21 August 2019 Generating high-performance polarization measurements with low-performance polarizers: demonstration with a microgrid polarization camera
Author Affiliations +
Abstract

Microgrid polarization cameras have historically experienced lower performance than is typical for monolithic polarization systems. Specifically, their polarizer elements have had a lower extinction ratio and a larger orientation error. We show how to use the calibrated parameters of a nonideal polarizer to modify the polarization measurement model, effectively allowing one to generate high-performance measurements from low-performance elements. We demonstrate the effectiveness of this approach on a commercial polarization camera and estimate the signal-to-noise ratio penalty for using nonideal polarizers in this camera as being 1.25  ×   versus a system using ideal polarizers.

A polarizer can be used to estimate the orientation and degree of linear polarization of light by rotating the polarizer to the four angles: 0 deg, 45 deg, 90 deg, and 135 deg and measuring the light intensity transmitted at each angle. In a polarization camera, this can be done in a single snapshot by combining the measurements of four adjacent pixels, each of which have micropolarizers oriented at each of the four orientation angles (Fig. 1).1 For incoherent light, we can use the Stokes vector s=(s0,s1,s2,s3)T to represent the input polarization state and the Mueller calculus to model the four intensities Iθ measured at the detector:

Eq. (1)

I0=12(s0+s1)+n0,I45=12(s0+s2)+n45,I90=12(s0s1)+n90,I135=12(s0s2)+n135,
for random noise values nθ. In the analysis to follow, we will assume all of the noise terms to have zero mean, i.e., Ii+ni=Ii and ni=0. This causes no difficulties for Poisson-distributed noise, since our definition of the signal I and noise n results in the mean value of the Poisson-distributed variable to be incorporated into I while n retains the zero-mean stochastic portion.

Fig. 1

A polarization camera showing a 2×2  pixels group from the detector array and their polarization filter orientations.

OE_58_8_080501_f001.png

The four equations of (1) have the matrix form

Eq. (2)

(I0I45I90I135)I=12(1100101011001010)W(s0s1s2s3)s+(n0n45n90n135)n
or I=Ws+n, where W is the measurement matrix. By inspection, we can obtain the input linear Stokes vector elements as

Eq. (3)

s^0=12(I0+I45+I90+I135),s^1=I0I90,s^2=I45I135,
or

Eq. (4)

(s^0s^1s^2)s^=12(111120200202)A(I0I45I90I135)I,
i.e., s=AI, where A is the analysis matrix. To evaluate the performance of a polarization measurement, we can use the measurement and analysis matrix elements to calculate the Stokes vector covariance matrix K as2,3

Eq. (5)

Kij=[k=03sk=03AiAjWk]Poisson noise part+[vd=03AiAj]Gaussian noise part,
where sk is the k’th Stokes vector element in units of photoelectrons and vd is the detector noise variance, also in units of photoelectrons. For an ideal polarizer, this set of four measurements produces variances in the Stokes vector components as3

Eq. (6)

var(s^0)=12s0+vd,var(s^1)=s0+2vd,var(s^2)=s0+2vd.
The equally weighted variance (EWV) is often used to summarize the measurement performance, such that in this case EWV=52s0+5vd.3

A nonideal linear polarizer is a diattenuator defined by its diattenuation D, its orientation α, and its efficiency η (the average transmission of the diattenuator for all input polarization angles):4

Eq. (7)

Mld=η(1DcDs0Dc1Ds212ηDcs0Ds12ηDcs1Dc200001D),
for c=cos(2α) and s=sin(2α). An ideal polarizer has D=1, η=1, and orientation α. The polarizer’s extinction ratio X is derived from the diattenuation as X=(1+D)/(1D).

Modeling each pixel of a polarization camera as a nonideal polarizer, we can calibrate the diattenuation Dn and orientation αn parameters at each pixel n, then use the Mueller calculus to estimate the polarization state by combining measurements Ii from each set of four neighboring pixels. This produces the measurement model5

Eq. (8)

(I0I2I3I4)I=12(1D0cos(2α0)D0sin(2α0)1D1cos(2α1)D1sin(2α1)1D2cos(2α2)D2sin(2α2)1D3cos(2α3)D3sin(2α3))W(s0s1s2)s.
The polarizer efficiency η does not appear here because it typically cannot be separated from the pixel’s internal quantum efficiency qint—the two appear together as the single factor qext=ηqint as the “external” quantum efficiency. Because we do not calibrate the camera’s quantum efficiency here, the equations below assume qext=1, so that the Stokes parameters are represented in units of photoelectrons rather than photons.

