27 January 2021 In situ terahertz monitoring of an ice ball formation during tissue cryosurgery: a feasibility test
Author Affiliations +
Abstract

Significance: Uncontrolled cryoablation of tissues is a strong reason limiting the wide application of cryosurgery and cryotherapy due to the certain risks of unpredicted damaging of healthy tissues. The existing guiding techniques are unable to be applied in situ or provide insufficient spatial resolution. Terahertz (THz) pulsed spectroscopy (TPS) based on sensitivity of THz time-domain signal to changes of tissue properties caused by freezing could form the basis of an instrument for observation of the ice ball formation.

Aim: The ability of TPS for in situ monitoring of tissue freezing depth is studied experimentally.

Approach: A THz pulsed spectrometer operated in reflection mode and equipped with a cooled sample holder and ex vivo samples of bovine visceral adipose tissue is applied. Signal spectrograms are used to analyze the changes of THz time-domain signals caused by the interface between frozen and unfrozen tissue parts.

Results: Experimental observation of TPS signals reflected from freezing tissue demonstrates the feasibility of TPS to detect ice ball formation up to 657-μm depth.

Conclusions: TPS could become the promising instrument for in situ control of cryoablation, enabling observation of the freezing front propagation, which could find applications in various fields of oncology, regenerative medicine, and THz biophotonics.

1.

Introduction

Terahertz (THz) pulsed spectroscopy (TPS) is known as a promising tool of biophotonics and medical diagnosis.14 It demonstrates strong potential of study and differentiate various conditions of tissues, both ex vivo and in vivo,59 and imaging and characterization of neoplasms in different localizations,1014 by means of endogenous markers of pathology, such as water content. However, high concentration of free and bound water in tissues restricts the penetration depth of THz waves by only several hundreds of microns.1 To overcome this limitation, different approaches can be applied. Among them are such techniques as tissue dehydration,15 paraffin-embedding,16 lyophilization,17 compression,18 immersion optical clearing,19,20 and freezing.2124 It was demonstrated that the latter one helps to increase the THz-wave penetration up to 1 mm, changing at the same time dielectric properties of tissues in THz range.23 Along with reduction of absorption, low temperatures alter the refractive index within certain limits, thus, yielding the dielectric contrast between tissues in different states. This opens further horizons of TPS applications, in particular, for monitoring of freezing depth in tissues during their cryosurgery.

Uncontrolled cryoablation of living tissues is a severe problem of modern cryosurgery, i.e., application of extreme cold aimed at rapid freezing and thus destroying or renewal of pathological tissues.2529 Unless cryoablation possesses several advantages, such as relative painlessness, hemostatic effect, short recovering of patients, and immunostimulating effect,25,30,31 it is associated with certain risks of damaging healthy tissues surrounding the pathology and with possibility of incomplete cell death.28 In particular, the use of cryosurgery for the treatment of brain tumors and epilepsy is currently limited by the possible risks for the patient.3234 Such guiding techniques as ultrasonography,35,36 magnetic resonance, and computer imaging3742 do not enable the monitoring of the tissue freezing in situ with high spatial resolution near the cryoprobe. Alternative approach is based on measuring of tissue temperature using thermocouples (TCs) combined with cryoprobe.43 However, it allows controlling of the tissue condition only at the cryoprobe–tissue interface. Therefore, cryosurgery still needs guiding approaches for the efficient control of ice ball formation.

In this work, we study the ability of TPS to solve this problem. We experimentally demonstrate the changes of THz time-domain signal during tissue freezing, using ex vivo bovine adipose specimen and laboratory spectrometer equipped with special sample holder. Processing of THz pulsed signal within 45 s demonstrates the ability to detect the movement of freezing front up to the depth of 657  μm. The results of this work reveal the potential of TPS to control the initial stage of ice ball formation during tissue cryoablation.

2.

