OCT. This interest is in part driven by the design considerations of common-path interferometry, especially since it does away with the need to pathlength-match the reference and sample arms. Common-path configurations have been proposed and demonstrated for both time domain OCT systems (see, for example, Ref. 26) and spectral domain OCT systems (see, for example, Refs. 10, 11, 12). TD common-path interferometers have similar topologies to their spectral domain counterparts, except that the spectral content of the interferometric signal is obtained using a TD autocorrelator as opposed to a SD spectrometer. While the phase stability of these TD systems has yet to be demonstrated, their phase stability is limited by the moving parts in the autocorrelator, just as the phase stability in Michelson-type TD OCT interferometers is limited by the linearity and repeatability of the reference arm group and phase delay. As such, SD common-path interferometry would be expected to have a higher level of phase stability. Additionally, since phase sensitivity is a function of amplitude sensitivity, the amplitude sensitivity advantage of SD OCT translates into a phase sensitivity advantage as well.

Phase-sensitive interferometry has the attractive ability to detect subwavelength motions of reflectors. When combined with broadband coherence interferometry, depth-indexed displacement and flow information can be obtained. As we demonstrated, the phase stability afforded by common-path SD OCT enables the acquisition of Doppler flow profiles of cytoplasmic streaming in an individual cell. This raises the exciting prospect of using SDPM to study cytoplasmic flow in a variety of cellular processes, including the development of neuronal polarity,27 cell migration,28 and a-p axis formation in embryos.29 The use of recently developed ultra-high-resolution sources30 and OCT imaging engines31, 32, 33 in SDPM will enable highly detailed cross-sectional flow maps in individual eukaryotic cells.

Acknowledgments

The authors would like to thank David Choma for fruitful discussions. This work was supported by National Institute of Health (NIH) Grants R24-EB00243 and R21 RR019769.

References

1. 

J. A. Izatt, M. D. Kulkarni, S. Yazdanfar, J. K. Barton, and A. J. Welch, “In vivo bidirectional color Doppler flow imaging of picoliter blood volumes using optical coherence tomography,” Opt. Lett., 22 (18), 1439 –1441 (1997). 0146-9592 Google Scholar

2. 

Z. P. Chen, T. E. Milner, D. Dave, and J. S. Nelson, “Optical Doppler tomographic imaging of fluid flow velocity in highly scattering media,” Opt. Lett., 22 (1), 64 –66 (1997). 0146-9592 Google Scholar

3. 

B. R. White, M. C. Pierce, N. Nassif, B. Cense, B. H. Park, G. J. Tearney, B. E. Bouma, T. C. Chen, and J. F. de Boer, “In vivo dynamic human retinal blood flow imaging using ultra-high-speed spectral domain optical Doppler tomography,” Opt. Express, 11 (25), 3490 –3497 (2003). 1094-4087 Google Scholar

4. 

C. K. Hitzenberger and A. F. Fercher, “Differential phase contrast in optical coherence tomography,” Opt. Lett., 24 (9), 622 –624 (1999). 0146-9592 Google Scholar

5. 

M. Sticker, M. Pircher, E. Gotzinger, H. Stattmann, A. F. Fercher, and C. K. Hitzenberger, “En face imaging of single cell layers by differential phase-contrast optical coherence microscopy,” Opt. Lett., 27 (13), 1126 –1128 (2002). 0146-9592 Google Scholar

6. 

D. P. Davé and T. E. Milner, “Optical low-coherence reflectometer for differential phase measurement,” Opt. Lett., 25 (4), 227 –229 (2000). 0146-9592 Google Scholar

7. 

T. Akkin, D. P. Davé, T. E. Milner, and H. G. Rylander, “Detection of neural activity using phase-sensitive low-coherence reflectometry,” Opt. Express, 12 (11), 2377 –2386 (2004). https://doi.org/10.1364/OPEX.12.002377 1094-4087 Google Scholar

8. 

C. Yang, A. Wax, M. S. Hahn, K. Badizadegan, R. R. Dasari, and M. S. Feld, “Phase-referenced interferometer with subwavelength and subhertz sensitivity applied to the study of cell membrane dynamics,” Opt. Lett., 26 (16), 1271 –1273 (2001). 0146-9592 Google Scholar

9. 