The general analytical form of A is quite complex, but if we assume that the orientation errors are small then we can use a Taylor expansion about their nominal values:

cos(2α0)=cos(2ϵ0)1,cos(2α1)=cos(90°+2ϵ1)2ϵ1,cos(2α2)=cos(180°+2ϵ2)1,cos(2α3)=cos(270°+2ϵ3)2ϵ3,
and
sin(2α0)=sin(2ϵ0)2ϵ0,sin(2α1)=sin(90°+2ϵ1)1,sin(2α2)=sin(180°+2ϵ2)2ϵ2,sin(2α3)=sin(270°+2ϵ3)1,
where ϵn represents the angular error about the nominal orientation. Likewise, each diattenuation can be represented as Di=1Δi, where the diattenuation errors Δi are assumed to be small. For this first-order approximation regime, the analysis matrix A in the Stokes vector estimation:

Eq. (9)

(s0^s^1s^2)s^=14(A00A01A02A03A10A11A12A13A20A21A22A23)A(I0I1I2I3)I
simplifies to

Eq. (10)

A00=2+Δ0Δ2+2ϵ12ϵ3,A01=2+Δ1Δ32ϵ0+2ϵ2,A02=2Δ0+Δ22ϵ1+2ϵ3,A03=2Δ1+Δ3+2ϵ02ϵ2,A10=4+Δ0+3Δ2+2ϵ12ϵ3,A11=Δ0Δ24ϵ0+2ϵ1+4ϵ22ϵ3,A12=43Δ0Δ2+2ϵ12ϵ3,A13=Δ0Δ2+4ϵ02ϵ1+4ϵ2+2ϵ3,A20=Δ1Δ3+2ϵ0+4ϵ12ϵ2+4ϵ3,A21=4+Δ1+3Δ32ϵ0+2ϵ2A22=Δ1Δ3+2ϵ04ϵ12ϵ24ϵ3,A23=43Δ1Δ32ϵ0+2ϵ2.
Here, s^1 indicates the calibration-corrected Stokes estimate, while s^i without a prime indicates the estimate assuming an ideal polarizer. In Eqs. (9) and (10), note that it is generally not useful to define α0 as anything but zero, since the system reference axis is a degree of freedom that we have to choose in most situations. In this case, we can set ϵ0=0.

To demonstrate this technique, we use a commercially available polarization camera to measure the Stokes vector elements of the light transmitted through a high-extinction-ratio Glan–Thompson polarizer as the polarizer is rotated through a set of angles from 0 deg to 180 deg. The polarization camera is a photonic lattice PI-110, having 1164×874  pixels of 4.65-μm size, 20-Hz frame rate, 12-bit depth, and 520±20-nm wavelength range. Figure 2 shows the resulting Stokes parameter estimates before (s^1 and s^2) and after (s^1 and s^2) applying the correction, corresponding to using the ideal polarization analysis matrix [Eq. (4)] or a corrected analysis matrix [Eq. (10)]. We can see that the corrected curves display a higher contrast than the uncorrected curves, so that the mean values of the Stokes parameter (averaged across all camera pixels) at each angle of the input polarizer are closer to the correct values. Looking closely at the curves, we can also see that the standard deviation of the corrected result is slightly larger than for the uncorrected measurement.

Fig. 2

The measured uncorrected linear Stokes parameters s^1 and s^2 and corrected parameters s^1 and s^2 obtained while rotating the input angle of polarization from 0 deg to 180 deg. Note that the Stokes parameters, here, are shown in their normalized form: s˜1=s1/s0 and s˜2=s2/s0. Each curve is obtained as an average of the results from all pixels of the camera, while the grayed regions astride each curve give the standard deviation of the result, also calculated using all pixels on the camera. The dashed curves indicate the measurement result using the ideal polarization analysis matrix [Eq. (4)] and the solid curves give the measurement with the corrected analysis matrix [Eq. (10)].

OE_58_8_080501_f002.png

To see why the corrected result has higher noise, we can use Eqs. (5), (8), and (10) in their analytical form to calculate the general equation for the variances. To keep the expressions as simple as possible, and because our experimental data show that the diattenuation errors are much larger than the angular errors in our system, we set the angular errors to zero (ϵi0). Thus, in the detector-limited noise regime (uniform Gaussian noise), the variances become