Experimental Setup

To experimentally test the feasibility of TPS to detect the freezing depth in biological tissues, we use a laboratory THz pulsed spectrometer, which was described in detail in Ref. 44. It operates in a reflection mode and has a maximum spectral operating range from about 0.05 to 4.0 THz and a maximum spectral resolution of about 0.002 THz. The THz wave features s-polarization. To avoid the impact of water vapors on the measurements, the spectrometer is purged with nitrogen gas. It is also equipped with a sample holder (Fig. 1). The latter consists of a reference window made of sapphire, which on the one hand is placed on the THz beam path in contact with the tissue sample [see Fig. 1(a)]. On the other hand, it is inserted in the metal frame connected with the reservoir filled with liquid nitrogen prior to the measurements [Fig. 1(b)]. Thus, due to high thermal conductivity of metal and sapphire, the window possesses low and stable temperature during the measurements.

Fig. 1

Schematic of the ex vivo TPS measurement of the tissue freezing: (a) a reflection-mode measurement unit of the THz pulsed spectrometer, PCA stands for a photoconductive antenna, GOAPM stands for a gold off-axis parabolic mirror; (b) a sample holder; (c) an example of the sample time-domain signal E(t) of the THz pulsed spectrometer, where A and B stand for pulses reflected from the interfaces free space–reference window and reference window–sample, respectively.

JBO_26_4_043003_f001.png

As a tissue sample, we use ex vivo bovine visceral adipose tissue because of its relatively homogeneous structure. The sample thickness is near 30 mm enabling prolonged duration of the freezing process. By two off-axis parabolic mirrors, the incident THz beam is focused on the interface between the reference window and the tissue; the reflected beam is collimated to the detector. Generation and detection of THz pulses are provided by two photoconductive antennas (PCAs). A typical form of the reflected THz time-domain signal is shown in Fig. 1(c). The sequence of pulses includes signal (A) reflected from free space–window interface and signal (B) reflected from the window–sample interface. Since we are interested in the latter one, the scanning range of the spectrometer is adjusted at this pulse (marked with red rectangle).

The measurements are performed at room temperature 24°C. However, the sample is preliminary heated to the temperature 37°C aimed at the slowing down the initial stage of freezing. After the contact between the tissue and the cooled window, THz pulsed signal is constantly recorded with the period of 15 s until the sample is completely frozen. For additional control of sample freezing, its temperature is measured by the inserted TCs.

3.

THz Pulsed Spectroscopy of a Freezing Tissue

3.1.

THz Dielectric Properties of a Sample

The sample refractive index was characterized before and after freezing. For this purpose, the complex amplitude of the incident THz field is considered to have the form E(t)=A0exp(iϕ), where A0 and ϕ are the amplitude and phase of the THz field, respectively. Assuming the negligible losses in the reference window and the normal incidence of THz beam, the signal amplitudes A1 and A2 of pulses reflected from the first and second interfaces [see wavelets (A) and (B) in Fig. 1(c)] are described by the Fresnel coefficients

Eq. (1)

A1=R01A0,

Eq. (2)

A2=T01R12T10A0,
where Rmk and Tmk are the Fresnel coefficients of amplitude reflection and transmission between m’th and k’th media, which for s-polarized wave are determined by the corresponding refractive indices nm and nk

Eq. (3)

Rmk=nmnknm+nk,

Eq. (4)

Tmk=2nknm+nk.

In this work, refractive indices for free space and reference window are n0=1 and n1=3.07, respectively, at 1-THz frequency.45 Estimating the relation |A2/A1| from the measured TPS waveforms and accounting the dispersion relation, the values of the sample refractive index n2 at 1-THz frequency were determined for two conditions of the tissue, i.e., nR=1.84±0.01 at room temperature and nF=1.37±0.01 after freezing. Thus, freezing of the adipose tissue leads to significant changes of its refractive index. The value nF is used further for estimation of the freezing depth z according to the relation z=tc/(2nF), where t is the temporal position of the detected freezing depth, c is speed of light in free space.

3.2.

Analysis of THz Signal Spectrograms

According to the measurement of sample temperature, the freezing depth of 4 mm was reached in <1  min. This depth significantly exceeds the typical values of THz penetration depth in frozen biological tissues.23 Thus, only several THz waveforms can be recorded during this stage of the analyzed process. To increase the number of detected signals, the accumulation time and discretization of the spectrometer should be reduced.