C. Fang-Yen, M. C. Chu, H. S. Seung, R. R. Dasari, and M. S. Feld, “Noncontact measurement of nerve displacement during action potential with dual-beam low-coherence interferometer,” Opt. Lett., 29 (17), 2028 –2030 (2004). https://doi.org/10.1364/OL.29.002028 0146-9592 Google Scholar

10. 

A. F. Fercher, C. K. Hitzenberger, G. Kamp, and S. Y. Elzaiat, “Measurement of intraocular distances by backscattering spectral interferometry,” Opt. Commun., 117 (1–2), 43 –48 (1995). https://doi.org/10.1016/0030-4018(95)00119-S 0030-4018 Google Scholar

11. 

A. B. Vakhtin, D. J. Kane, W. R. Wood, and K. A. Peterson, “Common-path interferometer for frequency-domain optical coherence tomography,” Appl. Opt., 42 (34), 6953 –6958 (2003). 0003-6935 Google Scholar

12. 

M. A. Choma, A. K. Ellerbee, C. H. Yang, T. L. Creazzo, and J. A. Izatt, “Spectral-domain phase microscopy,” Opt. Lett., 30 (10), 1162 –1164 (2005). https://doi.org/10.1364/OL.30.001162 0146-9592 Google Scholar

13. 

R. D. Allen, “Biophysical Aspects of Pseudopodia,” The Biology of Amoeba, Academic Press, New York (1973). Google Scholar

14. 

D. L. Taylor, Y. Wang, and J. M. Heiple, “Contractile basis of ameboid movement VII. The distribution of fluorescently labeled actin in living amebas,” J. Cell Biol., 86 (2), 590 –598 (1980). 0021-9525 Google Scholar

15. 

L. W. Janson and D. L. Taylor, “In vitro models of tail contraction and cytoplasmic streaming in amoeboid cells,” J. Cell Biol., 123 (2), 345 –355 (1993). 0021-9525 Google Scholar

16. 

S. W. Oh and K. W. Jeon, “Characterization of myosin heavy chain and its gene in amoeba proteus,” J. Eukaryot Microbiol., 45 (6), 600 –605 (1998). 1066-5234 Google Scholar

17. 

H. Miyoshi, Y. Kagawa, and Y. Tsuchiya, “Chaotic behavior in the locomotion of amoeba proteus,” Protoplasma, 216 (1–2), 66 –70 (2001). 0033-183X Google Scholar

18. 

M. Dominik, W. Klopocka, P. Pomorski, E. Kocik, and M. J. Redowicz, “Characterization of amoeba proteus myosin VI immunoanalog,” Cell Motil. Cytoskeleton, 61 (3), 172 –188 (2005). 0886-1544 Google Scholar

19. 

M. A. Choma, M. V. Sarunic, C. Yang, and J. A. Izatt, “Sensitivity advantage of swept-source and Fourier-domain optical coherence tomography,” Opt. Express, 11 (18), 2183 –2189 (2003). 1094-4087 Google Scholar

20. 

H. L. V. Trees, Detection, Estimation, and Modulation Theory, Wiley, New York (1968). Google Scholar

21. 

S. Yazdanfar, C. Yang, M. V. Sarunic, and J. A. Izatt, “Frequency estimation precision in Doppler optical coherence tomography using the Cramer-Rao lower bound,” Opt. Express, 13 (2), 410 –416 (2005). https://doi.org/10.1364/OPEX.13.000410 1094-4087 Google Scholar

22. 

M. D. Kulkarni, T. G. van Leeuwen, S. Yazdanfar, and J. A. Izatt, “Velocity-estimation accuracy and frame-rate limitations in color Doppler optical coherence tomography,” Opt. Lett., 23 (13), 1057 –1059 (1998). 0146-9592 Google Scholar

23. 

Y. H. Zhao, Z. P. Chen, C. Saxer, S. H. Xiang, J. F. de Boer, and J. S. Nelson, “Phase-resolved optical coherence tomography and optical Doppler tomography for imaging blood flow in human skin with fast scanning speed and high velocity sensitivity,” Opt. Lett., 25 (2), 114 –116 (2000). 0146-9592 Google Scholar

24. 