Eq. (11)

var(s^0)=vd8[8+Δ02+Δ12+Δ22+Δ322Δ0Δ22Δ1Δ3]vd,var(s^1)=vd4[8+3Δ02+3Δ22+2Δ0(Δ2+4)+8Δ2]2vd[1+Δ0+Δ2],var(s^2)=vd4[8+3Δ12+3Δ32+2Δ1(Δ3+4)+8Δ3]2vd[1+Δ1+Δ3],
where the variances are the diagonal elements of the Stokes vector covariance matrix, var(s^0)=K00, etc. The approximations in Eq. (11) assume small errors so that only the first-order terms are retained. In this approximation, var(s^0) is independent of the diattenuation errors Δi, while var(s^1) increases linearly with Δ0 and Δ2, and var(s^2) with Δ1 and Δ3, respectively. The Stokes parameters are often expressed in normalized form s˜i=si/s0, for which the expressions in (11) need to be modified. The normalized parameter variances are related to their unnormalized counterparts by6

Eq. (12)

var(s˜i)=1s02var(si)+si2s04var(s0),
which together with Eq. (11) gives

Eq. (13)

var(s˜1)=vds02(2[1+Δ0+Δ2]+s˜12),var(s˜2)=vds02(2[1+Δ1+Δ3]+s˜22),
and where we have replaced the corrected parameter means with the true parameter values: s0s0, etc.

To see how these variance equations behave quantitatively, we can look at an example group of 4 pixels on the polarization camera. For this group, we obtain calibrated diattenuations of

D0=0.8277,D1=0.7435,D2=0.8133,D3=0.7338,
and calibrated orientation angles of
α0=0  deg,α1=49.650  deg,α2=88.694  deg,α3=134.399  deg,
so that the differences from the nominal angles are 0 deg, 4.650 deg, 1.306  deg, and 0.601  deg. Converting the angular errors to radians for the ϵi values, we obtain the error parameters:

Eq. (14)

Δ0=0.1723,Δ1=0.2565,Δ2=0.1867,Δ3=0.2662,ϵ0=0,ϵ1=0.0811,ϵ2=0.0228,ϵ3=0.0105.
With these parameters, the variances (11) in the detector-limited noise regime (uniform Gaussian noise) become, via Eq. (5),

Eq. (15)

var(s^0)=1.00vd,var(s^1)=3.23vd,var(s^2)=3.01vd,
for detector noise variance vd. In comparison to Eq. (6), we see that the corrected nonideal measurement for this pixel group suffers an increase in the variances for measuring s1 and s2 by factors of 1.62 and 1.51, respectively, while the variance for s0 is basically unaffected. These correspond to a loss in signal-to-noise ratio (SNR) by factors of 1.27 in s1 and 1.23 in s2.

Using the same Eqs. (2), (5), and (10), we can obtain similar results for the Poisson noise regime, for which the variance equations become

Eq. (16)

var(s^0)12s0+18s1(Δ0Δ2)+18s2(Δ1Δ3),var(s^1)s0(1+Δ0+Δ2)+s1(Δ2Δ0),var(s^2)s0(1+Δ1+Δ3)+s2(Δ3Δ1),
where only the first-order terms are retained. As we expect, the variances of Eqs. (11) and (16) approach those of the ideal polarizer system3 as the diattenuation errors approach zero (Δi0).

For mixed Gaussian–Poisson noise, the Gaussian and Poisson expressions (11) and (16) are simply added together to get the total variance. Using the example diattenuation errors of Eq. (14), the Stokes parameter variances under Poisson noise become

Eq. (17)

var(s^0)=0.500s00.001s10.002s2,var(s^1)=1.542s0+0.092s1+0s2,var(s^2)=1.381s0+0s1+0.014s2.

The results of Eqs. (15) and (17) and Fig. 3 demonstrate an SNR penalty due to reduced diattenuation that is similar to that calculated by Tyo and Wei,7 where the authors concluded via numerical simulation that polarizer extinction ratios as low as 3 give a 3.5 dB (i.e., a factor of 2.2) penalty in SNR. Although the measurement system analyzed by Tyo and Wei is not the same as the one we consider, here, the SNR penalty is similar in magnitude. Our results also agree with those of Roussel et al.,8 who analyzed microgrid polarization cameras and concluded that diattenuations of 0.75 to 0.85 will increase the variance in the estimated degree of polarization by 1.6 to 3. What is different from Roussel et al. is that the analytic dependence of the Stokes parameter variance is expressed here in terms of the calibration parameters directly, allowing us to see more clearly the effect of misalignment or low diattenuation on the measurements.

Fig. 3

The increase (multiplying factor) in the normalized Stokes parameter standard deviations std(s^1) and std(s^2) with respect to the input polarization angle, where the increased noise is caused by applying the correction of Eq. (9). The dashed lines indicate the average standard deviation ratio taken across all angles.