Figure 2 shows the sequence of THz waveforms detected at (I) 0 s, (II) 15 s, (III) 30 s, and (IV) 45 s from the beginning of the sample cooling. At the initial moment t=0 of the ice front propagation, the temporal width of first THz pulse, caused by the window–tissue interface, is about 5 ps, which corresponds to 0.5  mm of the sample depth. Within this tissue layer, the second pulse caused by the existence of the ice front interface, is probably overlapped by the first one. Its amplitude is expected to be significantly smaller, since the dielectric contrast between frozen and unfrozen tissue is lower than that for the tissue and the reference window. Thus, it is difficult to estimate directly from the THz waveforms the moving position of the second pulse.

Fig. 2

The waterfall of THz time-domain signals reflected from the window–tissue interface and detected by the spectrometer during the tissue freezing. Signals I, II, III, and IV correspond to t=0, 15, 30, and 45 s moments of freezing, respectively.

JBO_26_4_043003_f002.png

To overcome this problem, we analyze the signal intensity spectrograms in addition to their temporal waveforms. Figures 3(a)3(d) show the spectrograms of the measured signals shown in the insets [Figs. 3(e)3(h)]. Comparing them between each other, it is possible to detect the temporal position of the pulse reflected from the ice front. Figures 3(i)3(k) show the difference ΔI(t) between the normalized spectrogram In(t)=I(t)/max[I(t)] at t=15, 30, and 45 s and the initial normalized spectrogram at t=0. Obviously, the first pulse is altering during the first 30 s, then its appearance remains stable. We can clearly observe the presence of the second pulse at t=15 and 30 s, which indicates the freezing depths of z=328  μm and z=657  μm, respectively; then, its amplitude becomes lower due to absorbtion, which does not allow us to detect it. Therefore, during 30 s starting from the tissue cooling, TPS provides an ability to analyze the ice ball formation and estimate the freezing depth. It should be mentioned that time-domain THz signals feature rather weak changes at the determined moments, not allowing us to evaluate them directly. To demonstrate this, the detected positions of the second pulses at t=15 and 30 s are marked with red rectangles on the THz waveforms [see Figs. 3(f) and 3(g)]. Due to the absence of second pulses at t=0 and 45 s, there are no marks on Figs. 3(e) and 3(h).

Fig. 3

Analysis of THz time-domain signals detected during the tissue freezing. (a)–(d) Signal intensity spectrograms, where white arrows indicate the positions of second pulses reflected from the freezing front; (e) and (f) the temporal amplitudes of the signals, where red rectangles mark the same positions; (a) and (e) correspond to the moment t=0; (b) and (f) 15 s; (c) and (g) 30 s; and (d) and (h) 45 s; (i)–(k) the differences ΔI(t) between the normalized spectrogram In(t)=I(t)/max[I(t)] at t=15, 30, and 45 s and the initial normalized spectrogram at t=0.

JBO_26_4_043003_f003.png

4.

Results and Discussion

The estimation of the tissue freezing depth is shown in Fig. 4, where the values obtained by TPS are combined with the direct measurements of the sample temperature. Figure 4(b) demonstrates the temperature change at two depths 4 and 11 mm from the window–tissue interface. The slowdown of the freezing process at the beginning can be explained by the initial cooling of the tissue from 37°C to 0°C. In our feasibility study, the maximal freezing depth observed by TPS is limited by the value 657  μm. However, we believe that it could be increased using THz time-domain spectrometer with more efficient performance, particularly, scanning rate, which would yield the reduction of the detection period and increasing of the temporal resolution.

Fig. 4

An estimation of the tissue freezing depth. (a) The combination of TPS measurements (points marked with red color) and data obtained by TCs (points marked with green color), here, TC stands for thermocouple; (b) temperature changes of the freezing tissue at depths of 4 and 11 mm from the reference window. Error bars represent ±2σ confidential interval of measurements.