M. Wojtkowski, A. Kowalczyk, R. Leitgeb, and A. F. Fercher, “Full range complex spectral optical coherence tomography technique in eye imaging,” Opt. Lett., 27 (16), 1415 –1417 (2002). 0146-9592 Google Scholar

25. 

M. V. Sarunic, M. A. Choma, C. H. Yang, and J. A. Izatt, “Instantaneous complex conjugate resolved spectral domain and swept-source OCT using 3×3 fiber couplers,” Opt. Express, 13 (3), 957 –967 (2005). https://doi.org/10.1364/OPEX.13.000957 1094-4087 Google Scholar

26. 

F. Feldchtein, J. Bush, G. Gelikonov, V. Gelikonov, and S. Piyevsky, “Cost effective, all-fiber autocorrelator based 1300nm OCT system,” Proc. SPIE, 5690 349 –355 (2005). 0277-786X Google Scholar

27. 

F. Bradke and C. G. Dotti, “Neuronal polarity: vectorial cytoplasmic flow precedes axon formation,” Neuron, 19 (6), 1175 –1186 (1997). 0896-6273 Google Scholar

28. 

A. J. Ridley, M. A. Schwartz, K. Burridge, R. A. Firtel, M. H. Ginsberg, G. Borisy, J. T. Parsons, and A. R. Horwitz, “Cell migration: integrating signals from front to back,” Science, 302 (5651), 1704 –1709 (2003). 0036-8075 Google Scholar

29. 

S. N. Hird and J. G. White, “Cortical and cytoplasmic flow polarity in early embryonic cells of Caenorhabditis elegans,” J. Cell Biol., 121 (6), 1343 –1355 (1993). 0021-9525 Google Scholar

30. 

W. Drexler, U. Morgner, F. X. Krtner, C. Pitris, S. A. Boppart, X. D. Li, E. P. Ippen, and J. G. Fujimoto, “In vivo ultrahigh-resolution optical coherence tomography,” Opt. Lett., 24 (17), 1221 –1223 (1999). 0146-9592 Google Scholar

31. 

B. Cense, N. A. Nassif, T. C. Chen, M. C. Pierce, S-H Yun, B. H. Park, B. E. Bouma, G. J. Tearney, and J. F. de Boer, “Ultrahigh-resolution high-speed retinal imaging using spectral-domain optical coherence tomography,” Opt. Express, 12 (11), 2435 –2447 (2004). https://doi.org/10.1364/OPEX.12.002435 1094-4087 Google Scholar

32. 

M. Wojtkowski, V. J. Srinivasan, T. H. Ko, J. G. Fujimoto, A. Kowalczyk, J. S. Duker, “Ultrahigh-resolution, high-speed, Fourier domain optical coherence tomography and methods for dispersion compensation,” Opt. Express, 12 (11), 2404 –2422 (2004). https://doi.org/10.1364/OPEX.12.002404 1094-4087 Google Scholar

33. 

R. A. Leitgeb, W. Drexler, A. Unterhuber, B. Hermann, T. Bajraszewski, T. Le, A. Stingl, and A. F. Fercher, “Ultrahigh resolution Fourier domain optical coherence tomography,” Opt. Express, 12 (10), 2156 –2165 (2004). https://doi.org/10.1364/OPEX.12.002156 1094-4087 Google Scholar
©(2006) Society of Photo-Optical Instrumentation Engineers (SPIE)
Michael A. Choma, Audrey K. Ellerbee, Siavash Yazdanfar, and Joseph A. Izatt "Doppler flow imaging of cytoplasmic streaming using spectral domain phase microscopy," Journal of Biomedical Optics 11(2), 024014 (1 March 2006). https://doi.org/10.1117/1.2193167
Published: 1 March 2006
Lens.org Logo
CITATIONS
Cited by 64 scholarly publications and 10 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Doppler effect

Interference (communication)

Optical coherence tomography

Reflectors

Signal to noise ratio

Doppler tomography

Interferometry

Back to Top