OE_58_8_080501_f003.png

In Fig. 2, we can see that the mean values of the corrected curves come close to the ideal behavior of a cosine curve oscillating between +1 and 1. The shaded regions surrounding each curve describe the values within one standard deviation about the mean, where we can expect to find the data most of the time. The noise fluctuations result primarily from the shot noise of the measurements, but are also influenced by any error in the calibration parameters. As we can expect for any linear estimation procedure, as the estimate approaches the ideal cosine curve, noise fluctuations cause some of the data at the peaks and valleys to extend beyond the physically valid region between +1 and 1.

From these results, we can see that most polarimeters can tolerate a surprisingly low diattenuation if calibrated well. As often is the case for indirect measurements, it is the accuracy with which one can calibrate the system measurement model that limits the measurement performance more than the inherent accuracy of the components themselves. If we work with low-extinction-ratio polarizers, we can still achieve high-accuracy measurements, but only if we take extreme care in the calibration and in ensuring the accuracy of our reconstruction.9 For example, we should keep in mind that the diattenuation value of a pixel is a spectrally dependent quantity, so that the measurement spectrum must be similar to the spectrum used during calibration in order for the results to be accurate.10

References

1. 

Y. Maruyama et al., “3.2-MP back-illuminated polarization image sensor with four-directional air-gap wire grid and 2.5-μm pixels,” IEEE Trans. Electron Devices, 65 2544 –2551 (2018). https://doi.org/10.1109/TED.2018.2829190 IETDAI 0018-9383 Google Scholar

2. 

F. Goudail, “Optimization of the contrast in active Stokes images,” Opt. Lett., 34 121 –123 (2009). https://doi.org/10.1364/OL.34.000121 OPLEDP 0146-9592 Google Scholar

3. 

N. Hagen and Y. Otani, “Stokes polarimeter performance: general noise model and analysis,” Appl. Opt., 57 4283 –4296 (2018). https://doi.org/10.1364/AO.57.004283 APOPAI 0003-6935 Google Scholar

4. 

R. A. Chipman, “Polarimetry,” Handbook of Optics, 2 22.1 –22.37 2nd ed.McGraw-Hill, New York (1995). Google Scholar

5. 

N. Hagen, S. Shibata and Y. Otani, “Calibration and performance assessment of microgrid polarization cameras,” Opt. Eng., 58 082408 (2019). https://doi.org/10.1117/1.OE.58.8.082408 Google Scholar

6. 

N. Hagen, “Statistics of normalized Stokes polarization parameters,” Appl. Opt., 57 5356 –5363 (2018). https://doi.org/10.1364/AO.57.005356 APOPAI 0003-6935 Google Scholar

7. 

J. S. Tyo and H. Wei, “Optimizing imaging polarimeters constructed with imperfect optics,” Appl. Opt., 45 5497 –5503 (2006). https://doi.org/10.1364/AO.45.005497 APOPAI 0003-6935 Google Scholar

8. 

S. Roussel, M. Boffety and F. Goudail, “Polarimetric precision of a micropolarizer grid-based camera in the presence of additive and Poisson shot noise,” Opt. Express, 26 29968 –29982 (2018). https://doi.org/10.1364/OE.26.029968 OPEXFF 1094-4087 Google Scholar

9. 

J. S. Tyo, “Design of optimal polarimeters: maximization of signal-to-noise ratio and minimization of systematic error,” Appl. Opt., 41 619 –630 (2002). https://doi.org/10.1364/AO.41.000619 APOPAI 0003-6935 Google Scholar

10. 

N. Hagen, “Flatfield correction errors due to spectral mismatching,” Opt. Eng., 53 (12), 123107 (2014). https://doi.org/10.1117/1.OE.53.12.123107 Google Scholar
CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Nathan Hagen, Shuhei Shibata, and Yukitoshi Otani "Generating high-performance polarization measurements with low-performance polarizers: demonstration with a microgrid polarization camera," Optical Engineering 58(8), 080501 (21 August 2019). https://doi.org/10.1117/1.OE.58.8.080501
Received: 3 June 2019; Accepted: 5 August 2019; Published: 21 August 2019
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications.
Advertisement
Advertisement
KEYWORDS
Polarization

Polarizers

Cameras

Calibration

Signal to noise ratio

Optical engineering

Sensors

RELATED CONTENT

The MIRC X 6 telescope imager key science drivers,...
Proceedings of SPIE (July 09 2018)
Performance assessment of microgrid polarization cameras
Proceedings of SPIE (September 06 2019)
Hyperspectral polarimetric imaging testbed
Proceedings of SPIE (June 01 2005)

Back to Top