JBO_26_4_043003_f004.png

Though the penetration depth of THz waves even in frozen tissue remains extremely low, the detection of the freezing depth by TPS could provide rather high precision of monitoring ice front propagation, enabling the observation and control of cryoablation and cooling of cell and artificial tissue structures, and agglomeration of cell spheroids. Thus, it could become a beneficial research instrument in regenerative medicine and microbiology.

THz technologies are still far from clinical application, suffering from a list of problems, such as low performance of THz sources and detectors, and low spatial resolution due to high absorbtion and scattering.46,47 However, the rapid progress of THz sources and optical components in the recent years 4852 stimulates the further improvements of THz instruments for biophotonics. A strong limitation is caused by the absence of efficient waveguides and fibers.53 Nevertheless, the sapphire shaped crystal waveguides, exploiting both anti-resonant reflecting optical and photonic crystal waveguidance principles, have certain perspectives for delivering THz pulses with relatively low losses and dispersion;44,54 thus, they could form the basis for further developments of THz instruments for biophotonics and medicine. In addition, sapphire features high thermal conductivity at cryogenic temperatures, being more efficient material for cryosurgery than metal (copper, titanium nickelide, and stainless steel), since it enables higher tissue freezing rate.53 Combination of these advantages opens future perspectives for the appearance of sapphire probes for controlled cryoablation.

Cryosurgery demonstrates high potential in oncology, in particular, to remove neoplasms and metastases of different localizations, such as liver,37 breast,55 lungs,40prostate,56 kidney,57 stomach,58 and skin.59 Considering the ability of THz technologies for tumor diagnosis along with currently developing THz instruments and methods for healthy and pathological tissues differentiation,2 the combination of THz preoperative characterization of tissue with further TPS-controlled cryoablation could become an effective tool for oncology.

It is worth mentioning that several techniques enable increasing of low penetration depth of THz waves into living tissues. Among them are methods of immersion optical clearing,19 which could also be combined with THz freezing depth detection. However, a list of hyperosmotic agents applied for tissue clearing act as cryoprotectors (e.g., glycerol, dimethyl sulfoxide, and propylene glycol), since they substitute the interstitial water. Thus, the particular approaches of their application for increasing of THz-wave penetration into tissues during their freezing, such as concentration of aqueous solutions and exposure time, are the subjects of further studies.

5.

Conclusion

The performed experimental study demonstrates the feasibility of in situ observation of the ice ball formation in tissues using TPS instrumentation. Applying the spectrogram analysis of THz time-domain signals recorded in the reflection-mode of THz pulsed spectrometer during tissue freezing, it is possible to detect the positions of the freezing depth up to 657μm. We believe that the described approach would be helpful for the development of novel approaches and instruments of controlled cryoablation and would be useful for various tasks of oncology, regenerative medicine, microbiology, and further progress of THz biophotonics.

Disclosures

The authors declare no conflict of interest.

Acknowledgments

This work was supported by the Russian Science Foundation (RSF), Project No. #19-79-10212.

Code, Data, and Materials Availability

All data generated or analyzed during this study are included in this article.

References

1. 

O. A. Smolyanskaya et al., “Terahertz biophotonics as a tool for studies of dielectric and spectral properties of biological tissues and liquids,” Prog. Quantum Electron., 62 1 –77 (2018). https://doi.org/10.1016/j.pquantelec.2018.10.001 PQUEAH 0079-6727 Google Scholar

2. 

K. I. Zaytsev et al., “The progress and perspectives of terahertz technology for diagnosis of neoplasms: a review,” J. Opt., 22 (1), 013001 (2020). https://doi.org/10.1088/2040-8986/ab4dc3 Google Scholar

3. 

J.-H. Son, S. J. Oh and H. Cheon, “Potential clinical applications of terahertz radiation,” J. Appl. Phys., 125 (19), 190901 (2019). https://doi.org/10.1063/1.5080205 JAPIAU 0021-8979 Google Scholar

4. 

H. Cheon, H. Yang and J. Son, “Toward clinical cancer imaging using terahertz spectroscopy,” IEEE J. Sel. Top. Quantum Electron., 23 (4), 1 –9 (2017). https://doi.org/10.1109/JSTQE.2017.2704905 IJSQEN 1077-260X Google Scholar

5. 

J. Wang et al., “In vivo terahertz imaging to evaluate scar treatment strategies: silicone gel sheeting,” Biomed. Opt. Express, 10 (7), 3584 –3590 (2019). https://doi.org/10.1364/BOE.10.003584 BOEICL 2156-7085 Google Scholar

6. 

G. Hernandez-Cardoso et al., “Terahertz imaging for early screening of diabetic foot syndrome: a proof of concept,” Sci. Rep., 7 42124 (2017). https://doi.org/10.1038/srep42124 SRCEC3 2045-2322 Google Scholar

7. 

N. Bajwa et al., “Non-invasive terahertz imaging of tissue water content for flap viability assessment,” Biomed. Opt. Express, 8 (1), 460 –474 (2017). https://doi.org/10.1364/BOE.8.000460 BOEICL 2156-7085 Google Scholar

8. 

Y. Cao et al., “Qualitative and quantitative detection of liver injury with terahertz time-domain spectroscopy,” Biomed. Opt. Express, 11 (2), 982 –993 (2020). https://doi.org/10.1364/BOE.381660 BOEICL 2156-7085 Google Scholar

9. 

E. P. Parrott et al., “Terahertz pulsed imaging in vivo: measurements and processing methods,” J. Biomed. Opt., 16 (10), 106010 (2011). https://doi.org/10.1117/1.3642002 JBOPFO 1083-3668 Google Scholar

10. 

D. Li et al., “Detecting melanoma with a terahertz spectroscopy imaging technique,” Spectrochim. Acta A, 234 118229 (2020). https://doi.org/10.1016/j.saa.2020.118229 Google Scholar

11. 

S. Oh et al., “Study of freshly excised brain tissues using terahertz imaging,” Biomed. Opt. Express, 5 (8), 2837 –2842 (2014). https://doi.org/10.1364/BOE.5.002837 BOEICL 2156-7085 Google Scholar

12. 

A. Gavdush et al., “Terahertz spectroscopy of gelatin-embedded human brain gliomas of different grades: a road toward intraoperative THz diagnosis,” J. Biomed. Opt., 24 (2), 027001 (2019). https://doi.org/10.1117/1.JBO.24.2.027001 JBOPFO 1083-3668 Google Scholar

13. 

P. C. Ashworth et al., “Terahertz pulsed spectroscopy of freshly excised human breast cancer,” Opt. Express, 17 (15), 12444 –12454 (2009). https://doi.org/10.1364/OE.17.012444 OPEXFF 1094-4087 Google Scholar

14. 

K. I. Zaytsev et al., “Highly accurate in vivo terahertz spectroscopy of healthy skin: variation of refractive index and absorption coefficient along the human body,” IEEE Trans. Terahertz Sci. Technol., 5 (5), 817 –827 (2015). https://doi.org/10.1109/TTHZ.2015.2460677 Google Scholar

15. 

Y. He et al., “Determination of terahertz permittivity of dehydrated biological samples,” Phys. Med. Biol., 62 (23), 8882 –8893 (2017). https://doi.org/10.1088/1361-6560/aa8ebe PHMBA7 0031-9155 Google Scholar

16. 

K. Meng et al., “Terahertz pulsed spectroscopy of paraffin-embedded brain glioma,” J. Biomed. Opt., 19 (7), 077001 (2014). https://doi.org/10.1117/1.JBO.19.7.077001 JBOPFO 1083-3668 Google Scholar

17. 

G. Png et al., “The impact of hydration changes in fresh bio-tissue on THz spectroscopic measurements,” Phys. Med. Biol., 53 (13), 3501 –3517 (2008). https://doi.org/10.1088/0031-9155/53/13/007 PHMBA7 0031-9155 Google Scholar

18. 

J. Wang et al., “THz in vivo measurements: the effects of pressure on skin reflectivity,” Biomed. Opt. Express, 9 (12), 6467 –6476 (2018). https://doi.org/10.1364/BOE.9.006467 BOEICL 2156-7085 Google Scholar

19. 

G. R. Musina et al., “Optimal hyperosmotic agents for tissue immersion optical clearing in terahertz biophotonics,” J. Biophotonics, 13 (12), e202000297 (2020). https://doi.org/10.1002/jbio.202000297 Google Scholar

20. 

A. Kolesnikov et al., “THz monitoring of the dehydration of biological tissues affected by hyperosmotic agents,” Phys. Wave Phenom., 22 169 –176 (2014). https://doi.org/10.3103/S1541308X14030029 Google Scholar

21. 

H. Hoshina et al., “Terahertz pulsed imaging of frozen biological tissues,” Appl. Phys. Lett., 94 (12), 123901 (2009). https://doi.org/10.1063/1.3106616 APPLAB 0003-6951 Google Scholar

22. 

Y. Sim et al., “Temperature-dependent terahertz imaging of excised oral malignant melanoma,” IEEE J. Biomed. Health, 17 (4), 779 –784 (2013). https://doi.org/10.1109/JBHI.2013.2252357 Google Scholar

23. 

Y. C. Sim et al., “Terahertz imaging of excised oral cancer at frozen temperature,” Biomed. Opt. Express, 4 (8), 1413 –1421 (2013). https://doi.org/10.1364/BOE.4.001413 BOEICL 2156-7085 Google Scholar

24. 

J. Y. Park et al., “Terahertz imaging of metastatic lymph nodes using spectroscopic integration technique,” Biomed. Opt. Express, 8 (2), 1122 –1129 (2017). https://doi.org/10.1364/BOE.8.001122 BOEICL 2156-7085 Google Scholar

25. 

E. Vansonnenberg et al., Tumor Ablation: Principles and Practice, Springer-Verlag, New York (2005). Google Scholar

26. 

P. Mazur, “Cryobiology: the freezing of biological systems,” Science, 168 (3934), 939 –949 (1970). https://doi.org/10.1126/science.168.3934.939 SCIEAS 0036-8075 Google Scholar

27. 

A. A. Gage and J. Baust, “Mechanisms of tissue injury in cryosurgery,” Cryobiology, 37 (3), 171 –186 (1998). https://doi.org/10.1006/cryo.1998.2115 CRYBAS 0011-2240 Google Scholar

28. 

J. Baust et al., “Mechanisms of cryoablation: clinical consequences on malignant tumors,” Cryobiology, 68 (1), 1 –11 (2014). https://doi.org/10.1016/j.cryobiol.2013.11.001 CRYBAS 0011-2240 Google Scholar

29. 

N. Korpan, Basics of Cryosurgery, Springer-Verlag, Wien (2001). Google Scholar

30. 

K. F. Chu and D. E. Dupuy, “Thermal ablation of tumours: biological mechanisms and advances in therapy,” Nat. Rev. Cancer, 14 (3), 199 –208 (2014). https://doi.org/10.1038/nrc3672 NRCAC4 1474-175X Google Scholar

31. 

F. Yuan, G. Zhao and F. Panhwar, “Enhanced killing of HepG2 during cryosurgery with Fe3O4-nanoparticle improved intracellular ice formation and cell dehydration,” Oncotarget, 8 (54), 92561 –92577 (2017). https://doi.org/10.18632/oncotarget.21499 Google Scholar

32. 

T. Tokiwa et al., “A palm-sized cryoprobe system with a built-in thermocouple and its application in an animal model of epilepsy,” IEEE Trans. Biomed. Eng., 66 (11), 3168 –3175 (2019). https://doi.org/10.1109/TBME.2019.2901530 IEBEAX 0018-9294 Google Scholar

33. 

C. Li et al., “MR imaging-guided cryoablation of metastatic brain tumours: initial experience in six patients,” Eur. Radiol., 20 (2), 404 –409 (2010). https://doi.org/10.1007/s00330-009-1554-8 Google Scholar

34. 

B. Martynov et al., “Stereotactic cryodestruction of gliomas,” Prog. Neurol. Surg., 32 27 –38 (2018). https://doi.org/10.1159/000469677 Google Scholar

35. 

I. G. Berim et al., “Radial probe ultrasound-guided cryobiopsy,” J. Bronchol. Interventional Pulmonol., 24 (2), 170 –173 (2017). https://doi.org/10.1097/LBR.0000000000000368 Google Scholar

36. 

M. Ahmed et al., “Percutaneous ultrasound-guided cryoablation for symptomatic plantar fibromas,” CardioVasc. Interventional Radiol., 41 298 –304 (2018). https://doi.org/10.1007/s00270-017-1801-3 CAIRDG 1432-086X Google Scholar

37. 

K. Zhou et al., “Combined magnetic resonance imaging and optical surgical navigation system guidance of percutaneous liver cryoablation in a porcine model,” Am. J. Transl. Res., 10 (5), 1531 –1538 (2018). Google Scholar

38. 

J. Lichter et al., “Real-time magnetic resonance imaging-guided cryoablation of the pulmonary veins with acute freeze-zone and chronic lesion assessment,” EP Europace, 21 (1), 154 –162 (2019). https://doi.org/10.1093/europace/euy089 Google Scholar

39. 

W. Fan et al., “Percutaneous computed tomography-guided cryoablation for recurrent retroperitoneal soft tissue sarcoma: a study of safety and efficacy,” Oncotarget, 7 (27), 42639 –42649 (2016). https://doi.org/10.18632/oncotarget.9476 Google Scholar

40. 

C. Prud’homme et al., “Image-guided lung metastasis ablation: a literature review,” Int. J. Hyperth., 36 (2), 37 –45 (2019). https://doi.org/10.1080/02656736.2019.1647358 IJHYEQ 0265-6736 Google Scholar

41. 

J. Pohlan et al., “Computed tomography thermography for ablation zone prediction in microwave ablation and cryoablation: advantages and challenges in an ex vivo porcine liver model,” J. Comput. Assist. Tomogr., 44 (5), 744 –749 (2020). https://doi.org/10.1097/RCT.0000000000001081 JCATD5 0363-8715 Google Scholar

42. 

Y. Yang et al., “Retrospective analysis of CT-guided percutaneous cryoablation for treatment of painful osteolytic bone metastasis,” Cryobiology, 92 203 –207 (2020). https://doi.org/10.1016/j.cryobiol.2020.01.012 CRYBAS 0011-2240 Google Scholar

43. 

M. Shurrab et al., “The cooling performance of a cryoprobe: establishing guidelines for the safety margins in cryosurgery,” Int. J. Refrig., 67 308 –318 (2016). https://doi.org/10.1016/j.ijrefrig.2016.03.007 IJRFDI 0140-7007 Google Scholar

44. 

G. M. Katyba et al., “Sapphire photonic crystal waveguides for terahertz sensing in aggressive environments,” Adv. Opt. Mater., 6 (22), 1800573 (2018). https://doi.org/10.1002/adom.201800573 2195-1071 Google Scholar

45. 

D. Grischkowsky et al., “Far-infrared time-domain spectroscopy with terahertz beams of dielectrics and semiconductors,” J. Opt. Soc. Am. B, 7 2006 –2015 (1990). https://doi.org/10.1364/JOSAB.7.002006 JOBPDE 0740-3224 Google Scholar

46. 

I. N. Dolganova et al., “The role of scattering in quasi-ordered structures for terahertz imaging: local order can increase an image quality,” IEEE Trans. Terahertz Sci. Technol., 8 (4), 403 –409 (2018). https://doi.org/10.1109/TTHZ.2018.2844104 Google Scholar

47. 

K. I. Zaytsev et al., “Overcoming the Abbe diffraction limit using a bundle of metal-coated high-refractive-index sapphire optical fibers,” Adv. Opt. Mater., 8 (18), 2000307 (2020). https://doi.org/10.1002/adom.202000307 2195-1071 Google Scholar

48. 

D. V. Lavrukhin et al., “Terahertz photoconductive emitter with dielectric-embedded high-aspect-ratio plasmonic grating for operation with low-power optical pumps,” AIP Adv., 9 (1), 015112 (2019). https://doi.org/10.1063/1.5081119 AAIDBI 2158-3226 Google Scholar

49. 

D. V. Lavrukhin et al., “Shaping the spectrum of terahertz photoconductive antenna by frequency-dependent impedance modulation,” Semicond. Sci. Technol., 34 034005 (2019). https://doi.org/10.1088/1361-6641/aaff31 SSTEET 0268-1242 Google Scholar

50. 

D. S. Ponomarev et al., “Enhanced terahertz emission from strain-induced InGaAs/InAlAs superlattices,” J. Appl. Phys., 125 (15), 151605 (2019). https://doi.org/10.1063/1.5079697 JAPIAU 0021-8979 Google Scholar

51. 

V. E. Ulitko et al., “Nanoporous SiO2 based on annealed artificial opals as a favorable material platform of terahertz optics,” Opt. Mater. Express, 10 (9), 2100 –2113 (2020). https://doi.org/10.1364/OME.402185 Google Scholar

52. 

A. A. Gavdush et al., “Proof of concept for continuously-tunable terahertz bandpass filter based on a gradient metal-hole array,” Opt. Express, 28 (18), 26228 –26238 (2020). https://doi.org/10.1364/OE.401608 OPEXFF 1094-4087 Google Scholar

53. 

G. Katyba et al., “Sapphire shaped crystals for waveguiding, sensing and exposure applications,” Prog. Cryst. Growth Charact. Mater., 64 (4), 133 –151 (2018). https://doi.org/10.1016/j.pcrysgrow.2018.10.002 PCGMED 0960-8974 Google Scholar

54. 

K. I. Zaytsev et al., “Terahertz photonic crystal waveguides based on sapphire shaped crystals,” IEEE Trans. Terahertz Sci. Technol., 6 (4), 576 –582 (2016). https://doi.org/10.1109/TTHZ.2016.2555981 Google Scholar

55. 

Z. Habrawi et al., “Cryoablation: a promising non-operative therapy for low-risk breast cancer,” Am. J. Surg., 221 127 –133 (2021). https://doi.org/10.1016/j.amjsurg.2020.07.028 AJOOA7 0096-6347 Google Scholar

56. 

K. Kinsman et al., “Whole-gland prostate cancer cryoablation with magnetic resonance imaging guidance: one-year follow-up,” CardioVasc. Interventional Radiol., 41 344 –349 (2018). https://doi.org/10.1007/s00270-017-1799-6 CAIRDG 1432-086X Google Scholar

57. 

S. Hebbadj et al., “Safety considerations and local tumor control following percutaneous image-guided cryoablation of T1b renal tumors,” CardioVasc. Interventional Radiol., 41 449 –458 (2018). https://doi.org/10.1007/s00270-017-1820-0 CAIRDG 1432-086X Google Scholar

58. 

W. Zheng et al., “A novel endoscopic visible full-thickness cryoablation device on stomach,” Sci. Rep., 10 5203 (2020). https://doi.org/10.1038/s41598-020-61595-x SRCEC3 2045-2322 Google Scholar

59. 

G. Mpourazanis et al., “The effectiveness of photodynamic therapy and cryotherapy on patients with basal cell carcinoma: a systematic review and meta-analysis,” Dermatologic Therapy, 33 e13881 (2020). https://doi.org/10.1111/dth.13881 Google Scholar

Biographies of the authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Arsen K. Zotov, Arsenii A. Gavdush, Gleb M. Katyba, Larisa P. Safonova, Nikita V. Chernomyrdin, and Irina N. Dolganova "In situ terahertz monitoring of an ice ball formation during tissue cryosurgery: a feasibility test," Journal of Biomedical Optics 26(4), 043003 (27 January 2021). https://doi.org/10.1117/1.JBO.26.4.043003
Received: 15 November 2020; Accepted: 7 January 2021; Published: 27 January 2021
Lens.org Logo
CITATIONS
Cited by 5 scholarly publications.
Advertisement
Advertisement
KEYWORDS
Terahertz radiation

Tissues

Tissue optics

Spectroscopy

Signal detection

Interfaces

Temperature metrology

Back to Top